UEi DTK2-N User manual

This manual suits for next models

1

Other UEi Thermometer manuals

UEi INF195C User manual

UEi

UEi INF195C User manual

UEi INF200 User manual

UEi

UEi INF200 User manual

UEi DT150 User manual

UEi

UEi DT150 User manual

UEi DT221 User manual

UEi

UEi DT221 User manual

UEi DTH51 User manual

UEi

UEi DTH51 User manual

UEi INF155 Scout 1 User manual

UEi

UEi INF155 Scout 1 User manual

UEi INF195 User manual

UEi

UEi INF195 User manual

UEi INF165 User manual

UEi

UEi INF165 User manual

UEi PDT660 User manual

UEi

UEi PDT660 User manual

UEi INF225 User manual

UEi

UEi INF225 User manual

UEi 550B User manual

UEi

UEi 550B User manual

UEi DT10K User manual

UEi

UEi DT10K User manual

UEi INF195C User manual

UEi

UEi INF195C User manual

UEi DT200 User manual

UEi

UEi DT200 User manual

UEi INF195C-N User manual

UEi

UEi INF195C-N User manual

UEi INF200 User manual

UEi

UEi INF200 User manual

UEi INF165C User manual

UEi

UEi INF165C User manual

UEi SCOUT INF185 User manual

UEi

UEi SCOUT INF185 User manual

UEi INF195C User manual

UEi

UEi INF195C User manual

UEi INF151 User manual

UEi

UEi INF151 User manual

Popular Thermometer manuals by other brands

Benetech GM900 instruction manual

Benetech

Benetech GM900 instruction manual

Parton PTD-400 manual

Parton

Parton PTD-400 manual

GAME 15900 owner's manual

GAME

GAME 15900 owner's manual

scigiene FP-180 instruction manual

scigiene

scigiene FP-180 instruction manual

Ametek Land UNO U1 600/1600C-L user guide

Ametek Land

Ametek Land UNO U1 600/1600C-L user guide

Delton 981A instruction manual

Delton

Delton 981A instruction manual

Taylor 5286498 instruction manual

Taylor

Taylor 5286498 instruction manual

Baker MM2P instruction manual

Baker

Baker MM2P instruction manual

Chino IR-FA Series instructions

Chino

Chino IR-FA Series instructions

Vive DMD1010 user manual

Vive

Vive DMD1010 user manual

Tarmo 320281 manual

Tarmo

Tarmo 320281 manual

Citizen CTD 504 instruction manual

Citizen

Citizen CTD 504 instruction manual

TFA 14.1500 instruction manual

TFA

TFA 14.1500 instruction manual

Gotemp THB4002 user guide

Gotemp

Gotemp THB4002 user guide

GREISINGER G 1202 Quick reference guide

GREISINGER

GREISINGER G 1202 Quick reference guide

Amprobe TMD-53W user manual

Amprobe

Amprobe TMD-53W user manual

Trotec BP10 operating manual

Trotec

Trotec BP10 operating manual

Radiance instruments RT616 Series product manual

Radiance instruments

Radiance instruments RT616 Series product manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

7CE CERTIFICATE
The product complies with EMC directive 2004/108/EC
Technical standard: Emission EN 61326-1:2006 Class B
EN 55011:2009/A1:2010 Group 1 Class B
Immunity EN 61326-1:2006 EN 61000-4-2:2009
EN 61000-4-3:2006 A2:2010
8BATTERY REPLACEMENT
• Turnofftheinstrument
• Removethebatterycompartment
• Changethebattery
• Replacethecompartmentcover
END OF LIFE
9LIMITEd WARRANTY
TheDTK2iswarrantedtobefreefromdefectsinmaterialsand
workmanshipforaperiodofthreeyearsfromthedateofpurchase.If
withinthewarrantyperiodyourinstrumentshouldbecomeinoperative
fromsuchdefects,theunitwillberepairedorreplacedatUEi’soption.
Thiswarrantycoversnormaluseanddoesnotcoverdamagewhich
occursinshipmentorfailurewhichresultsfromalteration,tampering,
accident,misuse,abuse,neglectorimpropermaintenance.Batteries
andconsequentialdamageresultingfromfailedbatteriesarenot
coveredbywarranty.
Anyimpliedwarranties,includingbutnotlimitedtoimpliedwarranties
ofmerchantabilityandfitnessforaparticularpurpose,arelimited
totheexpresswarranty.UEishallnotbeliableforlossofuseofthe
instrumentorotherincidentalorconsequentialdamages,expenses,or
economicloss,orforanyclaimorclaimsforsuchdamage,expensesor
economicloss.Apurchasereceiptorotherproofoforiginalpurchase
datewillberequiredbeforewarrantyrepairswillberendered.
Instrumentsoutofwarrantywillberepaired(whenrepairable)fora
servicecharge.Returntheunitpostagepaidandinsuredto:
8030SWNimbusAvenue,BeavertonOR97008
• Accessories,probesandbatteries(NotCoveredbyWarranty)
• Claimsarenotacceptableforimproperuse(includingadaptation
toparticularapplicationsnotforeseenintheinstructionmanual
orimpropercombinationwithincompatibleaccessoriesor
equipment,orforrepaircarriedoutbyunauthorizedpersonnel.)
DTK2
DTK2
75.9
°
f
7
7
5
5
.
5
.
.
9
.
.
.
9
9
9
°
°
f
f
f
T1
K
T1
T1
K
K
76.1°f
7
7
6.1
6.1
°
°
f
f
f
T2
K
T2
T2
K
K
Differential Thermometer
INSTRUCTION MANUAL
ENGLISH
Differential Thermometer
Temperature Units °F / °C
Sensor Type k-type
Range -328°~2498°F(-200°~1370°C)
Accuracy
Range Accuracy
-328°~-58°F(-200°~-50°C) ±(0.1%rdg+4°F(2.2°C))
-58°~1832°F(-50°~1000°C) ±(0.1%rdg+2°F(1°C))
1832°~2498°F(1000°~1370°C) ±(0.1%rdg+4°F(2.2°C))
Accuracy (after ice-
point calibration) 30°~120°F(-1.1~48.9°C) ±1°F(0.6°C)
Resolution 0.1°F/0.1°C
HOLD Yes
Max / Min T1 & T2
Differential (T1-T2) Yes
Operating / Storage
Operating Temperature 32~104°F(0~40°C)
Operating RH% <70%RH,Non-condensing
Storage Temperature -4~140°F(-20~60°C)
Storage RH% <80%RH,Non-condensing
Battery Type 1x9V(NEDA1604)
Battery Life (typical) 125 hours typical
6SPECIFICATIONS
CAUTION:Thissymbolindicatesthat
equipmentanditsaccessoriesshallbesubject
toaseparatecollectionandcorrectdisposal.
WARNING
Ifthesymbol““appearsontheLCD,
pleasereplacethebatteryimmediately.
Copyright©2013UEi.AllRightsReserved
06/13 17018
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
3LCd dISPLAY
1. PrimaryLCD:DisplayT1(1stthermocouplereading)orT1-T2
2. SecondaryLCD:DisplayT2
4OPERATION
• POWER ON /OFF
Note:
Make sure you have plugged in the thermocouple probe before
turning on.
PressandholdPWRbuttontoturnonthemeter.Whilethemeteris
on,pressPWRbuttontoturnoffthemeter.
• SLEEP MODE (AUTO POWER OFF)
Themeterwillturnoffautomaticallyafter15minutesifnobuttons
arepressed.Todisableautopowerofffunction,whenthemeter
isoff,pressandhold“HOLD”buttonthenpress“PWR”button
toturnonthemeter,theprimaryLCDshowsthefollowing.Sleep
modeisnowdisabled.
• TAKING MEASUREMENT
Connectthermocouple.Ifprobeisopenornotconnected,theLCD
willshow”---”.
• DIFFERENTIAL (T1 - T2)
Press"MODE"todisplaydifferential.Press"MODE"againto
returntonormaldisplay.
•MAX / MIN
PressMAX / MINbuttontochangeoperationmodefromlive
data->MAX->MIN->backtolivedata.Correspondingvaluewill
displayontheLCD.Thesevalueswillresetatpoweronorwhen
performingaresetonMAX/MIN.
Reset MAX/MIN:PressMax/Minbuttonfor2secondstoreset
MAX/MINanalysisvalue.TheMAX/MINvalueisupdatedto
currentreading.
• DATA HOLD
Freezedisplayofbasicmeasurementsordisable.PressHOLD
buttonagaintogobacktothetemperaturereading.
• CHANGE UNIT
PressUNITbuttontochangeunit˚Cor˚F.Themeterdefault
temperatureunitis˚F.Themeterwilldefaulttolastselectedscale.
1INTROdUCTION
ThankyouforpurchasingtheUEiDTK2dualinputKtype
differentialthermometer.Thisthermometerfeaturesabuilt-in
microprocessorandhighresolutionanalogtodigitalconverter
chip.Thethermocoupleproberespondsveryquicklytoawide
rangeoftemperaturemeasurements.Theindustrystandardmini
thermocouplejackacceptsawideselectionofprobetypes.
FEATURES
•Dualinputwithdifferential(T1-T2)
•Min/Maxcapture(T1&T2)
•Autopoweroffwithdisablefeature(15minutes)
•LargeLCDscreen
•Completewithbeadwiretemperatureprobesand9Vbattery
• ˚C/˚Fselectfunction
• Magneticmount
INCLUDES
•Meter
•9Vbattery
•Operationmanual
• ATT29AK-typebeadprobe(2)
• Pouch
2KEYPAd (CONTROLS)
T1
-
T2
HOLD MAX MIN
K
K
2
1
MODE
SET
PWR
MAX
MIN
UNIT
HOLD
PRESS“UNIT”TOSELECT˚C/˚F.
Freezedisplay
Press“MODE”tochangeoperationmodeT1orT1-T2.
Thecorrespondingreadingwilldisplayontheprimary
LCD.ThesecondaryLCDwillalwaysdisplayT2.
Pressandhold“PWR” to turn on the meter or
pressittoturnoffthemeter.Themeterwillturnoff
automaticallyafter15minutesifyoudon’tpressany
key.Press“HOLD”and“PWR”buttonstodisable
Autosleepmode.
Press“MAX/MIN”tochangeoperationmode
fromlivedata->Max.->Min.->backtolivedate.The
correspondingvaluewilldisplayontheLCD.Thevalue
iscalculatedsincepoweredonorreset.
Pressandholdtoenterice-pointcalibration.(Please
refertosection 5toensurecorrectcalibration.) PressUNIT
K
K
K
K
SingleInput:(TopView)
K
K
T2
K
K
MIN
K
K
MAX
K
K
HOLD
K
K
5AUTOMATIC ICE-POINT CALIBRATION
TheDTK2hasanautomaticice-pointcalibrationfunction.Please
followthesestepstocalibrateyourmetertothethermocouple
probes.
• ITEMS NEEDED:
1. DTK2Thermometer
2. Twok-typeprobes
3. Crushedice/watermixture
(distilledwaterrecommendedifavailable)
• CALIBRATION PROCEDURE
1. ConnectbothprobestotheDTK2
2. Placeprobetipsinthecrushedice/watermixture
3. PowerontheDTK2
4. PressandholdSETforfiveseconds.
a. Themeterwillenterice-pointcalibrationand0.0°Cwill
flashonbothT1andT2readings
5. Stirtheicebathwiththeprobessubmergedforoneminute
6. Continuestirringtheice/watermixtureuntilthereadings
returntonormalmeasurement(flashingwillstop)
Theice-pointcalibrationisnowcomplete.
DTK2
75.9
°
f
75.9
°
f
T1
K
T1
K
76.1°f
76.1°f
T2
K
T2
K
Differential Thermometer
WARNING! Alwaysconnectthermocouplebeforeturningthe
unitonformostaccuratereadings.
WARNING! Makesuretoplugsensorwithcorrectpolarity.
Donotbealarmediferraticreadingsappearondisplayprior
tonormalmode.
MODE
K
K
K
K
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com