U.S. Solid USS-DBS3-1 User manual

U.S. Solid®•USS-DBS Models • All Rights Reserved • Page 1
U.S. Solid® • 1200 E.55th Unit C, Cleveland, OH, 44103 • www.ussolid.com
USS-DBS
MODELS
OPERATING
MANUAL
ANALYTICAL BALANCES
This manual should be made available to all users of this
equipment. For best results, and for maximum durability of the
equipment, carefully read and follow all instructions.
ANALYTICAL BALANCE

U.S. Solid®•USS-DBS Models • All Rights Reserved • Page 2
U.S. Solid® • 1200 E.55th Unit C, Cleveland, OH, 44103 • www.ussolid.com
I. Getting Started
II. Specications
III. Panel Description
IV. Installation
V. Using the Balance
VI. Maintenance and Cleaning
Table of Contents

U.S. Solid®•USS-DBS Models • All Rights Reserved • Page 3
U.S. Solid® • 1200 E.55th Unit C, Cleveland, OH, 44103 • www.ussolid.com
Thank you for choosing U.S. Solid’s Analytical Balance. We hope you enjoy your pur-
chase. Please maintain the equipment according to all instructions within the manual.
This product was developed according to CE marking regulations while emphasizing
aesthetics and safety for the user. It has ISO certication, specically ISO 9001 The high
quality of the product, coupled with proper user care, should allow you to enjoy the
product for years.
Improper use of this equipment can cause accidents, including but not limited to:
electric discharges, circuit breaker malfunctions, re, etc. Please read the mainte-
nance section, where this is explained more fully.
FOR BEST RESULTS AND LONGEVITY IN PRODUCT
EFFICACY, READ THROUGH THIS MANUAL THOROUGHLY
BEFORE USING THE PRODUCT
Please bear in mind the following:
• This manual should be kept with the analytical balance, so it is
available to all users
• Be careful when handling the balance. Avoid any sudden
movements when in its vicinity. Be sure to not drop the balance,
nor drop any heavy/sharp objects on it. Keep liquids away from the
balance to avoid spills.
• Never take apart the balance to try and x yourself. This could ruin
the entire balance, and voids the product warranty. There is also risk
of injury when disassembling the balance.
• To prevent re and electrical issues, avoid dusty or overly dry
environments. In case of electrical issue or re, unplug the balance
immediately.
• During setup, installation, and use, contact your wholesaler with any
questions that arise.
• This equipment is protected under the Warranties and Consumer
Good Regulation (10/2003)
• Overhaul is not covered by the equipment warranty
• Changes made to the product by non-certied individuals will result
in a loss of the product warranty
• Accessories, including their loss, are not covered by the product
warranty, nor does the warrant cover piece’s deterioration over the
course of time.
• Keep your invoice, either for having the right to a warranty, or for
asking for warranty coverage. Manufacturer reserves the right to
modify or improve the manual or equipment.
II. Getting Started

U.S. Solid®•USS-DBS Models • All Rights Reserved • Page 4
U.S. Solid® • 1200 E.55th Unit C, Cleveland, OH, 44103 • www.ussolid.com
FEATURES
• Straing gauge technology
• Stainless steel pan
• Backlit LCD display
• Plastic draft shield
• Height-adjustable feet
• Taring
• Four dierent units of measure: g, lb, ct, oz
• Piece counting function
ALWAYS MAKE SURE YOU CALIBRATE YOUR SCALE BEFORE USING, AND NEVER
TOUCH THE INCLUDED WEIGHT WITH YOUR SKIN!

U.S. Solid®•USS-DBS Models • All Rights Reserved • Page 5
U.S. Solid® • 1200 E.55th Unit C, Cleveland, OH, 44103 • www.ussolid.com
II. Specications
Analytical balances oer fast and accurate mea-
surements of mass, using strain gauge technology for
fast, ecient measurement.
Technical Parameters:
MODEL: USS-DBS3-1 USS-DBS3-2 USS-DBS3-3 USS-DBS3-5
Max Capacity
in grams 100 200 300 500
Accuracy Class Class II
Min. Capacity 10 mg
Division 0.001g
Linearity ±0.002 g
Repeatability ±0.002 g
Operating
Temperature
5° C- 35° C, with uctuations of <1°C/Hour
41° F- 95° F with uctuations of <1.8° F/ Hour
Relative Hu-
midity 50%-85%
Power AC 110V-240V, DC 9V

U.S. Solid®•USS-DBS Models • All Rights Reserved • Page 6
U.S. Solid® • 1200 E.55th Unit C, Cleveland, OH, 44103 • www.ussolid.com
III. Panel Description
Panel Explanations
1. ON/OFF: Turns device on and o
2. UNIT: Select between ct, oz, lb, g
3. PCS: Piece counting mode
4. CAL: Calibration key
5. TARE/ZERO: Tare the balance

U.S. Solid®•USS-DBS Models • All Rights Reserved • Page 7
U.S. Solid® • 1200 E.55th Unit C, Cleveland, OH, 44103 • www.ussolid.com
IV. Installation
Choosing the installation location
When analytical balances are used in routine lab or industrial measure-
ment environments, the weighing speed will be much quicker, and the
results more accurate. The location should meet the following conditions:
1. Work room should be clean and dry
2. Balance should be placed on a solid, stable, plain at surface
3. Avoid locations in which the balance could be exposed to any of
the following:
a. Airow from air conditioners, fans, doors, or windows
b. Vibrations from surrounding or nearby equipment
c. Direct sunlight or radiation
4. Do not use the balance anywhere exposed to explosive, combus-
tible, or corrosive gases.
5. Do not use the balance in areas with high humidity or high levels
of dust
6. When moving from a cooler to a warmer place, the accuracy
and reliability of the scale will be inuenced by the moisture con-
densation inside the balance. In order to eliminate this inuence,
it is best to leave the scale unplugged in its new location for at
least two hours before next use.
7. Avoid extreme temperatures or exposure to direct sunlight or air
conditioners. Ideal working temperatures are between 5° C and
35° C (41°F and 95° F) with tempature uctuations of less than 1° C
(1.8° F) per hour.
8. Keep the balance clean.
9. Do not leave material on the balance when not in use.
Unpacking and inspection
Check that all of the items indicated below are included in the package
and that nothing has been damaged:
• Balance
• Pan
• Power Cable
• External Calibration weight
• Draft Shield Pieces
In case the balance or any component is damaged during transport, tell
your transportation agent immediately in order to process claims within
the proper time frame.

U.S. Solid®•USS-DBS Models • All Rights Reserved • Page 8
U.S. Solid® • 1200 E.55th Unit C, Cleveland, OH, 44103 • www.ussolid.com
V. Using the Balance
Level adjustment
Once the balance is placed at its location, check the bubble level; if bubble is not
well-centered, turn the adjustable feet so the bubble moves to the center and the
balance is properly leveled.
Start
1. Connect the balance to the power supply and press ON key on the panel.
Wait 30 minutes after powering on for best performance.
2. After warming up, calibrate the balance.
3. After calibration, the balance will enter weighing mode
4. To turn the balance o, press the OFF key on the panel and the display will shut
o. If the balance will be unused for an extended period of time, disconnect it
from the power supply.
Calibration
1. Press CAL key; the value of the necessary weight will ash on display
2. Put the calibration weight on the pan. Wait until the value of the calibration
weight stops ashing and is xed on the display (e.g. 100.0000 g, 200.0000g,
etc., depending on the model)
3. Remove the weight from the pan
4. The balance will enter weighing mode and will display 0.000g or 0.0000g
Note: Each balance includes the necessary weight to complete external calibra-
tion
Tare function
1. Put the receptacle on the weighing pan; its weight will be displayed
2. Press TARE key and the reading on the display will automatically be set to
0.000g or 0.0000g; tare is complete
3. If the receptacle is removed from the pan, the corresponding weight, with
negative value will be displayed.
4. Press the TARE key again and the value on display will be set to zero again
Weighing unit selection
Press UNIT key to select the desired weighing unit from g, ct, oz, and lb
Piece counting function
1. Press the PCS button
2. Press the CAL button repeatedly to cycle through dierent counts. This function
can be used with counts of 10, 20, 50, 100, 150, 200, 250, or 500 pieces.
3. Put the corresponding number of pieces on the pan, press the PCS button to
conrm.
4. Remove the pieces from the pan and the balance is ready to be used in piece
counting mode.
5. To return to weighing mode, simply press the PCS button again.

U.S. Solid®•USS-DBS Models • All Rights Reserved • Page 9
U.S. Solid® • 1200 E.55th Unit C, Cleveland, OH, 44103 • www.ussolid.com
VI. Maintenance and Cleaning
To get the best results and for product longevity, be sure to follow the
suggestions below.
Note: All of the suggestions and guidelines mentioned below and through-
out this manual will only be eective with continuous and careful mainte-
nance of the equipment
• Follow all processes outlined in the manual
• Make the manual available to all users of the equipment
• Prevent the balance from any sudden movements or falls, as well as
from direct sunlight or air ow. The balance is a precision instrument
and must be handled carefully.
• Balance must be plugged into a grounded electrical outlet and the
socket should be easily accessible to unplug in case of emergency
• Never unplug the balance by pulling on the wire, do so from the base
• Never use the balance in a wedged-in location, such as a shelf
• Never use other objects such as pens, pencils, etc. to press the but-
tons on the control panel; use only your ngers.
• Never place anything on the pan heavier than the maximum capaci-
ty of the balance, or the sensor could be damaged
• Do not submerge the balance, nor spill any liquids on it
• If any liquid does come into contact with the electrical parts of the
balance, immediately disconnect it and send it in for service to be
checked and adjusted if necessary
• Always use only original components and supplies. Other devices and
parts may appear similar, but can damage the equipment.
Cleaning
• Never use scourers or substances that can grate to clean metallic
parts. This include, but is not limited to, stainless steel, aluminum, coat-
ings, etc.. These can damage the balance and lead to early erosion
of eectiveness of the balance.
• -Use a lint-free cloth, dampened with soapy water, that does not con-
tain any abrasive surfaces, nor corrosive materials
ATTENTION!! If equipment is not properly
cleaned and cared for, technical service
will be unable to x or repair issues.

U.S. Solid®•USS-DBS Models • All Rights Reserved • Page 10
U.S. Solid® • 1200 E.55th Unit C, Cleveland, OH, 44103 • www.ussolid.com
Troubleshooting Guide
Problem Cause Solution
Display not working No power supply Plug in the adapter
Display shows
“ “
Object weight
below minimum
threshold, taring
issues exist, or bal-
ance is not properly
calibrated.
Make sure item
being weighed is at
least 10mg. Make
sure receptacles
have been tared,
and the balance
has been properly
calibrated.
Display shows
“ “
Item being weighed
is over maximum
allowable weight.
Make sure balance
is properly cali-
brated, and make
sure item is below
maximum allowable
weight.
Display shows “Err-1” Power issues with
the balance.
Machine may have
a poor power con-
nection. Check the
power in the back,
and the outlet.
Leave machine o
for seveeral minutes,
unplug fully, then re-
connect and begin
again.
Display shows “Err-2” Unstable weighing
conditions.
Recalibrate ma-
chine, make sure
balance is level,
and make sure item
being weighed is
fully stable.

U.S. Solid®•USS-DBS Models • All Rights Reserved • Page 11
U.S. Solid® • 1200 E.55th Unit C, Cleveland, OH, 44103 • www.ussolid.com
Note: According to the applicable legislation regarding “Non-
automatic weighing instruments”in which balances are included,
by means of writ dating from the 22nd of October, 1994 (BOE
1/3rd/95), these balances must not be used for:
• Commercial transactions
• Calculating taxes, taris, rates, indemnities, and other
similar canons
• Judicial surveys
• Pharmaceutical medicine preparations, as well as
analysis made in medical or pharmaceutical laboratories
• Determining the price or total amount in sale price or in
pre-packaged preparations
Instructions on environmental protection:
At the end of its life cycle, please do not dispose of
this equipment by throwing it in the usual garbage. Instead, hand
it over at a collection point for the recycling of electrical and
electronic appliances. It does not contain dangerous nor toxic
products for humans, but inadequate disposal could still damage
the environment. The materials are recyclable as mentioned.
By recycling material or through other forms or repurposing old
appliances, you are making an important contribution to the pro-
tection of our environment. Please inquire with your local com-
munity authorities for the proper disposal location.

U.S. Solid®•USS-DBS Models • All Rights Reserved • Page 12
U.S. Solid® • 1200 E.55th Unit C, Cleveland, OH, 44103 • www.ussolid.com
Other manuals for USS-DBS3-1
1
This manual suits for next models
3
Table of contents
Other U.S. Solid Scale manuals

U.S. Solid
U.S. Solid USS-DBS82 User manual

U.S. Solid
U.S. Solid USS-DBS3-1 User manual

U.S. Solid
U.S. Solid USS-DBS91 User manual

U.S. Solid
U.S. Solid USS-DBS49 User manual

U.S. Solid
U.S. Solid USS-DBS Series User manual

U.S. Solid
U.S. Solid USS-DBS81 User manual

U.S. Solid
U.S. Solid USS-DBS92 User manual

U.S. Solid
U.S. Solid 1004 User manual

U.S. Solid
U.S. Solid USS-DBS86 User manual

U.S. Solid
U.S. Solid USS-DBS93 User manual