Quick-Start Operating Guide
High- Ionizing Modules for Terra Desiccators
© Copyright 2010 Terra Universal Inc. All rights reserved. • Revised Sept. 2010 • Document No. 1800-39v2.0tm
Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020 2
1.0 Introduction
This manual provides information
on installing and operating your
Terra Universal Ionizing Module.
By studying this document
carefully, you can be assured of a
long, efficient service life from
your system.
2.0 Description
Terra’s Desiccator Ionizing Module incorporates pairs of ion emitters in each desiccator chamber that generate a
balanced stream of positive and negative ions to safely neutralize static charges on surfaces sensitive to electro-static
discharge. This module is designed for installation in a pressurized (purged) cabinet. The purge gas distributes the
positive and negative up to 48" from the emitter tips (see “Specifications” for static decay rates).
Because the Ionizing Module operates at low voltage (24 VAC), it is safe to operate and generates minimal RFI/EMI.
The stainless steel case contains the low-voltage power supply and titanium emitters, which can be replaced in a
matter of minutes.
Note: Refer to QuickStart document No. 1800-45 for ionizing equipment used for hoods, modular cleanrooms, and
glove boxes.
3.0 Installation
Carefully unpack your desiccator cabinet (refer to separate QuickStart for desiccator set-up). Report any shipping
damage immediately to the freight company.
The Ionization Module is installed in the desiccator plenum chamber before shipping. Use the low-voltage connector
and power cable included to connect the Module to the power transformer. Then connect the transformer to the
appropriate power supply (120VAC or 240VAC). A single transformer provides ample power for up to 20 emitter pairs.
Proprietary Notice
This manual pertains to proprietary devices manufactured by Terra
Universal, Inc. Neither this document nor any portion of it may be
reproduced in any way without prior written permission from Terra
Universal.
Terra Universal makes no warranties applying to information contained
in this manual or its suitability for any implied or inferred purpose.
Terra Universal shall not be held liable for any errors this manual
contains or for any damages that result from its use.
Safety Notice
CAUTION
Cautions are used when failure to observe instructions
could result in significant damage to equipment.
A thorough familiarity with all operating guidelines is essential to safe
operation of the product. Failure to observe safety precautions could
result in poor performance, damage to the system or other property, or
serious bodily injury or death.
The following symbols are intended to call your attention to two levels
of hazard involved in operation:
WARNING
Warnings are used when failure to observe instructions
or precautions could result in injury or death.
The information presented here is subject to change without notice.
Plenum holes
for gas
distribution

Quick-Start Operating Guide
High- Ionizing Modules for Terra Desiccators
© Copyright 2010 Terra Universal Inc. All rights reserved. • Revised Sept. 2010 • Document No. 1800-39v2.0tm
Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020 3
Once connected to power, the Module immediately emits balanced streams of positive and negative ions through
each emitter pair. For effective ion distribution, make sure that the bar is installed in a gas flow of at least 50
feet/minute.
4.0 Maintenance
The Ionizing Module requires little maintenance beyond occasional cleaning. Frequency of cleaning depends on the
cleanliness of air in the immediate vicinity of the Module.
If deposits are observed on the emitter tips,
1. Disconnect from system power.
2. Use a cleanroom swab moistened with a cleaning solution of 50% isopropyl alcohol and 50% de-ionized water to
remove deposits from the emitter points.
3. If deposits cannot be removed, or if the tips appear worn or bent, the Ionizing Module may not discharge
surfaces with optimal efficiency. In this case, replace the tips. Use a pencil eraser or needle-nosed pliers to
remove the old tip. Insert the new tip and gently press until it seats against the connector; no twisting is
necessary.
Cat. #
Replacement Titanium Emitters 2005-45
The emitter housing may be cleaned using alcohol or other mild detergent and de-ionized water.
5.0 Specifications
Emitter Points: Replaceable machined titanium, 3-year estimated life (field replaceable)
Input Power: 24 VAC – order transformer separately: No. 2005-09A (110VAC) or No. 2005-09A-
220 (220VAC)*
Emitter output: 2,000 V (one each positive and negative); ± 50 V charge balance
Power Indicator: Red LED
Ion Emission: Steady-state DC, 104mA current draw (44" model)
Air flow: 50 fpm (.030 m/s) minimum required at bar for proper operation
Static Decay Time: Discharge times directly under the IonBar™ average 28 seconds (from ± 1,000
volts to 100 volts), measured in a 4-foot laminar flow hood with air speed of 90
feet/minute.
Power Transformer*
Input Power: 120 VAC, 50/60Hz or 220VAC, 50 Hz
Output Power: 24 VAC, 1.0 Amp – drives up to 20 emitter pairs (low-voltage telephone cable, not
furnished, is required to interconnect IonBars)
Dimensions: 2.75" x 3" x 1.75"
Weight: 1 lb.
•Power transformer included with Ionizing Fan Filter Unit and Glove Box Ionizing Module

Quick-Start Operating Guide
High- Ionizing Modules for Terra Desiccators
© Copyright 2010 Terra Universal Inc. All rights reserved. • Revised Sept. 2010 • Document No. 1800-39v2.0tm
Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020 4
6.0 Warranty
Products Manufactured by Terra: Terra Universal, Inc., warrants products that it manufactures to be free from
defects for a period of 12 months for parts and 90 days for labor, commencing from the date of shipment. Terra’s
sole responsibility is to repair or replace, at its option, any part of the product that proves defective or malfunctioning
during this time limit. In some cases, components incorporated in Terra Universal products are covered by additional
warranties from component manufacturers; obtain specific information from Terra sales representatives. This
warranty is void if the equipment is abused or modified by the customer, is operated outside Terra’s operating
instructions or specifications, or is used in any application other than that for which it is specified. This warranty does
not include routine maintenance or service procedures, breakage of quartz baths after 60 days, shipping damage,
nor damage from misuse, intentional or unintentional abuse, neglect, natural disasters, or acts of God.
Products Manufactured by Others: Terra Universal, Inc., warrants that, to the best of its ability, Terra’s
representations of products that are manufactured by others reflect the manufacturer’s representations, subject to
change without notice. Sole warranty for these products is the original manufacturer’s warranty that is passed
forward to the purchaser and constitutes the customer’s sole remedy for these products. Detailed warranties for
distributed products are available through Terra sales representatives.
Freight Shortage or Damage: Upon receipt of any equipment from Terra Universal, Inc., customer shall
immediately unpack and inspect for damage or shortage. The customer shall not accept a damaged package or a
short shipment until the carrier makes a "damage or shortage" notation on both the carrier's and customer's copy of
the freight bill or delivery receipt. Service title passes when the shipment is loaded, so customer is responsible for
filing and collecting a freight claim. Any replacement products must be ordered and paid for separately. For Terra's
"Policy and Procedures for Returning Goods," see Terra's Internet site: www.TerraUniversal.com.
Generally, customers can improve the chance of collecting on a freight claim by following these procedures: 1)
formally requesting that the carrier inspect the shipment immediately upon suspecting damage or shortage to verify
condition; 2) notifying the carrier upon discovery of concealed damage and requesting an inspection within 15 days
of receipt, both in person or phone and following up via mail; 3) keeping the shipment as intact as possible, including
retaining original packaging materials and keeping the product as close to the original receiving location as possible;
4) holding salvage for disposition by the carrier.
All Claims: Terra Universal expressly disclaims all other warranties, expressed or implied or implied by statute,
including the warranties of merchantability or fitness for intended use. Terra Universal is not responsible for
consequential or incidental damages arising out of the purchase or use of the products supplied by Terra Universal.
Terra Universal is not liable for damage to facilities, other equipment, products, property or personnel of others, or of
their agents, suppliers, or affiliated parties, which is caused or alleged to have been caused by products supplied by
Terra Universal. In any event or series of events, Terra Universal’s total liability for any and all damages whatsoever
is limited to the lesser of the actual damages or the original invoice cost of the items alleged to have caused the
damage. The customer’s sole and exclusive remedy for any cause of action whatsoever is repair or replacement of
the non-conforming products or refund of the actual purchase price, at the sole option of Terra Universal. All claims
must be made in writing within 90 days of the date the product was shipped. Any claims not made within this time
limit shall be deemed waived by the customer. Terra Universal is not responsible for any additional costs of repair
caused by poor packaging or in-shipment damage during return.
Warranty Returns: All warranty returns must be authorized in advance by Terra Universal and approved under an
RMA. Unless approved in advance for good reason, all returns must be in original condition, including all manuals,
and must be packaged in original packaging materials. All returned goods are to be shipped to Terra Universal,
freight prepaid at customer’s expense. See Terra’s “Policy and Procedure for Returned Goods.”
Thank you for ordering from
Terra Universal!