manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Viavi
  6. •
  7. Microscope
  8. •
  9. Viavi P5000i User manual

Viavi P5000i User manual

Other manuals for P5000i

4

Other Viavi Microscope manuals

Viavi P5000i Parts list manual

Viavi

Viavi P5000i Parts list manual

Viavi FiberChek Parts list manual

Viavi

Viavi FiberChek Parts list manual

Viavi FVAm Series User manual

Viavi

Viavi FVAm Series User manual

Viavi P5000i Parts list manual

Viavi

Viavi P5000i Parts list manual

Viavi OneAdvisor ONA-800 Parts list manual

Viavi

Viavi OneAdvisor ONA-800 Parts list manual

Viavi FiberChek User manual

Viavi

Viavi FiberChek User manual

Viavi FVAi-2030 User manual

Viavi

Viavi FVAi-2030 User manual

Viavi FiberChek User manual

Viavi

Viavi FiberChek User manual

Viavi INX 760 User manual

Viavi

Viavi INX 760 User manual

Viavi P5000i Parts list manual

Viavi

Viavi P5000i Parts list manual

Popular Microscope manuals by other brands

Brütsch Rüegger FUTURO instruction manual

Brütsch Rüegger

Brütsch Rüegger FUTURO instruction manual

Elenco Electronics EDU-41003 instructions

Elenco Electronics

Elenco Electronics EDU-41003 instructions

Nikon SMZ25 instruction manual

Nikon

Nikon SMZ25 instruction manual

National DC3-163 instructions

National

National DC3-163 instructions

KERN Optics OPE-1 User instructions

KERN Optics

KERN Optics OPE-1 User instructions

OPTIKA MICROSCOPES B-150D Operation manual

OPTIKA MICROSCOPES

OPTIKA MICROSCOPES B-150D Operation manual

Prior Lux POL operating instructions

Prior

Prior Lux POL operating instructions

Olympus SZ61 instructions

Olympus

Olympus SZ61 instructions

Opitec Smartscoop 115615 instructions

Opitec

Opitec Smartscoop 115615 instructions

KERN OPO-1 operating instructions

KERN

KERN OPO-1 operating instructions

Topcon SP-1P user manual

Topcon

Topcon SP-1P user manual

AO Cycloptic Reference manual

AO

AO Cycloptic Reference manual

Leica DVM5000 Setup manual

Leica

Leica DVM5000 Setup manual

MshOt MZ62 user manual

MshOt

MshOt MZ62 user manual

Celestron 44345 instruction manual

Celestron

Celestron 44345 instruction manual

KERN Optics OZR-5 User instructions

KERN Optics

KERN Optics OZR-5 User instructions

VWR VisiScope 384 Series instruction manual

VWR

VWR VisiScope 384 Series instruction manual

Leica DI C800 User's manual & installation instructions

Leica

Leica DI C800 User's manual & installation instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

VIAVI//Public
P5000i Microscope
P5000i Microscope
QUICK CARD
https://www.viavisolutions.com/en-us/products/fiber-inspection-cleaning
Standard Tips
FBPT-COD-L
FBPT-COD-MTA
FBPT-BAP1
Fiber Inspection with the SmartOTDR
This Quick Card outlines how to use the P5000i Fiber Inspection Microscope to inspect
common single mode fiber end faces.
EQUIPMENT REQUIREMENTS
Figure 1: Equipment Requirements
►SmartOTDR Handheld Fiber Tester
►P5000i Fiber Microscope
►Bulkhead and Patch Cord Inspection
Tips matching the fiber under test
►Fiber optic cleaning tools
Figure 2: P5000i, Barrel Assembly and Tips
1. Attached appropriate Inspection Tip to the front of
the P5000i by threading the collar of the tip onto
the probe:
►Standard Patch cord tips and Bulkhead tips
connect to the Barrel Assembly (FBPT-BAP1)
by loosening and tightening the Retaining Nut
on the BarrelAssembly.
►Ribbon tips and the Corning OptiTap Bulkhead tip
(FBPT-COD-L) connect directly to the probe.
2. Insert the P5000i USB Connector into a USB Port
under a flap on top of the SmartOTDR
Figure 3: P5000i and SmartOTDR
Retaining
nut
Interchangeable
Inspection tip
Focus
Control
Magnification
control
USB
Connector
Barrel
Assembly
Test Button
USB Ports under flaps
CONNECT P5000i TO SmartOTDR
VIAVI//Public
P5000i Microscope
P5000i Microscope
QUICK CARD
2
1. Press the Home button to display the home screen.
2. Tap the Microscope icon.
3. Tap the Setup soft key .
4. Configure Tip Setting and Profile per the table below. If the Tip Selection is not shown on the
SmartOTDR, connect the SmartOTDR to a PC with FiberChekPRO™software using a USB to
MicroUSB cable and move the appropriate tip setting onto the SmartOTDR. For more information,
consult the FiberChekPRO user manual. To download FiberChekPRO to your PC, visit
http://fcpro.updatemyunit.net/.
SET UP THE P5000i
Connection Type Tip Profile Selection Tip Selection
LC Bulkhead SM UPC (IEC-61300-3-35 Ed. 2.0) Standard Tips (with BAP1)
Patch Cord SM UPC (IEC-61300-3-35 Ed. 2.0) Standard Tips (with BAP1)
SC-UPC Bulkhead SM UPC (IEC-61300-3-35 Ed. 2.0) Standard Tips (with BAP1)
Patch Cord SM UPC (IEC-61300-3-35 Ed. 2.0) Standard Tips (with BAP1)
SC-APC Bulkhead SM APC (IEC-61300-3-35 Ed. 2.0) Standard Tips (with BAP1)
Patch Cord SM APC (IEC-61300-3-35 Ed. 2.0) Standard Tips (with BAP1)
Corning
OptiTap Bulkhead SM APC (IEC-61300-3-35 Ed. 2.0) FBPT-COD-L
Patch Cord SM APC (IEC-61300-3-35 Ed. 2.0) Standard Tips (with BAP1)
Corning
OptiTip Bulkhead Ribbon, SM APC (IEC-61300-3-35 Ed. 2.0) Ribbon Tips
Patch Cord Ribbon, SM APC (IEC-61300-3-35 Ed. 2.0) Ribbon Tips
Figure 5:Test SetupFigure 4:Home Screen
VIAVI//Public
P5000i Microscope
P5000i Microscope
QUICK CARD
3
INSPECT THE BULKHEAD
1. Set up the P5000i as described on page 2.
2. Press the Back soft key to return to the
inspection screen.
3. Insert the Bulkhead inspection tip into the Bulkhead.
►For SC-APC Bulkheads, align the keys on the
inspection tip and the bulkhead connector.
►For OptiTap Bulkheads, align the arc and
flat pin in the inspection tip to the matching
insertions inside the OptiTap receptacle adapter.
►For OptiTip Patch Cord Inspection, slide
the patch cord adapter onto the main tip assembly.
Rotate wheel to view and test each fiber.
4. Adjust the Focus Control wheel of the P5000i
microscope to focus the optical fiber. The top vertical
bar on the screen indicates the level of focus
adjustment. The higher the level, the better the focus.
The screen will display a line in the focus bar where
the optimal focus lies for that image. Focus the image
until the bar is as close to the line as possible.
5. There are 2 ways to run the inspection test:
►Press the Test button on the P5000i.
►Tap the Test soft key on the SmartOTDR.
6. A green checkmark indicates a PASS result. A red X
indicates a FAIL result.
7. Tap the Live soft key on the SmartOTDR to return to
the Live fiber view.
8. If the test failed, clean the fiber end-face and re-run the
inspection test. Repeat until it passes.
Figure 7: SC APC Bulkhead
Figure 8: OptiTap Bulkhead
Figure 9: OptiTip Bulkhead
Figure 11: Pass/Fail Analysis
Figure 10: Live Inspection
Figure 6: LC Bulkhead
VIAVI//Public
P5000i Microscope
P5000i Microscope
QUICK CARD
© 2023 VIAVI Solutions, Inc,
Product specifications and descriptions in this
document are subject to change without notice.
Patented as described at viavisolutions.com/patents
Contact Us
To reach the VIAVI office nearest you,
visit viavisolutions.com/contact
+1 844 GO VIAVI
(+1 844 468-4284)
INSPECT THE PATCH CORD
1. Insert the Patch Cord into the Inspection tip:
➢For SC-APC Patch Cord inspection through an SC-SC Optical
Coupler, insert the Patch Cord into the SC APC optical coupler.
The Patch Cord, coupler, and tip are all keyed and the keys must
be aligned.
➢For OptiTap and SC-APC Patch Cord Inspection without an Optical
Coupler, insert the Patch Cord into the 2.5mm APC Patch Cord
inspection tip. Rotate fiber and align the connector key to the notch
on inspection tip for proper alignment.
➢For OptiTip Patch Cord Inspection, slide the bulkhead adapter onto
the main tip assembly. Rotate wheel to view and test each fiber.
2. Set up the P5000i as described on page 2.
3. Press the Back soft key to return to the
inspection screen.
4. Adjust the Focus Control wheel of the P5000i
microscope to focus the optical fiber.
5. There are 2 ways to run the inspection test:
►Press the Test button on the P5000i.
►Tap the Test soft key on the SmartOTDR.
6. A green checkmark indicates a PASS result.
A red Xindicates a FAIL result.
7. Tap the Live soft key on the SmartOTDR to return
to the Live fiber view.
8. If the test failed, clean the fiber end-face and re-run
the inspection test. Repeat until it passes.
Figure 12:
SC APC Patch Cord and Coupler
Figure 13: SC APC Patch Cord
Figure 14: OptiTap Patch Cord
Figure 15: OptiTip Patch Cord
Figure 16: Pass/Fail Analysis
Figure 15: Live Inspection