manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Tektronix
  6. •
  7. Test Equipment
  8. •
  9. Tektronix TDS3012B User manual

Tektronix TDS3012B User manual

University of Portland - p. 1 of 9 - Oscilloscope - TDS3012B.docx
Oscilloscope Tutorial
Tektronix TDS3012B and TDS3012
Figure 1: Oscilloscope with a 10X probe connected to the Probe Comp Terminals
Introduction
The oscilloscope, which displays voltage versus time, is an indispensable tool for testing and
troubleshooting circuits and taking electrical measurements. This tutorial will explain how to set
up the oscilloscope and take basic measurements.
Getting Started
Do the following steps to set up the oscilloscope.
1. Turn on the oscilloscope by pressing the power button located in the lower left-hand
corner. Wait for scope to boot-up. When a message shows up on the screen, press the
MENU OFF button.
2. When the oscilloscope powers up, it uses the settings that were in place when it was last
powered down. To return the oscilloscope to the factory settings, press the Save/Recall
button. Then press the button under the screen labeled Recall Factory Setup. Next press
the button on the right of the screen labeled OK Confirm Factory Init.
University of Portland - p. 2 of 9 - Oscilloscope - TDS3012B.docx
3. Connect a 10X high-impedance probe (see Figure 2) to the channel 1 BNC connector
which is labeled as CH 1. When connecting the probe, gently push the connector while
rotating it clockwise until it slides in, then turn it clockwise so that it snaps securely in
place. The 10X high-impedance probe should ONLY be connected to the oscilloscope,
not to any other equipment. (To remove the probe, turn it counter-clockwise to release,
then gently pull on the connector. Please do NOT pull on the cable.)
Figure 2: 10X High-Impedance Scope Probes (Two Different Models)
4. The alligator clip on the probe is the ground lead. In order to make sure that the probe is
working correctly, connect the ground lead to the lower connector of the Probe Comp
port as shown above in Figure 1.
5. If you gently push on the probe tip, a hook is exposed that can be connected to a wire.
Connect the probe tip to the upper Probe Comp terminal (see Figure 1). The Probe Comp
port generates square wave that oscillates between 0 and 5V.
6. Press the Autoset button, which selects certain settings automatically depending on the
input signal. The display should look similar to that in Figure 1.
7. Check that the oscilloscope is reading the Probe Comp port voltage correctly (see below
for details on how to read the voltage). If the voltage isn’t reading correctly, refer to the
section Probe Setup below.
8. Connect the probe tip to the voltage in the circuit that you want to measure. Connect the
alligator clip to the circuit ground.
Note that the ground lead of the oscilloscope is connected to earth ground internally by
the oscilloscope, so it should only be connected to ground in a circuit. If the ground lead
is connected anywhere else in a circuit, it will short that node to earth ground.
University of Portland - p. 3 of 9 - Oscilloscope - TDS3012B.docx
Reading the Display
The oscilloscope generates a graph with volts on the vertical axis and time on the horizontal axis.
This section will explain how to read the display in order to measure time and voltage.
Measuring Voltage
The location on the screen that represents ground (or 0 Volts) is marked by an arrow on the left
side of the screen (see Figure 3). Note that the screen is divided into 8 divisions vertically,
where the divisions are marked by dotted lines on the screen. The number of volts that each
division represents (or volts/division) is displayed in the lower left part of the screen. To
measure a voltage, we can count the number of divisions above or below the ground level and
multiply by the volts/division value. For example, in Figure 3, the maximum voltage of the
square wave is 2.5 divisions above the ground level, so the voltage is 2.5 divisions * 2
volts/division = 5 Volts. If the voltage is below the ground level, then the voltage is negative.
Figure 3: Oscilloscope Screen
Measuring Time
Note that the screen is divided into 10 divisions horizontally, where the divisions are marked by
dotted lines on the screen. We can measure the time between two parts of a signal by counting
the number of divisions in the horizontal direction and multiplying by the number of seconds per
division. In Figure 3, for example, there are 2.5 divisions in one period of the wave, so the
period is 2.5 divisions * 400 µS = 1 mS.
University of Portland - p. 4 of 9 - Oscilloscope - TDS3012B.docx
Adjusting the Display
Although the Autoset button automatically selects some of the settings for you depending on the
current input signal, it is often necessary to manually adjust the settings to get a useful display.
The oscilloscope has adjustments that allow you to set the vertical and horizontal scales
independently, and to slide the graph vertically and horizontally.
Adjusting the Vertical and Horizontal Scale
The vertical scale knob shown in Figure 4 adjusts the volts per division setting of the vertical
axis. This adjustment allows you to zoom in to observe small signals and zoom out for large
signals. Note that when you rotate the knob, the value of the volts/division is updated in the
lower left of the screen. In order to get an accurate voltage measurement, we need to set the
vertical scale so that the signal is as big as possible without going off the top and/or bottom of
the screen.
Figure 4: Vertical Scale Knob
The horizontal scale knob shown in Figure 5 adjusts the seconds per division setting of the
horizontal axis. This adjustment allows you to zoom the horizontal scale. Note that when you
rotate the knob, the value of the seconds/division is updated in the lower middle of the screen. In
order to get an accurate time measurement, we need to set the horizontal scale so that the display
shows one or two periods of the signal.
Figure 5: Horizontal Scale Knob
When the vertical position knob (Figure 6) is rotated, the signal slides up or down and the arrow
that marks the location of ground (0 Volts) is updated. The horizontal position knob (Figure 7)
slides the signal left or right.
University of Portland - p. 5 of 9 - Oscilloscope - TDS3012B.docx
Figure 6: Vertical Position Knob Figure 7: Horizontal Position Knob
Coupling
The oscilloscope has three coupling modes which determine how the signal is processed before it
is displayed. In order to select the coupling mode, press the Vertical Menu button (see Figure 8),
then press the Coupling button under the screen (Figure 9), and then select the coupling mode
with one of the buttons on the right of the screen (Figure 9).
Figure 8: Vertical Menu Button Figure 9: Coupling Button
•DC: In DC coupling mode, no processing is performed on the signal before it is
displayed, and the display represents the signal with both DC and AC parts. This is the
default mode, and is used in most cases.
•AC: The AC mode is used for signals that have a small AC signal added to a large DC
component, and you are only interested in the AC signal. This mode is often used when
testing transistor amplifiers. In AC coupling mode, the DC component is blocked, and
only the AC part of the signal is displayed. This mode allows you to expand the vertical
scale to zoom in on the small AC signal, but keep in mind that the displayed voltage is
the actual voltage minus the DC offset, NOT the actual voltage.
•GND: In ground mode, the input signal is replaced by ground, which causes the
oscilloscope to draw a line on the screen at the ground level (0 Volts). This mode is used
to check where ground is on the screen. (There is also an arrow on the left side of the
screen that marks the ground level.)
University of Portland - p. 6 of 9 - Oscilloscope - TDS3012B.docx
Triggering
The oscilloscope repeatedly draws the input voltage on the screen from left to right, then returns
to the left side and draws the signal again on top of the previous curve. Each time the
oscilloscope draws the signal, it is called a sweep. If the oscilloscope did not have a way to
arrange it so each sweep started at the same point in the period for periodic signals, the screen
would display multiple sweeps, which would interfere with taking measurements (see Figure 10).
Figure 10: Oscilloscope is not triggering properly
Triggering is a mechanism that causes the oscilloscope to pause the sweep at the left side of the
screen until the input voltage crosses a user-selected voltage called the trigger level. This causes
each sweep to start in the same point in the period for periodic signals, and leads to a stable
display.
If the input signal never crosses the trigger level, the oscilloscope will wait for about a second,
and then it will draw the sweep. In this case the display may appear to drift across the screen or
may have several sweeps superimposed on the screen at the same time. Either way, it is difficult
or impossible to make measurements if the oscilloscope is not triggering properly.
The trigger level knob controls the trigger level (see Figure 11). In order for the oscilloscope to
trigger properly, the trigger level must be set to a voltage that the input signal crosses.
Figure 11: Trigger Level Knob
University of Portland - p. 7 of 9 - Oscilloscope - TDS3012B.docx
It is often difficult to trigger properly on noisy signals. Fortunately, the oscilloscope has a mode
where it passes the trigger signal through a lowpass filter (LPF), which eliminates high
frequency noise. (This LPF only affects the trigger signal, not the displayed signal.) To activate
this mode, press the Trigger Menu button, then Coupling, then HF Reject.
If it is impossible to trigger on a signal for some reason, another option is to temporarily freeze
the display by pressing the Run/Stop button, which will also cause “Stop” to be displayed in the
upper left-hand corner of the display. Press the Run/Stop button again to restart the oscilloscope.
Measurements using the Cursors
The cursors can be used to measure time differences or voltage levels on the oscilloscope.
Measuring Time Difference
As mentioned above, one way to measure the time difference is to multiply the number of
horizontal divisions by the seconds/division value. This method is often quick and easy, but it is
sometimes difficult to measure the number of divisions accurately. An alternative approach is to
use the vertical cursors, which are sometimes more accurate. Activate the vertical cursors by
pressing the Cursor button (see Figure 12), and then pressing the V Bars button (Figure 13).
Figure 12: Cursor Button Figure 13: V Bars Button
Turning the knob (see Figure 14) moves one of the cursors. To move the other cursor, press the
Select button and then turn the knob. The time difference between the cursors is displayed after
the Δ symbol in the top right of the screen (see Figure 13). The time measurement will be most
accurate if the horizontal scale is adjusted so that the time difference is as wide as possible.
Figure 14: Cursor Knob and Select Button
University of Portland - p. 8 of 9 - Oscilloscope - TDS3012B.docx
Measuring Voltage
As mentioned above, one way to measure voltage is to multiply the number of vertical divisions
by the volts/division value. This method is often quick and easy, but it is sometimes difficult to
measure the number of divisions accurately. An alternative approach is to use the horizontal
cursors, which are sometimes more accurate. Activate the horizontal cursors by pressing the
cursor button (see Figure 15), and then pressing the H Bars button (Figure 16).
Figure 15: Cursor Button Figure 16: H Bars Button
Turning the knob (see Figure 17) moves one of the cursors and the voltage at the cursor level is
displayed in the upper right-hand corner of the screen after the “@” symbol. To move the other
cursor, press the Select button and then turn the knob. The voltage difference between the
cursors is displayed after the Δ symbol.
Figure 17: Cursor Knob and Select Button
University of Portland - p. 9 of 9 - Oscilloscope - TDS3012B.docx
Automated Measurements
The oscilloscope can also measure various values automatically. To activate one or more of
these measurements, press the Measure button, Select Measrmnt, and then the button next to the
desired measurement. To see the full list of automated measurements, click on the More button.
To turn a measurement off, press the Measure button, Remove Measrmnt, and then Measurement
1 button (or the button for the measurement you wish to remove).
Depending on the vertical and horizontal scale settings, the automated measurements are not
always accurate, so it is good practice to double-check these measurements by multiplying the
number of divisions by the scale or by using the using the cursors.
Displaying Two Signals Simultaneously
The oscilloscope can display two signals at the same time. To add a second signal, connect a
10X probe to the CH 2 input and connect it to the signal that you want to display. Press the blue
CH 2 button to turn on the channel 2 trace on the oscilloscope display. Pressing the CH 2 button
also causes the Vertical Scale and Position knobs to control the channel 2 trace on the screen
instead of the channel 1 trace. If you want to adjust the vertical scale or position of Channel 1,
you first switch back to Channel 1 by pressing the CH 1 button, and then turn the Vertical Scale
or Position knob. The horizontal scale is always the same for both channels and can be adjusted
when either channel is active. You can turn one of the channels off by pressing the CH 1 or CH
2 button and then pressing the OFF button.
Probe Setup
There is a connection on most (but not all) 10X probes that tells the oscilloscope that a 10X
probe is connected. Since the 10X probe divides the input signal by 10X, if the oscilloscope
detects a 10X probe is connected, it will multiply the signal by 10 times to compensate. If the
Probe Comp signal (see Figure 1) does not appear to be a square wave from 0 to 5 Volts, check
to see if the oscilloscope has detected the 10X probe correctly by pressing the Vertical Menu
button, and then press the Probe Setup button. The value should be 10X if a 10X probe is
connected to the input (and 1X if a coax cable is connected).

This manual suits for next models

1

Other Tektronix Test Equipment manuals

Tektronix 2235 Owner's manual

Tektronix

Tektronix 2235 Owner's manual

Tektronix 516 series User manual

Tektronix

Tektronix 516 series User manual

Tektronix MSO5000 Series User manual

Tektronix

Tektronix MSO5000 Series User manual

Tektronix OA 5000 Series User manual

Tektronix

Tektronix OA 5000 Series User manual

Tektronix 519 User manual

Tektronix

Tektronix 519 User manual

Tektronix MSO70000 Series Use and care manual

Tektronix

Tektronix MSO70000 Series Use and care manual

Tektronix TekScope THS710A User manual

Tektronix

Tektronix TekScope THS710A User manual

Tektronix 3S3 User manual

Tektronix

Tektronix 3S3 User manual

Tektronix 511A-AD User manual

Tektronix

Tektronix 511A-AD User manual

Tektronix 545A User manual

Tektronix

Tektronix 545A User manual

Tektronix 2445 User manual

Tektronix

Tektronix 2445 User manual

Tektronix KEITHLEY S530 User manual

Tektronix

Tektronix KEITHLEY S530 User manual

Tektronix 2445 User manual

Tektronix

Tektronix 2445 User manual

Tektronix 5000 Series User manual

Tektronix

Tektronix 5000 Series User manual

Tektronix TMS 540 User manual

Tektronix

Tektronix TMS 540 User manual

Tektronix TDS3000 Series Troubleshooting guide

Tektronix

Tektronix TDS3000 Series Troubleshooting guide

Tektronix P5210 User manual

Tektronix

Tektronix P5210 User manual

Tektronix TBS2000 Series User manual

Tektronix

Tektronix TBS2000 Series User manual

Tektronix MSO56 Operating and maintenance manual

Tektronix

Tektronix MSO56 Operating and maintenance manual

Tektronix 524AD series User manual

Tektronix

Tektronix 524AD series User manual

Tektronix MSO2000 Series User manual

Tektronix

Tektronix MSO2000 Series User manual

Tektronix RSA5100B series User manual

Tektronix

Tektronix RSA5100B series User manual

Tektronix P6042 PROBE User manual

Tektronix

Tektronix P6042 PROBE User manual

Tektronix 2440 Product information sheet

Tektronix

Tektronix 2440 Product information sheet

Popular Test Equipment manuals by other brands

Laversab 6500 user manual

Laversab

Laversab 6500 user manual

Gardner Bender GVT-392 owner's manual

Gardner Bender

Gardner Bender GVT-392 owner's manual

Megger TRS1 user manual

Megger

Megger TRS1 user manual

ION Tiger Select user manual

ION

ION Tiger Select user manual

Viavi OneAdvisor ONA-800 Quick Card

Viavi

Viavi OneAdvisor ONA-800 Quick Card

Albrecht SWR-20 manual

Albrecht

Albrecht SWR-20 manual

Veryx PktBlaster installation guide

Veryx

Veryx PktBlaster installation guide

Cutler-Hammer HWT-500 instructions

Cutler-Hammer

Cutler-Hammer HWT-500 instructions

TECALEMIT DE9419 user manual

TECALEMIT

TECALEMIT DE9419 user manual

KYORITSU 3165 instruction manual

KYORITSU

KYORITSU 3165 instruction manual

Consilium Salwico SE3000+ user guide

Consilium

Consilium Salwico SE3000+ user guide

JAVAC EVO-OS user manual

JAVAC

JAVAC EVO-OS user manual

Hazet 4650-5 Operation manual

Hazet

Hazet 4650-5 Operation manual

Test Equipment Depot Rapport 337 user manual

Test Equipment Depot

Test Equipment Depot Rapport 337 user manual

DSI Buxco FinePointe user manual

DSI

DSI Buxco FinePointe user manual

Kikusui tos6200 Operation manual

Kikusui

Kikusui tos6200 Operation manual

Uniks ISO-5 operating manual

Uniks

Uniks ISO-5 operating manual

Cedar DI-1M instruction manual

Cedar

Cedar DI-1M instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.