Aging SenTix®
6ba75845e02 12/2014
Prior to the next measurement, shortly rinse the electrode with the test
sample or deionized water.
Overnight or
longer
Insert the clean electrode into the watering cap filled with reference
electrolyte and shut the refilling opening.
Aging
Every ORP electrode undergoes a natural aging process. Extreme operating
conditions can considerably shorten the lifetime of the electrode. These are:
Strong acids or lyes, hydrofluoric acid, organic solvents, oils, fats, bro-
mides, sulfides, iodides, proteins
High temperatures
High changes in pH and temperature.
The warranty does not cover failure caused by measuring conditions and
mechanical damage.
Maintenance and cleaning
During operation, a small amount of reference electrolyte leaks through the
junction from the electrode into the test sample. If the level of reference
electrolyte becomes too low with time, refill it through the refilling opening.
Refilling the
reference
electrolyte
Refilling is very easy using a dropping bottle. Proceed as follows:
Cut off the tip of the dropping bottle at a right angle until the opening in the
tip can be seen
Open the refilling opening of the electrode
Press the tip of the dropping bottle into the refilling opening while turning
it slightly
Pump several small quantities of the reference electrolyte into the stem
using the dropper bottle
Electrode Reference electrolyte Model
(see page 8)
SenTix®ORP 900 3 mol/l KCl, Ag+-free KCl-250 (250 ml)
SenTix®ORP-T 900
During longer storing periods, salt sediments may develop on the
watering cap. They do not affect the measuring characteristics
and can easily be removed with deionized water when the elec-
trode is put into operation again.