Inspect Patch Cable:
For SC-APC Patch Cable Inspection through an SC-SC
Optical Coupler,
•Insert the patch cable into the SC APC optical coupler.
Note: The patch cable and coupler are both keyed. The
keys must be aligned.
•Insert the SC APC Bulkhead inspection tip into the SC APC
Optical Coupler.
Note: The tip is also keyed. The keys must be aligned.
For SC-APC Patch Cable Inspection without an SC-SC
Optical Coupler,
•Insert the patch cable into the 2.5mm APC Patch Cable
inspection tip.
•Rotate fiber and align the connector key to the notch on
inspection tip for proper alignment.
For OptiTap Drop Cable Inspection,
•Insert the drop cable into the 2.5mm APC Patch Cable
inspection tip.
•Rotate fiber and align the connector key to the notch on
inspection tip for proper alignment.
1. Focus the live fiber image on the screen using the P5000i’s
Focus Control knob. If dirty, clean the connector and
reinspect.
2. If it appears clean, press the TEST function key.
3. If it fails, clean the fiber end-face and re-run the inspection
test. Repeat until it passes.
4. When the test passes the display will highlighted in green.
5. Press the ZONES function key to verify all three zones are
green and pass. You are now ready to connect to the fiber
under test.