3-3TerminalConnection
●Input/OutputScrewTerminal
1)Powerconnection:POWER
Connectthepower(12to24VDC)to“POWER”ontheinput/output
terminalboard.
Sincethismeterisnotprovidewithapowerswitch,itisready
tooperateassoonasitisconnectedtothepowersupply.
2)Measured-InputTerminalConnection
●ThermocoupleType
Ⅰ)Connectthethermocouple+legtoAontheterminal,and
thethemcouple−legtoBonthesameterminal.
Ⅱ)Whencompensationwireswithshieldareused,connect
theshieldtoC(COM)ontheterminal.
Ⅲ)Neverapplyvoltagebeyondanallowablemaximumvoltage
totheinput(terminalsA,BandCOM)
●RTDType
Ⅰ)Connect3-wiresystemPt100ΩtoA,B,COMterminals.
Ⅱ)Forthe2-wiresystemshorttheterminalsBandCOM.
Ⅲ)NeverapplyinputotherthantheRTDtotheinputterminals.
4.Maintenanceandinspection
4-1Cautionformaintenance
Storethemeteratastoragetemperatureof−10℃to70℃
andarelativehumidityoflessthan60%.
Ifthemeterisusedatadustylocation,occasionallyremove
theinternalboardfromthecase,thenremovedustaccumulated
ontheboard.
(Otherwiseariseintemperaturesofinternalpartsmayshorten
amaterservicelife.)
Asthecaseandpanelaremoldedbyplastics,donotremove
stainsfromthemwithavolatileliquidsuchasthinner.
4-2Calibration
Inordertoassuretheinitialaccouracyforalongperiodof
time,calibratethemeterperiodicallyatanambient
temperatureof23℃±5℃andahumidityof35to85%RH.
Ifthemeterisoutofthespecifiedaccuracyasaresultof
calibration,contactus.
●ThermocoupleType
◎Conductcalibrationinthefollowingorderat23℃±5℃
(1)Removethefrontpanelofthismeter
(2)First,turnthepowerONwiththewiringmadeasshowninthe
aboveFigure,thenwarmupthemeterformorethan20minutes.
Note:Selectathermocoupleappropriatetothiscaribration.
(3)0℃CheckandAdjustment
Verifythatthedisplayshows0℃withreferencevoltage
generatoroutputsetto0.00mV.
(4)SpanAdjustment
Setreferencevoltagegeneratoroutputtoavaluethefullscale
(thefollowingtable)andifthedisplaydeviatesfromtheabove
value,turnthespanadjustmentVRuntilitdisplaysthevalue.
Sensor Display Inputvoltage(mV)
KC 1000℃ 41.276
JC 400℃ 21.848
●RTDType
Connectadecadetyperesistanceboxasshowninthefollowing
diagram
◎Calibratethemetertheordersfownbelow.
(1)Removethefrontpanelofthismeter.
(2)First,turnthepowerONwiththewiringmadeasshowninthe
aboveFigure,thenwarmupthemeterformorethan20minutes.
(3)ZeroAdjustment
TurnthezeroadjustmentVRuntilthedisplayshows00.0℃
forthePATypeand000℃forthePJTypewiththeresistance
boxsetto100Ω
(4)SpanAdjustment
Settheresistanceboxtoavalueasshowninthefollowingtable.
Ifthedisplaydeviatesfromtheabovevalue,turnthespan
adjustmentVRuntilitshownthevalue.
Model Display Pt100
Resistanceonresistancebox(Ω)
PA 199.0℃ 175.49
PJ 600℃ 313.71
5.Warranty
Thismeteriswarrantedforaperiodofoneyearfromdateof
delivery.Anydefectwhichoccursinthisperiodandisundoubtedly
causedbyWatanabeElectricIndustryfaultswillberemediedfree
ofcharge.
Thiswarrantydosenotapplytothemetershowingabuseordamage
whichhasbeenalteredorrepairedbyothersexceptasauthorized
byWatanabeElectricIndustry.
6.After-saleservice
Thismeterisdeliveredafterbeingmanufactured,testedand
inspectedunderstrictqualitycontrol.
However,ifanyproblemdoseoccur,contactyournearestWatanabe
ElectricIndustrysalesagentorWatanabeElectricIndustrygiving
asmuchinformationonproblemaspossible.
MODELAT-501Series UU-47009h