WaveMaster 8000HD Getting Started Guide 3
Power
* All PC peripherals and active probes connected to all channels.
WARNING. Secure the power cord with the cable clamp at all times.
Failure to do so creates a fire hazard.
Operating Environment Precautions
Observe all generally accepted safety procedures.
Use proper power cord. Use only the power cord shipped with this
instrument and certified for country of use. Secure the cord at all times.
Maintain ground.
The AC inlet ground is connected to the frame of the
instrument. Connect line cords only to outlets with safety ground contacts.
WARNING. Interrupting the protective conductor inside or outside
the oscilloscope, or disconnecting the safety ground terminal,
creates a hazardous situation. Intentional interruption is prohibited.
Connect and disconnect properly. Do not connect/disconnect probes or
cables while they are connected to a live voltage source.
Observe all terminal ratings. Do not apply a voltage to any input that
exceeds the maximum rating of that input. Refer to the markings next to
the terminals for maximum allowed values.
Use only within operating environment listed here. Do not use in wet or
explosive environments.
Use indoors only.
Keep product surfaces clean and dry.
Do not block the cooling vents. Leave a minimum six-inch (15 cm) gap
between the instrument and the nearest object. Keep the underside clear
of papers and other objects.
Exercise care when lifting and carrying. Unplug the instrument to move it.
Do not remove the covers or inside parts. Refer all maintenance to
qualified service personnel.
Do not operate with suspected failures. Check body and cables regularly.
If any part is damaged, cease operation immediately and secure the
instrument from inadvertent use.
OVERVIEW
Temperature 5 °C to 40 °C
Humidity Maximum RH 80% (non-condensing) up to 31 °C
decreasing linearly to RH 50% at 40 °C
Altitude Up to 10,000 ft (3,048 m) at or below 30 °C
AC Power 100-240 VAC (±10%) at 50/60 Hz (±10%)
Automatic AC Voltage Selection
Consumption Standard Models DBI Models
Nominal 1125 W 1175 W
Maximum* 1250 W 1300 W
Standby 15 W 15 W
Measuring Terminal Ratings
Measuring terminals have no rated measurement category (CAT) per EN IEC 61010-2-30:2021.
Measuring terminals are not intended to be connected directly to supply mains.
ProBus Inputs 50 Ω coupling ≤ 5 Vrms
1 MΩ coupling ≤ 400 Vpk max. (Peak AC ≤ 10 KHz + DC);
derating at 18 dB/decade from 10 KHz to 1 MHz;
5 Vpk max. above 1 MHz
ProLink Inputs 50 Ω coupling ±2 Vmax
ProAxial Inputs 50 Ω coupling ±2 Vmax
1.85 mm (DBI) Inputs 50 Ω coupling ±2 Vmax