Accuris ASPIRE User manual

Rev 5/5/16
Accuris Instruments / Benchmark Scientific
Ph: (908) 769-5555
(C) Benchmark Scientific, 2016
ASPIRE Laboratory Aspirator
USER MANUAL

2
THE ACCURIS ASPIRE LABORATORY ASPIRATOR is a portable instrument that operates on 115V~/60Hz or
230V~/50Hz power input. It is designed for the aspiration of liquid samples and suitable for removal of tissue
culture media and supernatant from laboratory plates, tubes and vessels.
The ASPIRE is housed in a heat-resistant, and electrically insulated plastic casing, incorporates a vacuum regulator
on the front panel, and also includes a polycarbonate autoclavable bottle and lid with overflow protection valve,
silicone tubing, hydrophobic filter, and handheld vacuum pipette.
GENERAL WARNING
READ THE INSTRUCTION MANUAL IN ITS ENTIRETY AND CAREFULLY BEFORE USE
THE INSTRUMENT SHOULD ONLY BE USED BY PROPERLY TRAINED LABORATORY PERSONNEL
THE INSTRUMENT’S PUMP UNIT MUST NOT BE DISASSEMBLED. FOR SERVICE ALWAYS CONTACT BENCHMARK
SCIENTIFIC
DO NOT LEAVE BLEACH OR BLEACH SOLUTION IN THE COLLECTION BOTTLE AS IT WILL DAMAGE THE
POLYCARBONATE MATERIAL.
IMPORTANT SAFETY INSTRUCTIONS
1. Check the condition of the unit before each use. The housing, collection bottle, and all components should be
carefully inspected for visual damage.
2. Check the main power cable and do not connect to power if damage is apparent;
3. Before connecting to power, check that the electric input data indicated on the data label and the type of plug
used, correspond to that of the electrical outlet and electrical supply
4. If the plug supplied with the appliance is incompatible with the power supply outlet, contact qualified staff for
replacement of the plug with a suitable type. The use of simple or multiple and / or extension adapters is not
generally recommended.
5. Always follow general safety requirements for electrical appliances:
Use original components and accessories provided by Benchmark Scientific to guarantee performance and
safety;
Do not operate the aspirator without the bacteriological filter in place between the collection bottle and
vacuum;
Never immerse the aspirator pump unit in water or other liquids;
Install the aspirator on a stable and flat surface;
Position the instrument in a way that the air vents on the back and sides are not obstructed;
Don’t use in the presence of flammable gases;
Don’t touch the device with wet hands and always prevent the appliance from coming into contact with
liquids;
Keep out of reach of children and unauthorized personnel;
Always use care when removing the power cord from the instrument;
Always use the instrument at a safe distance from heat sources;

3
6. For repairs, contact Benchmark Scientific technical services at 908-769-5555 and always use original spare
parts and accessories. Failure to comply with the
above can jeopardise the safety of the user of the device;
7. This laboratory instrument must be used exclusively for its intended applications as described in this
manual.
8. Any different use could be dangerous; the manufacturer will not be responsible for damage due to improper
use or connection to an electrical system not complying with current regulations;
9. Internal electrical and mechanical parts are not designed to be repaired by customers or end-users. Do not
open the pump unit, do not mishandle the electric / mechanical parts. Always contact Benchmark Scientific for
technical assistance
10. Using the device in environmental conditions different than those indicated in this manual may be dangerous to
the user and may cause damage to the instrument.
TECHNICAL SPECIFICATIONS
MODEL
ACCURIS ASPIRE LABORATORY ASPIRATOR
VOLTAGE
115V,50/60Hz or 230V, 50/60Hz
POWER CONSUMPTION
105VA / 184VA
FUSE
F 1 x 4A 250V / F 1 x 1.6A 250V
MAXIMUM SUCTION PRESSURE (without jar)
-75kPa (-0.75 Bar)
Adjustable from -75kPa (-0.75bar) to -10kPa (-0.10 bar)
MAXIMUM SUCTION FLOW (without bottle / filter)
15 l/min
WEIGHT
4.8 Lbs / 2.2 Kg
DIMENSIONS
250mm x 190mm x 160 mm
SICILICONE TUBE SIZE
Ø 6 x 10 mm
ACCURANCY OF VACUUM INDICATOR
± 5%
AMBIENT OPERATING CONDITIONS
Room temperature: 5 to 35°C
Room humidity percentage: 30 to 75% RH
Altitude: 0 to 2000m
STORAGE AND TRASPORT CONDITIONS
Room temperature: -40 to 70°C
Room humidity percentage: 10 to 100% RH
SYMBOLS
Class II isolation equipment
Warning, consult the instruction manual
Keep in a cool, dry place
Storage/transport temperature: -40 ÷ 70°C
Type B equipment
Fuse

4
~
Alternate Current
Hz
Mains Frequency
I
ON
O
OFF
Please note technical specifications may change without notice.
Guidance and manufacturer’s declaration – Electromagnetic Emissions
The ACCURIS ASPIRE LABORATORY ASPIRATOR is intended for use in the electromagnetic environment
specified below.
The customers or the user of the ACCURIS ASPIRE LABORATORY ASPIRATOR should assure that it’s used in
such an environment.
Emissions Test
Compliance
Electromagnetic environment -
guidance
Power disturbance
CISPR11
Group 1
THE ACCURIS LABORATORY
ASPIRATOR
RF emissions are very low and will
not cause interference in proximity
to any Electronic appliances.
Irradiated / Conducted emissions
CISPR11
Class [B]
THE ACCURIS LABORATORY
ASPIRATOR can be used in all
environments, including domestic
and those connected directly to the
public mains distribution that
supplies power to environments
used for domestic scopes.
Harmonic emissions IEC/EN
61000-3-2
Class [A]
Voltage fluctuations / flicker
emissions
IEC/EN 61000-3-3
Complies
Guidance and manufacturer’s declaration – Electromagnetic Immunity
THE ACCURIS LABORATORY ASPIRATOR is intended for use in the electromagnetic environment specified
below.
Users of this product should assure that it’s used in such an environment.
Immunity Test
Compliance
Electromagnetic environments -
guidance
Electrostatic discharge (ESD)
IEC/EN 61000-4-2
± 6kV on contact
± 8kV in air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast transient / burst
IEC/EN 61000-4-4
± 2kV power supply
Mains power quality should be that
of a typical commercial
environment.
Surge
IEC/EN 61000-4-5
± 1kV differential mode
Mains power quality should be that
of a typical commercial
environment.
Loss of voltage, brief voltage
interruptions and variations
IEC/EN 61000-4-11
5%UTfor 0.5 cycle
40%UTfor 05 cycle
70%UTfor 25 cycle
<5%UT
for 5 sec
Mains power quality should be that
of a typical commercial
environment.
Magnetic field
3A/m
The power frequency magnetic field

5
IEC/EN 61000-4-8
should be measured in the intended
installation location to assure that
it’s sufficiently low.
Conducted Immunity
IEC/EN 61000-4-6
3Vrms 150kHz to 80MHz
(for appliances that aren’t life -
supporting)
-
Irradiated Conducted
IEC/EN 61000-4-3
3V/m 80MHz to 2.5 GHz
(for appliances that aren’t life -
equipment)
-
Note UTis the value of the power supply voltage
ACCESSORIES SUPPLIED
DESCRIPTION
COMPLETE COLLECTION BOTTLE
2000ml
TUBING SET (3 pieces)
HANDHELD VACUUM CONTROLLER
2 ADAPTERS (PIPETTE TIP AND FINE
POINT METAL TUBE)
ANTIBACTERIAL FILTER
Antibacterial Filter: The filter is produced with (PTFE) hydrophobic material witch prevents fluids entering the
pump unit and also offers protection to prevent bacterial contamination of the pump unit. When the filter is wet, it’s
not possible to use the unit therefore the filter should be changed immediately.
In case of possible contamination or discoloration of the filter, change the filter immediately.
Do not use the instrument without the filter properly installed.
CLEANING OF ACCESSORIES
To clean the plastic housing of the device wear disposable protective gloves and clean with denatured alcohol or
hypochlorite solutions.
Washing and / or cleaning the autoclavable collection bottle can be carried out as follows:
Wear protective gloves and apron (glasses and face mask if necessary) to avoid contact with
contaminating substances;
Disconnect the bottle from the device
Disconnect all tubes from the jar and the filter
Empty and properly dispose of the contents in the collection bottle
Separate all parts of the cover (overflow valve, o-ring);
Thoroughly rinse all parts in cold water then soak in warm water (temperature not to exceed 60°C). Wash
thoroughly and if necessary use a non-abrasive brush to remove and deposits. Rinse in running warm water and
dry all parts with a soft cloth (non-abrasive).
Do not use leave bleach solutions in the collection bottle. If required for disinfection, use a 10%
bleach to water solution for cleaning, but then rinse thoroughly with water. Bleach can damage the
polycarbonate material of the collection bottle.
Table of contents
Other Accuris Laboratory Equipment manuals

Accuris
Accuris isoPURE Mini User manual

Accuris
Accuris SmartDoc 2.0 User manual

Accuris
Accuris SmartDoc 2.0 User manual

Accuris
Accuris Smartdoc 2.0 User manual

Accuris
Accuris AutoMATE 96 User manual

Accuris
Accuris SMARTREADER MULTIMODE User manual

Accuris
Accuris SmartDrop L NS1000 User manual

Accuris
Accuris MyGel Mini User manual
Popular Laboratory Equipment manuals by other brands

Belden
Belden HIRSCHMANN RPI-P1-4PoE installation manual

Koehler
Koehler K1223 Series Operation and instruction manual

Globe Scientific
Globe Scientific GCM-12 quick start guide

Getinge
Getinge 86 SERIES Technical manual

CORNING
CORNING Everon 6000 user manual

Biocomp
Biocomp GRADIENT MASTER 108 operating manual