ADE KWE Series User manual

OPERATING MANUAL
Stainless Steel Scales
Compact Scale in Stainless Steel Housing –
Series KWE
KWE-190107-Rev008-UM-en
ADE Germany GmbH & Co.KG
Hammer Steindamm 27 – 29
22089 Hamburg / GERMANY
+49 (0) 40 432776-0
Fax +49 (0) 40 432776-10
E-mail: info@ade-hamburg.de

1 KWE-190107-Rev008-UM-en
Contents
1.
Components and Features ................................................................................................. 2
Key Function ................................................................................................................ 2
2.
Installation and use of scale .............................................................................................. 2
2.1
Check carton contents ..................................................................................... 2
2.2
Positioning of the scale..................................................................................... 2
2.3
Power supply of the scale................................................................................. 3
2.4
Power immediate operation .............................................................................. 3
2.5
Power Charge Operation (only for models with rechargeable batteries): ............. 3
2.6
Switch on......................................................................................................... 4
2.7
Zero Operation ................................................................................................. 4
2.8
Displaying the voltage....................................................................................... 4
2.9
Tare Operation ................................................................................................. 4
2.10
Unit operation .................................................................................................. 5
2.11
Check weighing – Minimum and maximum values with acoustical alarm ............ 5
2.12
Counting Function ............................................................................................ 6
3.
Environment Parameter Setting......................................................................................... 6
3.1
Adjustment display precision ............................................................................ 7
3.2
Weighing unit (kg, g, lbs) .................................................................................. 7
3.3
Zero tracking.................................................................................................... 7
3.4
Stability Filter.................................................................................................... 7
3.5
Power Saving Mode / Standby (only for models with rechargeable batteries) ..... 8
3.6
Display Update Speed ...................................................................................... 8
3.7
Creep Tracking Range ...................................................................................... 8
3.8
Printer setting................................................................................................... 9
4.
Display Symbols.................................................................................................................. 9
5.
Display Symbols.................................................................................................................. 9
6.
Cleaning and Maintenance............................................................................................... 10
6.1
Cleaning......................................................................................................... 10
6.2
Storage and Transport.................................................................................... 10
6.3
Disposal of Waste .......................................................................................... 10
6.4
Warranty ........................................................................................................ 10
6.5
Safety Instructions.......................................................................................... 10
7.
Declaration of Conformity ................................................................................................ 11
8.
Technical Data .................................................................................................................. 12

KWE-190107-Rev008-UM-en 2
1. Components and Features
Key Function
Symbol Short Description Designation
[OFF]
Off-Button
[ON/ZERO]
On- and Zero-Button
[SET]
Tare-Button / Enter-Button
[TARE]
Set up und Configuration-Button
(in combination with another
button)
2. Installation and use of scale
Your ADE compact scale KWE in stainless steel housing is a quality product, which is designed
for a lot of different weighing performances for controlling, where approval is not necessary. The
scale has both-sided displays.
Four selectable graduations give the user different possibilities to adjust the displayed results to
the user’s intention.
The integrated program for piece counting offers the user easy piece counting.
In addition to that there can be stored minimum and maximum values combined with an
acoustical signal in the scale, e. g. for weighing always the same weights.
The stainless steel housing allows using the scale in humid working environments.
The scale is a precision instrument and must be treated accordingly. Therefore kindly read and
follow the Operating Manual in all respects.
2.1 Check carton contents
Compact Scale
Weighing platform
Operating manual including Certificate of Conformity
Rechargeable batteries (only for models with rechargeable batteries included)
2.2 Positioning of the scale
Make sure all four feet of the platform have direct contact to the surface.
The spirit level of the scale is an additional tool to make sure the scale is carefully positioned
and offers best results when the air bubble is centered.
The alignment of the scale must be checked and corrected every time the location of the scale
is changed.

3 KWE-190107-Rev008-UM-en
Place the metal weighing tray on the platform.
Connect the mains adapter to the scale.
2.3 Power supply of the scale
Power supply is by mains with directly connected power plug (230V) or with rechargeable
batteries (only for models with rechargeable batteries).
The battery pack is already installed in the base of the scale.
For the first loading of the rechargeable batteries please take the power line cable out of the
battery box inside the scale (at the bottom side of the scale) and connect it with the local
electricity grid. Complete first charging of power-packs needs approx. 10 hours. Please make
sure that the local voltage supply corresponds to the indication printed on the adapter.
2.4 Power immediate operation
Working with mains all three LCD („High“, „Mid“, „Low“) are on.
For models with rechargeable batteries:
•All three segments lighting = Power voltage of rechargeable batteries is more than 6.3V,
•Two segments lighting = Power voltage of rechargeable batteries is between 6V und
6.3V,
•One segment lighting = Power voltage of rechargeable batteries is under 6V.
2.5 Power Charge Operation (only for models with rechargeable batteries):
When the power voltage indicator shows just one segment, scale should be charged by the
charger in battery storehouse. If not, the scale would turn off in between 15 minutes.
Full charging needs about approx. 10 hours.
The battery should be charged at least once every 3 months.
If the running time gets more and more less, rechargeable batteries have to be replaced. Please
contact your local dealer.
Please take sure that used voltage is 230V/50Hz.

KWE-190107-Rev008-UM-en 4
2.6 Switch on
Press [ON]. The whole display illuminates with a serial of steps and ends with the screen
display „0“. After the [ZERO] indicator is lighting, the scale is ready for weighing.
2.6.1. Switch-off
Scale can be turned off forcibly by pressing [OFF].
2.6.2. Auto-off / Standby (only working with rechargeable batteries)
The scale has an automatic shut-off function, which can be activated or deactivated in
parameter setting.
Setting could be automatic shut-off after 10 minutes or deactivated shut-off. In this case
depending on the setting the backlight is always on or shut-off after 40 seconds showing “0”.
If you place something on the platform, the scale goes back in weighing modus.
2.6.3. Low voltage auto-off
If the voltage of the rechargeable batteries is less than 5.6V, the scales turn off automatically.
2.7 Zero Operation
When scale is on, [ON] is the key to zero. It is only available within 4% of the maximum
capacity; otherwise it will immediately display “HHHHHH“.
2.8 Displaying the voltage
When scale is on, it will display the voltage when pressing [ON] key for 3 seconds. Now if press
[ON] again, the scale will return to weighing mode.
2.9 Tare Operation
Place the container on the scale until the display indicates the weight number steadily.
Press [TARE], then the display indicates “0” and the TARE indicator is lighting.
Place the weighing objects into the container, and then scale shows the Net weight.
Move the container and the weighing objects off, and then scale will show the minus weight of
the container. Then press [TARE] again, the tare weight can be cleared, that the TARE indicate
out and the scale is set to zero.
If the display is not stable, no arrow is shown.

5 KWE-190107-Rev008-UM-en
2.10 Unit operation
When scale is on, press [TARE] for more than 3 seconds to change the weight unit between g,
kg and lbs. Requirement is a stable weight.
Selected unit will be stored, when scale is switched off.
2.11 Check weighing – Minimum and maximum values with acoustical alarm
For setting minimum and maximum values, scale has to be empty.
Now press both [SET] and [ON] together for 3 seconds.
Display will show “-no-“ (in delivery condition). Press [TARE], it will circularly display “-no-“,
“-in-”, “-out-“ and again “-no-“:
„-no” deactivates the alarm function.
„-in-” activates the alarm within the range.
„-out-“ activates the alarm out of the range.
When display shows “-in-“ or “-out-“, press [ON] to confirm the choice.
When “-in-“ or “-out-“ is confirmed now the minimum value can be chosen. The display shows
an arrow at “Under” and the first digit is lighting. The light shows the value set last time and the
digit which can be adjusted is blinking. Now press [SET] to change the digit, press [TARE] to
change the digit value, and press [ON] to confirm.
Then the setting of minimum value is done and the arrow at “Over” is lighting. Use the same
way to finish the setting of the maximum value.
When finished setting, it returns to weighing mode and the weighing alarm function is on.
If the weight is upper the maximum value, the “Over” indicator is lighting, while arrow to “Under”
is lighting, when weight is under minimum value.
If the weight is between minimum and maximum values the arrow next to “Accept” is lighting.
If activated (setting “-in-“ or “-out-“), control weighing is supported by an acoustic signal, which
sounds, if the weight is in between or out of the minimum and maximum values (as set in the
parameter; see Cap. 3.8).
For zeroing minimum and maximum values, set both values back to 000000 or deactivate
function by choosing “–no-”.
The settings will be stored in memory when scale is switched off.
If changing the weighing unit, this function will be closed and the current
setting value will be auto-cleared.
If changing the graduation, this function will be closed and the current setting
value will be cleared.

KWE-190107-Rev008-UM-en 6
2.12 Counting Function
Switch on the scale and put on the reference number of pieces.
This function can be activated only with loaded platform!
Press both [SET] and [ON] keys together for 3 seconds. Display shows “pcs_y”.
Press [TARE] to deactivate the counting function. Display shows “pcs_n”.
To confirm your choice press [ON].
When choosing „pcs_y“ press [ON]. Display will show “000000” and the first digit of reference
number of pieces is blinking. Now press [SET] to change the blinking digit, press [TARE] to
change the blinking digit value. Following this way enter the reference number of pieces put on
the platform and press [ON] to confirm. With setting the scale will store reference number and
will automatically turn into counting mode. Arrow at “PCS” will show activated counting mode.
Now place an arbitrary number of pieces on the weighing platform and display shows the
number.
The settings will be stored in memory when scale is switched off.
For returning to weighing mode press both [SET] and [ON] together for 3 seconds, and choose
“pcs_n” by pressing [TARE] Confirm your choice with [ON].
The setting will be stored and scale return to weighing mode. When scale is switched on next
time, it starts in weighing mode.
Piece counting function doesn’t work, if unit “lbs” is selected.
3. Environment Parameter Setting
The user has different parameters to adjust the scale to his preferences.
The following parameters can be adjusted:
PrE (Adjustment display precision)
UNT (Adjustment units)
ZEr (Zero tracking)
FIt (Stability Filter)
SAU (Power Saving Mode / Standby)
SPd (Display Update Speed)
drl (Setting for zero tracking)
Prt (Printer setting)
For activating parameter setting press [SET] and [TARE] on same time for 3 seconds. There will
appear menu „PrE“ to adjust the display precision.
To enter the following menus, press [ON] again and again. Within the menu you can
change adjustments by pressing [TARE]. Pressing [ON] confirms the choice, stores
setting and changes to next parameter.
By pressing [SET] you’ll return to weighing mode without retention of the setting.

7 KWE-190107-Rev008-UM-en
3.1 Adjustment display precision
Please start as described in the beginning of this chapter to enter this menu and „PrE“ is shown
in the display.
Now you can change between four given graduations as shown below by pressing [TARE].
Capacity: 3kg 6kg 15kg
Graduation Adjustment 0: 1g 2g 5g
Graduation Adjustment 1: 0.5g 1g 2g
Graduation Adjustment 2: 0.2g 0.5g 1g
Graduation Adjustment 3: 0.1g 0.2g 0.5g
3.2 Weighing unit (kg, g, lbs)
Please start as described in the beginning of chapter 3 to enter this menu and „UNT_“ is shown
in the display.
You can choose between the units by pressing [TARE]. Default is “UNT_0”.
Adjustment 0: g
Adjustment 1: kg
Adjustment 2: lbs
The setting will be stored when scale is switched off.
3.3 Zero tracking
Please start as described in the beginning of chapter 3 to enter this menu and „Zer_“ is shown
in the display.
In this menu the zero tracking can be adjusted, if there are weighing performances with weight
increasing, which are very slow.
There can be chosen between 11 adjustments (0 – 5 in half steps) by pressing [TARE] again
and again. Default is “Zer_2”.
Remark: If you have selected a higher graduation („PrE“; 3.1 Adjustment display precision),
select a high value for stability filter.
3.4 Stability Filter
Please start as described in the beginning of chapter 3 to enter this menu and „Flt_“ is shown in
the display.
To change the adjustment of the filter stability can be helpful if working environments make it
sensible, e. g. by draughts. If you chose a higher stability filter, it changes also the time to
stabilization the scale.
Please choose between four adjustments the stability filter by pressing [TARE]. Default is
“Flt_0”.
Adjustment 0: smallest filter = fast stabilization
Adjustment 1 + 2
Adjustment 3: biggest filter = slower stabilization
Remark: If you have selected a higher graduation („PrE“; 3.1 Adjustment display precision),
select a high value for stability filter.

KWE-190107-Rev008-UM-en 8
3.5 Power Saving Mode / Standby (only for models with rechargeable batteries)
Please start as described in the beginning of chapter 3 to enter this menu and „SAU“ is shown
in the display.
By pressing [TARE] the Standby function can be enabled or disabled. Default is “SAU_2”.
Adjustment 0: Standby disabled. Backlight of display is always on.
Adjustment 1: Standby enabled. After 40 seconds display is showing “0”, backlight
automatically switches off. By putting something on the weighing platform
again, backlight of display will switch on again.
Adjustment 2: Standby enabled. After 40 seconds display is showing “0”, backlight
automatically switches off. By putting something on the weighing platform
again, backlight of display will switch on again. If nothing will be put on the
weighing platform within ten minutes, the scale automatically switches off.
3.6 Display Update Speed
Please start as described in the beginning of this chapter to enter this menu and „SPd“ is
shown in the display.
By pressing [TARE] the speed of the displayed weighing results can be adjusted. Default is
“SPd_1”.
Adjustment 0: Speed low
Adjustment 1: Speed high
3.7 Creep Tracking Range
Please start as described in the beginning of chapter 3 to enter this menu and „drl_“ is shown in
the display.
There can be chosen between 5 adjustments (0 – 4) by pressing [TARE] again and again.
The higher the value the higher tracking range. Default is “drl_1”.

9 KWE-190107-Rev008-UM-en
3.8 Printer setting
Please start as described in the beginning of chapter 3 to enter this menu and „Prt_“ is shown
in the display.
Setting has to be “Prt_0”, because scale has no interface for connecting PC or printer.
4. Display Symbols
SP – x.x Version number
„xxxxxx“ Software number
“CAPxx.0” Capacity / Maximum load of the scale
„nnnnnn“ Scale is overloaded. The current weighing is higher than the valid
capacity.
“------“ Data is unstable
“LLLLLL” Output of load cell is lower.
“HHHHHH” Output of load cell is higher or turn on zero is higher than range.
- Lo - Low battery – Recharge the rechargeable battery immediately.
FULL Rechargeable battery has been charged fully.
“_Inlt_” Faulty weighing
5. Display Symbols
Fault Cause Solution
Faulty weighing result Parameter of load cells have
changed
Re-calibration
Display shows „HHHHHHHH“ Over zero range Re-calibration
Load cell or PCB damaged
Display shows „LLLLLLLL“ Load cell or PCB damaged
Rear display doesn’t show
anything
Parameter fault Correction of setting as
shown in the manual
Load cell damaged
Rechargeable battery is not
loading although loading is
shown
Rechargeable battery is
damaged
Replace rechargeable battery
Rechargeable battery is not
loading and no loading is
shown
Charging unit for
rechargeable battery is
damaged
Quickly discharging of
rechargeable battery
Old rechargeable battery Replace rechargeable battery
Display shows „_InIt_“ Faulty weighing Re-start the scale or Re-
calibration

KWE-190107-Rev008-UM-en 10
6. Cleaning and Maintenance
6.1 Cleaning
Before cleaning make sure the scale has been disconnected from the socket. For cleaning just
use a damp cloth or common disinfection products. Do not use any aggressive cleaning fluids.
Models without rechargeable battery can be cleaned under running water.
Please make sure before starting cleaning, if there’s a rechargeable battery inside.
Otherwise the scale or the rechargeable battery can be damaged.
If a rechargeable battery is inside, make sure no water or cleaning fluid penetrates
the scale.
It isn’t allowed to clean the scale with a high-pressure cleaner!
6.2 Storage and Transport
Keep all original parts and pieces for an eventual sending back of the scale in order to avoid any
possible damage during transport which automatically leads to no warranty.
6.3 Disposal of Waste
Never dispose the scale at the end of life to domestic waste.
Batteries and rechargeable batteries are not allowed in domestic waste and have
to be disposed fully discharged or with insulated poles. Otherwise there’s a danger
of an electrical short!
Please ask at your local administration for an environmentally correct disposal.
6.4 Warranty
A one year warranty period from date of purchase will be guaranteed for either repair or
replacement once it is proven that the defects are caused by poor material or poor
workmanship (please keep the receipt for presentation). All moveable parts such as batteries,
cables, plug-in power supply units, rechargeable batteries etc. are excluded. Warranty does not
apply to normal appearance of usage or sign of wear or any damages caused by accidents or
misuse. Products which have been opened by unauthorized persons are automatically out of
warranty.
Customers abroad are requested to contact their local dealers in the event of any warranty
becoming applicable.
6.5 Safety Instructions
Do NOT use the scale
:
•If the mains plug or cable shows visible damages
In this case contact your local dealer or service partner.

11 KWE-190107-Rev008-UM-en
Manufacturers Declaration of Conformity
This product where manufactured in accordance with the harmonised European Standards.
It corresponds to the regulations of below listed EC-directives:
– Low Voltage Directive 2014/35/EU
– EMC Directive 2014/30/EU regarding electromagnetic compatibility
– RoHS Directive 2011/65/EG
as appropriate amended.
This declaration loses their validity, in case modifications on the scales are made without manufactures
approval.
Hamburg, January 2019
ADE Germany GmbH & Co.KG
Hammer Steindamm 27 – 29, 22089 Hamburg
Germany
7. Declaration of Conformity
Manufacturer: ADE Germany GmbH & Co.KG
Hammer Steindamm 27 – 29
D - 22089 Hamburg
Germany
+49 (0) 40 432776-0
Fax +49 (0) 40 432776-10

KWE-190107-Rev008-UM-en 12
8. Technical Data
Type KWE3-IP65 KWE6-IP65 KWE15-IP65 / KWE15-IP68
Capacity 3 kg 6 kg 15 kg
Graduation (d) 0.1/0.2/0.5/1g 0.2/0.5/1/2g 0.5/1/2/5g
Min. Weight 20d
Maximum tare 100% maximum weight
Model KWE-IP65 (with rechargeable battery) KWE15-IP68
Weighing platform 220 x 180 mm 220 x 180 mm
Dimensions 230 x 290 x 115 mm 230 x 290 x 115 mm
Dead Weight 4.1 kg 3.8 kg
Connecting cable 1.00 m 2.00 m
Operating Temperature -10°C to + 40°C -10°C to + 40°C
Internal Protection IP65 IP68
Power Supply 230V mains 230V mains
and rechargeable battery 6,3V (6V4Ah) (refitting with rechargeable
battery not possible)
Display Large LCD-Display with second display on backside,
height of digit 20mm
Casing Stable stainless steel housing
Weighing platform stainless steel
Integrated Piece Counting
Free programmable high and low values for check weighing with acoustical support
4 different graduations selectable

13 KWE-190107-Rev008-UM-en

KWE-190107-Rev008-UM-en 14

15 KWE-190107-Rev008-UM-en
Händler:
This manual suits for next models
4