wtw MIQ User manual

ba64103e03 12/2002
MIQ/T2020
Operating Manual
C
MS
ESC OK
IQ SENSOR NET terminal

Accuracy when
going to press The use of advanced technology and the high quality standard of our
instruments are the result of continuous development. Consequently,
this may result in some differences between this operating manual and
your instrument. Also, we cannot guarantee that there are absolutely
no errors in this manual. Therefore, we are sure you will understand
that we cannot accept any legal claims resulting from the data, figures
or descriptions.
Note
The latest version of the present operating manual can be found in the
Internet under www.WTW.com.
Copyright ©Weilheim 2002, WTW GmbH & Co. KG
Reprinting - even as excerpts - is only allowed with the explicit written
authorization of WTW GmbH & Co. KG, Weilheim.
Printed in Germany.

IQ Sensor Net MIQ/T2020 List of contents
0 - 1
1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1.1 How to use this component operating manual . . . . . . . . 1-1
1.2 Features of the MIQ/T2020 terminal . . . . . . . . . . . . . . . 1-1
2 Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2.1 Authorized use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2.2 General safety instructions . . . . . . . . . . . . . . . . . . . . . . . 2-2
3 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.1 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.2 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.2.1 Docking and removing a terminal . . . . . . . . . . . 3-1
3.2.2 Docking the terminal on the
IQ SENSOR NET S184 system . . . . . . . . . . . . . . 3-3
3.2.3 Initialization . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4.1.1 Overview of the operating elements . . . . . . . . . 4-2
4.1.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
4.1.3 Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
4.1.4 Rotary switch . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
4.2 General operating principles . . . . . . . . . . . . . . . . . . . . . 4-6
4.2.1 Navigating in menus, lists and tables . . . . . . . . 4-7
4.2.2 Entering text or numerical values . . . . . . . . . . . 4-8
4.3 Software status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
5 Settings/Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
5.1 Selecting the language . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
5.2 Terminal settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
6 Maintenance and cleaning . . . . . . . . . . . . . . . . . . . . . 6-1
6.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
6.2 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
7 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1

List of contents IQ Sensor Net MIQ/T2020
0 - 2

IQ SENSOR NET MIQ/T2020 Overview
1 - 1
1Overview
1.1 How to use this component operating manual
Structure of the
IQ SENSOR NET
operating manual
Fig. 1-1 Structure of the IQ SENSOR NET operating manual.
The IQ SENSOR NET operating manual has a modular structure like the
IQ SENSOR NET itself. It consists of a system operating manual and the
operating manuals of all the components used.
Please file this component operating manual in the ring binder of the
system operating manual.
1.2 Features of the MIQ/T2020 terminal
The MIQ/T2020 terminal is an extension component for the
IQ SENSOR NET system. It can be docked on – and removed again from
– any free lid front of an MIQ module as a mobile control unit. The
docking is performed by a simple manual action without the need for
any tools.
IQ Sensor Net Betriebsanleitung
System-
Betriebs-
anleitung
(Ringordner)
Betriebs-
anleitung
IQ Sensor
Betriebs-
anleitung
MIQ Modul
Betriebs-
anleitung
MIQ Terminal
Komponenten-Betriebsanleitungen

Overview IQ SENSOR NET MIQ/T2020
1 - 2
Information that describes the behavior of the terminal after the docking
maneuver can be stored in each MIQ module. This information inclu-
des:
lThe name of the respective measurement location
lA selection of measured values that should be displayed at this
measurement location.
This function (local identity function) can be used, e.g. for sensors to
quickly find the calibration on site. The information is transmitted by the
two upper contacts on the back of the terminal.
Fig. 1-2 MIQ/T2020 terminal
1Graphical display
2Keys
3Rotary switch
4Contacts for voltage and communication
5Contacts for the local identity function
6Docking mechanism
C
MS
ESC OK
1
2
3
5
4
6
Front
Back

IQ SENSOR NET MIQ/T2020 Safety instructions
2 - 1
2 Safety instructions
This operating manual contains special instructions that must be
followed during the installation of the MIQ/T2020 terminal. Thus, it is
essential for the operator to read this component operating manual
before carrying out any work with the system. In addition to this manual,
the SAFETY chapter of the IQ SENSOR NET system operating manual
must be followed.
Always keep this component operating manual together with the sys-
tem operating manual and all other component operating manuals in
the vicinity of the IQ SENSOR NET system.
General safety
instructions
The following safety labels in the individual chapters of this operating
manual indicate different levels of danger:
Warning
indicates instructions that must be followed precisely in order to prevent
serious dangers to personnel.
Caution
indicates instructions that must be followed precisely in order to avoid
slight injuries to personnel or damage to the instrument or the environ-
ment.
Other labels
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. operating manu-
als.

Safety instructions IQ SENSOR NET MIQ/T2020
2 - 2
2.1 Authorized use
The authorized use of the MIQ/T2020 consists of its use as a control
unit in the IQ SENSOR NET.
Please keep to the technical specifications according to chapter 7
TECHNICAL DATA. Only operation according to the instructions in this
operating manual is authorized.
Any other use is considered to be unauthorized. Unauthorized use
invalidates any claims with regard to the guarantee.
2.2 General safety instructions
The MIQ/T2020 is constructed and inspected according to the relevant
guidelines and norms for electronic instruments (see chapter 7 TECHNI-
CAL DATA).
It left the factory in a safe and secure technical condition.
Function and operatio-
nal safety The failure-free function and operational safety of the MIQ/T2020 is
only guaranteed if the generally applicable safety measures and the
special safety instructions in this operating manual are followed during
its use.
The failure-free function and operational safety of the MIQ/T2020 is
only guaranteed under the environmental conditions that are specified
in chapter 7 TECHNICAL DATA.
Safe operation If safe operation is no longer possible, the MIQ/T2020 must be taken
out of operation and secured against inadvertent operation.
Safe operation is no longer possible if the MIQ/T2020:
lhas been damaged in transport
lhas been stored under adverse conditions for a lengthy period of
time
lis visibly damaged
lno longer operates as described in this manual.
If you are in any doubt, contact the supplier of your MIQ/T2020.

IQ SENSOR NET MIQ/T2020 Installation
3 - 1
3 Installation
3.1 Scope of delivery
lMIQ/T2020 terminal
lOperating manual
3.2 Commissioning
Note
Follow the installation instructions in the INSTALLATION chapter of the
system operating manual.
Up to three terminal components can be docked on an IQ SENSOR NET
system.
3.2.1 Docking and removing a terminal
The mobile MIQ/T2020 terminal is docked on the lid of any free
MIQ module. The terminal can be docked while the system is running
without the measurement being interrupted.
Docking the terminal
Fig. 3-1 Preparing the MIQ module for docking
1Pull the cover off the contacts (pos. 1 in Fig. 3-1) on the lid of
the MIQ module.
1

Installation IQ SENSOR NET MIQ/T2020
3 - 2
Note
For the temporary docking of a mobile terminal, do not remove the
whole contact cover. Leave it hanging on the side of the MIQ module.
After removing the terminal, replace the cover again to protect the
contacts.
Fig. 3-2 Docking the terminal
Removing the terminal To remove the terminal, press the locking lever and pull the terminal
forwards.
2Place the terminal on the module lid. To do so, insert the flap
on the back of the terminal into the slot on the door hinge of the
module. Then, pull back the interlocking lever (pos. 2 in Fig. 3-
2), place the terminal completely on the module and release
the interlocking lever.
2

IQ SENSOR NET MIQ/T2020 Installation
3 - 3
3.2.2 Docking the terminal on the IQ SENSOR NET S184 system
A free module is required to dock a mobile MIQ/T2020 terminal. With
the IQ SENSOR NET S184 system, there are the following options to
gain a free module:
lExpanding the system by an MIQ module, e.g. MIQ/IF232
lRemoving the MIQ/CR3 module from the preassembled MIQ/S184
module stack and mounting it separately (distributed mounting).
Materials required l1 x SNCIQ or SNCIQ/UG connecting cable
lWire end sleeves for 0.75 mm2wire cross-section with matching
crimping tool
l1 x cable gland with seal (scope of delivery of MIQ module).
Tools lCable stripping knife
lWire stripper
lPhillips screw driver
lSmall screw driver.
Note
General instructions concerning the system expansion and mounting
are given in the system operating manual.
Removing the MIQ/CR3
from the MIQ/S184
module stack and
mounting it separately
1In the case of process control via
lthe MIQ/CR3 module
lthe MIQ/IF232 module (with data server),
bring all instruments used for process control into a secure
condition (e.g. activate manual control).
2 Disassemble the MIQ/CR3 module from the IQ SENSOR NET
S184 system.
3Disassemble the contact carriers on the back of the MIQ/CR3
module.
4 Mount the MIQ/CR3 separately (distributed mounting).
5Mount the MIQ/C184 module on the MIQ/PS module.

Installation IQ SENSOR NET MIQ/T2020
3 - 4
3.2.3 Initialization
When the system is running, the terminal automatically starts to
operate as soon as contact is established to the IQ SENSOR NET. In the
first phase, the background illumination of the display switches on and
the terminal is initialized. This process can take several seconds. The
following display appears (Fig. 3-3):
Fig. 3-3 Display during the initialization process
The terminal then switches to the measured value display.

IQ SENSOR NET MIQ/T2020 Operation
4 - 1
4 Operation
4.1 Introduction
The MIQ/T2020 terminal is a control unit for the IQ SENSOR NET.
The terminal can be used to:
lDisplay measured values
lStart calibration
lPerform system and terminal settings
lDisplay messages.
Note
To guarantee data security within the IQ SENSOR NET, access to the
following functions is only possible from one terminal at the same time:
lOpening the Settings/Einstellungen menu
lStarting the calibration
lAssigning sensors or output modules as substitutes.
If one of these functions is active at another terminal, a note appears
on the display.

Operation IQ SENSOR NET MIQ/T2020
4 - 2
4.1.1 Overview of the operating elements
This section gives basic information on the operation of the IQ SENSOR
NET system via the MIQ/T2020.
The terminal is equipped with a large display for the clear presentation
of current measured values, the graph of measured values, status dis-
plays and message texts.
The IQ SENSOR NET system is operated using the 5 keys
m,c,s,e,gand the rotary switch d.
Fig. 4-1 View of the MIQ/T2020 terminal
1Display (see section 4.1.2 DISPLAY)
2Rotary switch (see section 4.1.4 ROTARY SWITCH)
35 keys m,c,s,e,g(see section 4.1.3 KEYS)
1
32
C
MSESC OK
MIQ/T2020
TERMINAL

IQ SENSOR NET MIQ/T2020 Operation
4 - 3
4.1.2 Display
The display contains the following information:
Fig. 4-2 View of the display
1Name of the display
2Name of the terminal
3Date
4Time
5Lock safety setting:
aopen (settings released)
–System settings can be changed
zclosed (settings locked)
–All system settings can only be read
6Error symbol h:
If the error symbol flashes, a new or unacknowledged error
message is present in the log book.
7Info symbol i:
If the info symbol flashes, a new or unacknowledged error
message is present in the log book.
8Number of the display
9Display range for measured values, menus, lists, etc.
10 Help line with operating instructions
2345 7
18
9
6
1
0

Operation IQ SENSOR NET MIQ/T2020
4 - 4
Note
A detailed description of the log book is given in the IQ SENSOR NET
system operating manual.
Measured value display The measured value display contains the following information for each
sensor/differential sensor:
Fig. 4-3 Display - measured value display
1Consecutive numbering of the measured values
2Main measured value
3Unit and measured variable of the main measured value
4Name of the sensor
5Secondary measured value with unit, e.g. temperature
1 2345

IQ SENSOR NET MIQ/T2020 Operation
4 - 5
Special displays
4.1.3 Keys
4.1.4 Rotary switch
Init Sensor is being initialized
lduring commissioning or
lif a new sensor is identified and this does not yet
provide any measured values
Set Sensor is being set
---- Invalid measured value
Cal Sensor is being calibrated
Error Sensor is inactive or defective
Hold Measured value is frozen
OFL Measuring range is undercut or exceeded (overflow)
Display
flashes
Sensor in maintenance condition
Key Function
mDisplays measured values
cStarts calibration
of the sensor selected in the measured value display
sOpens the Settings/Einstellungen menu
eChanges or aborts entries
in the higher menu levels without storing them
gConfirms selection
Rotary
switch Function
dHighlights and selects:
lMenu items
lEntries
lColumns or fields
lLetters or numerals

Operation IQ SENSOR NET MIQ/T2020
4 - 6
Note
The selection highlighted with the rotary switch is displayed as white
text on a black background (reverse video).
4.2 General operating principles
The operation of the IQ SENSOR NET is standardized and user-friendly.
lMake a selection with the rotary switch d
–Highlight individual elements in menus, lists and tables, e.g.
menu entries, list elements, columns or fields
–Select a setting in selection fields
–Select a character in text entry fields
lConfirm a selection with the gkey.
lBreak off an action and change to the next higher level with the e
key.
lStart a calibration procedure with the ckey.
lSwitch to the settings with the skey.
lChange to the measured value display and break off current actions
with the mkey.
Note
Short operating instructions are given in the help lines on the display.
Examples of the operating principles are given below:
lNavigation in menus, lists and tables (see chapter 4.2.1)
lEntering text and numerical values (see chapter 4.2.2)

IQ SENSOR NET MIQ/T2020 Operation
4 - 7
4.2.1 Navigating in menus, lists and tables
Fig. 4-4 100 - Settings/Einstellungen
1Open the Settings/Einstellungen menu with s.
Menus appear on the display in the form of a list, e.g. the
Settings/Einstellungen menu shown here.
2Select a menu item (e.g. System settings) with d.
Move the highlighting (reverse video display, black back-
ground) in the list of menu items by turning the rotary switch
d.
3 Confirm the menu item (e.g. System settings) with g.
The display (e.g. System settings) opens.
Confirm the selection and move to a new display by pressing
the gkey.
4Return to a higher level with e.
or:
Switch to the measured value display with m.

Operation IQ SENSOR NET MIQ/T2020
4 - 8
4.2.2 Entering text or numerical values
You can assign names to sensors, outputs, terminals and locations.
Example: Entering a sensor name:
Fig. 4-5 120 - Edit list of sensors
1Open the Settings/Einstellungen menu with s.
2 Select the Edit list of sensors menu item with d.
3Confirm the menu item (e.g.Edit list of sensors) with g.
The display (e.g.Edit list of sensors) opens. A column is high-
lighted.
4 Select the Sensor name column with d.
5Confirm the Sensor name column with g.
A sensor name is highlighted.
6 Select a sensor name with d.
7Confirm the selection with g.
The name of the selected sensor is edited.
This manual suits for next models
1