Wolfgang Warmbier PGT 120 User manual

PGT®120
Personnel Grounding Tester
Wrist Strap and Footwear Tester
Users's Manual

PGT120 Art.Nr. 7100.PGT120
2 / 13 2014-11-27
1 Table of contents
1TABLE OF CONTENTS....................................................................................... 2
2INTRODUCTION.................................................................................................. 3
2.1 Device return and environmentally compatible disposal........................... 3
3INSTALLATION................................................................................................... 4
4OPERATION........................................................................................................ 5
4.1 Wrist strap test................................................................................................ 5
4.2 Coil cord test................................................................................................... 5
4.3 Footwear test (single shoe) .......................................................................... 6
4.4 Wrist strap and footwear test ........................................................................ 6
4.5 Footwear in series .......................................................................................... 7
5CONFIGURATION............................................................................................... 8
6CONNECTORS.................................................................................................... 9
7WALL MOUNTING INSTRUCTIONS................................................................. 10
8SPECIFICATIONS ............................................................................................. 11
9PICTURES......................................................................................................... 12

PGT120 Art.Nr. 7100.PGT120
3 / 13 2014-11-27
2 Introduction
The Personnel Grounding Tester PGT®120 is an electronic test instrument for
checking personnel grounding systems such as wrist straps, coil cords and footwear.
The PGT®120 is suitable for compliance verification of the above products, according
to the IEC 61340-5-1 Edition 1.0 (2007-08) or ANSI/ESD S 20.20 –2007 (2007-03).
The unit operates with 3 independent measuring circuits for the left shoe, the right
shoe and the wrist strap. This makes it possible to measure all the values at the
same time
It is possible to enable or disable separately the measuring circuits.
The order of tests is random.
Footwear measurement can be configured to measure in series with hands free
for passenger gates.
Visual and audible test results and a dry relay contact for door opener
Use the optionally available "Calibration Unit " Part No. 7100.PGT120.CU to
check the unit Hi and Lo limit values
2.1 Device return and environmentally compatible
disposal
The instrument is a category 9 product (monitoring and control instrument) in
accordance with ElektroG (German Electrical and Electronic Device Law). This
device is not subject to the RoHS directive.
We identify our electrical and electronic devices (as of August 2005) in
accordance with WEEE 2002/96/EG and ElektroG with the symbol shown to
the right per DIN EN 50419.
These devices may not be disposed of with the trash.
Please contact our service department regarding the return of old devices.
If you use batteries or rechargeable batteries in your instrument or accessories
which no longer function properly, they must be duly disposed of in compliance with
the national regulations.
Batteries or rechargeable batteries may contain harmful substances or heavy metal
such as lead (Pb), cadmium (Cd) or mercury (Hg).
The symbol shown to the right indicates that batteries or rechargeable
batteries may not be disposed of with the trash, but must be delivered to
collection points specially provided for this purpose.
Pb Cd Hg

PGT120 Art.Nr. 7100.PGT120
4 / 13 2014-11-27
3 Installation
The Unit is for desktop or wall mounting. The optionally available wall mounting
frame (Part No. 7100.PGT120.WK) can be used to fix the unit to a wall.
The power is either supplied by a 9V battery or by a power supply. Insert the
batteries by opening the battery compartment flap on the bottom side of the unit.
Take care of the proper polarity. Close the battery compartment flap again.
It is strongly recommended to remove old or empty batteries to avoid battery
leakage.
For heavy use or footwear in series mode, we recommend using the external power
supply. Use only an original power supply connected to the "AC12V" socket on the
rear. On plugging in external power, the internal battery is disconnected.
The battery should be removed to avoid wear out.
Do not connect any conducting articles with PGT®120 exept original accessories
(power supply, battery and foot wear electrode) and the door opener.
Connect the foot electrode with the coloured marked plugs to the back of the unit for
footwear test.

PGT120 Art.Nr. 7100.PGT120
5 / 13 2014-11-27
4 Operation
This tester has no power switch. Pressing an electrode or activating "footwear in
series" activates the electrical circuit.
The measuring voltage is preset to 100V. Use the DIP switches 6+7 to adjust the
voltage to either 30V or 50V.
4.1 Wrist strap test
Settings:
Only wrist strap or OR is activated (DIP switch 1+2)
Put on the wrist strap and connect it via a coil cord to the snap or to the socket on the
left side of the unit.
Press the left electrode and keep it pressed. A peep signal indicates the start of
measurement. After a short measuring time the result is displayed.
OK
Green LED flashes
The measured value is o.k.
Hi-Fail
Red LED flashes,
audible signal
Above the resistance upper limit
Lo-Fail
Red LED flashes,
audible signal
Below the resistance lower limit
(not applicable if lower limit is disabled)
Release the electrode.
4.2 Coil cord test
Settings:
Only wrist strap or OR is activated (DIP switch 1+2)
To check only the coil cord, connect the coil cord to the 3mm snap located inside the
wrist strap symbol and to the 10mm snap or socket on the left side of the unit.
Press the left electrode and keep it pressed. A peep signal indicates the start of the
measurement. After a short measuring time the result is displayed.
OK
Green LED flashes
The measured value is o.k.
Hi-Fail
Red LED flashes,
audible signal
Above the resistance upper limit
Lo-Fail
Red LED flashes,
audible signal
Below the resistance lower limit
(not applicable if lower limit is disabled)
Release the electrode.

PGT120 Art.Nr. 7100.PGT120
6 / 13 2014-11-27
4.3 Footwear test (single shoe)
Settings:
Only footwear or OR is activated (DIP switch 1+2)
Stand on the foot electrode, then press the right electrode and keep it pressed. A
peep signal indicates the start of measurement. After a short measuring time the
result is displayed.
OK
Green LED flashes
The measured values of both shoes are o.k.
Hi-Fail
right
Red LED flashes,
audible signal
Right shoe above the resistance upper limit
Hi-Fail
left
Red LED flashes,
audible signal
Left shoe above the resistance upper limit
Lo-Fail
right
Red LED flashes,
audible signal
Right shoe below the resistance lower limit
(not applicable if lower limit is disabled)
Lo-Fail
left
Red LED flashes,
audible signal
Left shoe below the resistance lower limit
(not applicable if lower limit is disabled)
Release the electrode.
4.4 Wrist strap and footwear test
Settings:
AND function is activated (DIP switch1+2)
Put on the wrist strap and connect it via a coil cord to the snap or socket on the left
side of the unit.
Stand on the foot electrode, then press one electrode and keep it pressed. A peep
signal indicates the start of measurement. After a short measuring time the result is
displayed.
OK
Green LED flashes
All measured values are o.k.
Hi-Fail
Red LED flashes,
audible signal
Above the resistance upper limit
Lo-Fail
Red LED flashes,
audible signal
Below the resistance lower limit
(not applicable if lower limit is disabled)
Release the electrode.
The OK signal only appears when all measured values are within the limits.

PGT120 Art.Nr. 7100.PGT120
7 / 13 2014-11-27
4.5 Footwear in series
Settings:
Footwear in series activated (DIP switch 8)
To show that "Footwear in series" is activated the red LED's for Hi-Fail of Footwear
flash for a short time every 2s, while disspative shoes are not detected.
The footwear test can be accomplished hands free, without touching a electrode.
This is useful in combination with passenger handling gates. The resistance is
measured between the two shoes. This mode cannot directly indicate the faulty shoe.
Press the Shoe electrode on the instrument to identify it.
Stand with both feet onto the foot electrode. If the shoes are dissipative the
measurement starts automatically. After a short measuring time the result is
displayed and the connected gate will open.
OK
Green LED flashes
The measurement of the footwear in series is o.k.
Hi-Fail
Red LED flashes,
audible signal
Above the resistance upper limit for series
connection
Lo-Fail
Rote LED flashes,
audible signal
Below the resistance lower limit
(not applicable if lower limit is disabled)
You can step of the foot electrode.
Even if footwear in series is active, you can perform a test according chapter 4.1 to
chapter 4.4 by pressing a electrode, for example to identify a bad shoe.

PGT120 Art.Nr. 7100.PGT120
8 / 13 2014-11-27
5 Configuration
The unit can be configured with the DIP switches on the rear according to the table
below.
Standard settings are marked bold.
Switch 1
Switch 2
Test mode
OFF
OFF
"OR" (wrist strap or footwear test)
ON
OFF
Only footwear test
OFF
ON
Only wrist strap test
ON
ON
"AND" (wrist strap and footwear test)
Switch 3
Switch 4
Footwear upper limit
OFF
OFF
20 Mfor single shoe / 40 Mfor series
OFF
ON
35 Mfor single shoe / 70 Mfor series
ON
OFF
70 Mfor single shoe / 140 Mfor series
ON
ON
100 Mfor single shoe / 200 Mfor series
Switch 5
Lower limit
OFF
Lower limit disabled
ON
Lower limit enabled
Switch 6
Switch 7
Test voltage
OFF
OFF
30 V
OFF
ON
50 V
ON
--
100 V
Switch 8
Footwear test mode
OFF
test according to switch 1 + 2
ON
footwear in series active
Switch 9
Piep for footwear in series
OFF
At start of test
ON
At end of test if shoes OK
Switch 10
Door opener time
OFF
3s
ON
1s
Switch 11
Not used
OFF
-
ON
-
Switch 12
beeper
OFF
inactive
ON
active

PGT120 Art.Nr. 7100.PGT120
9 / 13 2014-11-27
6 Connectors
The connectors for the power supply, the foot electrodeand the door opener are
located on the rear side of the unit. Use a "RJ12" western modular plug to connect
the dry contact of the door opener.
The door opener relay is triggered and stays on for 3 seconds when the test result
indicates OK.
Normally Open Pin 3,4
Normally Closed Pin 2,3
1 65432 Door Opener
1 2 3 45 6
Connector viewContact function

PGT120 Art.Nr. 7100.PGT120
10 / 13 2014-11-27
7 Wall mounting instructions
(Part No. 7100. PGT120.WK)
Fix the wall mounting plate with the supplied dowels and screws and stick the self-
adhesive Velcro tapes according to the picture.
1. The surface of the plate and the bottom side of the PGT®120 have to be clean, dry
and free of grease.
2. Remove protecting foil of the velcro tapes and do not touch the sticky side.
3. Apply the velcro tapes according to the picture onto the mounting plate.
4. Remove the second protecting foil of the velcro tapes and press the PGT®120
agianst them.
5. After 24 Hour curing time the PGT®120 can be removed from the wall mounting
plate
Before removing the unit, please unplug all wires.
Hold the unit on both sides and pull it forwards. To fix it again, press it back onto the
velcro tapes.
Wall mounting plate
ca. 10mm
velcro tape

PGT120 Art.Nr. 7100.PGT120
11 / 13 2014-11-27
8 Specifications
Operating voltage:
9V E 6F22 battery
external power supply 230V / 50Hz
Operating conditions:
15 ... 40°C
up to 75% relative humidity,
non condensing
Storage conditions:
-10 ... 60°C
up to 85% relative humidity,
non condensing
Connectors:
Wrist strap
10mm snap, 4mm snap, 4mm
socket
Foot electrode
2 sockets 4mm
Door opener
Western socket 6 pin RJ-12
external 12VAC power supply (Use only for the original power
supply supplied with the instrument)
Measuring ranges:
Wrist strap
750k... 35M
Footwear - each shoe
100k... 100M
Footwear in series
200k... 200M
(Hands-free-Mode)
Tolerance
10%
Test voltage:
open circuit voltage
30V 10%
50V 10%
100V 10%
Signals:
Green LED
“OK”
Rote LED’s
and buzzer
“Hi-Fail“ or ”Lo-Fail”
Door opener
Dry contact “OK“
Contact ratings:
max. voltage
60V
max. current
2A
max. power
50 VA
Operating modes:
Single test ”OR“
Double test "AND"
Only wrist strap test
Only footwear test
Hands-free-Mode
Weight:
app. 500g
Dimensions:
150 x 200 x 63 mm
Serial number:
On the side of the unit
Complies with CE

PGT120 Art.Nr. 7100.PGT120
12 / 13 2014-11-27
9 Pictures
Press Press
Hi-Fail
Lo-Fail
OK
PGT 120
Personnel
Grounding
Tester
LED
"upper limit exceeded"
LED
"below low limit"
LED
"measured value
in the limit"
footwear test
male snap 3mm
"below low limit"
LED
"upper limit exceeded"
LED
wrist strap test
electrode for
wrist strap test electrode for
footwear test
connections for
wrist strap test

PGT120 Art.Nr. 7100.PGT120
13 / 13 2014-11-27
socket for
door opener
sockets for
foot electrode
socket for
external
power supply
Dip-Switch
AC 12 V Door opener
7 6 5 289
10 1
OFF
ON
mode
voltage
beep series
door opener
high limit
series
4 3
11
---------
12
beeper
low limits

Änderungen vorbehalten Subject to change without notice
WOLFGANG WARMBIER GmbH & Co. KG
Untere Gießwiesen21
D-78247 Hilzingen
Telefon +49 77 31 86 88-0
Telefax +49 77 31 86 88-30
e-mail: [email protected]
http://warmbier.com
Table of contents
Other Wolfgang Warmbier Test Equipment manuals
Popular Test Equipment manuals by other brands

Clas Ohlson
Clas Ohlson HL-205 instruction manual

Powertrain
Powertrain Walterscheid SW42 operating manual

UEi
UEi IRT807 instruction manual

Bosch
Bosch CRI SST Original instructions

Kyoritsu Electrical Instruments Works, Ltd.
Kyoritsu Electrical Instruments Works, Ltd. KEW3125A instruction manual

R&S
R&S RTE Series Getting started