ANALYTICAL BALANCE
USS-DBS
MODELS
OPERATING MANUAL
ANALYTICAL BALANCES

U.S. Solid®•USS-DBS Models • All Rights Reserved • Page 1
U.S. Solid® • 1200 E.55th Unit C, Cleveland, OH, 44103 • www.ussolid.com
U.S. Solid®•USS-DBS Models • All Rights Reserved • Page 10
U.S. Solid® • 1200 E.55th Unit C, Cleveland, OH, 44103 • www.ussolid.com
I. Getting Started
II. Specications
III. Panel Description
IV. Installation
V. Using the Balance
VI. Maintenance and Cleaning
Table of Contents Note: According to the applicable legislation regarding “Non-automatic
weighing instruments”in which balances are included, by means of writ
dating from the 22nd of October, 1994 (BOE 1/3rd/95), these balances must
not be used for:
• Commercial transactions
• Calculating taxes, taris, rates, indemnities, and other similar
canons
• Judicial surveys
• Pharmaceutical medicine preparations, as well as analysis made in
medical or pharmaceutical laboratories
• Determining the price or total amount in sale price or in pre-
Instructions on environmental protection:
At the end of its life cycle, please do not dispose of this equip-
ment by throwing it in the usual garbage. Instead, hand it over at a collec-
tion point for the recycling of electrical and electronic appliances. It does
not contain dangerous nor toxic products for humans, but inadequate
disposal could still damage the environment. The materials are recyclable as
mentioned. By recycling material or through other forms or repurposing old
appliances, you are making an important contribution to the protection of
our environment. Please inquire with your local community authorities for
the proper disposal location.

U.S. Solid®•USS-DBS Models • All Rights Reserved • Page 9
U.S. Solid® • 1200 E.55th Unit C, Cleveland, OH, 44103 • www.ussolid.com
U.S. Solid®•USS-DBS Models • All Rights Reserved • Page 2
U.S. Solid® • 1200 E.55th Unit C, Cleveland, OH, 44103 • www.ussolid.com
Thank you for choosing U.S. Solid’s Analytical Balance. We hope you enjoy your pur-
chase. Please maintain the equipment according to all instructions within the manual.
This product was developed according to CE marking regulations while emphasizing
aesthetics and safety for the user. It has ISO certific tion, specifically ISO 9001 The
high quality of the product, coupled with proper user care, should allow you to enjoy
the product for years.
Improper use of this equipment can cause accidents, including but not limited to:
electric discharges, circuit breaker malfunctions, fi e, etc. Please read the mainte-
nance section, where this is explained more fully.
FOR BEST RESULTS AND LONGEVITY IN PRODUCT
EFFICACY, READ THROUGH THIS MANUAL THOROUGHLY
BEFORE USING THE PRODUCT
Please bear in mind the following:
• This manual should be kept with the analytical balance, so it is available to all
users
• Be careful when handling the balance. Avoid any sudden movements when in
its vicinity. Be sure to not drop the balance, nor drop any heavy/sharp objects
on it. Keep liquids away from the balance to avoid spills.
• Never take apart the balance to try and fix ourself. This could ruin the entire
balance, and voids the product warranty. There is also risk of injury when
disassembling the balance.
• To prevent fi e and electrical issues, avoid dusty or overly dry environments. In
case of electrical issue or fi e, unplug the balance immediately.
• During setup, installation, and use, contact your wholesaler with any questions
that arise.
• This equipment is protected under the Warranties and Consumer Good
Regulation (10/2003)
• Overhaul is not covered by the equipment warranty
• Changes made to the product by non-certified individuals will esult in a loss of
the product warranty
• Accessories, including their loss, are not covered by the product warranty, nor
does the warrant cover piece’s deterioration over the course of time.
• Keep your invoice, either for having the right to a warranty, or for asking for
warranty coverage. Manufacturer reserves the right to modify or improve the
manual or equipment.
II. Getting Started
Troubleshooting Guide
Problem Cause Solution
Display not working No power supply Plug in the adapter
Display shows
“ “
Object weight below
minimum threshold,
taring issues exist, or
balance is not properly
calibrated.
Make sure item being
weighed is at least 10mg.
Make sure receptacles
have been tared, and
the balance has been
properly calibrated.
Display shows
“ “
Item being weighed is
over maximum allow-
able weight.
Make sure balance is
properly calibrated, and
make sure item is below
maximum allowable
weight.
Display shows “Err-1” Power issues with the
balance.
Machine may have a
poor power connection.
Check the power in the
back, and the outlet.
Leave machine off or
seveeral minutes, unplug
fully, then reconnect and
begin again.
Display shows “Err-2” Unstable weighing con-
ditions.
Recalibrate machine,
make sure balance is
level, and make sure
item being weighed is
fully stable.

U.S. Solid®•USS-DBS Models • All Rights Reserved • Page 3
U.S. Solid® • 1200 E.55th Unit C, Cleveland, OH, 44103 • www.ussolid.com
U.S. Solid®•USS-DBS Models • All Rights Reserved • Page 8
U.S. Solid® • 1200 E.55th Unit C, Cleveland, OH, 44103 • www.ussolid.com
FEATURES
• Straing gauge technology
• Stainless steel pan
• Backlit LCD display
• Plastic draft shield
• Height-adjustable feet
• Taring
• Four dierent units of measure: g, lb, ct, oz
• Piece counting function
ALWAYS MAKE SURE YOU CALIBRATE YOUR SCALE BEFORE USING, AND NEVER TOUCH THE
INCLUDED WEIGHT WITH YOUR SKIN!
VI. Maintenance and Cleaning
To get the best results and for product longevity, be sure to follow the suggestions
below.
Note: All of the suggestions and guidelines mentioned below and throughout this
manual will only be eective with continuous and careful maintenance of the equip-
ment
• Follow all processes outlined in the manual
• Make the manual available to all users of the equipment
• Prevent the balance from any sudden movements or falls, as well as from direct
sunlight or air fl w. The balance is a precision instrument and must be handled
carefully.
• Balance must be plugged into a grounded electrical outlet and the socket should
be easily accessible to unplug in case of emergency
• Never unplug the balance by pulling on the wire, do so from the base
• Never use the balance in a wedged-in location, such as a shelf
• Never use other objects such as pens, pencils, etc. to press the buttons on the
control panel; use only your finger .
• Never place anything on the pan heavier than the maximum capacity of the bal-
ance, or the sensor could be damaged
• Do not submerge the balance, nor spill any liquids on it
• If any liquid does come into contact with the electrical parts of the balance,
immediately disconnect it and send it in for service to be checked and adjusted if
necessary
• Always use only original components and supplies. Other devices and parts may
appear similar, but can damage the equipment.
Cleaning
• Never use scourers or substances that can grate to clean metallic parts. This
include, but is not limited to, stainless steel, aluminum, coatings, etc.. These can
damage the balance and lead to early erosion of eectiveness of the balance.
• -Use a lint-free cloth, dampened with soapy water, that does not contain any
abrasive surfaces, nor corrosive materials
ATTENTION!! If equipment is not properly
cleaned and cared for, technical service will be
unable to fix or epair issues.

U.S. Solid®•USS-DBS Models • All Rights Reserved • Page 7
U.S. Solid® • 1200 E.55th Unit C, Cleveland, OH, 44103 • www.ussolid.com
U.S. Solid®•USS-DBS Models • All Rights Reserved • Page 4
U.S. Solid® • 1200 E.55th Unit C, Cleveland, OH, 44103 • www.ussolid.com
II. Specific tions
Analytical balances oer fast and accurate measurements of
mass, using strain gauge technology for fast, effici t mea-
surement.
Technical Parameters:
MODEL: USS-DBS3-1 USS-DBS3-2 USS-DBS3-3 USS-DBS3-5
Max Capacity in
grams 100 200 300 500
Accuracy Class Class II
Min. Capacity 10 mg
Division 0.001g
Linearity ±0.002 g
Repeatability ±0.002 g
Operating
Temperature
5° C- 35° C, with flu tuations of <1°C/Hour
41° F- 95° F with flu tuations of <1.8° F/ Hour
Relative Humidity 50%-85%
Power AC 110V-240V, DC 9V
V. Using the Balance
Level adjustment
Once the balance is placed at its location, check the bubble level; if bubble is not well-centered,
turn the adjustable feet so the bubble moves to the center and the balance is properly leveled.
Start
1. Connect the balance to the power supply and press ON key on the panel. Wait 30 minutes
after powering on for best performance.
2. After warming up, calibrate the balance.
3. After calibration, the balance will enter weighing mode
4. To turn the balance o, press the OFF key on the panel and the display will shut o. If the
balance will be unused for an extended period of time, disconnect it from the power supply.
Calibration
3. Remove the weight from the pan
4. The balance will enter weighing mode and will display 0.000g.
Note: Each balance includes the necessary weight to complete external calibration
Tare function
1. Put the receptacle on the weighing pan; its weight will be displayed
2. Press TARE key and the reading on the display will automatically be set to 0.000g; tare is
complete
3. If the receptacle is removed from the pan, the corresponding weight, with negative value
will be displayed.
4. Press the TARE key again and the value on display will be set to zero again
Weighing unit selection
Press UNIT key to select the desired weighing unit from g, ct, oz, and lb
Piece counting function
1. Press the PCS button
2. Press the CAL button repeatedly to cycle through dierent counts. This function can be used
with counts of 10, 20, 50, 100, 150, 200, 250, or 500 pieces.
3. Put the corresponding number of pieces on the pan, press the PCS button to confi m.
4. Remove the pieces from the pan and the balance is ready to be used in piece counting
mode.
5. To return to weighing mode, simply press the PCS button again.
1. Press and hold down CAL key, when it shows ”CAL” on display, release the keys; then value
of the necessary weight will flash on display.
2. Put the calibration weight on the pan. Wait until the value of the calibration weight stops
flashing and is fixed on the display (e.g. 100.000 g, 200.000g, etc., depending on the model).

U.S. Solid®•USS-DBS Models • All Rights Reserved • Page 5
U.S. Solid® • 1200 E.55th Unit C, Cleveland, OH, 44103 • www.ussolid.com
U.S. Solid®•USS-DBS Models • All Rights Reserved • Page 6
U.S. Solid® • 1200 E.55th Unit C, Cleveland, OH, 44103 • www.ussolid.com
III. Panel Description
Panel Explanations
1. ON/OFF: Turns device on and o
2. UNIT: Select between ct, oz, lb, g
3. PCS: Piece counting mode
4. CAL: Calibration key
5. TARE/ZERO: Tare the balance
IV. Installation
Choosing the installation location
When analytical balances are used in routine lab or industrial measurement environ-
ments, the weighing speed will be much quicker, and the results more accurate. The
location should meet the following conditions:
1. Work room should be clean and dry
2. Balance should be placed on a solid, stable, plain fl t surface
3. Avoid locations in which the balance could be exposed to any of the follow-
ing:
a. Airfl w from air conditioners, fans, doors, or windows
b. Vibrations from surrounding or nearby equipment
c. Direct sunlight or radiation
4. Do not use the balance anywhere exposed to explosive, combustible, or
corrosive gases.
5. Do not use the balance in areas with high humidity or high levels of dust
6. When moving from a cooler to a warmer place, the accuracy and reliability of
the scale will be influen ed by the moisture condensation inside the balance.
In order to eliminate this influen e, it is best to leave the scale unplugged in
its new location for at least two hours before next use.
7. Avoid extreme temperatures or exposure to direct sunlight or air condition-
ers. Ideal working temperatures are between 5° C and 35° C (41°F and 95° F)
with tempature flu tuations of less than 1° C (1.8° F) per hour.
8. Keep the balance clean.
9. Do not leave material on the balance when not in use.
Unpacking and inspection
Check that all of the items indicated below are included in the package and that noth-
ing has been damaged:
• Balance
•Pan
• Power Cable
• External Calibration weight
• Draft Shield Pieces
In case the balance or any component is damaged during transport, tell your transpor-
tation agent immediately in order to process claims within the proper time frame.
Other manuals for USS-DBS3-1
1
This manual suits for next models
4
Table of contents
Other U.S. Solid Scale manuals

U.S. Solid
U.S. Solid USS-DBS81 User manual

U.S. Solid
U.S. Solid USS-DBS49 User manual

U.S. Solid
U.S. Solid USS-DBS82 User manual

U.S. Solid
U.S. Solid USS-DBS3-1 User manual

U.S. Solid
U.S. Solid USS-DBS92 User manual

U.S. Solid
U.S. Solid USS-DBS91 User manual

U.S. Solid
U.S. Solid USS-DBS83 User manual

U.S. Solid
U.S. Solid USS-DBS47 User manual

U.S. Solid
U.S. Solid USS-DBS86 User manual

U.S. Solid
U.S. Solid USS-DBS93 User manual