ABB FEF series User manual

User Guide OI/FEF/FEV/FEW–EN Rev. D
F, V, W Series
Electromagnetic flowmeter
Full-bore flow sensors
The perfect fit for all water
industry applications
Introduction
ABB’s full-bore FEF, FEV and FEW electromagnetic
flowmeter sensors are available with either an
AquaMaster 3 or a WaterMaster transmitter.
AquaMaster 3 and WaterMaster are a range of high
performance electromagnetic flowmeters for the
measurement of electrically conductive fluids and are
supplied as factory-configured and calibrated systems.
This User Guide provides end-user details for installation
and connection.
This User Guide should be used in conjunction with the
following publications:
– AquaMaster 3 FET200 electromagnetic flowmeter
transmitter User Guide (OI/FET200–EN)
– WaterMaster FET100 electromagnetic flowmeter
transmitter User Guide (OI/FET100–EN)
– WaterMaster FET100 electromagnetic flowmeter /
transmitter Hazardous areas ATEX / IECEx areas 2,
21 and 22 User Guide (OI/FET100/ATEX–EN)

The Company
We are an established world force in the design and manufacture of instrumentation for industrial process control,
flow measurement, gas and liquid analysis and environmental applications.
As a part of ABB, a world leader in process automation technology, we offer customers application expertise, service
and support worldwide.
We are committed to teamwork, high quality manufacturing, advanced technology and unrivalled service and
support.
The quality, accuracy and performance of the Company’s products result from over 100 years experience, combined
with a continuous program of innovative design and development to incorporate the latest technology.
Quality Control
The UKAS Calibration Laboratory No. 0255 is just one of the ten flow calibration plants operated by the Company
and is indicative of our dedication to quality and accuracy.
UKAS Calibration Laboratory No. 0255
0255

F, V, W S e rie s
Electromagnetic flowmeters | Full-bore flow sensors
OI/FEF/FEV/FEW–EN Rev. D 1
Contents
1 Safety ............................................................................................................................................... 2
1.1 Electrical Safety ............................................................................................................................................. 2
1.2 Symbols ........................................................................................................................................................ 2
1.3 Health & Safety ............................................................................................................................................. 3
2 Mechanical Installation .................................................................................................................... 4
2.1 Unpacking .................................................................................................................................................... 4
2.2 Installation Conditions ................................................................................................................................... 4
3 Electrical Installation ..................................................................................................................... 10
3.1 Grounding ................................................................................................................................................... 10
3.2 Cable Preparation (Remote Transmitter Only) .............................................................................................. 15
3.2.1 Sensor Cable Connections (Remote WaterMaster Transmitter Only) ............................................. 15
3.2.2 Sensor Cable Connections (Remote AquaMaster 3 Transmitter Only) ........................................... 16
3.3 Environmental Protection ............................................................................................................................ 17
4 Specification .................................................................................................................................. 18
5 Sensor dimensions ........................................................................................................................ 25
5.1 FEW – DN10 to 125 (3/8 to 5 in. NB) .......................................................................................................... 25
5.2 FEW – DN150 to 400 (6 to 16 in. NB) .......................................................................................................... 28
5.3 FEW – DN450 to 2400 (18 to 96 in. NB) ...................................................................................................... 30
5.4 FEV – DN40 to 200 (11/2 to 8 in. NB) ......................................................................................................... 36
5.5 FEF – DN80 to 600 (3 to 24 in. NB) ............................................................................................................. 38
6 Common accessories .................................................................................................................... 41
Notes ................................................................................................................................................... 42

F, V, W S e r i e s
Electromagnetic flowmeters | Full-bore flow sensors 1 Safety
2OI/FEF/FEV/FEW–EN Rev. D
1 Safety
Information in this manual is intended only to assist our customers in the efficient operation of our
equipment. Use of this manual for any other purpose is specifically prohibited and its contents are not to be
reproduced in full or part without prior approval of the Technical Publications Department.
1.1 Electrical Safety
This equipment complies with the requirements of CEI/IEC 61010-1:2001-2 'Safety Requirements for
Electrical Equipment for Measurement, Control and Laboratory Use' and complies with NIST and OSHA.
If the equipment is used in a manner NOT specified by the Company, the protection provided by the
equipment may be impaired.
1.2 Symbols
One or more of the following symbols may appear on the equipment labelling:
Warning – Refer to the manual for
instructions Direct current supply only
Caution – Risk of electric shock Alternating current supply only
Protective earth (ground) terminal Both direct and alternating current supply
Earth (ground) terminal The equipment is protected
through double insulation

F, V, W S e r i e s
Electromagnetic flowmeters | Full-bore flow sensors 1 Safety
OI/FEF/FEV/FEW–EN Rev. D 3
1.3 Health & Safety
Health and Safety
To ensure that our products are safe and without risk to health, the following points must be noted:
The safety requirements of this equipment, any associated equipment and the local
environment must be taken into consideration during installation.
Install and use this equipment and any associated equipment in accordance with the relevant
national and local standards.
The relevant sections of these instructions must be read carefully before proceeding.
Warning labels on containers and packages must be observed.
Installation, operation, maintenance and servicing must only be carried out by suitably trained
personnel and in accordance with the information given.
Normal safety precautions must be taken to avoid the possibility of an accident occurring when
operating in conditions of high pressure and / or temperature.
Chemicals must be stored away from heat, protected from temperature extremes and powders
kept dry. Normal safe handling procedures must be used.
When disposing of chemicals ensure that no two chemicals are mixed.
Product liability – advice and assistance provided without charge is given in good faith but
without liability.
Safety advice concerning the use of the equipment described in this manual or any relevant hazard
data sheets (where applicable) may be obtained from the Company address on the back cover,
together with servicing and spares information.
Warning.
System configuration must be carried out only by users or personnel with approved access rights
(user privileges).
Read all relevant sections of this guide before configuring the system or modifying system
parameters.
Install and use associated equipment in accordance with the relevant national and local
standards.

F, V, W S e r i e s
Electromagnetic flowmeters | Full-bore flow sensors 2 Mechanical Installation
4OI/FEF/FEV/FEW–EN Rev. D
2 Mechanical Installation
2.1 Unpacking
2.2 Installation Conditions
Fig. 2.1 Unpacking
Caution.
Inspect the flowmeter for damage before installation. Do NOT install a damaged or faulty
flowmeter.
When lifting the flowmeter, use the lifting hooks provided or sling under the body of the meter.
NEVER lift the flowmeter by the sensor cable terminal connection box – this will damage the
terminal connection box and invalidate the warranty.
Caution. Do NOT exceed the maximum working pressure marked on the flowmeter.
Fig. 2.2 Spillage
Caution. Do NOT install the flowmeter in an area where a spillage of any substance could damage the
flowmeter.

F, V, W S e r i e s
Electromagnetic flowmeters | Full-bore flow sensors 2 Mechanical Installation
OI/FEF/FEV/FEW–EN Rev. D 5
Fig. 2.3 Vibration
Caution. Do NOT install the flowmeter in a pipeline that may exert excessive movement and twisting
forces on the flowmeter, for example, vibration.
Fig. 2.4 Localized Heat
Caution. Do NOT subject the flowmeter to localized heat during installation. Remember – the flowmeter
is a measuring instrument.
Fig. 2.5 Siting
Note. When installing an integral flowmeter, allow enough room to read the transmitter's display and
data label.
Allow room to read display
or data plate

F, V, W S e r i e s
Electromagnetic flowmeters | Full-bore flow sensors 2 Mechanical Installation
6OI/FEF/FEV/FEW–EN Rev. D
Fig. 2.6 Within Temperature Limits
Caution. Ensure the flowmeter is operated within its specified temperature limits. Use flange seals
made from a material that is compatible with the fluid and fluid temperatures as required.
Fig. 2.7 Straight Pipe Requirements
Note. Ensure the flow direction in the pipeline corresponds to the identification plate. The flowmeter
measures the flowrate in both directions. Forward flow is the factory setting. Experience has shown that,
in most installations, a straight upstream section 3 x pipe diameters in length and a straight downstream
section 2 x pipe diameters in length are sufficient. However, wherever possible, straight upstream and
downstream sections 5 x 2 pipe diameters in length are recommended.
Fig. 2.8 Fluid Level
Note. For accurate and reliable operation, install the flowmeter sensor in a position where it will be
completely full when in operation. The flowmeter will operate when not full, but indicated readings will not
be accurate.
70 / 130 °C
(158 / 266 °F)
Max.
– 20 °C
(– 4 °F)
Min.
Min.
Min.
>2 Pipe Diameters
5 x Pipe Diameters
Flow Direction

F, V, W S e r i e s
Electromagnetic flowmeters | Full-bore flow sensors 2 Mechanical Installation
OI/FEF/FEV/FEW–EN Rev. D 7
Fig. 2.9 Above Ground
Caution. Do NOT install the flowmeter in a manner that will subject it to mechanical tension (torsion,
bending). If required, support the pipeline. Remember – the flowmeter is a measuring instrument.
Fig. 2.10 Temperature Difference
Caution. Do NOT install the flowmeter in a position that will expose it to direct sunlight. Provide
appropriate sun protection if required.
Fig. 2.11 Within Environmental Rating
Caution. Ensure the level of IP rating is correct for the required meter installation application. IP68
(NEMA 6) rated meters can be continuously submerged.
Supports
IP68 (NEMA 6)
Sensor Submersion: Continuous

F, V, W S e r i e s
Electromagnetic flowmeters | Full-bore flow sensors 2 Mechanical Installation
8OI/FEF/FEV/FEW–EN Rev. D
Fig. 2.12 Underground
Caution. When burying a flowmeter, ensure it can be found easily if required (for example, mark the
installation position with a post). Installing a protection plate above the meter is also recommended.
Note. For further advice when burying flowmeters, contact the ABB Service Organization.
Fig. 2.13 Cable Routing
Caution. Install all flowmeter cabling neatly. Installation within a conduit is recommended. Install the
cabling or conduit with a U-bend below the terminal connection box height to prevent water ingress into
the flowmeter sensor by capillary action.
Protection Plate
Recommended
Backfill

F, V, W S e r i e s
Electromagnetic flowmeters | Full-bore flow sensors 2 Mechanical Installation
OI/FEF/FEV/FEW–EN Rev. D 9
Fig. 2.14 Gasket Fitting
Caution. Use correctly sized gaskets. Do NOT fit gaskets that will extend into the flow area – the
turbulence caused will adversely affect flowmeter accuracy.
Fig. 2.15 Separation of Sensors
Caution. If flowmeters are installed in adjacent pipelines, ensure they are at least 0.7 m (2.3 ft.) apart to
prevent the electromagnetic field from one meter affecting the other.
Fit Gaskets
Gaskets Same Size as Pipe
0.7 m (2.3 ft) Min.

F, V, W S e r i e s
Electromagnetic flowmeters | Full-bore flow sensors 3 Electrical Installation
10 OI/FEF/FEV/FEW–EN Rev. D
3 Electrical Installation
3.1 Grounding
Caution. For safety reasons and optimum performance, the flowmeter, pipelines and medium must be
correctly bonded and grounded according to regulations.
Do not ground cathodically-protected pipelines to an external earth.
Note.
The flow sensor must not be connected to a ground spike.
For bonding connections use ³ 4mm2(< 10AWG) cable.
Arrangement Number of Fluid Contact Rings
AquaMaster 3 Transmitter 2
WaterMaster Transmitter 1 minimum
Insulating Pipe 2
Table 3.1 Fluid Contact Rings
Fig. 3.1 All Metal Pipe (Including Lined Metal Pipe)
Fluid contact ring for lined pipes Flow Direction

F, V, W S e r i e s
Electromagnetic flowmeters | Full-bore flow sensors 3 Electrical Installation
OI/FEF/FEV/FEW–EN Rev. D 11
Fig. 3.2 Metal Pipe with Flange Adaptor
Fig. 3.3 Flanged Metal Pipe to Plastic Pipe
Fig. 3.4 Metal Pipe with Plastic Make-up Insert
Plastic Pipe
Metal Pipe Fluid Contact Ring
Plastic Pipe
Metal Pipe Fluid Contact Ring

F, V, W S e r i e s
Electromagnetic flowmeters | Full-bore flow sensors 3 Electrical Installation
12 OI/FEF/FEV/FEW–EN Rev. D
Fig. 3.5 All-plastic Pipe
Fig. 3.6 Pipelines with Cathodic Protection
Caution. Do NOT ground cathodically-protected systems.
Fluid Contact Rings
Fluid Contact Rings
Bonding Cables (Supplied)

F, V, W S e r i e s
Electromagnetic flowmeters | Full-bore flow sensors 3 Electrical Installation
OI/FEF/FEV/FEW–EN Rev. D 13
Fig. 3.7 AquaMaster 3 Transmitter Mounted in a Chamber (Battery Version Shown)
Fig. 3.8 AquaMaster 3 Transmitter Mounted in a Cabinet (Battery Version Shown)

F, V, W S e r i e s
Electromagnetic flowmeters | Full-bore flow sensors 3 Electrical Installation
14 OI/FEF/FEV/FEW–EN Rev. D
Fig. 3.9 WaterMaster Remote Transmitter Mounted in a Roadside Cabinet
Power Supply

F, V, W S e r i e s
Electromagnetic flowmeters | Full-bore flow sensors 3 Electrical Installation
OI/FEF/FEV/FEW–EN Rev. D 15
3.2 Cable Preparation (Remote Transmitter Only)
To prepare the cable for connection at the transmitter and sensor terminal blocks:
1. Remove the outer cable insulation and Mylar® wrap.
2. Ensure the drain wire is sleeved.
3. Cut the cable connection wires to the lengths shown.
3.2.1 Sensor Cable Connections (Remote WaterMaster Transmitter Only)
Caution. Maintain Environmental Protection at all times – See Section 3.3, page 17.
Caution.
Make connections only as shown.
Twist the screen wire of D1 / TFE + D2 with the outer screen drain wire and sleeve them green /
yellow.
Ensure the seal and mating surfaces are clean to maintain environmental rating.
Conduit connections must provide cable entry sealing.
Ensure cable glands are tightened after wiring. Do not overtighten the plastic cable glands to
avoid destroying their sealing properties. Initially, tighten finger-tight, then a further 1/2to 3/4turn
using a suitable spanner or wrench.
Fig. 3.10 Sensor Cable Connections – WaterMaster Transmitter
Screen to
Internal Earth
for NPT Variants
*Inner Wire
**For Cathodically Protected Systems connect the drain wire to terminal PE.
**Drain Wire (Twisted with Screen Wire
from D1/TFE – Orange and D2 – Yellow)
for M20 Versions connect to PE
S1 Violet (Screen)
Cut cables to
60 mm (2.35 in)
E1 Violet (*Signal)
E2 Blue (*Signal)
S2 Blue (Screen)
3 Green (Sleeve)
D2 Yellow
D1/TFE Orange
PE (Screen)
M2 Red
M1 Brown

F, V, W S e r i e s
Electromagnetic flowmeters | Full-bore flow sensors 3 Electrical Installation
16 OI/FEF/FEV/FEW–EN Rev. D
3.2.2 Sensor Cable Connections (Remote AquaMaster 3 Transmitter Only)
Caution.
Twist the three screen wires together and sleeve them.
Keep cable pairs twisted.
Make connections only as shown.
Maintain Environmental Protection at all times.
Conduit connections must provide cable entry sealing.
Fig. 3.11 Sensor Cable Connections – AquaMaster 3 Transmitter
Drain / Screen
to Internal Earth
for NPT Gland
Cut cables to
60 mm (2.35 in)
7 Violet
6 Blue
5 Green
4 Yellow
3 Orange
2 Red
1 Brown
Drain / Screen Wire for M20 Gland

F, V, W S e r i e s
Electromagnetic flowmeters | Full-bore flow sensors 3 Electrical Installation
OI/FEF/FEV/FEW–EN Rev. D 17
3.3 Environmental Protection
Fig. 3.12 Potting the Sensor Terminal Box
Warning.
Potting materials are toxic. Read the manufacturers' instructions carefully before preparing the
potting material and use suitable safety precautions.
Power up and check all connections before potting.
The remote sensor terminal box connections must be potted immediately on completion to
prevent the ingress of moisture.
Do not overfill or allow the potting material to come into contact with 'O' rings or grooves.
Do not let potting material enter conduit (if used).

F, V, W S e r i e s
Electromagnetic flowmeters | Full-bore flow sensors 4 Specification
18 OI/FEF/FEV/FEW–EN Rev. D
4 Specification
WaterMaster optimized full-bore meter / full-bore meters flow performance (m3/h)
Standard Calibration – 0.4 % Class 2 High Accuracy Calibration – 0.2 % Class 1
DN Q4 Q3 Q0.4% Q2 Q1 Q0.2% Q2 Q1
10 3.1 2.5 0.167 0.013 0.008 0.31 0.02 0.012
15 7.88 6.3 0.42 0.032 0.02 0.79 0.05 0.03
20 12.5 10 0.67 0.05 0.032 1.25 0.08 0.05
25 20 16 1.1 0.08 0.05 2 0.13 0.08
32 31.25 25 1.67 0.13 0.08 3 0.20 0.13
40* 50 40 4.2 0.2 0.13 6 0.32 0.2
50* 79 63 4.2 0.32 0.20 7.9 0.5 0.32
65* 125 100 6.7 0.5 0.32 12.5 0.8 0.5
80* 200 160 10.7 0.81 0.51 16 1.3 0.8
100* 313 250 16.7 1.3 0.79 25 2 1.25
125* 313 250 16.7 1.3 0.79 25 2 1.25
150* 788 630 42 3.2 2.0 63 5 3.2
200* 1,250 1,000 67 5.1 3.2 100 8 5
250 2,000 1,600 107 8.1 5.1 160 13 8
300 3,125 2,500 167 12.7 7.9 250 20 12.5
350 5,000 4,000 267 20.3 12.7 400 32 20
400 5,000 4,000 267 20.3 12.7 400 32 20
450 7,875 6,300 420 32 20 630 50 32
500 7,875 6,300 420 32 20 630 50 32
600 12,500 10,000 667 51 32 1000 80 50
700 20,000 16,000 1600 102 64 1600 160 100
750 20,000 16,000 1600 102 64 1600 160 100
30 in (760) 20,000 16,000 1600 102 64 1600 160 100
800 20,000 16,000 1600 102 64 1600 160 100
900 31,250 25,000 2500 160 100 2500 250 156
1000 31,250 25,000 2500 160 100 2500 250 156
42 in 31,250 25,000 2500 160 100 2500 250 156
1100 31,250 25,000 2500 160 100 2500 250 156
1200 50,000 40,000 4000 256 160 4000 400 250
1350 78,750 63,000 6300 403 252 6300 630 394
1400 78,750 63,000 6300 403 252 6300 630 394
1500 78,750 63,000 6300 403 252 6300 630 394
60 in (1500) 78,750 63,000 6300 403 252 6300 630 394
1600 78,750 63,000 6300 403 252 6300 630 394
1650 78,750 63,000 6300 403 252 6300 630 394
1800 125,000 100,000 10000 640 400 10000 1000 625
1950 125,000 100,000 10000 640 400 10000 1000 625
2000 125,000 100,000 10000 640 400 10000 1000 625
2200 200,000 160,000 16000 1024 640 16000 1600 1000
2400 200,000 160,000 16000 1024 640 16000 1600 1000
* OIML R49 Certificate of Conformance to Class 1 and Class 2, with OIML R49 and MID versions available.
Note. OIML R49–1 allows Class 1 only for meters with Q3³ 100 m3/h. Meters outside this range have been tested and
conform to Class 1.
This manual suits for next models
2
Table of contents
Other ABB Measuring Instrument manuals

ABB
ABB COPA-XE Training manual

ABB
ABB SensyMaster FMT230 User manual

ABB
ABB FXE4000 Training manual

ABB
ABB SR100A User manual

ABB
ABB Comem eOLI22 User manual

ABB
ABB GLA232 Series User manual

ABB
ABB LS4000 User manual

ABB
ABB Endura AZ20 series User manual

ABB
ABB AX400 Series User manual

ABB
ABB ScreenMaster RVG200 Original operating manual

ABB
ABB XSeriesG4 User manual

ABB
ABB FSM4000 Series User manual

ABB
ABB NGC8200 User manual

ABB
ABB Relion RES670 User manual

ABB
ABB AWT210 User manual

ABB
ABB GLA133 Series User manual

ABB
ABB MicroGuard solution User manual

ABB
ABB CoreSense M10 User manual

ABB
ABB Sensyflow FMT200-D User manual

ABB
ABB ProcessMaster FEP630 series User manual