Teledyne TIENet 360 LaserFlow Ex Operating instructions


Foreword
This instruction manual is designed to help you gain a thorough understanding of the operation of
the equipment. Teledyne ISCO recommends that you read this manual completely before placing
the equipment in service.
Although Teledyne ISCO designs reliability into all equipment, there is always the possibility of a
malfunction. This manual may help in diagnosing and repairing the malfunction.
If a problem persists, call or e-mail Teledyne ISCO technical support for assistance. Simple
difficulties can often be diagnosed over the phone. For faster service, please have your serial
number ready.
If it is necessary to return the equipment to the factory for service, please follow the shipping
instructions provided by technical support, including the use of the Return Material
Authorization (RMA) specified. Be sure to include a note describing the malfunction. This will
aid in the prompt repair and return of the equipment.
Teledyne ISCO welcomes suggestions that would improve the information presented in this manual
or enhance the operation of the equipment itself.
Teledyne ISCO is continually improving its products and reserves the right to change product
specifications, replacement parts, schematics, and instructions without notice.
Contact Information
Customer Service
Phone: (800) 228-4373 (USA, Canada, Mexico)
(402) 464-0231 (Outside North America)
Fax: (402) 465-3022
Technical Support
Phone: Toll Free (800) 775-2965 (Syringe Pumps and Liquid Chromatography)
Email: [email protected]
Return equipment to: 4700 Superior Street, Lincoln, NE 68504-1398
Other Correspondence
Mail to: P.O. Box 82531, Lincoln, NE 68501-2531
Email: [email protected]
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.


1-1 Basic LaserFlow system (showing both AC and battery-powered options) . . . . . 1-2
1-2 Optional submerged functionality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
1-3 Remove protective window cap before installation . . . . . . . . . . . . . . . . . . . . . . . . 1-8
2-1 Ultrasonic level sensor beam angle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2-2 TIENet Device terminal strips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
2-3 Installing cable with a cord-grip fitting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
2-4 TIENet Device terminal connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
2-5 Attach wired terminal strip to case board socket . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
2-6 Insert the cable reference tubing into the case board reference port . . . . . . . . . . 2-6
2-7 Position and secure the cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
2-8 Preparing the LaserFlow connector receptacle . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
2-9 Secure caps and connect 310 ultrasonic plug to LaserFlow . . . . . . . . . . . . . . . . . . 2-8
2-10 Preparing the 2160 connector receptacle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
2-11 Secure caps and connect LaserFlow to 2160 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
2-12 Overall dimensions including wall mounting hardware (part 1) . . . . . . . . . . . 2-11
2-13 Overall dimensions including wall mounting hardware (part 2) . . . . . . . . . . . 2-12
2-14 Overall dimensions including temporary mounting hardware (part 1) . . . . . . 2-13
2-15 Overall dimensions including temporary mounting hardware (part 2) . . . . . . 2-14
2-16 Sensor installation constants . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
2-17 Typical manhole installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
2-18 Wall mounting kit (Numbers correspond with Table 2-1) . . . . . . . . . . . . . . . . . 2-17
2-19 Wall Mount: Vertical clamp and vertical clamp with tube coupler inserted . . 2-18
2-20 Wall Mount: Sensor carrier installed on tube coupler . . . . . . . . . . . . . . . . . . . . 2-19
2-21 Wall Mount: Place the sensor on the carrier . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-20
2-22 Turn the locking handle clockwise until sensor is locked into place (shown with op-
tional sensor retrieval arm, described in Section 2.8) . . . . . . . . . . . . . . . . . . . . . 2-21
2-23 Temporary mounting kit (Numbers correspond with Table 2-2) . . . . . . . . . . . . 2-22
2-24 Temporary Mount: Remove stationary foot . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23
2-25 Temporary Mount: Install coupler, replace stationary foot . . . . . . . . . . . . . . . . 2-23
2-26 Temp Mount: Securing the coupler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-24
2-27 Temporary Mount: Install sensor carrier/knuckle onto 24" pipe . . . . . . . . . . . . 2-24
2-28 Installing the sensor carrier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-25
2-29 Temporary Mount: Place the sensor on the carrier . . . . . . . . . . . . . . . . . . . . . . 2-26
2-30 Turn the locking handle clockwise until sensor is locked into place (shown with the
optional sensor retrieval arm, described in Section 2.8) . . . . . . . . . . . . . . . . . . . 2-26
2-31 Temporary Mount: Optional elbow pipe for offset . . . . . . . . . . . . . . . . . . . . . . . 2-27
2-32 Suggested placement of Laserflow sensor when using temporary mounting hard-
ware and the sensor retrieval bar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-28
2-33 Optional Sensor Retrieval Arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-29
2-34 Mounting the sunshade . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-30
2-35 Mounted sunshade . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-31
3-1 Character grid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
3-2 Menu Tree: LaserFlow Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3-3 Measurement Setup: Sensor input settings
(optional 350 AV Sensor also shown) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
3-4 Blanking distance for LaserFlow with built-in sensor or with remote ultrasonic sen-
sor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
3-5 Measurement setup: Advanced settings for LaserFlow sensor . . . . . . . . . . . . . . . 3-8
3-6 350 Level adjustment screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
3-7 Submerged functionality: Initial 350 Level setting . . . . . . . . . . . . . . . . . . . . . . . 3-10
3-8 Measurement setup: Advanced settings for 350 AV sensor . . . . . . . . . . . . . . . . . 3-12
3-9 360 Velocity grid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
3-10 Selection of velocity measurement points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
3-11 Ultrasonic level adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
4-1 Activate connected sensors in Flowlink . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4-2 Activating connected sensors in Flowlink . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
4-3 Measurements tab: Displays real-time measurements . . . . . . . . . . . . . . . . . . . . . 4-3
4-4 Drop down list of measurements for 2160 Module . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
4-5 360 Level Measurement tab in Flowlink . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5

4-6 Blanking distance for LaserFlow with built-in sensor or 310 remote ultrasonic sensor
4-6
4-7 360 Velocity Measurement tab in Flowlink . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
4-8 360 Velocity Tab: Advanced settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
4-9 . Remote distance measurement with the 310 remote ultrasonic sensor . . . . . . 4-10
4-10 Remote distance measurement with the 310 remote ultrasonic sensor and horizon-
tal sensor mount bracket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
4-11 LaserFlow placement using Parallel to Channel Slope option . . . . . . . . . . . . . 4-12
4-12 LaserFlow placement using Slope Percentage option . . . . . . . . . . . . . . . . . . . . 4-13
4-13 350 Level Measurement tab in Flowlink . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
4-14 Submerged functionality: Initial 350 Level setting . . . . . . . . . . . . . . . . . . . . . . 4-15
4-15 350 Velocity Measurement tab in Flowlink . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-16
4-16 350 Velocity tab: Advanced settings for optional TIENet 350 AV sensor . . . . . 4-17
4-17 Selection of velocity measurement points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18
5-1 Remove large screws and nose piece (3 on each side) . . . . . . . . . . . . . . . . . . . . . . 5-3
5-2 Location and orientation of cross-pins for mounting block installation . . . . . . . . 5-4
5-3 Cleaning the laser window: Moisten Kimwipe tissue w/ Rain-X . . . . . . . . . . . . . . 5-6
5-4 350 AV sensor with mounting plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
5-5 Mounting the integrated 350 AV sensor on the LaserFlow . . . . . . . . . . . . . . . . . . 5-8
5-6 Connecting the 350 AV sensor to the LaserFlow . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
5-7 Reassembling the LaserFlow w/ AV sensor installed . . . . . . . . . . . . . . . . . . . . . . 5-10
B-1 Do not install LaserFlow with built-in sensor in a free-flowing outfall or in a drop
manhole. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-5
B-2 Correct setup when using the LaserFlow module with remote ultrasonic sensor and
horizontal mount bracket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-6
B-3 Placement of LaserFlow with built-in sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-6
B-4 Placement of LaserFlow with remote ultrasonic sensor . . . . . . . . . . . . . . . . . . . . B-7
C-1 Signature flow data: Selecting the .ddp file(s) . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-2
C-2 Signature flow data: Importing the .ddp file . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-2
C-3 Identifying error codes in the 360 Velocity data set . . . . . . . . . . . . . . . . . . . . . . . C-3

TIENet™ 360 LaserFlow™ Velocity Sensor
Safety
v
TIENet™ 360 LaserFlow™ Velocity Sensor
Safet y
General Warnings
WARNING
Caution: Use ofcontrols or adjustments or performance of
procedures other than those specified hereinmay result in
hazardous radiation exposure.
Wavelength: 655 nm
Maximum Power: 5 mW
Before installing, operating, or maintaining this equipment, it is
imperative that all hazards and preventive measures are fully
understood. While specific hazards may vary according to
location and application, take heed of the following general
warnings:
WARNING
Avoid hazardous practices! If you use this instrument in
any way not specified in this manual, the protection
provided by the instrument may be impaired.

TIENet™ 360 LaserFlow™ Velocity Sensor
Safety
vi
Hazard Severity Levels This manual applies Hazard Severity Levels to the safety alerts,
These three levels are described in the sample alerts below.
CAUTION
Cautions identify a potential hazard, which if not avoided, may
result in minor or moderate injury. This category can also warn
you of unsafe practices, or conditions that may cause property
damage.
WARNING
Warnings identify apotentially hazardous condition, which
if not avoided, could result in death or serious injury.
DANGER
DANGER–limitedtothe mostextremesituations
to identify an imminent hazard, which if not
avoided, will result in death or serious injury.

TIENet™ 360 LaserFlow™ Velocity Sensor
Safety
vii
Hazard Symbols The equipment and this manual use symbols used to warn of
hazards. The symbols are explained below.
Hazard Symbols
Warnings and Cautions
The exclamation point within the triangle is a warning sign alerting you of
important instructions in the instrument’s technical reference manual.
The lightning flash and arrowhead within the triangle is a warning sign alert-
ing you of “dangerous voltage” inside the product.
Symboles de sécurité
Ce symbole signale l’existence d’instructions importantes relatives au
produit dans ce manuel.
Ce symbole signale la présence d’un danger d’électocution.
Warnungen und Vorsichtshinweise
Das Ausrufezeichen in Dreieck ist ein Warnzeichen, das Sie darauf
aufmerksam macht, daß wichtige Anleitungen zu diesem Handbuch
gehören.
Der gepfeilte Blitz im Dreieck ist ein Warnzeichen, das Sei vor “gefährlichen
Spannungen” im Inneren des Produkts warnt.
Advertencias y Precauciones
Esta señal le advierte sobre la importancia de las instrucciones del manual
que acompañan a este producto.
Esta señal alerta sobre la presencia de alto voltaje en el interior del
producto.

TIENet™ 360 LaserFlow™ Velocity Sensor
Safety
viii

vii
TIENet™ 360 LaserFlow™
Velocity Sensor
Table of Contents
Section 1 Introduction
1.1 Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1.2 Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1.2.1 Measurement Technology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1.2.2 Cabling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1.3 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1.4 Optional Submerged Functionality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
1.5 Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
1.6 Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
1.6.1 Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
1.7 Unpacking Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
1.7.1 Protective Window Cap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Section 2 Preparation and Installation
2.1 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2.1.1 Site Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2.2 Sensor Installation Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2.2.1 Ultrasonic Beam Angle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2.2.2 High Water Levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2.2.3 Submersion and Fouling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2.2.4 Humidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2.2.5 Surface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
2.2.6 Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
2.2.7 Waves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
2.2.8 Wind . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
2.3 Installation Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
2.4 Connecting the Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
2.4.1 Cable Connection: Signature Flow Meter . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
2.4.2 Remote Ultrasonic Cable Connection:
310 Ultrasonic Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
2.4.3 Cable Connection: 2160 LaserFlow Module . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
2.5 Sensor Installation Tools and Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
2.5.1 Tools Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
2.5.2 Sensor and Mounting Equipment Dimensions . . . . . . . . . . . . . . . . . . . . . 2-10
2.5.3 Sensor Positioning and Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
2.6 Wall Mounting Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17
2.6.1 Wall Mount Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18
2.7 Temporary Mounting Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22
2.7.1 Temporary Mounting Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23
2.7.2 Offset Carrier from Cargo Bar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-27
2.8 Optional Sensor Retrieval Arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-28
2.9 Optional Sunshade . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-30
2.10 Remote TIENet™ Model 310 Ultrasonic Level Sensor . . . . . . . . . . . . . . . . . . . . 2-31

TIENet™ 360 LaserFlow™ Velocity Sensor
Table of Contents
viii
Section 3 Setup with Signature® Flow Meter
3.1 Configuring the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.1.1 Updating the Device List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.2 Measurement Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
3.2.1 360 Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
3.2.2 360 Velocity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
3.2.3 LaserFlow Velocity Advanced Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
3.2.4 Adjust 350 Level (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
3.2.5 350 Velocity (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
3.2.6 350 Advanced Settings (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
3.3 360 Velocity Grid. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
3.4 Sensor Positioning and Adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
3.4.1 Centering the LaserFlow Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
3.4.2 Laser Velocity Measurement Point Selection . . . . . . . . . . . . . . . . . . . . . . 3-14
3.4.3 Ultrasonic Level Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
3.5 Contact Teledyne Isco . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
Section 4 Setup with 2160 LaserFlow Module
4.1 Activating Connected Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4.2 Measurement Display and Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
4.2.1 360 Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
4.2.2 360 Velocity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
4.2.3 Advanced Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
4.2.4 350 Level (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
4.2.5 350 Velocity (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-16
4.3 Sensor Positioning and Adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17
4.3.1 Centering the LaserFlow Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17
4.3.2 Laser Velocity Measurement Point Selection . . . . . . . . . . . . . . . . . . . . . . 4-18
Section 5 Maintenance and Optional Kits
5.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
5.2 Firmware Updates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
5.2.1 With Signature Flow Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
5.2.2 With 2160 LaserFlow Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
5.3 Cleaning the Sensor Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
5.4 Accessing the LaserFlow’s Nose Piece . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
5.4.1 Removing the Nose Piece . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
5.4.2 Replacing the Nose Piece . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
5.5 Velocity Readings and the Laser Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
5.6 Cleaning the Laser Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
5.7 Installing the Optional Submerged Measurement Kit . . . . . . . . . . . . . . . . . . . . . . 5-7
5.7.1 Installation Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
5.8 Installing the Remote TIENet™ 310 Ultrasonic Level Sensor. . . . . . . . . . . . . . . 5-10
Appendix A Replacement Parts
A.1 Replacement Parts Diagrams and Listings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
A.2 LaserFlow Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
A.3 Wall Mounting Hardware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-8
A.4 Temporary Mounting Hardware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-10
A.5 Replacement Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-12
A.6 Sensor Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-12
Appendix B Installation Checklist

TIENet™ 360 LaserFlow™ Velocity Sensor
Table of Contents
ix
B.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
B.2 Equipment and Recommended ToolsLaptop computer . . . . . . . . . . . . . . . . . . . . . B-2
B.3 Site Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-2
B.4 Recommended Essential Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-5
B.5 Site Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-5
B.6 Sensor Positioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-8
Appendix C Velocity Error Codes
C.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-1
C.2 Importing Data Dump (.ddp) Files
(Signature Only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-1
C.3 Viewing Velocity Error Codes in Flowlink . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-3
Appendix D Material Safety Data Sheets
List of Figures
1-1 Basic LaserFlow system (showing both AC and battery-powered options) . . . . . 1-2
1-2 Optional submerged functionality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
1-3 Remove protective window cap before installation . . . . . . . . . . . . . . . . . . . . . . . . 1-8
2-1 Ultrasonic level sensor beam angle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2-2 TIENet Device terminal strips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
2-3 Installing cable with a cord-grip fitting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
2-4 TIENet Device terminal connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
2-5 Attach wired terminal strip to case board socket . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
2-6 Insert the cable reference tubing into the case board reference port . . . . . . . . . . 2-6
2-7 Position and secure the cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
2-8 Preparing the LaserFlow connector receptacle . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
2-9 Secure caps and connect 310 ultrasonic plug to LaserFlow . . . . . . . . . . . . . . . . . . 2-8
2-10 Preparing the 2160 connector receptacle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
2-11 Secure caps and connect LaserFlow to 2160 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
2-12 Overall dimensions including wall mounting hardware (part 1) . . . . . . . . . . . 2-11
2-13 Overall dimensions including wall mounting hardware (part 2) . . . . . . . . . . . 2-12
2-14 Overall dimensions including temporary mounting hardware (part 1) . . . . . . 2-13
2-15 Overall dimensions including temporary mounting hardware (part 2) . . . . . . 2-14
2-16 Sensor installation constants . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
2-17 Typical manhole installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
2-18 Wall mounting kit (Numbers correspond with Table 2-1) . . . . . . . . . . . . . . . . . 2-17
2-19 Wall Mount: Vertical clamp and vertical clamp with tube coupler inserted . . 2-18
2-20 Wall Mount: Sensor carrier installed on tube coupler . . . . . . . . . . . . . . . . . . . . 2-19
2-21 Wall Mount: Place the sensor on the carrier . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-20
2-22 Turn the locking handle clockwise until sensor is locked into place (shown with op-
tional sensor retrieval arm, described in Section 2.8) . . . . . . . . . . . . . . . . . . . . . 2-21
2-23 Temporary mounting kit (Numbers correspond with Table 2-2) . . . . . . . . . . . . 2-22
2-24 Temporary Mount: Remove stationary foot . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23
2-25 Temporary Mount: Install coupler, replace stationary foot . . . . . . . . . . . . . . . . 2-23
2-26 Temp Mount: Securing the coupler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-24
2-27 Temporary Mount: Install sensor carrier/knuckle onto 24" pipe . . . . . . . . . . . . 2-24
2-28 Installing the sensor carrier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-25
2-29 Temporary Mount: Place the sensor on the carrier . . . . . . . . . . . . . . . . . . . . . . 2-26
2-30 Turn the locking handle clockwise until sensor is locked into place (shown with the
optional sensor retrieval arm, described in Section 2.8) . . . . . . . . . . . . . . . . . . . 2-26
2-31 Temporary Mount: Optional elbow pipe for offset . . . . . . . . . . . . . . . . . . . . . . . 2-27
2-32 Suggested placement of Laserflow sensor when using temporary mounting hard-
ware and the sensor retrieval bar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-28

TIENet™ 360 LaserFlow™ Velocity Sensor
Table of Contents
x
2-33 Optional Sensor Retrieval Arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-29
2-34 Mounting the sunshade . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-30
2-35 Mounted sunshade . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-31
3-1 Character grid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
3-2 Menu Tree: LaserFlow Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3-3 Measurement Setup: Sensor input settings
(optional 350 AV Sensor also shown) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
3-4 Blanking distance for LaserFlow with built-in sensor or with remote ultrasonic sen-
sor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
3-5 Measurement setup: Advanced settings for LaserFlow sensor . . . . . . . . . . . . . . . 3-8
3-6 350 Level adjustment screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
3-7 Submerged functionality: Initial 350 Level setting . . . . . . . . . . . . . . . . . . . . . . . . 3-9
3-8 Measurement setup: Advanced settings for 350 AV sensor . . . . . . . . . . . . . . . . . 3-11
3-9 360 Velocity grid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
3-10 Selection of velocity measurement points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
3-11 Ultrasonic level adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
4-1 Activate connected sensors in Flowlink . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4-2 Activating connected sensors in Flowlink . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
4-3 Measurements tab: Displays real-time measurements . . . . . . . . . . . . . . . . . . . . . 4-3
4-4 Drop down list of measurements for 2160 Module . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
4-5 360 Level Measurement tab in Flowlink . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
4-6 Blanking distance for LaserFlow with built-in sensor or 310 remote ultrasonic sensor
4-6
4-7 360 Velocity Measurement tab in Flowlink . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
4-8 360 Velocity Tab: Advanced settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
4-9 . Remote distance measurement with the 310 remote ultrasonic sensor . . . . . . 4-10
4-10 Remote distance measurement with the 310 remote ultrasonic sensor and horizon-
tal sensor mount bracket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
4-11 LaserFlow placement using Parallel to Channel Slope option . . . . . . . . . . . . . 4-12
4-12 LaserFlow placement using Slope Percentage option . . . . . . . . . . . . . . . . . . . . 4-13
4-13 350 Level Measurement tab in Flowlink . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
4-14 Submerged functionality: Initial 350 Level setting . . . . . . . . . . . . . . . . . . . . . . 4-15
4-15 350 Velocity Measurement tab in Flowlink . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-16
4-16 350 Velocity tab: Advanced settings for optional TIENet 350 AV sensor . . . . . 4-17
4-17 Selection of velocity measurement points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18
5-1 Remove large screws and nose piece (3 on each side) . . . . . . . . . . . . . . . . . . . . . . 5-3
5-2 Location and orientation of cross-pins for mounting block installation . . . . . . . . 5-4
5-3 Cleaning the laser window: Moisten Kimwipe tissue w/ Rain-X . . . . . . . . . . . . . . 5-6
5-4 350 AV sensor with mounting plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
5-5 Mounting the integrated 350 AV sensor on the LaserFlow . . . . . . . . . . . . . . . . . . 5-8
5-6 Connecting the 350 AV sensor to the LaserFlow . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
5-7 Reassembling the LaserFlow w/ AV sensor installed . . . . . . . . . . . . . . . . . . . . . . 5-10
B-1 Do not install LaserFlow with built-in sensor in a free-flowing outfall or in a drop
manhole. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-5
B-2 Correct setup when using the LaserFlow module with remote ultrasonic sensor and
horizontal mount bracket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-6
B-3 Placement of LaserFlow with built-in sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-6
B-4 Placement of LaserFlow with remote ultrasonic sensor . . . . . . . . . . . . . . . . . . . . B-7
C-1 Signature flow data: Selecting the .ddp file(s) . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-2
C-2 Signature flow data: Importing the .ddp file . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-2
C-3 Identifying error codes in the 360 Velocity data set . . . . . . . . . . . . . . . . . . . . . . . C-3

1-1
TIENet™ 360 LaserFlow™
Velocity Sensor
Section 1 Introduction
1.1 Description The TIENet™ 360 LaserFlow sensor is an Area Velocity flow
measurement device that remotely measures flow in open
channels with non-contact Laser Doppler Velocity Sensing and
non-contact Ultrasonic Level Sensing technologies. The sensor
uses advanced technology to measure velocity with a laser beam
at single or multiple points below the surface of the wastewater
stream.
0612
Laser Flow Module
OR
AC Powered Sites Battery Powered Sites
Area Velocity
Level
2160 Module
Signature Meter
The LaserFlow sensor is ideal for a
broad range of wastewater
monitoring applications. It is
compatible with both the Teledyne
Isco Signature® Flow Meter and
the Teledyne Isco 2160 LaserFlow
Module, depending on the type of
installation.
To operate with the LaserFlow, the
Signature requires software
version 1.21 or later.

TIENet™ 360 LaserFlow™ Velocity Sensor
Section 1 Introduction
1-2
Figure 1-1 Basic LaserFlow system (showing both AC and battery-powered options)
1.2 Design The LaserFlow sensor consists of a housing with two mea-
surement technologies working together.
1.2.1 Measurement
Te chn ol og y
The laser velocimeter has a laser diode that serves as both
transmitter and receiver of a laser light beam. The
Doppler-shifted signal frequency is used to determine flow
stream velocity.
The ultrasonic level transducer determines the stream’s level
by emitting an ultrasonic pulse and measuring the time it takes
for the echo to return from the stream’s surface. The transducer
is both pulse transmitter and echo receiver.
1.2.2 Cabling The LaserFlow sensor is available with a 10m or 23m standard
length cable. Custom cable length is also available. For greater
distances, external connection via conduit, and connection of
additional TIENet™ devices, the TIENet™ Expansion Box is
available.
Distance can be increased by installing the TIENet™ Expansion
Box closer to the installed sensor and adding a custom-length
TIENet™ cable between the box and the Signature® flow meter.
The maximum recommended distance between the LaserFlow
and the Signature® flow meter is 45.7 meters (150 feet). The
maximum recommended distance between the LaserFlow and
the 2160 LaserFlow module is 22.9 meters (75 feet).
1.3 Operation The laser velocimeter uses a laser beam to generate a source
frequency light, which is focused at points below the surface of
the flow stream. The light is scattered back to the laser. The
returned light is frequency shifted due to the Doppler effect and
the motion of the flow. The lens that focused the laser light below
the surface of the flow stream now focuses the returned light
back into the laser. The source light and shifted light frequencies
are mixed to determine a Doppler shift, which is then used to cal-
culate flow stream velocity.
The LaserFlow is able to move the laser beam transverse to the
flow in order to obtain readings at multiple points in the flow,
with automatic compensation to maintain precise focus at all
times.
The ultrasonic level transducer emits multiple ultrasonic
pulses per second. Between pulses, the transducer switches from
transmitter to receiver. When the transducer receives the echo
from the water’s surface, the sound energy is converted into an
electrical signal. The signal is then amplified and processed by
the flow meter into an “echo-received” signal. The time between
the transmitted pulse and the echo-received signal is propor-
tional to the distance between the transducer and the liquid
surface, which is then translated into a level reading. Because
the speed of the pulse through the air varies with temperature,
compensation is built-in. A temperature sensor inside the

TIENet™ 360 LaserFlow™ Velocity Sensor
Section 1 Introduction
1-3
LaserFlow measures ambient temperature. The microprocessor
program automatically compensates for speed-of-sound changes
caused by air-temperature changes.
The LaserFlow operates with no deadband from the mea-
surement point for both level and velocity measurement.
1.4 Optional Submerged
Functionality
During submerged conditions, flow measurement continues
without interruption with the optional TIENet™ 350 AV sensor,
which combines continuous wave Doppler to measure area
velocity with a differential pressure transducer to measure level.
The 350 AV sensor is factory-installed on the bottom of the
LaserFlow sensor, and is also available as a kit for installation in
an existing system. Refer to Section 5.7 for kit installation
instructions.
Figure 1-2 Optional submerged functionality
Level
(pressure transducer)
Area Velocity
(CWD)

TIENet™ 360 LaserFlow™ Velocity Sensor
Section 1 Introduction
1-4
1.5 Technical
Specifications
Table 1-1 provides technical specifications for the LaserFlow.
Table 1-1 LaserFlow Sensor Specificationsa
Size (H x W x D) 38.01 x 26.21 x 56.7 cm 14.96 x 10.3 x 22.32 in
Cable Lengthsb10 or 23 m 32.8 or 75.5 ft
The maximum recommended distance between the Laser-
Flow and the Signature® flow meter is 45.7 meters (150
feet). The maximum recommended distance between the
LaserFlow and the 2160 LaserFlow module is 23 meters
(75.5 feet).
Weight (LaserFlow w/ built-in sensor, 10 m cable)
Sensor w/o built-in sensor, 10 m cable
LaserFlow w/ built-in sensor, 350 AV Sensor
8.7 kg 19.2 lbs
8.25 kg 18.2 lbs
9 kg 20 lbs
Materials Conductive Carbon Filled ABS Plastic, SST, Conductive
Kynar®c, Anodized Aluminum, UV Rated PVC
Enclosure (self-certified) IP68 (Submerged @ 2 m depth for 72 hrs)
Certifications CE EN61326; FDA CDRH 21CFR1040; IEC 60825-1
Laser Class Class 3R
Laser Wavelength 655 nm
Laser Exit Angle 45° ±3°
Power (from connected flow meter/module) Input voltage range: 8 to 26VDC; 12VDC Nominal
Electrical Connections Signature Flow Meter Screw Terminal
2160 LaserFlow Module TIENet™Plug
Temperature Range Operating: -20 to 60 °C -4 to 140 °F
Storage: -40 to 60 °C -40 to 140 °F
Flow Accuracy ±4% of Reading (Typical under normal flow conditions)
Velocity Measurement
Technology Non-Contact, Subsurface Laser Doppler Velocity
(U.S. patent 8,339,584 and patents pending)
Measurement Range -4.6 to + 4.6 m/sec -15 to +15 ft/sec
Number of velocity readings per measurement 1 to 15 (selectable measurement points)
Maximum distance: liquid surface to bottom of sensor 3m (10 ft)
Minimum depth for velocity measurement 0.01m (0.5 in)
Direction Selectable Bi-Directional Measurementd
Minimum Velocity 0.15 m/s 0.5 ft/s
Deadband Zero deadband from bottom of LaserFlow sensord
Accuracy ±0.5% of reading ±0.03 m/s (0.1 ft/s)

TIENet™ 360 LaserFlow™ Velocity Sensor
Section 1 Introduction
1-5
Level Measurement
Technology Non-Contact Ultrasonic Signal
Measurement Range 0 to 3 m (0 to 10 ft) from measurement point
Accuracy @ 22 °C ±0.006m (0.02 ft) at < 1ft level change;
±0.012m (0.04 ft) at > 1ft level change
Ultrasonic Temperature Coefficient ± 0.0002 x D (m) per °C (± 0.00011 x D (ft) per °F)
(Where D is the distance from transducer to liquid surface)
Ultrasonic Beam Angle 10° (5° from center line)
Ultrasonic Signal 50KHz
Deadband Zero deadband from bottom of LaserFlow sensore
a. All specifications are subject to change without notice.
b. Custom cable lengths also available.
c. Kynar® is a registered trademark of Arkema, Inc.
d. Turbidity > 20 NTU. Distance < 48 inches.
e. Deadband for 310 remote ultrasonic sensor varies due to the type of mounting hardware.
Table 1-1 LaserFlow Sensor Specificationsa (Continued)

TIENet™ 360 LaserFlow™ Velocity Sensor
Section 1 Introduction
1-6
1.6 Accessories Accessories used in sensor installation are briefly described
below. Refer to the next section for ordering information.
1.6.1 Ordering Information Options and accessories can be purchased by contacting Teledyne
Isco’s Customer Service Department.
Teledyne Isco
Customer Service Dept.
P.O. Box 82531
Lincoln, NE 68501 USA
Phone: 800 228-4373
402 464-0231
FAX: 402 465-3022
E-mail: IscoInf[email protected]
TIENet™ Expansion Box......................................................................................................... 60-4307-023
Cord grip fitting, 3/4" NPT, for TIENet™ cable ....................................................................... 209-0073-12
Integrated 350 AV Kit for surcharge functionality................................................................. 60-4354-017
Signature Flow Meter Exterior desiccator
(required for use with 330 and 350 TIENet™devices)....................................................... 60-4354-019
Silica gel desiccant, 1.5-lb container ....................................................................................... 099-0011-03
Note
Guidelines for Area Velocity reference line support and net-
working with the expansion box can be downloaded in the
Teledyne Isco application note Signature Flow Meter Expan-
sion Box, available on the company website. Contact the fac-
tory for more information.
LaserFlow Instruction Manual................................................................................................ 69-4363-043
Signature Flow Meter Instruction Manual............................................................................. 69-4303-070
Signature Bubbler Flow Meter Instruction Manual .............................................................. 69-4333-004
Permanent Wall Mounting Hardware Kit............................................................................... 60-4364-003
Refer to Section 2.6.1 Wall Mount Installation for additional information.
Temp Mount Cargo Bar, 48-55" Variable Range..................................................................... 60-4364-032
Temp Mount Cargo Bar, 54-66" Variable Range..................................................................... 60-4364-034
Temp Mount Cargo Bar, 63-84" Variable Range..................................................................... 60-4364-035
Temp Mount Cargo Bar, 83-114" Variable Range................................................................... 60-4364-036
Temp Mount Hardware Assembly ........................................................................................... 60-4364-038
Temp Mount Kit, 48-55" Variable Range ................................................................................ 62-4364-032
Temp Mount Kit, 54-66" Variable Range ................................................................................ 62-4364-034
Temp Mount Kit, 63-84" Variable Range ................................................................................ 62-4364-035
Temp Mount Kit, 83-114" Variable Range .............................................................................. 62-4364-036
Temp Mount Elbow Pipe for Offset ......................................................................................... 60-4368-016
Refer to Section 2.7.1 Temporary Mounting Installation for additional information.
Horizontal Sensor Mount Bracket........................................................................................... 60-4368-020
Sensor retrieval arm for sensor retrieval, replacement, and locking
(Maximum extension of 23 feet) ....................................................................................... 60-4364-033
Other manuals for TIENet 360 LaserFlow Ex
1
Table of contents
Other Teledyne Accessories manuals

Teledyne
Teledyne 452 User manual

Teledyne
Teledyne PP007 User manual

Teledyne
Teledyne TIENet 310 Ex Operating instructions

Teledyne
Teledyne Citadel CTD-ES User manual

Teledyne
Teledyne BOA Spot ID Series User manual

Teledyne
Teledyne Everywhereyoulook CCD261-04 User manual

Teledyne
Teledyne TIENet 350 Operating instructions

Teledyne
Teledyne UFO-130 User manual

Teledyne
Teledyne 2150 Operating instructions

Teledyne
Teledyne HVD3000 User manual