
EQUIPMENT REQUIRED:
A.T.S. 1200 calibration hose (supplied).
Calibrated 0 to 500 mm Hg pressure meter.
Adjustable 0 to 500 mm Hg pressure source.
CAUTION: The following steps must be taken in the exact
order to calibrate the unit. Failure to do so may result in
incorrect pressure readings while the unit is in operation.
MAIN CUFF CALIBRATION
The Main cuff transducer is always calibrated first.
Below is a step-by-step procedure for calibrating the
Main cuff transducer. The procedure continues with the
Second cuff. Both cuff transducers should always be
calibrated. The calibration will not be complete unless
both cuff transducer calibrations are performed.
1. To enter the calibration mode, press and hold the MAIN
CUFF INFLATE and DEFLATE buttons while powering
the unit ON. The unit will enter the calibration mode
when the PRESSURE and TIME display momentarily
displays “CAL” “MODE”. After the “CAL” “MODE”
is displayed, the unit will display the software revision
level. The software revision level can be recorded for
future reference.
2. After displaying the software revision level, the unit will
display “MAIN” “CUFF” in the PRESSURE and TIME
displays respectively. This is to indicate that the unit is
now ready to calibrate the Main cuff transducer.
3. The unit will now display “0” in the PRESSURE display
and alternating “CAL” and “0” in the TIME display.
Throughout this procedure, the TIME display will
indicate the pressure in which the user is calibrating.
4. For zero, allow the port to be open to atmospheric
pressure so the unit can sense the zero point (i.e. when
setting the zero point, nothing should be connected to
the cuff port). By pressing the PRESSURE button, the
unit will calibrate the pressure. The unit will beep to let
the user know the set point was taken.
5. Connect the calibration hose, calibrated pressure meter
and adjustable pressure source to the Main Cuff port.
See Figure 3 for more details.
6. Once the zero point is calibrated, press the MAIN CUFF
INFLATE button to advance the unit to the next pressure
level. The unit’s TIME display will now be alternating
between “CAL” and “50”. Apply 50 mm Hg to the unit’s
Main cuff port. Once the pressure has stabilized,
press the PRESSURE button so the unit can calibrate the
50 mm Hg point. The unit will beep to let the user
knowthe set point was taken.
7. Once the 50 mm Hg point is calibrated, press the MAIN
CUFF INFLATE button to advance the unit to the next
pressure level. The unit’s TIME display will now be
alternating between “CAL” and “250”. Apply 250 mm Hg
to the unit’s Main cuff port. Once the pressure has
stabilized, press the PRESSURE button so the unit can
calibrate the 250 mm Hg point. The unit will beep to let
the user know the set point was taken.
8. Once the 250 mm Hg point is calibrated, press the MAIN
CUFF INFLATE button to advance the unit to the next
pressure level. The unit’s TIME display will now be
alternating between “CAL” and “475”. Apply 475 mm Hg
to the unit’s Main cuff port. Once the pressure has
stabilized, press the PRESSURE button so the unit can
calibrate the 475 mm Hg point. The unit will beep to let
the user know the set point was taken.
9. At this point, use the MAIN CUFF DEFLATE button
to back the pressure level down to the point where the
TIME display is alternating between “CAL” and “0”.
Then, press the MAIN CUFF INFLATE and the
MAIN CUFF DEFLATE buttons simultaneously
to exit the Main cuff calibration and to enter the
Second cuff calibration.
SECOND CUFF CALIBRATION
The Second cuff transducer is always calibrated last.
Below is a step-by-step procedure for calibrating the
Second cufftransducer. The Second cuff transducer
calibration is performed the same way the Main cuff
transducer calibration. Following the completion of the
Second cufftransducer calibration, be certain to perform
the techniques described to save the calibration in non-
volatile memory.
10. Once the unit enters the Second cuff transducer
calibration, the unit will beep and momentarily display
“SCND” “CUFF” in the PRESSURE and TIME window
respectively. This is to indicate that the unit is now ready
to calibrate the Second cuff transducer.
11. The unit will now display “0” in the PRESSURE display
and alternating “CAL” and “0” in the TIME display.
Throughout this procedure, the TIME display will
indicate the pressure in which the user is calibrating.
12. For zero, allow the port to be open to atmospheric
pressure so the unit can sense the zero point (i.e. when
setting the zero point, nothing should be connected to
the cuff port). By pressing the PRESSURE button, the
unit will calibrate the pressure. The unit will beep to let
the user know the set point was taken.
13. Connect the calibration hose, calibrated pressure meter
and adjustable pressure source to the Second Cuff port.
See Figure 3 for more details.
14. Once the zero point is calibrated, press the SECOND
CUFF INFLATE button to advance the unit to the next
pressure level. The unit’s TIME display will now be
alternating between “CAL” and “50”. Apply 50 mm Hg
to the unit’s Second cuff port. Once the pressure has
stabilized, press the PRESSURE button so the unit can
calibrate the 50 mm Hg point. The unit will beep to let
the user know the set point was taken.
16
A.T.S. 1200 TOURNIQUET SYSTEM