Woodpecker DTE Implant-X User manual

The pictures are only for reference. The nal interpretation rights belong to Guilin Woodpecker Medical Instrument Co., Ltd. Woodpecker reserves the right to change
the design of the equipment, the technique, ttings, instruction manual and the content of the original packing list at any time without further notice.
The product were authorized appearance patent, counterfeit will be sued.
Guilin Woodpecker Medical Instrument Co., Ltd.

Thank you for purchasing Woodpecker Dental Implant Device Implant-X. In order to guarantee the correct operation, it is
recommended to read this instruction manual carefully before operation. For convenient reading, it is recommended to put it where is
available at any time.
Device Type
1. Type of protection against electric shock: Class I equipment with internal power supply
2. Degree of protection against electric shock: B type applied part
3. Recommended disinfection method: See section 6 Cleaning, disinfection, and sterilization
4. Waterproof protection is in line with the current version IEC60529: host IPX1,foot pedal IPX6.
5. Degree of safety application in the presence of a ammable anesthetic mixture with air, oxygen, or nitrous oxide: Equipment
cannot be used in the presence of a ammable anesthetic mixture with air, oxygen, or nitrous oxide.
6. Operating mode: Continuous operation
Precautions
1. Please read these precautions before operation and operate in correct way.
2. The following icons is for ensuring safe operation, preventing you or others from being hurt. These icons are classified by
degree of risk, degree of damage and severity. All indicators should be highly concerned. Please obey the instruction.
Classication Degree of risk, degree of damage and severity
Dangers Indicating potential personal injury or bodily injury
Warnings Indicating potential slight injury or bodily injury
Precautions Indicating instructions to be observed for ensuring safety

Contents
1. Product introduction........................................................................................................1
2. Accessories description...................................................................................................3
3. Control of host and foot pedal .......................................................................................4
4. Installation ......................................................................................................................7
5. Operation ......................................................................................................................10
6. Clean, disinfection, and sterilization ............................................................................19
7. Error code and solution (error alarm interface) ............................................................21
8. Storage and maintenance .............................................................................................21
9. Symbols.........................................................................................................................23
10. Specications ..............................................................................................................24
11. After-sales service .......................................................................................................25
12. Environment protection .............................................................................................25
13. Statement ....................................................................................................................25
14. Guarantee ....................................................................................................................25
15. EMC-Declaration of comformity................................................................................26

1
1. Product introduction
1.1 Precautions
Danger
1. To prevent electric shock, do not use wet hands to pull the power cord; be sure to prevent the control circuit from water; use a
grounded electrical outlet.
2. Keep it away from explosives and combustibles, with special care not to use this machine for patients who use nitrous oxide
anesthesia.
3. This equipment may be used only by specialized and suitably trained personnel such as surgeons. The application place of the
device is dental clinic or hospital.If correctly used, this equipment does not give rise to side eects. Improper use, on other hand, will
give rise to transmission of heat to the tissues.
Warnings
1. To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth.
2. In the presence of electromagnetic interference environment, the planter may be malfunctioning. Do not install the Dental
Implanting Device near equipment that releases magnetic waves. When using ultrasonic vibrating equipment or electrode knife
nearby, close the switch on the control panel.
3. Not to position the device to make it dicult to operate the disconnection device.
4. Implant-X requires special precautions for EMC and needs to be installed and put into service according to the EMC
environment.
5. Device with electromagnetic launcher will aect the normal operation of Implant-X, do not run both devices at the same time.
6. Implant-X cannot be used in operating rooms containing potentially ammable gas mixtures.
7. To avoid possible injury to human or damage to the device, make sure that the motor handpiece (hereinafter simply referred to
as the motor) is completely parked when replacing the planting tool. And the replacement shall be conducted by foot pedal controller.
8. Severe impact, such as dropping, will lead to damage to the implanting device.
9. During the work of peristaltic pump, the water pipe cannot be excessive bending or knotting, otherwise the pipe may fracture.
10. Do not attempt to disassemble the control panel, foot control or motor.
11. Dental handpieces (hereinafter referred to as handpieces ) should be cleaned, lubricated and disinfected immediately after use.
12. Do not lubricate the motor. Lubricating oil can cause overheating, resulting in damage to the motor. Control panel and multi-
function pedals cannot be disinfected.
13. Do not clean control panel with dissolving solution.

2
14. The motor cable cannot be removed from motor.
15. Switch o electrical power after each use.
Precautions
1. If you need to repair and purchase spare parts, please contact the authorized supplier.
2. It is recommended to use the original pre-disinfection disposable water pipe combination.
3. The accuracy of torque monitoring depends on the accuracy of the handpiece installed on the micro motor. If the handpiece
produced by other manufacturers is used, the actual torque value may not be displayed correctly. To ensure that the actual torque
matches the displayed torque, please use the matched handpiece.
4. Please read this instruction manual before operation and master parts of functions.
5. Check the operating status of Implant-X before use and conrm that there is no abnormal condition.
6. Test the Implant-X before operation to ensure correct operation.
7. If there is a permanent malfunction (excessive vibrations, noise, heat production, etc.) on the Implant-X, please immediately
close it and return it to the authorized dealer.
8. If the frequency of use is very high, please consider storing some spare parts.
9. Please cut o the power before cleaning the control panel with a damp cloth.
10. Dispose water pipe after operation with the method of disposing medical waste.
11. The operation mode of Implant-X is continuous operation mode, i.e., there will be 10 minutes pause after 3 minutes' operation.
If there is no system overheating, it will prevent the patients, users or the third party from hurt. The user should be responsible for the
use and shutdown of system.
This instruction manual is intended to indicate the safety requirements, installation procedures, proper methods of use and proper
maintenance of the equipment. If you encounter any unexpected problems, please contact Service Center of Guilin Woodpecker
Medical Instrument Co., Ltd.
The manufacturer will not be responsible for any personal injury or property damage caused by device tampering or modication
conducted by unauthorized person.
Woodpecker reserves the right to change the design of the equipment, the technique, ttings, instruction manual and the content of
the original packing list at any time without further notice. The pictures are only for reference. The nal interpretation rights belong to
Guilin Woodpecker Medical Instrument Co., Ltd.
Guilin Woodpecker Medical Devices Co., Ltd. will continue to update its products, thus bring changes in device components. If
there is any dierence between your manual and the description on your product, please contact the authorized distributor or after-
sales service center of Guilin Woodpecker Medical Instrument Co., Ltd. for explanation.

3
This manual is strictly prohibited from being used in any way other than installation, use and maintenance of the equipment.
1.2 Contraindications and precautions
1.2.1 The hemophilia patient is forbidden to use this equipment.
1.2.2 The patients and doctors with heart pacemaker are forbidden to use this equipment.
1.2.3 Heart disease patients and children should be cautious to use the equipment.
1.2.4 Patients with oral and maxillo-facial infection, oral mucosal diseases, periapical disease, gingivitis, periodontitis, or mouth
neoplasm should be cautious to use this equipment.
1.2.5 Patients with allergic constitution and drug allergy history are forbidden to use this equipment.
1.2.6 People with mental disorders should be cautious to use this equipment.
1.2.7 Patients with severe systemic infection or systemic diseases such as the diseases of heart, liver, kidney, hematopoietic system,
digestive system and endocrine system should be cautious to use this equipment.
8. Pregnant women, lactating women, and women have a plan of birth should be cautious to use this equipment.
1.3 Scope of use
This product is intended for use in dental surgery, thus other uses are not allowed. There will be potential danger if it is used for
other purposes!
1.4 Safety requirement
Guilin Woodpecker Medical Instrument Co., Ltd. will NOT be responsible for any direct or indirect damages and losses under the
following conditions:
The equipment is used for any purpose that is not mentioned in the scope of use.
The operator does not follow the steps and requirements in instruction manual to use the device.
The cabling system of the room where the device is used does not meet the appropriate standards and the appropriate requirements.
Assemble, operate and repair the device without authorization of manufacturer.
The environment in which the device is located or stored does not meet the requirements mentioned in technical requirements
section of the instruction manual.
2. Accessories description
Please refer to packing list for device conguration.

4
3. Control of host and foot pedal
3.1 Control of host keys
3.1.1 Working interface and control of keys

5
1. Light key Motor LED switch; circular touch to control the on/o of LED.
2. Store key Store key; click to store setting specications.
3. Water control key Click to select water volume. Six water gear, including 00%, 20%, 40%, 60%, 80%, and 100%
4. Forward/reverse rotation
Used to select the direction of rotation; direction changes for each pressing.
5. Speed ratio key
Used to set the gear ratio with handpiece; repeatedly pressing till the LCD correctly displays gear ratio with handpiece.
6. Torque control key
Used to set torque range of motor; “+” for increase, while “-” for decrease. Press and hold to accelerate the speed change.
7. Speed control key
Used to set rotating speed of motor; “+” for speed up, while “-” for slow down. Press and hold to accelerate the speed change.
8. Program key
Touch the icons to choose corresponding programs. Please refer to the Clause 5.1 for the functions of each program.
3.1.2 Interface of restoring the factory settings
Whether to restore the
factory settings?
Yes
No
Figure 1 Restore factory setting interface
While starting up, press the foot pedal at the same time. The factory reset menu will pop up as shown in the picture. When "YES"
is selected, the saved parameters will be cleared and the original factory setting parameters will be restored. When "No" is selected,
the factory settings will not be restored and the system will boot normally.

6
3.1.3 Error alarm interface
Error 01
The foot is not connected!
Please check whether the
foot is in good contact.
Figure 3 Error alarm interface
As shown in Figure 3, the warning 0x indicates alarm number. Please refer to section 7 Error code and solution for specic number
and the corresponding content.
3.2 Foot pedal control
1. Water volume adjustment button
Used for choosing 6 cooling water ow levels. The water level is incremented each time the button is stepped. At the maximum

7
level 6, pressing once to loops back to level 1.
2. Program switchover button
Used for choosing needed programs. The program is changed each time the button is stepped. At the Program 8, pressing once to
loops back to Program 1.
3. Forward / reverse rotating direction
Used for changing the rotating direction of contra-angle. The direction will change after stepping.
4. Speed control foot pedal
Used to start/stop the motor and control the speed during operation. The operating speed of motor is controlled by the foot of
operator. After lifting the foot, the displayed data will change to the maximum setting value.
4. Installation
4.1 Safety requirements during installation
Danger: Equipment is installed on the premise that the installation must meet the appropriate standards and related electrical
safety requirements.
Danger: Never install the device in an explosion-hazardous area and the device must not be operated in areas with ammable
gases (anesthetic mixture, oxygen, etc.).
Danger: Installation site should be able to protect device from shocks and splashing of water or other liquids.
Danger: Do not install the device near or above a heat source. It must be installed in a well-ventilated area with sucient
space around it, especially the exhaust fan and back.
Warning: Do not directly expose the parts to the sun or UV light source.
Warning: The device is movable. Be careful while handling it.
Warning: Before connecting the cord to the device, make sure the joint is dry. If necessary, dry it with air gun.

8
4.2 Accessories connection
Figure 4
4.2.1 Installation of foot pedal:
Connect the foot pedal plug to the foot pedal socket and tighten those two xing screws (Figure 4 - E)
4.2.2 Power cord installation:
Plug the power cord output into the power supply socket of the device (Figure 4 - D)
4.2.3 Installation of infusion bottle holder:
Insert the infusion bottle holder into the xing hole on the right rear of the shell; (Figure 4 - A)
4.2.4 Installation of infusion bottle:
Hang the infusion bottle (The infusion bottle contains purchased normal saline injection.) on its holder.
4.2.5 Installation of motor:
Plug the tail cord of motor into the output socket on the front of the device (Note: align the red marking point). (Figure 4 - B)
Figure 5

9
4.2.6 Peristaltic pump tube installation:
A.Turn the peristaltic pump knob clockwise to the "OPEN" indicator and open the pump head (Figure 5-A).
B.Place the tube in the impeller of the peristaltic pump (Figure 5 - B).
C.Turn the peristaltic pump knob clockwise to the "CLOSE" position and close the pump head (Figure 5 - D).
4.2.7 Complete machine eect after installations of all accessories: (Figure 6)
Figure 6
4.2.8 Switch on (Figure 4 - C) ; start to use the machine as it display normally:
Step on the pedal after the parameters such as speed, torque and water are set properly. The device starts to work. When the pedal
is released, the device stops working.

10
5. Operation
5.1 Program
5.1.1 Choice of programs
Implant-X owns 8 programs. There are two ways to choose program:
1. Touch the corresponding icons on the screen.
2. Step on the “Program switchover” button on the foot pedal.
5.1.2 Function description of programs
The function of each program is as shown as follow:
Icon Function Description
Positioning Accurate positioning on the alveolar bone by using a positioning drill.

11
Hole-drilling Determine the direction and depth of hole-drilling.
Hole-
broadening Determine the diameter of the hole.
Tapping Make a thread on the hole to match the implant.
Implanting Implant dental implants into alveolar bone.
Lock the
abutment screw Screw the nut onto the dental implant.
User dened
mode
Change straight handpiece, contra-angle with dierent speed ratios for dierent
dental procedures.
Cleaning Water discharging without motor rotation is convenient for ushing.
5.1.3 Factory Settings
Before delivery, several parameters mainly including speed, torque, speed ratio and water output have been set according to the
actual application. These parameters can be changed within the range of parameters specied in the current program.

12
The range of dierent parameters and their factory settings are as shown in the table below:
Icon Function Speed/ rpm Torque/ N·cm Speed ratio Water output/%
Positioning 200-2500 1000(D) 5-80 35(D) 16:1,20:1, 27:1,20:1(D) 40
Hole-drilling 200-2500 800(D) 5-80 35(D) 16:1,20:1, 27:1,20:1(D) 40
Hole-broadening 200-2500 600(D) 5-80 35(D) 16:1,20:1,27:1,20:1(D) 40
Tapping 15-100 20(D) 5-80 35(D) 16:1,20:1, 27:1,20:1(D) 40
Implanting 15-100 20(D) 5-80 35(D) 16:1,20:1, 27:1,20:1(D) 0
Lock the abutment
screw 15-100 20(D) 5-15 10(D) 16:1,20:1, 27:1,20:1(D) 0
User dened mode 15-40000 1200(D) 5-80 45(D) 1:1,1:2,1:3, 1:5,16:1,
20:1,27:1, 20:1(D) 40
Cleaning —— —— —— 80
Note: the letter “D” stands for default value.

13
5.2 Default parameter adjustment
Within specied range, the adjustable parameters are as follow:
1. Maximum speed
2. Maximum torque
3. Water output
4. Speed ratio
5.2.1 Maximum speed adjustment
Touch the “Speed” (+, -) key to adjust motor speed. The speed will change each time after touching the “Speed” key. Long press
the “Speed” key to accelerate the change of speed setting value.

14
5.2.2 Maximum torque adjustment
Touch the “Torque” (+, -) key to adjust maximum torque output of motor. The torque will change each time after touching the
“Torque” key. Long press the “Torque” key to accelerate the change of torque setting value.
5.2.3 Water output adjustment

15
Touch the “Water volume” key on the screen to adjust. There are 6 water levels. The water level will change to the next level after
each touch.
Step on the blue “Water volume adjustment” button to adjust water volume.
5.2.4 Speed ratio adjustment
Adjust by pressing the "Speed Ratio" button to match the gear ratio of the handpiece to be used.

16
5.3 Motor rotating direction adjustment
Touch the key shown above to change the rotating direction of motor.
Step on the “Forward/reverse rotation” during operation to change the rotating direction of motor.

17
5.4 Motor LED adjustment
Touch the “LED” button to for setting to determine the on or o state of LED while stepping on foot pedal. The state of LED will
change once after each touch. Only the device with LED owns this function.
5.5 Save the parameters
Table of contents
Other Woodpecker Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Biosense
Biosense Breath Ketone Monitor user manual

RECK
RECK MOTOmed loop instruction manual

Ci:z. Labo
Ci:z. Labo Esthe Pro CII instruction manual

COOK Medical
COOK Medical Zenith Spiral-Z AAA Instructions for use

Quantel Medical
Quantel Medical Pocket II E PCK2 MT AN Service manual

medi
medi Genumedi E+motion Instructions for use

Stryker
Stryker Power-PRO XT 6500 Operation & maintenance manual

laerdal
laerdal Resusci Anne Directions for use

BIOTRONIK
BIOTRONIK Galeo Pro Instructions for use

Shimadzu
Shimadzu UD 150L-30 EX installation guide

FujiFilm
FujiFilm SonoSite Edge II User guide supplement

Welch Allyn
Welch Allyn KleenSpec 790 Series Directions for use