5
6
STEP 5 – TAKE MEASUREMENTS
Always begin with handle in SAFE position. Ensure
that the handle is in the SAFE position by firmly
tapping down on the handle without engaging the
trigger.
We recommend 2 minutes for measurement time and
1 minute for background time. (Setup menu)
The gauge can be set to receive moisture values from
an outside device or by keypad entry. You can also
choose to perform the moisture measurement before
or after the density measurement. Press Setup, choose
4. Moisture Input, then choose your method /
order desired.
Prepare the test site as described in step 3. If using
the moisture probe, drive the drill rod to at least 15
cm (6 in).
PERFORMING DENSITY MEASUREMENTS
1
1.
.
Place the gauge on the test site with the
handle in the Safe position and press
〈ENTER/START〉. Follow the prompts on
the screen to either lower the handle into
the prepared hole to the desired measure
depth or to wait for the 15-second
initialization phase to complete (this will
occur approx. every 10 min. or as
programmed). Then lower handle and press
〈ENTER/START〉.
2
2.
.
After count time, (when prompted) select
the Active background count or choose to
take a New background count. If taking a
new background count, lower the source
rod to background position and press
〈ENTER/START〉. A new Background count
should be performed at a new jobsite and
when the soil material properties change.
3
3.
.
After the measurement, raise the handle.
The gauge displays the measurement
results. To store the reading, press 〈STORE〉.
After taking readings, lift the gauge from the test site
by the source rod handle. This returns the source rod
to the SAFE position. When not taking readings,
always keep the source rod in the SAFE position.
PERFORMING MOISTURE MEASUREMENTS
1.
Press <Enter/Start> to begin.
2.
When prompted by the 4590 gauge, insert
moisture probe into the hole until the base
plate contacts soil surface.The probe should
fit tightly in the hole, some resistance should
be felt when lowering it.
3.
Power the probe ON (green light illuminates).
The blue light will remain solid when the
BlueTooth connection is established.
4.
Press Enter on gauge keypad to perform
moisture reading. Wait a few seconds then
press Enter on the keypad again to accept the
moisture reading. See the Guide for the Model
6760 Moisture Probe for more details on
operation.
5.
Carefully remove the probe and set aside.
USING THE OFFSET FUNCTION
The gauge readings can be adjusted using an offset.
The gauge applies the offset to measurements until
the offset is disabled or the gauge is turned off. The
gauge provides both density and trench offsets.
Press 〈OFFSET〉to access the Offset menu.
MORE INFORMATION
The user manual provides more information about the
gauge’s functions and use.
Model 4590 Quick Reference Guide PN 128500
Edition 1.2, May 2021
Model 4590
Surface Density Gauge
E
Gauge
QUICK
REFERENCE
GUIDE
Troxler Electronic Laboratories, Inc.
3008 E. Cornwallis Rd. •P.O. Box 12057
Research Triangle Park, NC 27709
Phone: 1.877.TROXLER
Outside the USA: +1.919.549.8661
Fax: +1.919.549.0761
www.troxlerlabs.com
2
3
4
Note- Be sure no other nuclear sources (gauges)
are within 10 m (30 ft.) of EGauge while in use.
PR
:
#.#
pcf
1:
0.0
2:
0.0
3:
0.0
4:
0.0
5:
New
6:
Disable
CAUTION
Safety glasses must be worn during this
procedure.
STEP 1 – START THE GAUGE
To turn on the gauge, press the power switch. After
briefly displaying the model number, software version,
and serial number, the gauge performs a self-test,
followed by a display test. The gauge then enters a
300-sec. warm-up then displays the Ready screen.
STEP 2 – SET UP THE GAUGE
SET UNITS
The gauge can display measurement results in either
U.S. units (pcf) or metric (SI) units (kg/m3or g/cm3). To
select the units, press 〈SETUP〉to display the Setup
menu. Press 〈5〉to display Options. Then press 〈3〉to
display the Units menu. Select the new units using the
corresponding number key. Then press 〈ESC〉to return
to the Ready screen.
TARGET VALUES
To select or change a Proctor value, press 〈TARGET〉.
The gauge displays the Target Value menu.
To enter a new target value, press 〈5〉. At the prompt,
use the number keys to enter the desired target value.
To select a stored target value, press the number key
that matches the displayed target value.
After entering a new value press 〈ENTER/START〉. The
gauge displays the value entered and asks if you want
to store the value. To store the value, press 〈YES〉and
select one of the four locations. Storing a new value in
an occupied location will erase the old value. Press
〈NO〉if storing the value is not desired. This will enable
the value and return to the Ready screen. This value
will remain enabled until changed or disabled.
STEP 3 – PREPARE THE TEST SITE
To ensure measurement accuracy, properly prepare
the test site before taking gauge measurements or
standard counts.
1
1.
.
Locate a smooth site on the compacted soil
free from any large holes, cracks, or debris.
Place the scraper plate on the surface and
press down slightly or scrape lightly if needed
to smooth the surface.
2
2.
.
As shown in Chapter 4 of the user manual, put
the drill rod through the extraction tool and
then through the guide tube on the scraper
plate.
3
3.
.
Step on the scraper plate and hammer the
drill rod to at least 50 mm (2 in.) deeper than
the desired test depth. The drill rod
increments include the additional depth. If
only a standard count is to be performed,
make a 50 mm (2 in.) deep hole (minimum).
4
4.
.
Remove the drill rod by pulling straight up
on the extraction tool. Do not loosen the drill
rod by moving it from side to side. Trace the
plate to mark the test area.
STEP 4 – TAKE A STANDARD COUNT
To compensate for the source decay, adjust for
background influences and to check proper operation
of the gauge, take a standard count each day that the
gauge is used. It is recommended that the standard
count be performed at the first test site at a particular
job site.
Keep any other nuclear gauge or radioactive source at
least 10 m (30 ft.) from test site.
1
1.
.
Prepare the test site as described in Step 3 of
this guide and place the gauge on the marked
area. The prepared hole is required to perform
the second step of the standard count (DSB
count)
2
2.
.
Press 〈STD〉. The gauge displays the last standard
count. To take a new standard count, press 〈YES〉.
With the gauge in the standard count (safe)
position, press 〈ENTER/START〉to begin the
standard count.
3
3.
.
When prompted, lower the gauge handle to
the background position (slightly below the
safe position) to begin the second phase of the
standard count. Press 〈ENTER/START〉as
prompted.
Troxler recommends keeping a daily log of the
standard count results. An example log is provided in
the appendix of the user manual.
Note- This gauge does not require a reference
block to take the standard count. Ensure that the
source rod is in the SAFE position by firmly
tapping down on the handle of the source rod.

Other manuals for EGauge 4590

1

Other Troxler Measuring Instrument manuals

Troxler Enhanced RoadReader Plus 3451 User manual

Troxler

Troxler Enhanced RoadReader Plus 3451 User manual

Troxler 3440-L User manual

Troxler

Troxler 3440-L User manual

Troxler RoadReader 3430 User manual

Troxler

Troxler RoadReader 3430 User manual

Troxler RoadReader Plus 3450 User manual

Troxler

Troxler RoadReader Plus 3450 User manual

Troxler 3216 User manual

Troxler

Troxler 3216 User manual

Troxler EGauge 4590 User manual

Troxler

Troxler EGauge 4590 User manual

Troxler 4640-B Guide

Troxler

Troxler 4640-B Guide

Troxler EGauge Combo 4540 User manual

Troxler

Troxler EGauge Combo 4540 User manual

Troxler 3400-B Series User manual

Troxler

Troxler 3400-B Series User manual

Troxler EGauge Combo 4540 User manual

Troxler

Troxler EGauge Combo 4540 User manual

Troxler RoadReader 3430 User manual

Troxler

Troxler RoadReader 3430 User manual

Troxler TroxAlert User manual

Troxler

Troxler TroxAlert User manual

Troxler 3430 Plus User manual

Troxler

Troxler 3430 Plus User manual

Troxler RoadReader 3430 Guide

Troxler

Troxler RoadReader 3430 Guide

Troxler 3241 Series User manual

Troxler

Troxler 3241 Series User manual

Troxler PaveTracker Plus User manual

Troxler

Troxler PaveTracker Plus User manual

Popular Measuring Instrument manuals by other brands

LaserLiner LiveTester Plus operating instructions

LaserLiner

LaserLiner LiveTester Plus operating instructions

BEKA BA334E Application guide

BEKA

BEKA BA334E Application guide

X-Rite Ci51 user guide

X-Rite

X-Rite Ci51 user guide

Public Safety Equipment KUSTOM SIGNALS ProLaser 4 Operator's manual

Public Safety Equipment

Public Safety Equipment KUSTOM SIGNALS ProLaser 4 Operator's manual

schmersal SRB 200X2 operating instructions

schmersal

schmersal SRB 200X2 operating instructions

Fluke PM6681 Service manual

Fluke

Fluke PM6681 Service manual

PCE Americas PCE-318 manual

PCE Americas

PCE Americas PCE-318 manual

General DCS1700 user manual

General

General DCS1700 user manual

Bushnell GOLF Pro X2 user manual

Bushnell GOLF

Bushnell GOLF Pro X2 user manual

Goodwe HK3000 Quick installation instructions

Goodwe

Goodwe HK3000 Quick installation instructions

PQ Plus UMD 710A Quick start manual

PQ Plus

PQ Plus UMD 710A Quick start manual

HBM T10F KF1 operating manual

HBM

HBM T10F KF1 operating manual

BSS Audio Case Economic Operation manual

BSS Audio

BSS Audio Case Economic Operation manual

Gossen MetraWatt SECULIFE ES TECH operating instructions

Gossen MetraWatt

Gossen MetraWatt SECULIFE ES TECH operating instructions

Dinel ULM-53 series instruction manual

Dinel

Dinel ULM-53 series instruction manual

PIE PIECAL 541 operating instructions

PIE

PIE PIECAL 541 operating instructions

Oval EL4131 instruction manual

Oval

Oval EL4131 instruction manual

Phynix SURFIX PRO X instruction manual

Phynix

Phynix SURFIX PRO X instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.