YOKOGAWA 701938 User manual

Use
r’s
Manual
Thank you for purchasing the 200 MHz Passive Probe 701938. To ensure correct use, please read
this manual thoroughly before beginning operation. After reading the manual, keep it in a convenient
location for quick reference whenever a question arises during operation.
Model 701938
200 MHz Passive Probe (10:1)
The instrument comes with the following manuals. Please keep them in a safe place.
Manual Title Manual No Notes
200 MHz Passive Probe (10:1)
User's Manual
IM701938-01E This manual.
Safety Instruction Manual IM 00C01C01-01Z1 Safety manual (European languages)
Inquiries PIM113-01Z2 List of worldwide contacts
Notes
• The contents of this manual are subject to change without prior notice as a result of continuing
improvements to the product’s performance and functionality. The figures given in this manual may
differ from those that actually appear on your product.
• Every effort has been made in the preparation of this manual to ensure the accuracy of its
contents. However, should you have any questions or find any errors, please contact your nearest
YOKOGAWA dealer.
• Copying or reproducing all or any part of the contents of this manual without the permission of
YOKOGAWA is strictly prohibited.
The following symbols are used in this manual.
Improper handling or use can lead to injury to the user or damage to the instrument.
This symbol appears on the instrument to indicate that the user must refer to the
user’s manual for special instructions. The same symbol appears in the corresponding
place in the user’s manual to identify those instructions. In the manual, the symbol is
used in conjunction with the word “WARNING” or “CAUTION.”
WARNING Calls attention to actions or conditions that could cause serious or fatal injury to the
user, and precautions that can be taken to prevent such occurrences.
CAUTION Calls attentions to actions or conditions that could cause light injury to the user or
damage to the instrument or the user’s data, and precautions that can be taken to
prevent such occurrences.
Note Calls attention to information that is important for proper operation of the instrument.
Safety Precautions
This product is designed to be used by a person with specialized knowledge.
Make sure to comply with the safety precautions mentioned hereafter when handling the probe.
YOKOGAWA assumes no responsibility for any consequences resulting from failure to comply with
these safety precautions. Also, read the User’s Manual of the measuring instrument thoroughly so that
you are fully aware of its specifications and handling, before starting to use the probe.
This manual is part of the product and contains important information. Store this manualin a safe place
close to the instrument so that you can refer to it immediately. Keep this manual until you dispose of
the instrument.
The following symbols are used on this instrument.
Handle with care. Refer to the user’s manual or service manual. This symbol appears on
dangerous locations on the instrument which require special instructions for proper handling
or use. The same symbol appears in the corresponding place in the manual to identify those
instructions.
Make sure to comply with the following safety precautions in order to prevent accidents
such as an electric shock which impose serious health risks to the user and damage to the
instrument.
WARNING
Grounding of the measuring instrument
The protective grounding terminal of the measuring instrument must be connected to ground.
Ground lead of the probe
Make sure to connect the ground lead of the probe to the grounding potential.
Handling of the passive probe
Do not touch the probe’s input terminal or the probe itself with wet hands.
Do not operated with suspected failures
If you suspect that there is damage to this probe, contact your nearest YOKOGAWA dealer
or sales representative.
Observe maximum working voltage
When the oscilloscope’s input coupling is AC, DC voltage of the same electric potential
as the probe’s input is applied to the oscilloscope’s input. Make sure not to exceed the
oscilloscope’s maximum input voltage.
Do not operate in wet/damp conditions
To avoid electric shock, do not operate this probe in wet or damp conditions.
Do not operate in explosive atmosphere
To aviod injury or fire hazard, do not operate this probe in an explosive atmosphere.
Do not disassemble or modify
Do not disassemble or modify the product. YOKOGAWA assumes no liability if you
disassemble or modify the product.
Avoid exposed circuitry
To avoid injury, remove jewelry such as rings, watches, and other metallic objects. Do not
touch exposed connections and components when power is present.
Damaged Signal Cable
If the signal cable is torn and the inner metal is exposed or if a color different from the outer
sheath appears, stop using the cable immediately.
CAUTION
Maximum input voltage
Do not apply any voltages exceeding the maximum input voltage to the probe.
French
AVERTISSEMENT
Mise à la terre de l’instrument de mesure
S’assurer de connecter la mise à la terre protectrice de l’instrument de mesure.
Fil de terre de la sonde
S’assurer de connecter le fil de terre de la sonde au potentiel de mise à la terre.
Manipulation de la sonde passive
Ne pas toucher le terminal d’entrée de la sonde ou la sonde elle-même avec des mains
mouillées.
Ne pas utiliser en cas de défaillances suspectées
Si vous suspectez que la sonde est endommagée, contactez votre revendeur ou
représentant commercial YOKOGAWA.
Respecter la tension d’entrée maximum
Lorsque le couplage d’entrée de l’instrument de mesure est en CA, la tension en CC du
même potentiel électrique que l’entrée de la sonde est appliquée à l’entrée de l’instrument
de mesure. S’assurer de ne pas dépasser la tension d’entrée maximum de l’instrument de
mesure.
Ne pas utiliser dans des conditions humides
Afin d’éviter un choc électrique, ne pas utiliser cette sonde dans des conditions humides.”
Ne pas utiliser dans une atmosphère explosive
Afin d’éviter des risques de blessures ou d’incendie, ne pas utiliser cette sonde dans une
atmosphère explosive.
Ne pas démonter ou modifier
Ne pas démonter ou modifier le produit. YOKOGAWA se dégage de toute responsabilité si
vous démontez ou modifiez le produit.
Éviter les circuits exposés
Afin d’éviter des blessures, retirer les bijoux comme les bagues, montres et autres objets
métalliques. Ne pas toucher les connexions et les composants exposés après mise sous
tension.
Câble de signal endommagé
Si le câble de signal est déchiré et que le métal intérieur est exposé ou si une couleur
différente de la gaine externe est visible, arrêter immédiatement d’utiliser ce câble.
ATTENTION
Tension d’entrée maximum
Ne pas appliquer à la sonde de tension dépassant la tension d’entrée maximum.
Waste Electrical and Electronic Equipment (WEEE)
(EU WEEE Directive valid only in the EEA* and UK WEEE Regulation in the UK)
This product complies with the WEEE marking requirement. This marking indicates
that you must not discard this electrical/electronic product in domestic household
waste. When disposing of products in the EEA or UK, contact your local Yokogawa
office in the EEA or UK respectively.
* EEA: European Economic Area
UKCA Marking
This product complies with the UKCA (UK Conformity Assessed) marking.
Authorized Representative in the EEA
Yokogawa Europe B.V. is the authorized representative of Yokogawa Test & Measurement Corporation
for this product in the EEA. To contact Yokogawa Europe B.V., see the separate list of worldwide
contacts, PIM 113-01Z2.
Disposal
When disposing of YOKOGWA products, follow the laws and ordinances of the country or region where
the product will be disposed of.
1 Description
701938 is a 200 MHz passive probe with probe ID pin and attenuation ratio of 10:1. This probe can
be used for oscilloscopes with input impedances of 1 MΩ.
2 Appearance
yellow
green
magenta
cyan
red
orange
light blue
purple
1
5
23
4
6
7
9
8
Trimmer
adjustment
screw
10 pcs
Standard Accessories
1
2
3
4
5
Name Part No.
Ground lead
Pinchers tip
IC test tip
Ground attachment
Marker tip
B9852BD
B9852BE
B9852BF
B9852BG

Optional Accessories
(Sold Separately)
6
7
8
9
Name Model
Mini clip adapter
BNC adapter
Printed circuit
board adapter
Solder-in adapter
700971
700972
366945
366946
3 Operation
Use adequate attachment suitable for the point to measure. Before using the probe, adjust its
capacitance by tuning the trimmer.
Adjustment
1.
Connect the probe connector to the input of the oscilloscope, and connect the tip of the probe to
the CAL signal output terminal.
2.
Change the T/Div and the V/Div to get the display shown below. And tune the trimmer to get the
correct waveform.
Over Compensation Inadequate CompensationCorrect Waveform
7th Edition: November 2021 (YMI)
All Rights Reserved. Copyright © 2008 Yokogawa Electric Corporation
All Rights Reserved. Copyright © 2012 Yokogawa Test & Measurement Corporation
Printed in Japan
IM 701938-01E 1/2
IM 701938-01E
7th Edition
WARNING
• Make sure to avoid an electric shock when connecting the probe to the object of
measurement. Do not remove the probe from the measuring instrument after the object
of measurement is connected.
• When disconnecting the probe BNC output connector, first turn OFF the power to the
circuit under measurement. Then, disconnect the probe from the high voltage parts of the
circuit under measurement.
CAUTION
Use a soft cloth to clean the dirt. Prevent damage to the probe. Avoid immersing the probe,
using abrasive cleaners, and using chemicals contains benzene or similar solvents.
French
AVERTISSEMENT
• S’assurer d’éviter un choc électrique lors de la connexion de la sonde à l’objet de la
mesure. Ne pas retirer la sonde de l’instrument de mesure après avoir connecté l’objet
de la mesure.
• Lors de la déconnexion du connecteur de sortie BNC de la sonde, mettre d’abord HORS
tension le circuit faisant l’objet de la mesure. Puis déconnecter la sonde des parties à
haute tension du circuit faisant l’objet de la mesure.
ATTENTION
Utiliser un chiffon doux pour nettoyer la sonde. Faire attention de ne pas casser la sonde.
Ne pas immerger la sonde dans un liquide ni utiliser de nettoyants abrasifs sur la sonde. Ne
pas utiliser de benzène ni d’autres solvants sur la sonde.
Note
Accurate measurement may not be possible near objects with strong electric fields (such as
cordless equipment, transformers, or circuits with large currents).
4 Specifications
Item Specifications
Probe length Approx. 1.5 m
Connector type BNC
Input resistance 10 MΩ ±2%
In conjunction with an oscilloscope with an input impedance of
1 MΩ ±1%.
Input capacitance 13.5 pF (typical)1
In conjunction with an oscilloscope with an input impedance of
1 MΩ ±1%.
Attenuation ratio 10:1 ±2%
In conjunction with an oscilloscope with an input impedance of
1 MΩ ±1%.
Frequency bandwidth (≤−3 dB ) DC to 200 MHz
Subject to change depending on type of oscilloscope used.
Rise time 1.75 ns max. (typical)1
Subject to change depending on type of oscilloscope used.
Max input voltage 2 600 V (DC + ACpeak) or 424 Vrms
Operating temperature 5 to 40°C
Operating maximum relative
humidity
80% RH at a temperature of up to 31°C, decreasing linearly to
50% RH at 40°C if the temperature is 31°C or higher.
Operating altitude 3000 m or less
Storage temperature −20 to 70°C
Storage humidity 90%RH or less (-20 to +55°C)
45%RH or less(+55 to +70°C)
Storage altitude 4600 m or less
Matching input capacitance Approx. 15 to 25 pF
Safety standards Complying standards
EN61010-031
Measurement category II 3: 600 V (DC + ACpeak)
Pollution degree 2 4
Environmental standards 5 EU RoHS Directive compliant
1 Typical (or average) value; not guaranteed.
2 The maximum allowable input decreases depending on the frequency. Refer to the derating curve.
3 This equipment is for measurement category II (CAT II). Do not use it with measurement category III (CAT
III), nor measurement category IV (CAT IV). When using devices or accessories with different measurement
categories, the lower measurement category applies. See below for definitions of measurement categories.
Measurement Category Definition
Measurement category “O
(Other)”
Measurement category O (Other) applies to measurement of a
circuit that is not connected directly to the main power source.
Measurement category II
(CAT II)
CAT II applies to measurement of electrical equipment that is
powered through a fixed installation such as a wall outlet wired to
a distribution board and measurement on such wiring.
Measurement category III
(CAT III)
CAT III applies to measurement at the distribution level, that is,
building wiring, fixed installations.
Measurement category IV
(CAT IV)
CAT IV applies to measurement at the primary supply level, that
is, overhead lines, cable systems.
4 Pollution degree applies to the degree of adhesion of a solid, liquid, or gas which deteriorates withstand
voltage or surface resistivity. Pollution degree 2 applies to normal indoor atmospheres (with only non-
conductive pollution).
5 For conformity to environmental regulations and/or standards other than EU, contact your nearest
YOKOGAWA office (PIM113-01Z2).
Input voltage derating
Frequency (Hz)
0.1 M
10
100
1000
600
1 M 10 M 100 M
1000 M
Max input voltage (V)
14.6V
IM 701938-01E 2/2

Other YOKOGAWA Measuring Instrument manuals

YOKOGAWA 701925 PBDH0500 User manual

YOKOGAWA

YOKOGAWA 701925 PBDH0500 User manual

YOKOGAWA MT300 Guide

YOKOGAWA

YOKOGAWA MT300 Guide

YOKOGAWA WT500 User manual

YOKOGAWA

YOKOGAWA WT500 User manual

YOKOGAWA AQ6376E User manual

YOKOGAWA

YOKOGAWA AQ6376E User manual

YOKOGAWA 701947 User manual

YOKOGAWA

YOKOGAWA 701947 User manual

YOKOGAWA TR User manual

YOKOGAWA

YOKOGAWA TR User manual

YOKOGAWA RAGN Manual

YOKOGAWA

YOKOGAWA RAGN Manual

YOKOGAWA 96060 User manual

YOKOGAWA

YOKOGAWA 96060 User manual

YOKOGAWA FC400G User manual

YOKOGAWA

YOKOGAWA FC400G User manual

YOKOGAWA 960 30 User manual

YOKOGAWA

YOKOGAWA 960 30 User manual

YOKOGAWA 2014 00 User manual

YOKOGAWA

YOKOGAWA 2014 00 User manual

YOKOGAWA IR800G User manual

YOKOGAWA

YOKOGAWA IR800G User manual

YOKOGAWA AQ2160-02 User manual

YOKOGAWA

YOKOGAWA AQ2160-02 User manual

YOKOGAWA CW500 User manual

YOKOGAWA

YOKOGAWA CW500 User manual

YOKOGAWA WT1803 User manual

YOKOGAWA

YOKOGAWA WT1803 User manual

YOKOGAWA PH71 User manual

YOKOGAWA

YOKOGAWA PH71 User manual

YOKOGAWA UV700G User manual

YOKOGAWA

YOKOGAWA UV700G User manual

YOKOGAWA 51011 User manual

YOKOGAWA

YOKOGAWA 51011 User manual

YOKOGAWA UPM101 User manual

YOKOGAWA

YOKOGAWA UPM101 User manual

YOKOGAWA MV100 User manual

YOKOGAWA

YOKOGAWA MV100 User manual

YOKOGAWA PR300 User manual

YOKOGAWA

YOKOGAWA PR300 User manual

YOKOGAWA AQ6376 User manual

YOKOGAWA

YOKOGAWA AQ6376 User manual

YOKOGAWA AQ6370C User manual

YOKOGAWA

YOKOGAWA AQ6370C User manual

YOKOGAWA WT1801E User manual

YOKOGAWA

YOKOGAWA WT1801E User manual

Popular Measuring Instrument manuals by other brands

Panasonic LP-V10U-C Specification sheet

Panasonic

Panasonic LP-V10U-C Specification sheet

BIRD IQC5000B Operation manual

BIRD

BIRD IQC5000B Operation manual

TESTO 445 instruction manual

TESTO

TESTO 445 instruction manual

Seitron PRESSOTEST 200 manual

Seitron

Seitron PRESSOTEST 200 manual

Hanna Instruments HI 993310 instruction manual

Hanna Instruments

Hanna Instruments HI 993310 instruction manual

Michelin 12290 Operation instructions

Michelin

Michelin 12290 Operation instructions

Endress+Hauser Proline Promag 55 PROFIBUS PA operating instructions

Endress+Hauser

Endress+Hauser Proline Promag 55 PROFIBUS PA operating instructions

Hanna Instruments HI88703-02 instruction manual

Hanna Instruments

Hanna Instruments HI88703-02 instruction manual

THORLABS SA210 Series user guide

THORLABS

THORLABS SA210 Series user guide

PulseTech 390PT operating instructions

PulseTech

PulseTech 390PT operating instructions

VXI VM2601 user manual

VXI

VXI VM2601 user manual

MULTISPAN PI 38 manual

MULTISPAN

MULTISPAN PI 38 manual

MRC BCT-4308 Operation manual

MRC

MRC BCT-4308 Operation manual

Kroeplin C0 Series user manual

Kroeplin

Kroeplin C0 Series user manual

Electrex FEMTO 96 installation guide

Electrex

Electrex FEMTO 96 installation guide

Bruker D8 Discover Standard operating procedure

Bruker

Bruker D8 Discover Standard operating procedure

TIMEDOX Silver Plus quick start guide

TIMEDOX

TIMEDOX Silver Plus quick start guide

METREL EurotestIM instruction manual

METREL

METREL EurotestIM instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.