QuickStart: FLIR E30-E60, E30bx-E60bx
© 2011 The Snell Group / 322 N. Main St. / Suite 8 / Barre, VT 05641 / 800.636.9820 / www.thesnellgroup.com
Flir E30-E60, E30bx-E60bx Page 2
7. PAUSE/SAVE IMAGE: The [Trigger] is used to Pause (preview) or Save a live image.
NOTE: the trigger can be programmed to function in different ways. Check the following setting to make
sure that the trigger is set to Pause/Save.
Press the [Menu-Back] button to display the menu, use the [Navigation Pad] to select (Settings),
and then press the [OK] button. Make sure the trigger is set to Pause/Save.
You can now Pause and preview a live image by briefly pulling and releasing the trigger. (the image can be
readjusted while paused if you wish)
To Save the paused image pull and hold the trigger longer for longer than 1 second..
8. RECALL IMAGE: Press the [View Saved Images] button.
Use the [Navigation Pad] to highlight the desired image then press the [Auto/Manual-OK] button.
The recalled image can now be readjusted by pressing the [Auto/Manual-OK] button to show the menu.
9. PALETTE: Press [Menu-Back] then use the [navigation Pad] to highlight (Palette) then press
[Auto/manual-OK] button
Now use the [Navigation Pad] to select the desired Palette then press [Auto/manual-OK].
10. RANGE: In this series of cameras, it depends on the model if the camera has more than one temperature range.
Models E50 and E60 have two ranges but the camera “Auto Ranges” dependant on the object being viewed, no
operator selection is provided.
11. EMISSIVITY: Press the [Menu-Back] button now use the [navigation Pad] to select (Object Parameters)
then press [Auto/manual-OK] button..
Use the [Navigation Pad] to highlight Emissivity then press [Auto/Manual-OK].
Use the [Navigation Pad] to change the value then press [Auto/Manual-OK].
12. BACKGROUND: (Reflective temperature) Press the [Menu-Back] button, now use the [Navigation Pad] to
select (Object Parameters) then press [Auto/Manual-OK] button.
Use the [Navigation Pad] to highlight Reflective Temperature then press
[Auto/Manual-OK].
Use the [Navigation Pad] to change the value then press [Auto/Manual-
OK].
E30, E30bx, E40, E40bx
E50, E50bx
E60, E60bx
IFOV =2.72 mRad (Theoretical with 120x160 Detector and standard 25˚x19˚lens) Detect 1in. target @ 30ft.
Fi
1
IFOV =1.82 mRad (Theoretical with 240x180 Detector and standard 25˚x19˚lens) Detect 1in. target @ 45ft.
IFOV =1.36 mRad (Theoretical with 320x240 Detector and standard 25˚x19˚lens) Detect 1in. target @ 61ft.