Advantech AVAS-402 User manual

AVAS-402
Medical Video recorder

Instructions for the User
The document combines text and illustrations, providing a comprehensive
overview of the system. The information is presented as a sequential series
of actions, allowing the user to learn directly how to use the device.
The text provides explanations and instructs the user step-by-step in the
practical use of the product, with short, clear instructions in an
easy-to-follow sequence.
Definitions
Warning! A WARNING statement provides important information about
a potentially hazardous situation which, if not avoided, could
result in death or serious injury.
Caution! A CAUTION statement provides important information about a
potentially hazardous situation which, if not avoided, may
result in minor or moderate injury to the user or patient or in
damage to the equipment or other property.
Note! A NOTE provides additional information intended to avoid
inconveniences during operation.
Safety Instructions
1. Strictly follow these Instructions for Use; please read these safety
instructions carefully.
2. Please keep this User Manual for later reference; any use of the
product requires full understanding and strict observation of all
portions of these instructions. Observe all WARNINGS and
CAUTIONS as rendered throughout this manual and on labels on the
equipment.
3. Repair of the device may only be carried out by trained service
personnel. Advantech recommends that a service contract be
obtained with Advantech Service and that all repairs also be carried
out by them. Otherwise the correct functioning of the device may be
compromised.
Warning! Because of the danger of electric shock, never remove the
cover of a device while it is in operation or connected to a
power outlet.
4. If one of the following situations arises, have the equipment checked by service
personnel:
The power cord or plug is damaged.
Liquid has penetrated into the equipment.
The equipment has been exposed to moisture.
The equipment does not work well, or you cannot get it to work according to
the user manual.
The equipment has been dropped and damaged.

The equipment has obvious signs of breakage.
5. Disconnect this equipment from any AC outlet before cleaning. Use a damp
cloth. Do not use liquid or spray detergents for cleaning and keep this
equipment away from humidity.
Caution! To avoid short-circuiting and otherwise damaging the device, do not
allow fluids to come in contact with the device. If fluids are accidentally
spilled on the equipment, remove the affected unit from service as soon
as possible and contact the service personnel to verify that patient
safety is not compromised.
6. Put this equipment on a reliable surface during installation. Dropping it or letting
it fall may cause damage. For plug-in equipment, the power outlet socket must
be located near the equipment and must be easily accessible.
Caution! To prevent overheating, do not cover the openings or place the device
in direct sunlight or near radiant heaters.
7. Make sure the voltage of the power source is correct before connecting the
equipment to the power outlet. Position the power cord so that people cannot
step on it. Do not place anything over the power cord. If the equipment is not
used for a long time, disconnect it from the power source to avoid damage by
transient over voltage.
Caution! Do not leave this equipment in an uncontrolled environment where the
storage temperature is below -20°C (-4°F) or above 60°C (140°F).
This may damage the equipment.
8. If your computer does not keep the correct time or the BIOS configuration has
been reset to default, the battery may have no charge.
Caution! Do not replace battery yourself. Please contact a qualified technician or
your retailer.
The computer is provided with a battery- powered, real-time clock circuit.
There is a danger of explosion if battery is incorrectly replaced. Replace
only with same or equivalent type recommended by the manufacturer.
Discard used batteries according to the manufacturer’s instructions.
Caution! The battery charging indicator is not included with this device. It will be
added to the finished system assembly and be shown with the
completed system.
9. Classification:
1). Supply Class I adapter
2). No applied part
3). Continuous Operation
4). Not AP or APG category
Warning! This device is not suitable for use in the presence of flammable anesthetic
mixture with air, oxygen, nitrous oxide, or for life support systems.
11. Environmental protection: follow national requirements to dispose of unit.
12. Maintenance: to properly maintain and clean the surfaces, use only the

approved products or clean with a dry applicator.
Caution! When servicing the device, always use replacement parts that are
qualified to Advantech standards. Advantech Digital Healthcare cannot
war-rant or endorse the safe performance of third-party replacement
parts for use with our medical device.
13. Make sure the user does not allow contact between SIP/SOPs and the patient
at the same time.
14. When networking with electrical devices, the operator is responsible for
ensuring that the resulting system meets the requirements set forth by the
following standards:
–EN 60601-1 (IEC 60601-1)
Medical electrical equipment
Part 1: General requirements for safety
–EN 60601-1-1 (IEC 60601-1-1)
Medical electrical equipment
Part 1-1: General requirements for safety
Collateral standard: Safety requirements for Medical electrical systems
–EN 60601-1-2 (IEC 60601-1-2)
Medical electrical equipment
Part 1-2: General requirements for safety
Collateral standard: Electromagnetic compatibility; Requirements and tests
15. Accessory equipment connected to analog and digital interfaces must be in
compliance with the respective nationally harmonized IEC standards (i.e. IEC
60950 for data processing equipment, IEC 60065 for video equipment, IEC
61010-1 for laboratory equipment, and IEC 60601-1 for medical equipment.)
-402 User Manual iv

Furthermore all configurations shall comply with the system standard IEC
60601-1- 1. Anyone who connects additional equipment to the signal input part
or signal output part is configuring a medical system, and is therefore,
responsible that the system complies with the requirements of the system
standard IEC 60601-1-1. The unit is for exclusive interconnection with IEC
60601-1 certified equipment in the patient environment and IEC 60XXX certified
equipment out-side of the patient environment. If in doubt, consult the technical
services department or your local representative.
Caution! Use suitable mounting apparatus to avoid risk of injury.
16. Grounding reliability can only be achieved when the equipment is connected
to an equivalent receptacle marked "Hospital Only" or "Hospital Grade".
17. Use a power cord that matches the voltage of the power outlet, which has been
approved and complies with the safety standard of your particular country.
Note! Environmental protection
Follow national requirements to dispose of unit.
18. “WARNING - Do not modify this equipment without authorization of the
manufacturer.”
19. “WARNING – To avoid risk of electric shock, this equipment must only be
connected to a supply mains with protective earth.
20. “CAUTION: This adapter Sinpro Power: Type: HPU101-107 is a forming part of the
medical device”.
21. Remove the power cord to fully turn off the device.
22. Installation is only to be carried out by trained and authorized personnel.
23. CAUTION! WARNING! If this device was not used in the manner which manufacturer
suggested, the end system shall evaluate for further compliance.
Explanation of Graphical Symbols
IEC 60878 and ISO 3864-B.3.6 : Warning: dangerous voltage
ISO 7000-0434 : Caution, consult ACCOMPANYING DOCUMENTS.
IEC 60417 -5009 : STAND-BY.
IEC 60417-5032 : Alternating Current
IEC 60417-5021 : Equipotentiality.

ISO 7010-M002: Follow instructions for us

Disposing of Old Products
Within the European Union
EU-wide legislation, as implemented in each member state,
requires that waste electrical and electronic products carrying the
mark shown at left must be disposed of separately from normal
household waste. This includes monitors and electrical
accessories, such as signal cables or power cords. When you need
to dis-pose of your display products, please follow the guidance of
your local authority, or ask the shop where you purchased the
product, or if applicable, follow any agreements made between you
and the provider.
The mark on electrical and electronic products only applies to the
current European Union Member States.
FCC Class B
This equipment has been tested and found to comply with the limits for a Class B
digital device, pursuant to Part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against harmful
interference when the equipment is operated in a residential environment. This
equipment generates, uses and can radiate radio frequency energy. If not installed
and used in accordance with this user manual, it may cause harmful interference to
radio communications.
Note that even when this equipment is installed and used in accordance with this
user manual, there is still no guarantee that interference will not occur. If this
equipment is believed to be causing harmful interference to radio or television
reception, this can be determined by turning the equipment on and off. If interference
is occur-ring, the user is encouraged to try to correct the interference by one or more
of the following measures:
Reorient or relocate the receiving antenna
Increase the separation between the equipment and the receiver
Connect the equipment to a power outlet on a circuit different from that to which
the receiver is connected
Consult the dealer or an experienced radio/TV technician for help
Warning! Any changes or modifications made to the equipment which are not
expressly approved by the relevant standard’s authority could void your
authority to operate the equipment.

List of Accessories
Before installing your Point-of-Care Terminal, ensure that the
following materials have been received:
AVAS-402
China RoHs note x1
Warning! No user serviceable parts inside; refer servicing to
qualified personnel. Only the accessories indicated
on the list of accessories above have been tested and
approved to be used with the device. Accordingly it is
strongly recommended that only these accessories
be used in conjunction with the specific device.
Otherwise the correct functioning of the device may
be compromised.
Additional Information and Assistance
Contact your distributor, sales representative, or Advantech's
customer service center for technical support if you need additional
assistance. Please have the following information ready before you
call:
Product name and serial number
Description of your peripheral attachments
Description of your software (operating system, version,
application software, etc.)
A complete description of the problem
The exact wording of any error messages
This equipment is a source of electromagnetic waves. Before
use please, make sure that there are not EMI sensitive
devices in its surrounding which may mal-function therefore.
Environmental protection
Follow national requirements to dispose of unit.
Manufacturer:
Advantech Co., Ltd.
No.1, Alley 20, Lane 26, Rueiguang Road Neihu District, Taipei,
Taiwan 114, R.O.C.
TEL: (02) 2792-7818
Distributed in Europe by:
Advantech Europe GmbH Kolberger Straße 7
D-40599 Düsseldorf, Germany
Tel: 49-211-97477350
Fax: 49-211-97477300
Visit the Advantech websites at www.advantech.com or
www.advantech.com.tw if you need more information.
General Information

1.1 Introduction
AVAS-402 is a video recorder that supports multi-channel
recording at one time. It has video capture I/O, video display I/O
and other I/Os. Video capture I/O includes HDMI, which supports
video data up to 4k resolution and @60 frame rate, as well as
Y/PbPr, CVBS, S-Video, SDI, HDMI, DVI-I which supports video
data up to FHD resolution and @60 frame rate. There could be
three I/Os for video display: DVI, Display Port, and HDMI. Dual-link
2560 x 1600 resolution and @60 frame rate is supported for DVI-D,
while both HDMI and Display Port support maximum resolution at
4096x2160 (4K)@60Hz frame rate
Other I/Os include RJ-45 ports, USB 3.0 ports and USB 2.0 ports,
RS-232 port, and Audio out/in 3.5mm stereo jack connectors.
With the SATA 3.0 interface, the AVAS-402 is a user-friendly
computer that system integrators can use this highly integrated
multimedia system to easily implement into your applications.
Apart from that, AVAS-402 provides SDKs and APIs for system
integrators to program their video applications efficiently. In other
chapters, the SDKs and APIs provided will be described in more
detail.
Intended use - AVAS-402 is intended to serve as a medical video
recorder for integration with hospital systems. AVAS-402 is
designed for general purpose used in medical image or video
related application in the hospital environment, for data collection
and for dis-playing information via connecting a display.
It should not be used as a life-support system.
The latest version of this user manual is available
for download from
http://support.advantech.com.tw/support/
1.2 Specifications
System
CPU
Intel® Core™ i7-7700 3.6GHz
Memory
16GB, Up to 32GB DDR4 2133MHz
Storage
1TB, based on customers’requirement
System input
rating
Input Voltage
100-240 AC
Video
Input Resolution
Up to 4K
Output Resolution
Up to 4K
Compression
Format
MPEG2 H.264, H265
Audio
Audio
2 (Mic-in/ Line-out) (Mic-in could be configured as "Line-in"
under Win OS.)
IO Port
Video Display
1 x DVI, 1 x Display Port, 1 x HDMI (can be optional)
Video Capture
4Kp60: 1 x HDMI
FHD (max): Y/PbPr, CVBS, S-Video, SDI, HDMI, DVI-I (can be
optional)

COM
1 x RS232
USB Ports
6 x USB 3.0 ports
4 x USB 2.0 ports
LAN
2 x RJ-45
Audio
2 x Audio out/in 3.5mm stereo jack connectors
Power Supply
Input Voltage
100-240V AC, 47-63Hz, 500W
Certification
UL/CB 60601-1, CE &FCC Class B certified
Environment
Temperature
0 ~ 40°C(Operating)
-10 ~50°C (Storage/Transportation)
Humidity
10%-90% @ 40°C (non-condensing) (Operating)
5 ~90% (non-condensing) (Storage/ Transportation)
Pressure
700-1060 hPa (Operation)
500-1060 hPa (Storage/Transportation)
Physical
Characteristics
Dimensions
(W x H x D)
329.5mmX320mmX145mm
Weight
4.65Kg
1.3 Dimensions
Dimensions: 329.5 x 320 x 145 mm
Figure 1.1 Dimensions of the AVAS-402

Figure 1.2 AVAS-402 Front Panel
(1) Power button (blue light)
(2) USB 3.0
Figure 1.3 AVAS-402 Rear Panel
(3) RS232
(4) DVI-D
(5) Display Port
(6) HDMI
(7) USB 2.0
(8) USB 3.0
(9) RJ-45
(10) Headphone
(11) MIC

1.3.1 Optional Add-on Cards
Capture cards
Part Number
Description
DVP-7611HE
1ch HDV/SDI/CVBS/YPrPr/S-video
PCIe capture
card
DVP-7011UHE
1ch 4K HDMI 2.0 Video capture Card
There are SDKs and APIs for video capturing and recording which allow system
integrators to implement into their use applications. After installing the capture cards
to AVAS-402, software programmers can easily program their customized software
application. For SDKs and APIs contents, you can download the related documents
from Advantech official website.
Graphic card
Part Number
Description
GFX-NG1050TF16-5D1
GTX1050TI 4G PCIEX16
DVI+HDMI+DP FHFL
1.3.2 Cleaning and Disinfecting
During normal use of AVAS-402 the device may become dirty and should be
regularly cleaned.
Steps:
1. Prepare cleaning detergent
2. Wipe AVAS-402 with a clean cloth that has been moistened in the detergent.
3. Wipe thoroughly with a clean cloth.
Suggested cleaning detergent:
1. 10% Sodium HyAVAShlorite 90% water solution.
2. A solution of 90% water, 6% benzyl alcohol, and 3% Hydrogen Peroxide
3. Glutaraldehyde (2.4%)
Caution!
○Do not immerse or rinse the AVAS or its peripherals. If you accidentally
spill liquid on the device, disconnect the unit from the power
source. Contact your IT support department regarding the continued
safety of the unit before placing it back in operation-Do not
spray cleaning agent on the chassis.
○Do not use disinfectants that contain phenol.
○Do not autoclave or clean the AVAS or its peripherals with strong
aromatic, chlorinated, ketone, ether, or ether solvents, sharp tools
or abrasives. Never immerse electrical connectors in water or other
liquids.
1.4 Operating Principle
The device provides input through I/O ports mainly on the rear cover.
The device computes the input data with its processing unit and then
output the generated data to accessories or other devices through its
I/O ports or through its LAN connections. The device is able to store
data in its storage, and when the device is turned off, still maintain the

data in the memory units of the storage.

1.5 Intended User Profile
Intend user profile:
Age: 18 to 65
Weight: not relevant
Health: not relevant
Nationality: Global
Patient state: patient will not be the operator.
Part of the body or type of tissue applied to or interacted with: hands,
expected contact time shall be less than 1 min, and will not touch AVAS-402
so frequently once it is set up.
Education level: at least 8 years intensive reading experience (school)
Knowledge:
Minimum –read and understand “westernized Arabic” numerals when written
in Arial font
–can distinguish: every parts of body as described in user manual
–trained and authorized by manufacturer only.
To be considered as trained and authorized, they must complete the training
course of the manufacturer; see document number AVAS-402_User
Manual_Rev 1.0 for qualification method, when considered necessary by the
manufacturer, technician shall be called back for retraining and annual training
is also considered necessary.
Language understanding: English, whenever other languages are required,
professional translation company shall translate and review by the
manufacturer, see SOP document number: SOP_Writing_Guidelines-ed.3
Experience: Mentally and physical competent, specific medical training to
understand basic knowledge for symbols.
Permissible impairments:
–Mild reading vision impairment or vision corrected to log MAR 0,2 (6/10 or
20/32)
–Average degree of aging-related short term memory impairment
–impaired by 40 % resulting in 60 % of normal hearing at 500 Hz to 2 kHz
Operating Principle:
The device provides input through I/O ports, and accessories through USB
ports or its LAN connections. The device computes the input data with its
processing unit and then output the generated data to accessories or other
devices through its I/O ports or through its LAN connections. The device is
able to store data in its storage, and when the device is turned off, still
maintain the data in the memory units of the storage.
2 System Setup
2.1 A Quick Tour of the AVAS-402
Before you start to set up the AVAS-402, take a moment to become familiar
with the locations and purposes of the controls, drives, connections and
ports, which are illustrated in the figures below.
When you place the AVAS-402 upright on the desktop, its front panel
appears as shown in Figure 2.1.

2.1.1 Front View
Figure 2.1 Front View of the AVAS-402
Front View
(1) Power Button w/ indicator blue light when it is switched on
(2) USB 3.0
2.1.2 Rear View
Most of I/O ports ofAVAS-402 are located at the rear cover, including Display
port, COM port, HDMI port, DVI-D port, Ethernet ports, USB ports and so on.
Apart from the I/O ports shown in below fig 2.2, additional I/O ports are also
included (not shown here) as add-on cards have been assembled to
AVAS-402 for capturing and recording video.
Figure 2-2: Rear view of AVAS-402
(3) RS-232
(4) DVI-D out
(5) Display Port out
(6) HDMI out
(7) USB 2.0 X 2
(1)

(8) USB 3.0 X 6
(9) Lan Ports (RJ-45) X 2
(10) Audio out (Headphone)
(11) Audio in (MIC)
NOTE: Equipotential terminal need link to
hospital ground/earth system before
system boot to protect operator and system
.

2.2 Installation Procedures
2.2.1 Connecting the Power Cord
The AVAS-402 can only be powered by connecting the power cord to its
power supply (FSP500M-80PA), and pressing the power button on the front
cover. Be sure to always handle the power cords by holding the plug ends
only.
Follow these procedures in order:
1. Connect the female end of the power cord to the 3-pin AC male plug of
AVAS-402.
2. Connect the 3-pin male plug of the power cord to an electrical outlet.
Figure 2.7 Connecting the power cord
2.2.2 Connecting the Ground Pin
Step 1. System ready and find the Equipotential Terminal on rear side of AVAS.
An Equipotential Terminal is provide to optionally connect to a hospital
ground/earth system
Figure 2-8: AVAS-402 Equipotential Terminal Pin
Step 2. Prepare the Grounding cable and the other terminal link to hospital
ground/earth system.

Figure 2-9: Grounding cable with connector
Step 3. Grounding cable plug with AVAS-402 Equipotential Terminal (See Figure 2-8)
2.3 Running the BIOS Setup Program
Your AVAS-402 was probably set up and configured by your dealer prior to delivery.
You may still find it necessary to use the BIOS (Basic Input-Output System) setup
program to change system configuration information, such as the current date and
time or your type of hard drive. The setup program is stored in read-only memory. It
can be accessed either when you turn on or reset the AVAS-402, by pressing the “F2
or Del” key on your keyboard immediately after powering on the computer.
The settings you specify with the setup program are recorded in a special area of
memory called CMOS RAM. This memory is backed up by a battery so that it will
not be erased when you turn off or reset the system. Whenever you turn on the
power, the system reads the settings stored in CMOS RAM and compares them to
the equipment check conducted during the power on self-test (POST). If an error
occurs, an error message will be displayed on screen, and you will be prompted to
run the setup program.
2.4 Installing System Software
Recent releases of operating systems from major vendors include setup programs
which load automatically and guide you through hard disk preparation and
operating system installation. The guidelines below will help you determine the
steps necessary to install your operating system on the AVAS hard drive.
Note! Some distributors and system integrators may have
already pre-installed system software prior to shipment of your AVAS-402.
If required, insert your operating system's installation or setup diskette into the
external diskette drive until the release button pops out.
The BIOS supports system boots up directly from the CD-ROM drive. You may
also insert your system installation CD-ROM disk into your external CD-ROM
drive.
Power on or reset the system by pressing the “Esc”or Del” key to boot into BIOS
menu and adjust the boot device sequence.
You can also press F7 key when booting; a bootable device popup menu will
appear, you can select bootable device that you want. AVAS-402 will
automatically load the operating system from the diskette or CD-ROM.

If you are presented with the opening screen of a setup or installation program,
follow the instructions on screen. The setup program will guide you through
preparation of your hard drive, and installation of the operating system.
2.5 Installing the Drivers
After installing your system software, you will be able to set up the Chipset,
Graphics, Ethernet, and Audio functions from your own external CD-ROM drive.
All the drivers can be downloaded via Advantech official website.
The standard automatic installation procedures for installing the Chipset,
Graphics, Audio, and Ethernet are described in Chapter 3.
Troubleshooting
When system behaves abnormally, such as:
1. Failure to power on
2. Failure to power off
3. AC power in and all switches ON, but system doesn't power on
Contact your distributor, sales representative, or Advantech's customer service
center for technical support if you need additional assistance. Please have the
following information ready before you call:
Product name and serial number
Description of your peripheral attachments
Description of your software (operating system, version, application software,
etc.)
A complete description of the problem
The exact wording of any error messages
Symptoms, photo or video if available.
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
Emissions Test
Compliance
Electromagnetic Environmental
Guidance
AVAS-402 uses RF
RF emissions
energy only for its internal function.
Group 1
Therefore, its RF emissions
CISPR 11
are very low and are not likely to
cause any interference in nearby
electronic equipment.
RF emissions
Class B
AVAS-402 is suitable
CISPR 11
for use in all establishments,
Harmonic emissions
Class A
including domestic establishments
IEC 61000-3-2
and those directly connected to
the public low-voltage power sup-
Voltage fluctuations/
ply network that supplies buildings
flicker emissions
Not applicable
used for domestic purposes.
IEC 61000-3-3
Recommended Separation Distances Between
Portable and Mobile RF Communications Equipment and the AVAS-402
AVAS-402 is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the model AVAS-212 can help
prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the model AVAS

-402 as recommended below, according to the maximum output power of the
communications equipment.
Rated Maximum
Output Power of
Separation Distance According to Frequency of Transmitter
Transmitter
m
W
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2,5 GHz
d = 1,2 P
d = 1,2 P
d = 2,3 P
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance din meters (m) can be estimated using the equation in the table above
applicable to the frequency of the transmitter, where Pis the maximum output power rating
of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Table of contents
Other Advantech DVR manuals

Advantech
Advantech DVR-520 User manual

Advantech
Advantech ARK-2250V User manual

Advantech
Advantech DVS-500 User manual

Advantech
Advantech ARK-2151V User manual

Advantech
Advantech DMR2008 User manual

Advantech
Advantech 530 User manual

Advantech
Advantech ARK-2121V User manual

Advantech
Advantech MIC-710IVA User manual

Advantech
Advantech ARK-2151S User manual