2- Cleaning of Placido rings
Cleaning the placido rings should be done between each
patient.
Clean the Placido rings with alcohol-free wipes disinfectant.
For this specific area, using solutions with alcohol could
damage the device.
Please use a cleaning solution without
alcohol
The screen must be wiped gently without
applying excessive force.
Only use the below recommended
solutions. Other solutions may damage or
discolour the device.
LUNEAU TECHNOLOGY OPERATIONS
2
Rue
Roger
Bonnet-27340
PONT
DE
L
’ARCHE
-
FRANCE
-
T
el.
+
33
232
989
132
• [email protected] • www. luneautech.com
Wipes - Instructions for use please read carefully !
01 02 03
Take a wipe out of its packaging and close Clean thoroughly by hand wiping. Throw away the wipe.
the packaging tightly after use. Leave the product on for 5 minutes. Change wipe for each medical
Do not rinse. surface or device.
3- Cleaning of screen
The screen should not be seen or touched by the patient.
Clean the screen regularly with Cidalkan wipes pr other
alternative.
4- Cleaning of the external
body of the device
Proceed to a regular cleaning of the exterior of the
device.
The patient will probably touch the device during the
exam.
Before cleaning the unit, turn it off and unplug it from
the electric outlet.
To clean the plastic surface of the device gently wipe
the top, bottom, and front surfaces.
CMJN (print) : 80 / 0 / 7 / 0
RVB : 0 / 178 / 227
Pantone : 306 C
HEX : #00b2e3
CMJN (print) : 67 / 9 / 100 / 0
RVB : 97 / 167 / 48
Pantone : 369 C
HEX : #61a730
Avoid to give the printed results to your
patients.
REF. 369113 Cidalkan wipes incl. alcohol.
disinfectant
Standards CE 0459
This document is not contractual - FC001163 - ind 00- April 2020