IM 01F07A03-03EN
4
<2. Explosion Protection Type> 4
• Power Supply:
(Integral Flowmeter and Remote Transmitter)
10.5 to 42 V DC (Communication and I/O
Code: JA or JB)
• Current I/O:
(Integral Flowmeter and Remote Transmitter)
Output: 3.6 to 21.6 mA
Input: ≤ 21.6 mA
• Pulse Output:
(Integral Flowmeter and Remote Transmitter)
Output: ≤ 42 V DC, ≤ 120 mA
• Dielectric Strength:
(Remote Transmitter)
1500 V AC r.m.s., 1 min, 5 mA
Terminals: SUPPLY+, SUPPLY−, DOUT+,
DOUT−, AIN+ and AIN− to
BROWN, RED, ORANGE,
YELLOW, GREEN and BLUE
500 V AC r.m.s., 1 min, 5 mA
Terminals: BROWN, RED, ORANGE,
YELLOW, GREEN and BLUE to
Earth terminal
• Control Drawing: Read Section 2.3.
*1: The ambient temperature for the process temperature
under −50°C is shown as Figure 1.
−50
−50
Process temperature °C
Ambient temperature °C
−196
−20
Operating range
Figure 1 The ambient temperature for the process
temperature under −50°C
Intrinsically safe (Certication Code: KS2,
Communication and I/O Code: JA or JB)
• Type of Gas Atmosphere Protection:
(Integral Flowmeter)
II 1 G Ex ia IIC T4...T1 Ga
(Remote Sensor)
II 1 G Ex ia IIC T6...T1 Ga
(Remote Transmitter)
II 1 G Ex ia IIC T4 Ga
• Ambient Temperature: *2
(Integral Flowmeter)
−55°C ≤ Ta ≤ +60°C
(Remote Sensor)
T6, T80°C:−55°C ≤ Ta ≤ +40°C
T5, T95°C to T1, T440°C:−55°C ≤ Ta ≤ +60°C
(Remote Transmitter)
−55°C ≤ Ta ≤ +80°C
• Process Temperature: *2
(Integral Flowmeter and Remote Sensor)
T6: −196°C to +80°C / [+78°C]
T5: −196°C to +95°C / [+93°C]
T4: −196°C to +130°C / [+128°C]
T3: −196°C to +195°C / [+193°C]
T2: −196°C to +290°C / [+288°C]
T1: −196°C to +440°C / [+438°C]
[ ] : Built-in temperature sensor
• Atmospheric Pressure: 80 kPa to 110 kPa
• Enclosure:
IP66/IP67 in accordance with only EN 60529
• Pollution Degree: 2
• Overvoltage Category: I
• Power Supply:
(Integral Flowmeter and Remote Transmitter)
10.5 to 30 V DC
• Current I/O:
(Integral Flowmeter and Remote Transmitter)
Output: 3.6 to 21.6 mA
Input: ≤ 21.6 mA
• Pulse Output:
(Integral Flowmeter and Remote Transmitter)
Output: ≤ 30 V DC, ≤ 80 mA
• Dielectric Strength:
(Integral Flowmeter)
500 V AC r.m.s., 1 min, 5 mA
Terminals: SUPPLY+, SUPPLY−, DOUT+,
DOUT−, AIN+ and AIN− to Earth
terminal
Terminals: SUPPLY+ and SUPPLY− to
DOUT+, DOUT−, AIN+ and AIN−
Terminals: DOUT+, and DOUT− to AIN+ and
AIN−
(Remote Transmitter)
500 V AC r.m.s., 1 min, 5 mA
Terminals: SUPPLY+, SUPPLY− to DOUT+,
DOUT−, AIN+ and AIN−
Terminals: DOUT+ and DOUT− to AIN+ and
AIN−
Terminals: SUPPLY+, SUPPLY−, DOUT+,
DOUT−, AIN+, AIN−, BROWN, RED,
ORANGE, YELLOW, GREEN and
BLUE to Earth terminal
Terminals: SUPPLY+, SUPPLY−, DOUT+,
DOUT−, AIN+ and AIN− to BROWN,
RED, ORANGE, YELLOW,
GREEN and BLUE
• Control Drawing: Read Section 2.3.
*2: The ambient temperature for the process temperature
under −55°C is shown as Figure 2.