Zico DT-5500 User manual

1 2
DIGITAL INSULATION TESTER
INSTRUCTION MANUAL
SAFETY INFORMATION
Read the following safety information carefully before
attempting to operate or service the meter.
To avoid damages to the instrument do not apply the
signals which exceed the maximum limits shown in the
technical specifications tables.
Do not use the meter or test leads if they look damaged.
Use extreme caution when working around bare
conductors or bus bars.
Accidental contact with the conductor could result in
electric shock.
Use the meter only as specified in this manual;
otherwise, the protection provided by the meter may be
impaired.
Read the operating instructions before use and follow
all safety Information.
Caution when working with voltages above 60V DC or
30V AC RMS. Such voltages pose a shock hazard.
Before taking resistance measurements or testing
acoustic continuity, disconnect circuit from main power
supply and all loads from the circuit.

3 4
Safety symbols:
Caution refer to this manual before using the meter.
Dangerous voltages.
Meter is protected throughout by double insulation or
reinforced insulation.
When servicing, use only specified replacement parts.
CE Comply with EN-61010-1
1. SPECIFICATIONS
1-1 General Information
Environment conditions:
①Installation Categories II
②Pollution Degree 2
③Altitude up to 2000 meters
④Indoor use only
⑤Relatively humidity 80% max.
⑥Operation Ambient 0~40ºC
Maintenance & Clearing:
①Repairs or servicing not covered in this manual should
only be performed by qualified personnel.
②Periodically wipe the case with a dry cloth. Do not use
abrasives or solvents on this instruments.
Display: Large LCD with dual display
Measurement Range: 200Ω, 200kΩ, 200MΩ/100V,
200MΩ/100V, 2000MΩ/100V,
750V/ACV, 1000V/DCV.
Sampling Rate: 2.5 times per second.
Zero Adjustment: Automatic adjustment.
Over Range Indicator: Number 1 of highest digit is
displayed.
Low Battery Indication: The is displayed when the
battery Voltage drop below the
operating voltage.
Operating Temperature: 0ºC to 40ºC (32ºF to 104ºF) and
Humidity below 80% RH
Storage Temperature: -10ºC to 60ºC (14ºF to 140ºF) and
Humidity below 70% RH
Power source: DC9V (6x1.5V Size “AA” battery or Equivalent)
Dimensions: 200(L) x 92(W) x 50(H) mm
Weight: Approx 700g include battery
Accessories: Test leads, 6pcs battery, Carrying case, manual.

5 6
1-2 Electrical Specifications
Accuracies are specified in the way:
±(…% of reading +…digits) at 23ºC±5ºC,below 80% RH.
OHMS
Range Resolution Accuracy Max. open
Circuit
Voltage
Overload
Protection
200Ω
0.1Ω
+(1%+2)
4.5V
250Vrms
200kΩ0.1kΩ3.0V
Continuity Beeper
Range Resolution Operation
Resistance
Max.open
Circuit
Voltage
Overload
Protection
•))) 0.1ΩResistance≤40Ω4.5V 250Vrms
Short circuit current
≤200mA
DC Voltage
Range Resolution Accuracy Input
Impedance
Overload
Protection
1000V 1V +(0.8%+3) 10MΩ1000Vrms
AC Voltage (40Hz~400Hz)
Range Resolution Accuracy Input
Impedance
Overload
Protection
750V 1V +(1.2%+10) 10MΩ750Vrms
Meg OHMS
Range
Resolution
Accuracy
200MΩ/100V
0.1MΩ
+(3.5%+5)
200MΩ/100V
0.1MΩ
0~1000MΩ/100V
1MΩ
1000~2000MΩ
/100V
+(5%+5)

7 8
2. PARTS & CONTROLS
1Digital Display
2Data Hold Button
3Lock Button
4Backlight Button
5Test Button
6Rotary Function switch
7VΩJack
8COM input jack
9Pothook
10 Battery Cover
3. BATTERY REPLACEMENT
3-1 How to connect test leads.
a). On MΩRange: Connect the red test lead into the “VΩ”
terminal and the black lead into the “COM” terminal.
b) On 200Ωand ACV Range: Connect the red test lead into
the “VΩ” terminal and the black lead into terminal “COM”
3-2 Battery Check-UP & Replacement
a) As battery power is not sufficient. LCD will display .
Replacement of 6 pcs new batteries, type 1.5V size “AA”
is Required.
b). Place back the battery cover and four the screws.
3-3 Test leads check
Set the range select switch to the 200Ωrange. With the tip
and alligator clip of the test leads connected. The indicator
should read 00.0Ω. When the leads are not connected the
display will read infinity indicated by “1”. This will ensure that
test lead are under working condition.
4. INSULATION RESISTANCE
MEASUREMENTS
a). Measurements at 200MΩ/100V This is the voltage used
for the majority of insulation resistance tests on normal
installation requirement. To measure insulation resistance,
press the test button to power on the tester. The LCD will

9 10
displayed the insulation resistance. Section VII indicated
that subdivision of large installations might be necessary
because of the large number of parallel insulation
resistance. In such a case, an installation may be divided
into sections, each being separately tested. Each section
must have not less than fifty outlets, an outlet being a
switch, socket, lighting point etc. A switched socket
counts as one outlet. The minimum acceptable insulation
resistance is 1MΩ. For a large installation, the
capacitance of the insulation will be high, and it will take
longer for it to become charged by the direct testing
voltage. Care must be taken not to take a reading until
there is a steady reading, indicating that the charging
process is complete.
Note: The charge stored in the insulation will be discharged
automatically when the test button is released. Be
careful not to turn the range switch knob whilst the test
button is pressed, or the instrument will be damaged.
b). Measurements at 2000MΩ/100V
Some specifications require testing at 100V. This voltage
must also be selected where the supply voltage of the
installation is between 100V. First, set the range switch to
100V and then proceed as indicated in a above for 100V
testing. The above note also applies to testing at 100V. In
addition the following applies.
Note: Make sure that the circuit under does not include
components Which will be damaged by the 100V applied.
Many normal components of an installation are likely to
be damaged if tested at 100V. Examples are power factor
correction capacitors, low voltage mineral insulated
cables, electronic light dimmers, electronic ballasts and
starters for fluorescent lamps etc…
c). Lock power on Feature
For hands free operation a lock power on feature is
incorporated on the press to test button. Set LOCK button
to lock test voltage, Pressing it again will tester off.
5. LOW RESISTANCE (CONTINUITY)
MEASUREMENTS
a). Set the range switch to 200Ω∙))) Position
b). Connect the red test lead to the V Ωterminal and black to
the COM terminal.
c). Connect the tips of the test leads to both ends of the
circuit under test. read resistance in Ωon the LCD.
d). When the impedance on circuit is below approximately
40Ω.It will indicate by a continuous beeper.
6. AC/DC VOLTAGE MEASUREMENTS
a). Set the range switch to ACV or DCV position
b). Connect red test lead to “V Ω” terminal and black test lead
to terminal “COM”.
c). Connect test prods of test leads IN PARALLEL to the
circuit being measured.

11 12
d). Read the voltage value on LCD.
7. POWER TOOLS AND AMALL
APPLIANCES
This test would also apply to other similar equipment that has
a line cord. For double insulated power tools, the
megohmmeter lead shown connected to the housing would
be connected to some metal part lf the tool(e..g chuck,blade).
Note:The switch of the device must be in the “ON”position
and the main power should be disconnected.
MOTORS
AC-Disconnect the motor from the line by disconnecting the
wires at the motor terminals or by opening the main switch.
If the main switch is used and the motor also has a starter
then the starter must be held, by some means, in the “ON”
position. In the latter case, the measured resistance will
include the resistance of the motor, wire and all other
components between the motor and the main switch. If a
weakness is in dicated, the motor and other components
should be checked individually. If the motor is disconnected
at the motor terminals, connect one megohmmeter lead to
the grounded motor housing and the other lead to One of the
motor leads. DC-Disconnect the motor from the line. To test
the brush rigging,field coils and armature connect one
megohmmeter lead to the grounded motor housing and the
other lead to the brush on the commutator. If the resistance
measurement indicates a weakness,raise the brushes off the
commutator and separately test the armature, field coils and
brush rigging by connecting one megohmmeter lead to each
of them individually, leaving the other connected to the
grounded motor housing. The above also applies to DC
Gemerators.

13
CABLES
Disconnect the cable from the line. Also disconnect opposite
end to avoid errors due to leakage from other equipment.
Check each conductor to ground and /or lead sheath by
connecting one megohmmeter lead to a ground and /or lead
sheather and the other megohmmeter lead to each of the
conductors in turn. Check insulation resistance between
conductors by connecting megohmmeter leads to conductors
in pairs.
Table of contents
Other Zico Test Equipment manuals
Popular Test Equipment manuals by other brands

Jadak
Jadak PHOTO RESEARCH A-TAKT V-7HS user manual

CCTVMATE.COM
CCTVMATE.COM LCD330 Instruction book

Cirris
Cirris 4200 Series user manual

Huazheng Electric Manufacturing
Huazheng Electric Manufacturing HZ-3140A manual

EZ Digital
EZ Digital OS-3060D Operation manual

ALIGNER
ALIGNER CCD2.0WIFI Installation and manual

The Diagnostic Box
The Diagnostic Box TDB760 operating manual

Wolfgang Warmbier
Wolfgang Warmbier PGT130.DT user manual

MSA
MSA FlameGard 5 Test Lamp instruction manual

Pratt Safety Systems
Pratt Safety Systems SETESTKIT instruction manual

American DJ
American DJ Dr. DMX User instructions

Amprobe
Amprobe MEGATEST 5000 owner's manual