Abaxis VetScan HM5 User manual

Quick
Reference
Guide
For Veterinary Use Only
HematologyAnalyzer

Running a Blank
Running a Blank
1
1. From the Home Screen, touch Measure
then Run. If no blank has been run in the
last 12 hours, an expired blank message
will appear. Touch OK to start the blank.
2. If no warning message appears, Accept
the blank and proceed to running patient
samples.
3. If blank values are high, a pop-up warning
will appear. Touch OK, to view the blank
results and the corresponding Warning
flag (top right of screen).
If Warning flags are present, refer
to the troubleshooting section of this
guide or the Operator‘s Manual to resolve
the warning.
Blanks must be run every 12 hours, after a Soak Cleaning has been performed, after a
reagent pack change, and after the instrument has been restarted. It is not necessary
to run blanks on days that samples are not run. Additionally, blanks need to be run after
self tests and Deep Cleaning.
Run
2
Blank results are expired!
Press OK to measure a Blank first
There are parameters with high
blank values! Check the User’s Manual
Troubleshooting section for guidance
and then repeat the blank

Running Samples
Before running a patient sample, take
a few seconds to visually inspect the
sample for clots.
Hold the test tube at eye level and
slowly invert the tube. If a clot
is found, the sample should be
discarded and a new sample should
be drawn.
Even small sized clots will affect the
patient results and could damage
your HM5.
Running a Sample
Running a Sample

Running a Sample
1. From the Home Screen, touch the
Measure icon.
2. Touch Run.
3. Select the appropriate species Type.
Optionally, input additional patient
information. Touch Run.
4. Mix as directed and ensure the bubble
moves from the top to the bottom of the
sample tube to ensure all of the sample
is mixing with the anticoagulant. Remove
the cap of the tube, place the tube in the
tube holder and touch Run.
1
3
4
Run
2
Running a Sample

1. From the Home Screen, touch the
Maintenence icon.
2. Touch the Quality Control.
3. Select the appropriate control level to match
the vial being used.
4. Touch Measure.
5. Touch Run QC.
6. If the lot number displayed matches the
control value sheet, check that the target
and gap values match the screen and
touch Accept.
If the lot number does NOT match,
touch New.
3
1
Controls should be run when reagent packs are changed, after a Soak Cleaning has been
performed, and to check the calibration of the analyzer when results are in question.
Remove the control from the refrigerator and allow to warm to room temperature for
15 minutes before use. Mix control tubes thoroughly immediately prior to use. Roll 10X
between hands before gently inverting 10-15X.
Controls should be discarded if expired or if they have been open for more than 14 days.
Running Controls

7. From the New QC Lot screen, select the method
for entering the control target and ranges:
• Enter Manually allows the user to type
in the target and range values from the
control package insert sheet.
• Load from USB Drive allows the user to
upload the values downloaded onto a
USB drive from http://www.abaxis.com/
veterinary/products/vetscan-hm5
• Scan Barcode allows the user to scan
the barcode on the control package
insert sheet using a 2D barcode scanner
to upload the values.
8. Accept the new values.
9. Confirm that the values entered are correct.
10. Follow the mixing instructions, remove
cap, and place tube in the sample adapter.
Touch Run.
11. Run the control 2-3 times to make sure
values are similar from run to run. If all
values reported are within reference range,
the analyzer is in calibration and samples
may be run. If values are repeatedly out of
range, consult Abaxis Technical Support
before calibrating.
(See Calibrating the Analyzer)
10
7
Running Controls

Changing the
Reagent Pack-Caps
Calibration should be performed when the quality control runs are showing values
consistently out of range and it has been determined that the analyzer is not in need
of cleaning and the quality control material itself is not degraded. Use only the Abaxis
Normal HM5 Control for calibration. Allow the control to come to room temperature by
setting it on the counter for 15 minutes before use. Do not calibrate with controls that are
expired or have been open for more than 14 days.
3
1. From the Home Screen, touch
Maintenance, Calibration then Measure.
2. Accept settings for Prediluted [No] and
Cal Sampling Depth [-2mm].
3. If calibration is being performed with the
same lot number last used for quality
control, touch Copy. If not, manually enter
the target values from the control package
insert. Touch Accept.
4. Run Cal three or more times in a row,
following the mixing instructions displayed
in between each run. The analyzer will
calculate the average of the 3+ sample runs.
5. Touch Calibrate at the bottom of the
screen after the 3+ runs have completed.
6. Accept the new coefficients displayed on
the Calibration Results Screen. If any error
messages pop-up during calibration, call
Abaxis Technical Support.
7. Run Quality Control again 2-3 times to make
sure the calibration was successful and all
values are now within range.
1
Calibrating the Analyzer

Changing the
Reagent Pack-QCA
Materials needed:
HM5 Reagent Pack, HM5 Normal Control and latex or nitrile gloves.
If “Change Reagent Pack” is used when installing new reagents, the analyzer will warn the
user when there are fewer than 10 tests left in any given bottle. This feature will also track
the number of samples, number of blanks, install date and the volume remaining in each of
the bottles.
2
1. From the Home Screen, touch the
Maintenance icon.
2. Touch Change Reagent Pack.
3. Follow the detailed directions on the
screen for changing the reagents. Touch
Next to advance through the various steps.
4. Run Quality Control to ensure the Analyzer
is within calibration.
1
Changing the Reagent Pack

Changing the
Reagent Pack-QCA
The wash head should be cleaned once a week. Warnings will alert users when a
wash head cleaning is due. The warnings will differ depending on whether Scheduled
Maintenance is Enabled or Disabled.
To Clean the Wash Head for troubleshooting, touch the
Maintenance icon from the Home screen, then Cleaning, then
Clean Wash Head. On-screen instructions will guide the user
through the cleaning.
If Scheduled Maintenance is Enabled, a message will appear when a cleaning is due.
Touch Clean, then follow the on-screen instructions. Alerts will start appearing 3 days
before the cleaning is due. Maintenance is required on the due date.
Maintenance (Weekly)

Calibrating the
Analyzer
Soak Cleaning should be performed every 14 days. When the software is upgraded to v2.3
or higher, the Soak Clean Interval will automatically be defaulted to 14 days.
If Scheduled Maintenance is Enabled, a message will appear when a cleaning is due. Touch
Next, then follow the on-screen instructions. Alerts will start appearing 3 days before the
cleaning is due. Maintenance is required on the due date.
Instructions will guide the user through the cleaning. This cleaning takes 10 minutes; after the
cleaning is completed run a Blank, and if the Blank is acceptable, proceed with Quality control.
If your HM5 analyzer is not set up for Soak Cleanings every 14 days, refer to HM5
Operator’s manual.
Maintenance (14 days)

Deep Cleaning Cleaning & Maintenance Procedures
This cleaning is meant to be used as an emergency procedure when a high HCT trend is
noticed: and it is not part of the regularly scheduled maintenance.
Please ensure that your HM5 is on software version 2.3 or higher.
1. Deep Cleaning. Go to Maintenance> Cleaning> Deep Cleaning.
1a. Remove the cap
and place the VetScan
HemaClean tube into
the tube holder in the front
of the analyzer.
Touch Next to start
the cleaning. It takes
approximately 20 minutes
to complete.

Deep Cleaning Cleaning & Maintenance Procedures
2. Quality control.
Go to Maintenance> Quality Control> QC Normal. Make sure the control has come to
room temperature. Mix well by rolling the tube flat between your hands 10 times (A)
and then slowly invert the tube 10-15 times (B). Run the well mixed HM5 control 2-3
times and ensure the results are within range to confirm the analyzer is ready to be
used.
2a. If 2 out of 3 QC runs fail, proceed to Calibration.
Go to Maintenance> Calibration> Measure> Accept, then enter in the values
from the Package insert. Accept> Run Cal. After a successful calibration is
completed, go to Maintenance> Quality Control> QC Normal and run QC again.
Note: If a warning message pops up or the Quality control fails, call Abaxis Technical
Support.
AB10x
10x
5X

Blank Warning Flags
Flag Meaning Solution
pPLT blank is high Visually inspect the wash head and clean if necessary; run 2-3
automatic self cleanings, then repeat the blank. If the warning
persists, run Soak Cleaning and the repeat the Blank.
XEOS blank high Visually inspect the wash head and clean if necessary: check
Diluent (green) and Lyse2 (orange) lines for loose connections;
prime Diluent and/or Lyse2 if bubbles are present or were
introduced while checking the tubing; repeat the blank.
bRBC blank is high Visually inspect the wash head and chamber below it for
blood debris; clean if necessary. Run a Soak Cleaning, then an
automatic self-cleaning; repeat the Blank.
More thorough troubleshooting procedures can be found in the Abaxis HM5 Operator‘s Manual.
Warning Flags (Partial List)

Results Warning Flags
Flag Meaning Solution
EEmpty sample Ensure the sampling depth is appropriate for the tube being
used and that the tube is filled at least 1/2 full. If not, redraw a
new sample.
M,N Lyse delivery issue
or very
high WBC count
Check Lyse (yellow) tube for kinks and loose connections at
the bottle cap or dip tube and back of the analyzer and verify
the Lyse line is connected to the correct port on the analyzer
and the correct bottle from the pack. Prime Lyse if bubbles are
present. If a high WBC value is suspected, view manual smear
to confirm high WBC.
c,C,Q Cellular clumping Run 2 automatic self cleanings. Redraw then rerun if the draw
was difficult. Run Soak Cleaning if error continues.
LPlatelet clumping,
Lyse resistance or
Macroplatelets
If the patient is a King Charles Cavalier Spaniel or other breed
with known predisposition to macro-platelets, perform a
manual smear for the WBC and PLT. If the draw was difficult,
the platelets are low or there were delay in filling the tube,
redraw and rerun. Check Lyse tube for kinks or bubbles. If no
Lyse issues found and the warning persists, rerun the sample
with an increased Lyse +0.1 or +0.2.
WLarge Platelet
clumps in cat
samples
Redraw sample, being sure to immediately transfer the sample
into the tube and invert completely 10-15 times or more.
YEOS clogging Ensure Lyse 2 (orange line) is connected properly to the back
of the analyzer and to the correct reagent bottle. If connections
are good but bubbles are present, prime Lyse 2 twice and re-run
sample. Run automatic self cleanings and re-run well-mixed sample.
Warning Flags (Partial List)

1. Order the CBC in the Practice
Management Software.
2. From the Home Screen, touch the
Measure icon then Work List.
OR
Touch the blood drop icon at the top of
the screen.
Work List requires the use of additional VetScan connectivity products which enable
bi-directional communication between the HM5 and compatible veterinary practice
management software systems.
12
Using Work List

3. Select a patient from the list by checking
the associated box.
4. Touch Run.
5. Follow the mixing directions on the screen.
Touch Run. When the CBC is complete, the
animal sample just run will no longer be
displayed on the Work List.
3
Using Work List


Abaxis, Inc.
3240 Whipple Road
Union City, CA 94587
USA
Tel +1 800 822 2947
Fax +1 510 441 6150
www.abaxis.com
Abaxis and VetScan are registered trademarks of Abaxis, Inc. © Abaxis 2018. 790-7028 Rev. E
800 822 2947
For Technical Support 24 hours a day/7 days a week:
Other manuals for VetScan HM5
4
Table of contents
Other Abaxis Measuring Instrument manuals

Abaxis
Abaxis VetScan HM5 User manual

Abaxis
Abaxis VetScan vs2 User manual

Abaxis
Abaxis VetScan HM5 User manual

Abaxis
Abaxis i-STAT Alinity v User manual

Abaxis
Abaxis VetScan HM5 User manual

Abaxis
Abaxis VetScan i-STAT 1 User manual

Abaxis
Abaxis VetScan UA10 User manual

Abaxis
Abaxis VetScan SA User manual

Abaxis
Abaxis Piccolo Xpress AB-801-110-0000 User manual

Abaxis
Abaxis VetScan HM5 User manual