Safety instructions StirrOx G
6ba25310d07 08/2018
2 Safety instructions
This operating manual contains special instructions that must be followed
during the operation of the D.O. sensor.
Always keep this operating manual in the vicinity of the sensor.
Special user
qualifications
The membrane cap of the D.O. sensor is filled with a small amount of an
alkaline electrolyte solution. All maintenance work that requires dealing with
the electrolyte solution must only be carried out by persons who know how to
safely deal with chemicals.
General safety
instructions
The individual chapters of this operating manual use safety labels like the one
below to indicate danger:
CAUTION
indicates instructions that must be followed precisely in order to avoid
slight injuries or damage to the instrument or the environment.
3 Commissioning
Scope of delivery D.O. sensor StirrOx G, filled with electrolyte solution and ready to operate
Calibration and storage vessel, OxiCal®-ST
ZBK-ST accessory kit, comprising:
– 3 exchange membrane caps, WP-ST
– Electrolyte solution, ELY/G
– Cleaning solution, RL/G
– Polishing strip, SF 300
Operating manual
Getting the sensor
ready for
measuring
Connect the 8-pin plug to the Oxi input and the 2-pin plug to the stirrer
connection of the meter (see operating manual of the meter). The sensor is
immediately ready to measure. It is not necessary to polarize the sensor.
The membrane cap that is mounted on the sensor for delivery
serves mainly as a transport protection. Depending on the duration
of the transport and storage period, it may have a shortened oper-
ational lifetime. If the measuring system cannot be calibrated (error
message of the instrument), please proceed according to section
REPLACING THE ELECTROLYTE SOLUTION AND MEMBRANE CAP.