VWR VWRI611-2258 User manual

MAN_VWR_M_MW_L_LW_EN_R_6.02_Rev1 03.01.19
Top Loading Balances
User Manual
Europen Catalogue Numbers:
VWRI611-2258
VWRI611-3395
VWRI611-3327
VWRI611-3478
VWRI611-2565
VWRI611-3328
VWRI611-2273
VWRI611-2288
VWRI611-2610
VWRI611-3397
VWR611-2259
VWRI611-3391
VWRI611-3378
VWRI611-3367
VWRI611-3361
VWR611-3122
VWRI611-3459
VWRI611-3471
VWRI611-2296
VWRI611-2309
VWR611-2260
VWRI611-2274
VWRI611-3469
VWRI611-3346
VWRI611-3388
VWR611-3123
VWRI611-3371
VWRI611-2289
VWRI611-3545
VWRI611-2320
VWR611-2261
VWRI611-3269
VWRI611-3277
VWRI611-2298
VWRI611-3375
VWRI611-3343
VWRI611-2275
VWRI611-3348
VWRI611-3379
VWRI611-2321
VWR611-3261
VWRI611-3373
VWRI611-3472
VWRI611-2299
VWRI611-3383
VWRI611-2262
VWRI611-3270
VWRI611-3296
VWRI611-3370
VWRI611-3386
VWRI611-3458
VWRI611-2292
VWRI611-2300
VWRI611-3262
VWRI611-2276
VWRI611-3384
VWRI611-3374
VWR611-2263
VWRI611-3392
VWRI611-3543
VWRI611-2301
VWRI611-3546
VWRI611-2277
VWRI611-2600
VWR611-3360
VWR611-3263
VWRI611-3396
VWRI611-3355
VWRI611-2319
VWRI611-3326
VWRI611-2278
VWRI611-2548
VWRI611-3359
VWR611-2264
VWRI611-3298
VWRI611-3464
VWRI611-3480
VWRI611-3351
VWRI611-3271
VWRI611-2293
VWRI611-2629
VWR611-3264
VWRI611-2279
VWRI611-3353
VWRI611-3398
VWR611-2265
VWRI611-3354
VWRI611-3544
VWRI611-2302
VWRI611-3350
VWRI611-3272
VWRI611-3466
VWRI611-3347
VWRI611-2272
VWRI611-3283
VWRI611-3462
VWRI611-2565
Version 1 Issued 14/04/15
Soft. R_6.02

Legal Address of Manufacturer
Europa
VWR International bvba
Researchpark Haasrode 2020
Geldenaaksebaan 464
B-3001 Leuven
+ 32 16 385011
http://be.vwr.com
Country of origin
Italy
Intended use
Precision balances are used for accurate weighings in laboratory, industry, universities,
etc…

1
INDEX
1INSTALLATION INSTRUCTION............................................................................................3
2STORAGE..............................................................................................................................4
3WEIGHING PAN ASSEMBLY ...............................................................................................5
4KEYBOARD AND DISPLAY..................................................................................................6
5OVERVIEW ............................................................................................................................7
5.1 FRONT PART OF THE BALANCE............................................................................................. 7
5.2 REAR AND BOTTOM PART OF THE BALANCE ........................................................................ 8
5.3 REAR AND BOTTOM PART OF THE BALANCE,MODELS VERSION SQUARE PAN.................... 9
5.4 IONISER (ONLY FOR MODELS -ION).................................................................................. 10
6WEIGHING...........................................................................................................................12
6.1 STAND BY.......................................................................................................................... 12
6.2 SIMPLE WEIGHING.............................................................................................................. 12
7CALIBRATION.....................................................................................................................13
7.1 EXTERNAL CALIBRATION................................................................................................... 13
7.2 INTERNAL CALIBRATION.................................................................................................... 14
7.2.1 Autocalibration (AUT-CAL)......................................................................................... 15
7.2.2 Internal calibration (I-CAL) [locked for verified balances] ....................................... 16
7.2.3 External calibration (E-CAL) [locked for verified balances]..................................... 16
7.2.4 Technical calibration (TEC-CAL) [locked for verified balances].............................. 17
7.3 VERIFICATION.................................................................................................................... 18
8TARE FUNCTION ................................................................................................................19
8.1 MANUAL TARE FUNCTION.................................................................................................. 20
8.2 OPTIONAL ALPHANUMERIC EXTERNAL KEYBOARD........................................................... 21
9WEIGHT UNITS....................................................................................................................23
10 PC COMMUNICATION SETTING........................................................................................26
11 PRINTER SERIAL COMMUNICATION SELECTION..........................................................27
12 TRANSMISSION SPEED SELECTION...............................................................................28
13 AUTOZERO FUNCTION......................................................................................................29
14 FILTERS SELECTION.........................................................................................................30
15 STABILITY FUNCTION .......................................................................................................31
16 DISPLAY CONTRAST REGULATION................................................................................32
17 DISPLAY BACKLIGHT SETUP...........................................................................................33
18 AUTO POWER-OFF FUNCTION.........................................................................................34
19 SELECTION OF WORKING MODE WITH TABLET...........................................................35
20 PIECECOUNTING FUNCTION............................................................................................37
20.1 VISUALIZATION OF TOTAL AND UNIT WEIGHT OF PIECES................................................... 38
20.2 MANUAL INSERTION OF THE UNIT AVERAGE WEIGHT........................................................ 39
20.3 AUTOMATIC UPDATE OF UNIT WEIGHT .............................................................................. 40

2
21 DENSITY DETERMINATION OF A SOLID OR A LIQUID..................................................41
21.1 DENSITY DETERMINATION OF A SOLID............................................................................... 41
21.2 DENSITY DETERMINATION OF A LIQUID............................................................................. 43
22 MAXIMUM LOAD DETERMINATION FUNCTION..............................................................45
23 PERCENTAGE WEIGHING FUNCTION .............................................................................46
23.1 MODE WITH WEIGHT REFERENCE....................................................................................... 46
23.2 MODE WITH MANUAL INSERTION OF THE REFERENCE WEIGHT.......................................... 47
24 ANIMAL WEIGHING FUNCTION ........................................................................................49
25 TOTALIZING........................................................................................................................50
26 THRESHOLD FUNCTION....................................................................................................52
26.1 BOTH THRESHOLDS SET ..................................................................................................... 53
26.2 LOWER THRESHOLD SET ONLY........................................................................................... 53
26.3 HIGHER THRESHOLD SET ONLY.......................................................................................... 53
27 RS232 INTERFACE FEATURES.........................................................................................54
28 ERROR CODES...................................................................................................................60
29 MAINTENANCE AND CARE...............................................................................................61
30 QUICK GUIDE TO BALANCE PARAMETERS SETUP......................................................62
31 QUICK GUIDE TO THE USE OF THE BALANCE PROGRAMS........................................63
32 BALANCE TECHNICAL FEATURES....................................................................................64
33 OPTIONAL ACCESSORIES...................................................................................................65
34 TECHNICAL SERVICE...........................................................................................................65
35 WARRANTY............................................................................................................................66
36 COMPLIANCE WITH LOCAL LAWS AND REGULATIONS.................................................66
37 EQUIPMENT DISPOSAL........................................................................................................66

3
1 Installation instruction
WARNING:
Please follow carefully these steps for installing and use the new balance
before starting your work routine. A way of use of the instrument different
from this user manual will not guarantee the instrument’s safety anymore.
Remove the balance, calibration weight( if included) and accessories from the
cartoon and check for any visible damage of the instrument.
Do not install the balance in a place with air flows, heavy thermal changes and
vibrations.
Do not use the balance in blast risk environment.
The humidity rate of the balance environment must be between 45% and 75%
Place the weighing pan and the support pan on the balance (see par. 3).
Level the balance using the level bubble and levelling feet located underneath the
case
Connect power supply to connector 2 located on the rear panel of the unit (see
par. 5).
ATTENTION
In models with internal calibration remove the screw in the left lower
part of the balance (see par.5)

4
Insert power unit into AC outlet, which shall be easily accessible; after few
seconds the balance will automatically switch on.
Balance Warm-up:
Wait 8 hours from switch one for Analytical balanaces (0,0001g)
Wait 30 minutes for other models
Then calibrate the balance using the appropriate mass (if supplied), following
instructions (par.7).
It is strongly suggested not to disconnect the balance from the electrical
mains and to use the ON/OFF button to stand-by the balance when you
finished using it
Calibrate again the balance every time it’s moved to another place
Check balance calibration periodically.
We recommend not dropping heavy objects on balance pan, in order to avoid
damages.
Assistance service must be effected by specialized staff and the spare parts used
must be original. Therefore, it is necessary to apply to the reseller who sold the
equipment.
2 Storage
Storage temperature: +5 °C…+40°C
Storage humidity 45% - 75%.
Keep package in the case the balance should be sent back to the factory for service.
Disconnect the cables and the accessories for avoiding damages during
transportation.
Do not place the balance in extreme temperature and humidity condition and avoid
the balance to take strong hit.

5
3 Weighing pan assembly

6
4 Keyboard and display
*Stability indicator
O Zero indicator
%Percentage weighing
PC Piece counting
Battery charge indicator
Insert data mode
HUpper threshold
L Lower threshold
DS Density measure mode
ct, Measure unit
ozt, lb, GN, dwt, Kg, mg
Standby (OFF/ON) or escape
ESC.
TARE or zero button.
Selection CONFIRM or SEND
data to printer.
Balance setup MENU button.
Balance CALIBRATION button.
ESC
ESC

7
5 Overview
5.1 Front part of the balance
1
2
3
4
5
6
Fig.1
Balance with circular and three-panes windshield
Name and function
1
Cap of windshield
2
Levelling bubble
3
Windshield
4
LCD display
5
Function buttons
6
Levelling feet

8
5.2 Rear and bottom part of the balance
Fig.2
Rear and bottom part of the balance
Name and function (Fig.2)
1
Weighing pan
2
9 poles (pin) female connector for RS232 interface for printer/PC/external keyboard
3
Levelling feet
4
connector for power supply
5
Screws for balance closing. In models with 4 feet it is necessary first to remove the
two rear feet (one fixed and one moving) to access to the closing screws
6
Hook for under-balance weighing. Hook not available in internal calibration models
and for verified balances. Note: please remove the cap to access the hook
7
Protection screw for autocalibration motor (only for models with internal calibration).
ATTENTION: remove this screw by hand as soon as the balance is removed from
its package
2
3
3
1
4
4
5
5
5
6
6
5
5
5
5
5
7

9
5.3 Rear and bottom part of the balance, models Version Square Pan
Fig.3
Rear and bottom part of the balance
Name and function (Fig.3)
1
Weighing pan
2
9 poles (pin) female connector for RS232 interface for printer/PC/external keyboard
3
Levelling feet
4
connector for power supply
5
Screws for balance closing. In models with 4 feet it is necessary first to remove the
two rear feet (one fixed and one moving) to access to the closing screws
6
Hook for under-balance weighing. Hook not available in internal calibration models
and for verified balances. Note: please remove the cap to access the hook
7
Protection screw for autocalibration motor (only for models with internal calibration).
ATTENTION: remove this screw by hand as soon as the balance is removed from
its package
8
Adjustable rear
9
Fixed rear feet
2
3
3
1
4
4
8
5
5
6
6
5
5
5
5
5
7
9

10
5.4 Ioniser ( only for models -ION)
Using the ioniser:
Keys descriptions.
1 Activation button ioniser in continuous mode. (Auto power off after 8 hours of
operation)
2 Activation button ioniser in timed mode. (Operation for 2 minutes)
N.B. to switch from one function or to turn off the device, press either one of the two
keys.
LED green: powered ioniser.
Solid red LED: ioniser activated in continuous mode.
Flashing red LED: ioniser activated in time.
1
2

11
The ionizer is mounted in place of the rear glass in the models with
resolution 0.0001 version -ION as in the photo.
Technical data Ioniser
Distance “sample ion source”
c. 5 - 40 cm
Ozone concentration
0 ~ 0.05 ppm (2cm from ion source)
Ambient conditions
0 –50°C, 20 ~ 80% air humidity (non-condensing)
AC adapter (primary)
AC 100-240V, 50/60Hz
Rated electric power supply
DC 12V, 500mA
Pollution degree
2
Overvoltage category
Category II
Altitude
Up to 2000m
Installation site
Device may only be used indoors

12
6 Weighing
For verified Balances models at switch on it will appear “WARM UP” on
display and countdown will start
After having connected the balance to AC outlet, it will perform an internal circuits test,
after that the balance will set itself in stand-by mode.
6.1 Stand By
From “STAND BY” mode:
Press ON/OFF button to bring balance to work conditions.
Press again ON/OFF button to return to “STAND BY” condition.
6.2 Simple weighing
Load the sample to weigh on the pan and read the value on display as soon as the
stability symbol ж(star) appears
The digit that represents the verification division “e” is rounded with a frame

13
7 Calibration
Electronic balances take mass measurements making use of gravity (g). Difference of
latitude in geographic areas and altitude will vary gravity acceleration value (g).
Therefore, for accurate measurements, the balance must be adjusted to the local
environment. This adjustment is accomplished by calibration function.
7.1 External calibration
For verified Balances models the External calibration function is locked
Calibration is accomplished by pressing CAL button.
1. Press CAL button when pan is empty, dashes are displayed on the display.
2. When calibration weight value starts to flash, load the weight on the pan.
3. The display will stop flashing, indicating calibration weight value.
Once the calibration is effected will be shown the value of the calibrated weight and
the current unit of measure.
4. Unload calibration weight from the pan.
The balance is ready for weighing operations.
NOTE: if there is interference during calibration process, an error message will be displayed.

14
Moreover, it is possible to calibrate the balance with a calibration weight higher than the
one set by default:
1. Press and keep CAL button pressed with empty pan until the acoustic alarm
stops, then release the button. On display it will be visualized the string "-CAL-",
followed by flashing string "LOAD".
2. Load on the pan a weight equal higher or lower than default calibration weight; the
balance will recognize it as valid weight if equal or higher than calibration weight
as long as it is a whole number in comparison with the most meaningful digit of
calibration weight .
e.g.: if calibration weight is 200g, it will be possible to calibrate the balance with
values from 100g 200g, 300g, 400g up to the highest limit of balance weighing
range.
The message “LOAD” on display will stop flashing. Once calibration has been
effected, the value of calibrated weight will be displayed.
3. Unload calibration weight. The balance is ready for weighing operations.
NOTE: if there is interference during calibration process, an error message will be displayed.
7.2 Internal calibration
* only for models with internal calibration “-i”
2 Sec

15
In these balance models there are 4 calibration modes:
From display zero condition, press and keep pressed the MENU button until the acoustic
alarm gets mute, then release the button. The message “unitS” will be visualized on
display, press then MENU button until you visualize “Calib” on display. Press PRINT to
confirm.
1. Select the calibration mode you wish by pressing MENU button in sequence:
AUT-CAL: auto calibration
I-CAL: internal calibration*
E-CAL: external calibration*
TEC-CAL: technical calibration*
2. Press PRINT button to confirm “AUT-CAL”, “I-CAL”, “E-CAL”.
To confirm “TEC-CAL” keep pressed the PRINT button until the acoustic alarm gets
mute.
3. After selection, the balance returns to calibration menu. Press and keep pressed
MENU button until the acoustic alarm gets mute, then release the button. Balance is
again ready for weighing operations.
7.2.1 Autocalibration (AUT-CAL)
This is Factory setting for verified balances models
The balance self-calibrates when the temperature variation exceeds the factory preset
value and at factory preset time intervals, through the internal reference mass, and only if
the balance pan is empty.
When the balance needs to perform the Automatic calibration, the display will show the
following message:
A 25-seconds countdown will start during which you can decide if:
*Locked for verified balances models

16
Stop the automatic-calibration procedure by pressing the “ON/OFF button that will
be delayed of 5 minutes
or
Let the countdown finish so that the automatic calibration starts
Note: during the countdown do NOT load nothing on the plate!
In this mode, it is also possible to carry out the calibration with the internal reference
mass by pressing the CAL button at any moment, first ensuring that no weight is loaded
on the plate.
1. Press CAL button with empty pan.
The message “CAL” will be displayed and balance calibration will be effected
automatically.
2. At the end of calibration, balance returns to normal weighing conditions.
If calibration is not finished due to vibrations or air flows, then the message “CAL
bUt” is displayed. Press again CAL button. If the problem repeats then select external
calibration and contact the product supplier.
7.2.2 Internal calibration (I-CAL) [locked for verified balances]
The balance calibrates itself with internal reference mass ONLY on user command by
pressing CAL button.
Before effecting internal calibration, be sure that there’s no weight loaded on the balance
pan.
7.2.3 External calibration (E-CAL) [locked for verified balances]
The balance is calibrated through an external reference mass (NOT provided with
models with internal calibration).
(Follow procedures described at paragraph 7.1)

17
7.2.4 Technical calibration (TEC-CAL) [locked for verified balances]
This function allows storing the value of internal reference mass whenever checking or
assistance actions require it.
1. After having selected the TEC-CAL calibration mode, press CAL button at empty
pan. It will be displayed “CAL”.
2. When the value of
calibration weight start flashing on display, load the weight on to the balance pan.
3. Wait the acoustic alarm and that the displayed calibrated weight value stops flashing,
then unload the weight from balance pan.
4. When string “0.000”is displayed continuously, then press and keep pressed the
PRINT button. This starts the internal weight value automatic acquisition and store.
During the acquisition cycle, the display will show “TEC-MEM”.
5. After having stored the value of internal calibration weight, balance returns to normal
weighing conditions.
6. Return to calibration menu as described at paragraph 6.2 and set the desired
calibration mode: internal, automatic or external..
ATTENTION : this procedure must be effected only using E2-class reference
masses.

18
7.3 Verification
General introduction:
According to EU directive 90/384/EEC balances must be verified if they are used for
legally controlled area as follows:
a) For commercial transactions if the price of goods is determined by weighing.
b) For the production of medicines in pharmacies as well as for analyses in the
medical and pharmaceutical laboratory.
c) For official purposes.
d) For manufacturing final packages.
In cases of doubt, please contact your local trade in standard.
Verification instructions
An EU type approval exists for balances described in their technical data as
verifiable. If a balance is used where obligation to verify exists as described above, it
must verified and re-verified in regular intervals.
Re-verification of a balance is carried out according to the respective national
regulations. The legal regulation of the country where the balance is used must be
observed!
After verification the balance is sealed at the indicated positions.
Verification of the balance is invalid without the "seal".
Balances with obligation to verify must be taken out of operation if:
The weighing result of the balance is outside the error limit.
Therefore, in regular intervals load balance with known test weight (ca.
1/3 of the max. load) and compare with displayed value.
The reverification deadline has been exceeded.
This manual suits for next models
95