manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Zico
  6. •
  7. Measuring Instrument
  8. •
  9. Zico ZI-9940 User manual

Zico ZI-9940 User manual

OPERATING INSTRUCTION
ZI-9940
200A AC CLAMP METER
Safety
International Safety Symbols
This symbol, adjacent to another symbol or terminal, indicates the user must refer to
the manual for further information.
This symbol, adjacent to a terminal, indicates that, under normal use, hazardous
voltages may be present
Double insulation
SAFETY NOTES
Do not exceed the maximum allowable input range of any function
Do not apply voltage to meter when resistance function is selected.
Set the function switch OFF when the meter is not in use.
WARNINGS
Set function switch to the appropriate position before measuring.
When measuring volts do not switch to current/resistance modes.
When changing ranges using the selector switch always disconnect the test leads from the
circuit under test.
Do not exceed the maximum rated input limits.
CAUTIONS
Improper use of this meter can cause damage, shock, injury or death. Read and understand
this user manual before operating the meter.
Always remove the test leads before replacing the battery.
Inspect the condition of the test leads and the meter itself for any damage before operating the
meter. Repair or replace any damage before use.
Use great care when making measurements if the voltages are greater than 25VAC rms or
35VDC. These voltages are considered a shock hazard.
Remove the battery if the meter is to be stored for long periods.
Always discharge capacitors and remove power from the device under test before performing
Diode, Resistance or Continuity tests.
Voltage checks on electrical outlets can be difficult and misleading because of the uncertainty
of connection to the recessed electrical contacts. Other means should be used to ensure that
the terminals are not "live".
If the equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
Input Limits
Function
Maximum Input
A AC
200A
V DC, V AC
600V DC/AC
Resistance, Diode, Continuity Test
600V DC/AC
Meter Description
1. Current clamp
2. Clamp trigger
3. Rotary Function swith
4. LCD display
5. MAX Hold/ Backlight button
6. Mode select button
7. COM input jack
8. V Ωjack
9. Battery Cover
1. AC DC AC (alternating current) and DC (direct currrent)
2. Minus sign
3. 1.8.8.8 2000 count (0 to 1999) measurement reading
4. AUTO AutoRange mode
5. MAX MAX Hold mode
6. Diode test mode
7. •))) Audible Continuity
8. HOLD Data Hold mode
9. BAT Low Battery icon
10. m,V,A,K,M,ΩUnits of measure list
Specifications
Function
Range & Resolution
Accuracy (% of reading)
AC Current
200.0 AAC
± (2.5% + 10 digits)
DC Voltage
200.0 mVDC
± (0.5% + 5 digits)
2.000 VDC
± (1.2% + 3 digits)
20.00 VDC
200.0 VDC
600 VDC
± (1.5% + 3 digits)
AC Voltage
2.000 VAC
± (1.5% + 3 digits)
20.00 VAC
200.0 VAC
600 VAC
± (2.0% + 4 digits)
Resistance
200.0 
± (1.0% + 4 digits)
2.000K
± (1.5% + 2 digits)
20.00K
200.0K
2.000M
± (2.0% + 3 digits)
20.00M
± (3.0% + 5 digits)
Clamp size Opening 0.7" (18mm) approx
Diode Test Test current of 0.3mA typical; Open circuit voltage 1.5V DC
typical.
Continuity Check Threshold <120; Test current < 1mA
Low Battery Indication “ BAT” is displayed
Overrange Indication “OL “ is displayed
Measurements Rate 2 per second, nominal
Input Impedance 7.8M(VDC and VAC)
kM
BAT
Display 3-1/2 digits (2000 counts) LCD
AC Current bandwidth 50/60Hz (AAC)
AC Voltage bandwidth 50/400Hz (VAC)
Operating Temperature 14 to 122oF (-10 to 50oC)
Storage Temperature -14 to 140oF (-30 to 60oC)
Relative Humidity 90%(0oC to 30oC); 75%(30oC to 40oC); 45%(40oC to 50oC)
Altitude Operating: 3000m; Storage 10,000m
Over voltage Category II 600V/ Category III 300V
Battery Two 1.5V “AAA” Batteries
Auto OFF approx. 15 minutes
Dimensions/Weight 164x65x32mm/175g
Safety For indoor use and in accordance with Overvoltage Category II,
Pollution Degree 2. Category II includes local level, appliance,
portable equipment, etc., with transient overvoltages less than
Overvoltage Cat. III
Operation
NOTICES: Read and understand all warning and precaution statements listed in the safety
section of this operation manual prior to using this meter. Set the function select
switch to the OFF position when the meter is not in use.
AC Current Measurements
WARNING: Ensure that the test leads are YES
disconnected from the meter before making current NO
clamp measurements.
Set the Function switch to the 200AAC range.
If the range of the measured is not known,
select the higher range first then move to the lower
range if necessary.
1. Press the trigger to open jaw. Fully enclose one conductor
to be measured.
2. The clamp meter LCD will display the reading.
AC/DC Voltage Measurements
1. Insert the black test lead into the negative COM terminal and the
red test lead into the positive Vterminal.
2. Set the function switch to the V position.
3. Select AC or DC with the MODE button.
4. Connect the test leads in parallel to the circuit under test.
5. Read the voltage measurement on the LCD display.
Resistance and Continuity Measurements
1. Insert the black test lead into the negative COM terminal and the
red test lead into the positive terminal.
2. Set the function switch to the •))) position.
3. Use the multifunction MODE button to select resistance.
4. Touch the test probe tips across the circuit or component under
test. It is best to disconnect one side of the device under test so
the rest of the circuit will not interfere with the resistance reading.
5. For Resistance tests, read the resistance on the LCD display.
6. For Continuity tests, if the resistance is < 120, a tone will sound.
Diode Measurements
1. Insert the black test lead banana plug into the negative COM jack and the red test
lead banana plug into the positive diode jack.
2. Turn the rotary switch to the •))) position.
3. Press the MODE button until “ “ appears in the display.
4. Touch the test probes to the diode under test. Forward voltage will indicate 0.4V to
0.7V. Reverse voltage will indicate “OL”. Shorted devices will indicate near 0mV
and an open device will indicate “OL” in both polarities.
MAX Hold
To hold the highest reading on the LCD, press the MAX hold button. The MAX hold button is
located on the left side of the meter (bottom button). The meter reading will not change as
readings change, rather it will only display the highest reading encountered since the MAX hold
button was pressed. Press the MAX hold button again to return to normal operation.
Backlight
The backlight function illuminates the display and is used when the ambient light to too low to
permit viewing of the displayed readings. Press the button for one second to turn the
backlight on and press the button a second time to turn the backlight off.
Battery Replacement
1. Remove the one rear Phillips head screw
2. Open the battery compartment
3. Replace the Requires two “AAA”batteries (UM4 R03)
4. Re-assemble the meter
www.zicotech.com
Red Black Black Red
Probe Probe Probe Probe
Forward test Reverse test

Other Zico Measuring Instrument manuals

Zico ZI-2090 User manual

Zico

Zico ZI-2090 User manual

Zico ZI-9943 User manual

Zico

Zico ZI-9943 User manual

Zico ZI-6550 User manual

Zico

Zico ZI-6550 User manual

Zico ZI-9940 User manual

Zico

Zico ZI-9940 User manual

Zico ZI-7825 User manual

Zico

Zico ZI-7825 User manual

Zico ZI-9981 User manual

Zico

Zico ZI-9981 User manual

Zico ZI-9941 User manual

Zico

Zico ZI-9941 User manual

Zico ZI-9100 User manual

Zico

Zico ZI-9100 User manual

Zico ZI-7820 User manual

Zico

Zico ZI-7820 User manual

Zico ZI-9980 User manual

Zico

Zico ZI-9980 User manual

Zico ZI-9983 User manual

Zico

Zico ZI-9983 User manual

Zico ZI-6940 User manual

Zico

Zico ZI-6940 User manual

Popular Measuring Instrument manuals by other brands

YOKOGAWA PX8000 user manual

YOKOGAWA

YOKOGAWA PX8000 user manual

MEASUREMENT TECHNOLOGY LCI-90 installation manual

MEASUREMENT TECHNOLOGY

MEASUREMENT TECHNOLOGY LCI-90 installation manual

baxter ML-132580 Operator's & installation manual

baxter

baxter ML-132580 Operator's & installation manual

Piusi SUZZARA BLUE DEF METER owner's manual

Piusi

Piusi SUZZARA BLUE DEF METER owner's manual

Bently Nevada 3300/40 Operation manual

Bently Nevada

Bently Nevada 3300/40 Operation manual

Xylem YSI MultiLab 4010P-1 Operation manual

Xylem

Xylem YSI MultiLab 4010P-1 Operation manual

General DM8200 user manual

General

General DM8200 user manual

Universal Laser Systems ULS-25PS Operation manual

Universal Laser Systems

Universal Laser Systems ULS-25PS Operation manual

Pyxis ST-525 user manual

Pyxis

Pyxis ST-525 user manual

Kimans HYDRASONIC S8 user manual

Kimans

Kimans HYDRASONIC S8 user manual

Camille Bauer SINEAX DM5000 Device handbook

Camille Bauer

Camille Bauer SINEAX DM5000 Device handbook

Sequoia LISST-200X user manual

Sequoia

Sequoia LISST-200X user manual

Novalynx 110-WS-25P quick start

Novalynx

Novalynx 110-WS-25P quick start

Ludlum Measurements 3100 manual

Ludlum Measurements

Ludlum Measurements 3100 manual

Perfect 10 Satellite Distributing Birdog manual

Perfect 10 Satellite Distributing

Perfect 10 Satellite Distributing Birdog manual

Salutron ComBi D3 user manual

Salutron

Salutron ComBi D3 user manual

McCrometer FlowConnect FC500 instructions

McCrometer

McCrometer FlowConnect FC500 instructions

SONOPAN LMC-10 instruction manual

SONOPAN

SONOPAN LMC-10 instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.