
Instruction Manual Version Page
Q-Flow qflow_D5_0 © Vögtlin Instruments GmbH 15
Notes on the configuration and specification of variable area flowmeters
When designing variable area flowmeters, misunderstandings often occur and,
consequently, misuse in practice. We would be grateful if you would take a few minutes to
read these instructions and complete the specification sheet.
Pressure and temperature dependence
Variable area flowmeters are highly dependent on the temperature and pressure in the
measuring tube. So that we can design the "right" device for you, we need additional
information from you.
Please select the appropriate installation situation 1-8 for your application from the
following pages.
Pressure conditions
For flow meters with integrated control valve, we need the differential pressure via the
control valve so that the correct valve size (KV value) can be selected. Furthermore, the
pressure in the measuring tube is decisive for a correct display.
Control valve at input (see installation situations 3+4)
The most common application is operating mode 3. On the output side there is no or only
a low back pressure (e.g. gassing of an open vessel or the influence of the pipeline). In
this case, we determine 1 bar abs (ambient pressure) in the measuring tube. Even if the
upstream pressure at the pressure reducer is now changed, the same flow can be
expected at the same scale height.
Control valve at outlet (see installation situations 5 to 7)
In the following operating modes, the valves are set to the output:
Processes with changing back pressure. Here we recommend that you set the
upstream pressure to at least twice the back pressure. This guarantees a constant
flow even with unstable back pressure. However, it is important that the set
pressure corresponds to the pressure marked on the scale or the pressure defined
in the chart.
During suction operation, ambient air is sucked through the flow controller with a
vacuum pump. (e.g. for air samplers) The inlet is open to atmosphere. In this case
we take 1 bar abs as a basis.
Reference pressure and reference temperature
The volume indicated on the scale or in the chart always refers to 0°C and 1013 mbar abs
(according to DIN 1343).
WWW.TEESING.COM | +31 70 413 07 50