manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Abaxis
  6. •
  7. Measuring Instrument
  8. •
  9. Abaxis VetScan HM5 User manual

Abaxis VetScan HM5 User manual

Other manuals for VetScan HM5

4

Other Abaxis Measuring Instrument manuals

Abaxis Piccolo Xpress AB-801-110-0000 User manual

Abaxis

Abaxis Piccolo Xpress AB-801-110-0000 User manual

Abaxis VetScan UA10 User manual

Abaxis

Abaxis VetScan UA10 User manual

Abaxis VetScan vs2 User manual

Abaxis

Abaxis VetScan vs2 User manual

Abaxis VetScan SA User manual

Abaxis

Abaxis VetScan SA User manual

Abaxis VetScan HM5 User manual

Abaxis

Abaxis VetScan HM5 User manual

Abaxis VetScan HM5 User manual

Abaxis

Abaxis VetScan HM5 User manual

Abaxis VetScan HM5 User manual

Abaxis

Abaxis VetScan HM5 User manual

Abaxis VetScan i-STAT 1 User manual

Abaxis

Abaxis VetScan i-STAT 1 User manual

Abaxis VetScan HM5 User manual

Abaxis

Abaxis VetScan HM5 User manual

Abaxis i-STAT Alinity v User manual

Abaxis

Abaxis i-STAT Alinity v User manual

Popular Measuring Instrument manuals by other brands

Powerfix Profi 278296 Operation and safety notes

Powerfix Profi

Powerfix Profi 278296 Operation and safety notes

Test Equipment Depot GVT-427B user manual

Test Equipment Depot

Test Equipment Depot GVT-427B user manual

Fieldpiece ACH Operator's manual

Fieldpiece

Fieldpiece ACH Operator's manual

FLYSURFER VIRON3 user manual

FLYSURFER

FLYSURFER VIRON3 user manual

GMW TG uni 1 operating manual

GMW

GMW TG uni 1 operating manual

Downeaster Wind & Weather Medallion Series instruction manual

Downeaster

Downeaster Wind & Weather Medallion Series instruction manual

Hanna Instruments HI96725C instruction manual

Hanna Instruments

Hanna Instruments HI96725C instruction manual

Nokeval KMR260 quick guide

Nokeval

Nokeval KMR260 quick guide

HOKUYO AUTOMATIC UBG-05LN instruction manual

HOKUYO AUTOMATIC

HOKUYO AUTOMATIC UBG-05LN instruction manual

Fluke 96000 Series Operator's manual

Fluke

Fluke 96000 Series Operator's manual

Test Products International SP565 user manual

Test Products International

Test Products International SP565 user manual

General Sleep Zmachine Insight+ DT-200 Service manual

General Sleep

General Sleep Zmachine Insight+ DT-200 Service manual

Sensa Core Lacto Spark user manual

Sensa Core

Sensa Core Lacto Spark user manual

VOLTCRAFT UTS-1980 operating instructions

VOLTCRAFT

VOLTCRAFT UTS-1980 operating instructions

EBCHQ 94915 Operation manual

EBCHQ

EBCHQ 94915 Operation manual

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Balluff

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Pulsar IMP Lite Series instruction manual

Pulsar

Pulsar IMP Lite Series instruction manual

ETCR 6800 user manual

ETCR

ETCR 6800 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Vetscan HM5 Reagent Pack
Installation Guide & Package Insert
Technical Support Phone Number 1-800-822-2947
©January 2010, Abaxis Inc., Union City, CA 94587 Abaxis Part No. 770-9000
INTENDED USE: the VetScan HM5 reagent pack is intended for
veterinary hematology use with the VetScan HM5 analyzer.
COMPONENTS INCLUDED:
Diluent (green cap)
Lyse (yellow cap)
Rinse (white cap)
Lyse2 (white/orange cap)
Cleaner (blue cap)
Thermal paper roll (one)
STORAGE:
Store reagent pack at room temperature (68-86° F). Do not use after
the expiration date listed on top of carton. Do not freeze reagents.
Do not use if reagents have been previously frozen. Reagents are
stable for 26 weeks after opening, if stored at optimum conditions.
Allow your reagent pack to equilibrate to room temperature for 24 hours
prior to use to prevent the formation of micro-bubbles that may affect
your results.
CHANGING THE REAGENT PACK
Before changing the reagent pack, be sure to read the instructions
carefully. You will need the following:
Distilled water (approximately 100-200 mls, or
about 1 cup) – do not use tap water
5 ml of household bleach
(approximately 5-6% hypochlorite)
HM5 cleaning tube
HM5 reagent pack
Two clean containers/cups to hold the water
and bleach
HM5 Normal control (Abaxis part no. 770-9029)
Prepare a 1:20 dilution of bleach.1.
Mix 5 ml household bleach with 95 ml ofa.
distilled water.
Perform a bleach cleaning (2. Utilities menu ,
Maintenance –Bleaching), see HM5 User’s
Manual, section 7.3 for details.
Reconnect the reagent tubing to the analyzer.3.
Unscrew the reagent caps from the old reagent4.
pack. Connect them to the new reagent bottles.
NOTE: Take care not to touch the drop down
tubing (tubing that is inside the bottles) or the
reagents may become contaminated. Be sure
to match the correct caps to the color-matched
bottles.
Press5. New for new reagent pack.
The software will ask if you are changing thea.
reagent pack, or using an old one.
Prime the reagent lines: follow the menus below:6.
Utilitiesa. key – Maintenance –Priming –Prime
All
Reset the Reagent Status: follow the menus7.
below:
Utilitiesa. key – Maintenance –Reagent Status
Pressb. Change and the settings will be reset to
100%. Press Accept.
If you have moved the reagent pack’s location8.
from the previous pack’s location, you will need to
reset the reagent sensors. Follow the menu below:
Utilitiesa. –Settings –Fluid Sensors –Calibrate
Sensors
Run a control sample.9.
Be sure to follow the procedure in the HM5a.
User’s manual when running a control. Re-
calibrate the instrument if the control results
do not match the expected values.
Your instrument is ready to go!10.
Doc. Number & Rev 770-7017 Rev B