2. Press SELECT to choose STATISTCAL VIEW
3. Use the UP and DOWN buttons to scroll through the AVERAGE, MINIMUM, MAXIMUM, NUMBER OF
DATA, and SDEV (Standard Deviation) values for the stored readings.
4. ‘NO DATA’ will display if no readings are available for the meter to analyze. Stored readings will clear
when the meter powers down unless the GROUP feature is used (refer to GROUP function explanation
later in this section).
5. Press the BACK and then the ESC soft‐keys to return to the normal operating mode.
OPTIONS Menu
1. Press the MENU button to access the programming menu
2. Use the DOWN ARROW button to scroll down to OPTIONS
3. Press SELECT to choose OPTIONS
4. Use the UP and DOWN buttons to scroll to the MEASURE MODE, WORKING MODE, PROBE USED, UNIT
SETTINGS, BACKLIGHT, LCD STATISTICS, AND AUTO POWER OFF parameters. Use the SELECT soft‐key to
select the desired parameter. Each parameter is detailed below:
Measure Modes
Select CONTINUOUS or SINGLE under MEASURE MODES in the OPTIONS Menu using the arrow buttons
and the SELECT soft‐key.
In the CONTINUOUS measurement mode the meter displays a running average of readings as they are
taken. Note that the audible measurement ‘beep’ is not active in this mode.
In the SINGLE measurement mode measurements are taken one at a time. Single mode measurement
readings are accompanied by an audible tone.
Working Modes
Select DIRECT or GROUP 1, 2, 3, or 4 under WORKING MODES in the OPTIONS Menu using the arrow
buttons and the SELECT soft‐key.
In DIRECT mode, individual readings are logged to memory. When power is switched off or if the meter
is switched to GROUP mode, all DIRECT readings will be cleared. However, the statistical analysis data
will remain. The statistical analysis utility can evaluate up to 80 readings. When the memory is filled,
new readings will replace old readings. Lastly, this mode has its own calibration and alarm limit values.
In GROUP mode each group memory can store a maximum of 80 readings and 5 statistical values.
Calibration and alarm limit values can be individually set and stored for each group. When the memory
is filled, measurements will continue to be taken but readings will no longer log (previously logged
readings are not affected); in addition, statistical data will no longer update. If desired, the group data,
statistical values, calibration data, and alarm limit values can be deleted using Delete under the
programming menu.
Used Probe
Select AUTO, Fe, or No Fe under USED PROBE in the OPTIONS Menu using the arrow buttons and the
SELECT soft‐key.
In the AUTO mode, the meter automatically activates the probe measurement method (ferrous or non‐
ferrous) based on the metal substrate that is being measured. When the probe is placed on a magnetic
substrate it will work in the magnetic induction mode. If the probe is placed on a non‐ferrous metal it
will work in the eddy current mode.
In the Ferrous (Fe) Mode the Magnetic induction measurement mode is activated.
In the Non‐Ferrous (No Fe) Mode the eddy current measurement mode is activated.
Units of measure selections
Select mm, µm, or mils under UNIT SETTING in the OPTIONS Menu using the arrow buttons and the
SELECT soft‐key (mm = millimeters; µm = micrometers; mils = mm*2.54/100)