Prior to using the device on a patient, the user should become familiar
instructions for use and individual examination methods to be used as well as the
indications/contraindications, warnings and application information. Additional
sources of information about the application should be followed.
All patient inlets are protected against defibrillation voltages.
ECG electrodes must be applied at a distance of at least 15 cm from defibrillator
electrodes.
There is no hazard if the CardioPort Four is used on cardiac pacemaker patients
or if other stimulation devices are also used. However a certain degree of
caution should be used. The stimulation devices should only be used at an
adequate distance from the lead electrodes. In cases of doubt, the patient
should be disconnected from the recording device.
Electrostimulation devices and/or cardiac pacemakers can cause the ECG signal
to be distorted.
For example, overloading the ECG amplifier during or directly after a
defibrillation is shown by red lines instead of the ECG waves.
The heartbeat and heart rate display can be impaired by the pulse from a
cardiac pacemaker or by atrial fibrillation.
is not intended for use during HF surgery.
is not suitable for direct application on the heart.
Prior to the examination, patients should be informed of the objectives and course
the examination as well as their cooperative participation.
The power supply of the computer must correspond to the international device
standard IEC 60950 or European standard EN 60950.
Prior to use, it should be ensured that the necessary isolation barriers between
CardioPort Four and the mains as well as any connected networks are in place.
If the computer to which the CardioPort Four is connected via a USB cable is not
a medical PC and does not have any equivalent isolation, a USB isolator with an
isolation barrier of at least 1.5 kV must be used between the CardioPort Four
and the computer.
If the computer to which the CardioPort Four is connected via a USB cable is
connected to a network, there must be an insulation barrier of at least 1.5 kV
between the CardioPort Four and the network.
If CardioPort Four is wirelessly connected to the computer, the wireless
connection is the necessary isolation barrier.
Magnetic and electrical fields can affect the function of the device.
do not operate CardioPort Four in the vicinity of devices which generate strong
electromagnetic fields (X-ray or diathermy equipment, MRI machines, transformers,
motors). Please keep a safe distance of several meters.
is not suitable for use in areas with an explosive, flammable or
combustive atmosphere.
To avoid the risk of electric shock, the USB connection must be disconnected from
the mains supply before performing any cleaning or maintenance activities.
Inspect the device before use. If there is any damage, it must not be used.