
Peak Hold Function (AC/DC Voltage measurements only)
The PEAK HOLD function is used with the MAX/MIN function to capture
voltage max and min peaks in the range of 10 to 100mS (milliseconds).
The voltage range is 0.5V to 600V AC or DC.
1. Press the MAX/MIN button. MIN will appear on the LCD display.
2. Press the PEAK button. MIN/PEAK will appear on the LCD display.
The meter will now capture minimum readings.
3. Press the MAX/MIN button again to display the MAX/PEAK icon.
The meter will now capture maximum readings.
4. Press and hold the MAX/MIN button for at least 2 seconds to exit the
PEAK hold mode.
Relative Mode (AC/DC Voltage measurements only)
The relative mode displays the difference between the measured
value and a stored reference value.
1. Press REL to store the reading presently indicated on the display
as the reference value; the "REL" indicator will appear.
2. Subsequent readings will represent the difference between the
stored reference value and the measured value.
3. Press the REL button again to return to normal operation
MAX/MIN Function (AC/DC Voltage measurements only)
1. Press the MAX/MIN button to enter the MAX/MIN mode. The ‘MIN’ icon
will appear on the LCD display and the meter will display and hold only
the minimum value. The value will be held until a new minimum value
is measured.
2. Press the MAX/MIN button again and ‘MAX’ will appear on the LCD
display. The meter will display and hold only the maximum value. The
value will be held until a new maximum value is measured.
3. Press the MAX/MIN button a 3rd time, MAX/MIN will appear on the
LCD display. In this mode the meter will display the real time
readings but will continue to storemaximum and minimum values for
later recall. Use the MAX/MIN button to check the MIN and MAX
values.