IM 12A01A02-20E 4th Edition : Oct. 01, 2015-00
SensorInputCircuit
Type of Measuring Module pH, SC and DO ISC
MaximumVoltage(Uo) 11.76V 11.76V
MaximumCurrent(Io) 116.5mA 60.6mA
MaximumPower(Po) 0.3424W 0.178W
ExternalCapacitance(Ca,Co) 100nF 100nF
ExternalInductance(La,Lo) 1.7mH 8mH
Note1: Inanysafetybarrierused,theoutputcurrentmustbelimitedbyaresistor“R”suchthatImaxout=Uz/R.
Note2: ThesafetybarriershallbeFMEntity-Approvedassociatedapparatus/barrierwhere:
BarrierVoc,Uo≤30V;
BarrierIsc,Io≤100mA;
BarrierPo≤0.75W;
BarrierCa,Co≥13nF+Ccable;
BarrierLa,Lo≥Lcable
Note3: Whenusingnonisolationbarrierconnect(*1)toISearthingsystem.
Note4: pHandSCSensor(s)areofapassivetypetoberegardedas‘simpleapparatus’sameas06ATEX0218X,
06ATEX0219,IECExKEM06.0052X,FM3028779,06ATEX0220X,06ATEX0221,IECExKEM06.0053Xortheone
individuallycertiedwithrelevantparameters.
Note5: ISCSensor(s)areISC40Sof00ATEX1067Xortheoneindividuallycertiedwithrelevantparameters.
Note6: DOSensor(s)areofapassivetypetoberegardedas‘simpleapparatus’ortheoneindividuallycertiedwith
relevantparameters.
Note7: Measuringmodule2maynotmounted.AsforISCmodule,onlyonecanbemounted.
Note8: InstallpertheNationalElectricalCode(NFPA70)
Note9: WARNING-PotentialelectrostaticcharginghazardElectrostaticchargemaycauseanexplosionhazard.Avoidany
actionsthatcausethegenerationofelectrostaticcharge,e.g.,rubbingwithadrycloth.
Note10:
AsanalternativetoinstallingtheFLXA21inDivision2usingClassI,Division2wiringmethods,theFLXA21may
beinstalledinDivision2usingnonincendiveeldwiringinaccordancewiththeNationalElectricalCode(NFPA
70)usingthesameparametersidentiedforintrinsicallysafeentityinstallations.TheAssociatedNonincendive
ApparatusshallhavenonincendiveeldwiringconnectionswhichareFMApprovedforuseintheClassI,Division2
location.
CSA
Control Drawing (for 4-20mA type)
Measuring module 1
Measuring module 2
(Refer to Note 7)
Sensor 1
Sensor 2
Housing Assembly
Supply
-
Supply
+
→
Non Hazardous Location
Hazardous Location ←
(*1)
Refer to Note
Safety Barrier
-
+
Intrinsically Safe
Group IIC, Zone 0
Class I, Division 1
Non-incendive
Class I, Division 2,
Groups A, B, C, D
2 wire analyzer
Not use safety barrier but
CSA certified equipment
use in Non-incendive
Note: The measuring module on this
drawing means the sensor module
on this General Specifications.
Electricalparameters(Intrinsicallysafe)
HousingAssembly
Supplyandoutputcircuit(terminals+and-)
Ui(Vmax)=30V,Ii(Imax)=100mA,Pi(Pmax)=0.75W,Ci=13nF,Li=0mH
Measuringmoduleinputcircuit(CN2orCN3onBackboard)
Uo(Vt,Voc)=13.65V,Io(It,Isc)=50mA,Po=0.372W,Co(Ca)=80nF,Lo(La)=7.7mH
pHmodule,SCmoduleandDOmodule
Ui(Vmax)=13.92V,Ii(Imax)=50mA,Pi(Pmax)=0.374W,Ci=40nF,Li=2.9mH
Sensorinputcircuit(terminals11through19)
Uo(Vt,Voc)=11.76V,Io(It,Isc)=116.5mA,Po=0.3424W,Co(Ca)=100nF,Lo(La)=1.7mH
ISCmodule
Ui(Vmax)=13.92V,Ii(Imax)=50mA,Pi(Pmax)=0.374W,Ci=40nF,Li=7.7mH
Sensorinputcircuit(terminals11through17)
Uo(Vt,Voc)=11.76V,Io(It,Isc)=60.6mA,Po=0.178W,Co(Ca)=100nF,Lo(La)=8mH
Installationrequirementsbetweenhousingassemblyandsafetybarrier
Uo≤UiIo≤IiPo≤PiCo≥Ci+CcableLo≥Li+Lcable
Voc≤VmaxIsc≤ImaxCa≥Ci+CcableLa≥Li+Lcable
Uo,Io,Po,Co,Lo,Voc,Isc,CaandLaareparametersofbarrier.