<2. SPECIFICATIONS>
2-4
IM 11G02N01-01E
Auto calibration starts by opening the auto-calibration-remote-start input terminals after short-
circuiting them for 1.5 seconds or longer.
Auto zero calibration:
Auto zero calibration is carried out periodically at the preset cycle.
This cycle is independent of “Auto calibration” cycle.
Whenzerocalibrationgasandsolenoidvalveforopening/closingthecalibrationgasowlineare
prepared externally by the customer, the zero calibration will be carried out at the preset auto zero
calibrationtimingwiththesolenoidvalvedrivecontact(forzerocalibration)turnedon/o
.
Auto zero calibration cycle setting;
Auto zero calibration cycle is set.
Setting is variable within 1 to 99 hours (in increments of 1 hour) or 1 to 40 days (in
increments of 1 day)
Gasowtimesetting;
Thetimeforowingzerogasinautozerocalibrationisset.
Settable 60 to 900 seconds (in increments of 1 second)
Upper/lower limit alarm:
Alarm contact output turns on when measurement value reaches the preset upper or lower
limit of alarm value.
Contacts close when the instantaneous value of each component becomes larger than the
upper alarm limit value or smaller than the lower alarm limit value.
Instrument error contact output:
Contacts close at an occurrence of analyzer error No. 1, 3 or 10.
Calibration error contact output:
Contacts close at occurrence of manual or auto calibration error (any of errors No. 4 to 9).
Auto calibration status contact output:
Contacts close during auto calibration.
Pump ON/OFF contact output:
Duringmeasurement,contactsclose.Whilecalibrationgasisowing,contactsopen.
Contacts are connected in power supply of pump, and stop the sample gas while calibration
gasisowing.
OptionalFunctions(availablewithspecifyingoptionalcode)
O2correction:(-/K)
Conversion of measured NO, SO2and CO gas concentrations into values at reference O2
concentration.
Correction formula: C = 21 – On
21 – Osx Cs
Where:
C: Sample gas concentration after O2correction
Cs:
Measured concentration of sample gas
Os:Measured O2concentration
(limit setting: 1 to 20%O2, default 17% )
On:Reference O2concentration (value changeable by setting: 0 to 19%O2, default 4%)
The result of calculation is indicated and output in an analog output signal.
Average value after O2correction and O2average value calculation (-/K):
The result of O2correction or instantaneous O2value can be outputted as an average value in
the determined period of time.
Used for averaging is the moving average method in which sampling is carried out at intervals
of 30 seconds.
(Output is updated every 30 seconds. It is the average value in the determined period of time
just before the latest updating.)
Averaging time is settable within 1 to 59 minutes (in increments of 1 minute) or 1 to 4 hours (in
increments of 1 hour).
Average value reset (-/K):
The above-mentioned output of average value starts from the initial state by opening the
average-value-resetting input terminals after short-circuiting them for 1.5 seconds or longer.
Output is reset by short-circuiting and restarted by opening.
CO concentration peak count alarm (-/A):
(available only for CO + O2measurement)
Alarm output turns on according to the preset concentration and count.
Whenever the instantaneous value of CO exceeds the preset concentration value, count