manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Abbott
  6. •
  7. Medical Equipment
  8. •
  9. Abbott CentriMag User manual

Abbott CentriMag User manual

PL-0370 Rev 00 / December 2019 / Page 1
ID Alarm/
Alert
Text Mes-
sage System Status & OperatorResponse
S1 Alarm POWER ON
TEST FAIL
The Pump will not start. An audible alarm will sound, which
cannot be muted. Switch the Console OFF and ON again. If
the alarm reappears, switch to the backup Console, Motor
and Flow Probe, record the alarm message and contact
your local Abbott Medical representative.
S2 Alarm
SYSTEM
FAULT
(Run-Time
System
Failure)
The Pump will stop. An audible alarm will sound, which
cannot be muted. Clamp the return tubing and switch to
the backup Console, Motor and Flow Probe according the
procedure described in Section 2. Resume support. Record
the alarm message and contact your local Abbott Medical
representative.
M1 Alarm MOTOR
STOPPED
The Pump will stop. An audible alarm will sound, which
can be muted for 60 seconds.Clamp the return tubing
and switch to the backup Console, Motor and Flow Probe
according the procedure described in Section 2. Resume
support. Record the alarm message and contact your local
Abbott Medical representative.
M2 Alarm
MOTOR
DISCON-
NECTED
The Pump will stop. An audible alarm will sound, which can
be muted for 60 seconds.
During setup of the System:
Press the alarm acknowledge button and check that the Mo-
tor connector is fully inserted into the back of the Console.
During support:
Press the alarm acknowledge button and check that the
connector of the Motor is fully inserted into the back of the
Console. Resume support. If the visual alarm message
does not disappear, clamp the return tubing and switch to
the backup Console, Motor and Flow Probe according the
procedure described in Section 2. Resume support. Record
the alarm message and contact your local Abbott Medical
representative.
M3 Alarm PUMP NOT
INSERTED
System will not start. An audible alarm will sound, which can
be muted for 60 seconds. Press the alarm acknowledge
button. Insert or reinsert the pump and secure it with
the locking screw. If the alarm repeats, switch to backup
Console, Motor and Flow Probe according the procedure
described in Section 2.
M4 Alarm MOTOR
ALARM
An audible alarm will sound and the System will continue
to operate. Press the alarm acknowledge button, if the
visual alarm message does not disappear, clamp the return
tubing, stop the pump and switch to the backup Console,
Motor and Flow Probe according the procedure described
in Section 2. Resume support. Record the alarm message
and contact your local Abbott Medical representative.
M5 Alert
SET PUMP
SPEED NOT
REACHED
Check the Pump ow. If pump ow is satisfactory, reduce
set speed while insuring that ow is maintained. Press
the alarm acknowledge button. If the alert repeats, clamp
the return tubing, stop the pump and switch to the backup
Console, Motor and Flow Probe according the procedure
described in Section 2. Resume support. Record the alert
message and contact your Abbott Medical representative.
If the pump ow is not satisfactory, clamp the return tubing
and switch to the backup Console, Motor and Flow Probe
according to the procedure described in Section 2. Resume
support. Record the alert message and contact Abbott Med-
ical representative.
ID Alarm/
Alert
Text Mes-
sage System Status & OperatorResponse
B1 Alert
BATTERY
MODULE
FAIL
The Console battery will not function. An audible alarm
will sound. Switch to the backup Console, Motor and Flow
Probe according to the procedure described in Section 2.
Resume support. Record the alert message and contact
your Abbott Medical representative.
B2 Alert
BATTERY
BELOW
MINIMUM
The Pump will stop after a very short time. Plug the Console
into a AC power outlet to charge the battery. If no AC outlet
is available, switch to the backup Console, Motor and Flow
Probe according to the procedure described in Section 2.
Resume support.
F1 Alert
FLOW
PROBE
DISCON-
NECTED
Check the Flow Probe connection on back of the Console.
If necessary, reconnect the Flow Probe connector to the
back of the Console. Press the alarm acknowledge button.
Switch to the backup Flow Probe, if the alert message
repeats.
S3 Alert SYSTEM
ALERT
Press the alarm acknowledge button, if the message does
not disappear, clamp the return tubing, stop the pump
and switch to the backup Console, Motor and Flow Probe
according the procedure described in Section 2. Resume
support. Record the alert message and contact your local
Abbott Medical representative.
F2 Alert
FLOW
SIGNAL
INTER-
RUPTED
(Flow rate
sensor error)
Manually disconnect, reposition and reconnect the
Flow Probe transducer to the tubing. Press the alarm
acknowledge button. Switch to the backup Flow Probe,
if the alert message repeats. If problem still persists after
switching to the backup Flow Probe, stop the pump and
switch to a backup Console. Follow the instructions
described in Section 2. Resume support. Record the alert
message and contact your Abbott Medical representative.
F3 Alert
FLOW
BELOW
MINIMUM
(Low Flow)
Check for physiologic cause or circuit obstruction.
Check minimum ow set point. Do not increase RPM
without conrming that adequate blood volume is available.
Common cause of this alert is inadequate blood volume at
the drainage cannula site for the desired pump ow.
F4 Alert
FLOW
ABOVE
MAXIMUM
Reduce Pump speed and check for cause.
P1 Alert
PRESSURE
1 DISCON-
NECTED
Check the electrical connections on the pressure 1 trans-
ducer and recalibrate. If the problem persists disconnect,
reconnect, and recalibrate the transducer. Consider chang-
ing the transducer and cable if the problem persists.
P2 Alert
PRESSURE
2 DISCON-
NECTED
Check the electrical connections on the pressure 2 trans-
ducer and recalibrate. If the problem persists disconnect,
reconnect, and recalibrate the transducer. Consider chang-
ing the transducer and cable if the problem persists.
ID Alarm/
Alert
Text Mes-
sage System Status & OperatorResponse
P3 Alert
PRESSURE
SYSTEM
FAIL
The pressure monitoring system will not function. If pressure
monitoring is needed then change to the backup Console,
Motor and Flow Probe. Switch to the backup System
according the procedure described in Section 2. Resume
support. Record the alert message and contact your Abbott
Medical representative.
P4 Alert
PRESSURE
1 BELOW
MINIMUM
Check for and resolve a physiological or mechanical cause.
Ensure appropriate pressure alarms are set. Consider re-
calibrating the transducer, if alert cannot be explained by
conventional troubleshooting.
P5 Alert
PRESSURE
2 BELOW
MINIMUM
Check for and resolve a physiological or mechanical cause.
Ensure appropriate pressure alarms are set. Consider re-
calibrating the transducer, if alert cannot be explained by
conventional troubleshooting.
P6 Alert
PRESSURE
1 ABOVE
MAXIMUM
Check for and resolve a physiological or mechanical cause.
Ensure appropriate pressure alarms are set. Consider re-
calibrating the transducer, if alert cannot be explained by
conventional troubleshooting. Consider reducing RPM to
reduce the pressure if appropriate.
P7 Alert
PRESSURE
2 ABOVE
MAXIMUM
Check for and resolve a physiological or mechanical cause.
Ensure appropriate pressure alarms are set. Consider re-
calibrating the transducer, if alert cannot be explained by
conventional troubleshooting. Consider reducing RPM to
reduce the pressure if appropriate.
M6 Alert MOTOR
OVER TEMP
Switch to backup Console, Motor and Flow Probe accord-
ing to the procedure described in Section 2. Verify that the
backup Motor stands free and is not covered (e.g. blankets).
B3 Alert
BATTERY
CHARGER
FAIL
Press the alarm acknowledge button. If the alert message
repeats, switch to backup System as described in Section
2. If this alert is associated with BATTERY MODULE FAIL
then carry out the procedure associated with that alarm.
B4 Alert
BATTERY
MAINTE-
NANCE
REQUIRED
Do not use the Console. Perform battery maintenance ac-
cording to the Battery Maintenance Procedure listed in the
CentriMag Circulatory Support System Operation Manual.
B5 Alert LOW
BATTERY
Plug the Console into a AC power outlet to charge the
battery. If no AC outlet is available, switch to the backup
Console, Motor and Flow Probe according to the procedure
described in Section 2.
Resume support.
B6 Alert ON BAT-
TERY
Verify that the user wants the Console to be on battery. If so,
carefully monitor the status of the battery charge indicator,
while using the System on battery.
Reconnect to AC-outlet, as soon as possible.
CentriMagTM Circulatory Support System
Alarms and Alerts Quick Reference Guide
1 RESPONSE TO SYSTEM ALARMS OR ALERTS
When a System alarm or alert condition occurs, an alarm or alert tone sounds and
text message appears on the display of the 2nd Generation CentriMag Primary
Console. Most alarms and alerts require some action on the user’s part to correct
the cause. The following table can be used to determine how to respond to an
alarm or alert condition.
PL-0370 Rev 00 / December 2019 / Page 2
2 SWITCHING TO BACKUP HARDWARE
Switch all components (Console, Motor, Flow Probe and cables) simultaneously and
then perform troubleshooting on the non-functioning System when it is no longer
being used for patient support.
The steps to switch to a backup Console, Motor and Flow Probe set are as follows:
© 2019 Abbott. All Rights Reserved.
Continue clamping the tubing while lowering the RPM to zero. Turn the main
Console’s power switch o󰀨. Turn on the backup System, and make sure that it
is properly assembled with the backup Motor.
Unthread the Pump retaining screw on the Motor by turning the screw counter-
clockwise several revolutions until the screw tip is clear of the locking groove
on the Pump.
Place the Pump in the backup Motor receptacle (the Pump will drop into place
in one of 3 orientations).
Rotate the Pump counterclockwise until it stops.
Thread the Pump retaining screw to secure the Pump in place by turning the
screw clockwise until it stops. Conrm that the retaining screw is visible in one
of the notches on the side of the pump. If the retaining screw is not visible in a
notch, loosen the retaining screw, remove the pump, remount, rotate the Pump
counterclockwise, and secure by advancing the retaining screw.
Pumping operation may now be re-established via the backup Console
by increasing the RPM to about 1000 RPM. Gradually increase RPM while
unclamping the Pump outow tubing. Continue to gradually increase the RPM
to achieve the desired ow. Pumping operation should be returned to the last
operating condition of the Pump. Be sure to reset options and alarms to match
those selected before hardware exchange.
CentriMagTM Circulatory Support System
Alarms and Alerts Quick Reference Guide
NOTE: To be available for emergency use, the backup Console, Motor and Flow
Probe must be transported with and in proximity of the main Console and Motor,
and plugged into AC power, with the main power switch OFF. The battery power for
the backup Console must be periodically assessed by turning the main power switch
ON, checking the status of the battery, and then turning the main power switch OFF.
In addition, the expiration date and maintenance due date of the backup Console’s
Battery, must be checked on a periodical basis.
For additional information, refer to the CentriMag Circulatory Support System
Operation Manual.
1
3
4Rotate the Pump body clockwise until the grooves in the Pump match
the Motor. Lift the Pump from the receptacle.
5
6
7
8
If the Pump has been OFF for more than ve minutes without ade-
quate anticoagulation, or if there has been a Motor Overheating con-
dition, it will be necessary to replace the Pump and circuit, including
cannulae.
2
Clamp the return tubing (outlet side).
1
2
3
4
5
6
7
8

Other manuals for CentriMag

2

Other Abbott Medical Equipment manuals

Abbott Heartmate 3 User manual

Abbott

Abbott Heartmate 3 User manual

Abbott CardioMEMS HF User manual

Abbott

Abbott CardioMEMS HF User manual

Abbott Whitestar Signature User manual

Abbott

Abbott Whitestar Signature User manual

Abbott Heartmate 3 User manual

Abbott

Abbott Heartmate 3 User manual

Abbott CARDIOMEMS User manual

Abbott

Abbott CARDIOMEMS User manual

Abbott DIGIVAL LFR-000 User manual

Abbott

Abbott DIGIVAL LFR-000 User manual

Abbott FreeStyle Precision Pro User manual

Abbott

Abbott FreeStyle Precision Pro User manual

Abbott Absorb GT1 Bioresorbable Vascular Scaffold... User manual

Abbott

Abbott Absorb GT1 Bioresorbable Vascular Scaffold... User manual

Abbott NT000iX User manual

Abbott

Abbott NT000iX User manual

Abbott Advisor  FL Circular Mapping Catheter, Sensor... User manual

Abbott

Abbott Advisor FL Circular Mapping Catheter, Sensor... User manual

Abbott XIENCE Sierra User manual

Abbott

Abbott XIENCE Sierra User manual

Abbott St. Jude Medical Patient Controller 3875 User manual

Abbott

Abbott St. Jude Medical Patient Controller 3875 User manual

Abbott StarClose SE User manual

Abbott

Abbott StarClose SE User manual

Abbott FreeStyle Libre 3 User manual

Abbott

Abbott FreeStyle Libre 3 User manual

Abbott myMerlin Installation instructions

Abbott

Abbott myMerlin Installation instructions

Abbott CentriMag User manual

Abbott

Abbott CentriMag User manual

Abbott AMPLATZER Sizing Balloon II Series User manual

Abbott

Abbott AMPLATZER Sizing Balloon II Series User manual

Abbott CARDIOMEMS HF User manual

Abbott

Abbott CARDIOMEMS HF User manual

Abbott HeartMate Touch User manual

Abbott

Abbott HeartMate Touch User manual

Abbott AMPLATZER TorqVue LP User manual

Abbott

Abbott AMPLATZER TorqVue LP User manual

Abbott FreeStyle Libre2 User manual

Abbott

Abbott FreeStyle Libre2 User manual

Abbott i-STAT Alinity User manual

Abbott

Abbott i-STAT Alinity User manual

Abbott e i-STAT Alinity User manual

Abbott

Abbott e i-STAT Alinity User manual

Abbott AMPLATZER TorqVue Series User manual

Abbott

Abbott AMPLATZER TorqVue Series User manual

Popular Medical Equipment manuals by other brands

BTL 5000 Shockwave Series user manual

BTL

BTL 5000 Shockwave Series user manual

Ultrasound Technologies PD1+ combi operating instructions

Ultrasound Technologies

Ultrasound Technologies PD1+ combi operating instructions

Promeba PA-179 instruction manual

Promeba

Promeba PA-179 instruction manual

Gima 27793 quick start guide

Gima

Gima 27793 quick start guide

Charder HM202P user manual

Charder

Charder HM202P user manual

Covidien Force FX Service manual

Covidien

Covidien Force FX Service manual

Ossur Equalizer Air Walker Instructions for use

Ossur

Ossur Equalizer Air Walker Instructions for use

Amico ICE 25 Installation and operation instruction manual

Amico

Amico ICE 25 Installation and operation instruction manual

Ambu aScope 2 quick guide

Ambu

Ambu aScope 2 quick guide

Drager Savina 5.0 Repair Procedure

Drager

Drager Savina 5.0 Repair Procedure

GETINGE GROUP Maquet CARDIOSAVE IABP Quick reference guide

GETINGE GROUP

GETINGE GROUP Maquet CARDIOSAVE IABP Quick reference guide

Arjo HUNTLEIGH Sonicaid Team3 Instructions for use

Arjo

Arjo HUNTLEIGH Sonicaid Team3 Instructions for use

NSK Ti-Max Z Z-SG45L Operation manual

NSK

NSK Ti-Max Z Z-SG45L Operation manual

Saadat Dena 350 Service manual

Saadat

Saadat Dena 350 Service manual

SMARTfit Mini user manual

SMARTfit

SMARTfit Mini user manual

Tomey SP-100 Operation manual

Tomey

Tomey SP-100 Operation manual

FIOR & GENTZ e-Cast quick start guide

FIOR & GENTZ

FIOR & GENTZ e-Cast quick start guide

medi Lumbamed basic Instructions for use

medi

medi Lumbamed basic Instructions for use

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.