
2-4 OPERATION
3. After the slurry cup is loaded into the cell, the potentiometer mechanism (pot. mech.) is
pushed onto the slurry cup paddle shaft and the test cell contact pins. Attach the short bail
to the top of the potentiometer and lower the pot mech into the test cell. When properly
engaged, the top of the paddle shaft will be flush with the top of the torque measurement
potentiometer bearing. Remove the bail.
4. Check to be certain that the slurry cup and pot mech are properly engaged. Turn the
Motor switch to ON. No rubbing noise should be heard.
The Model 8040D10 is supplied with two types of O-ring seals for the cylinder plug. (See
drawing 08-0280, item 18.)
Caution: Selection of the proper O-ring to match the test conditions is critical.
• The viton O-ring (C09762) is suitable only for low temperature/pressure tests below
20,000 psi (138 Mpa), or 275°F (135°C).
• The metal O-ring (P-4080) is suitable for testing at any rated temperature or pressure.
5. Close the pressure cylinder by swinging the Swivel Arm Assembly and plug, vertically
above the cylinder, lowering the plug until the tapered threads engage. Screw the plug
down until it is firmly engaged. In order to assure that the cylinder will operate at the
maximum rated working pressure and temperature, we recommend that you work the
plug down until the line up mark on the plug matches the mark on the cylinder. Never run
a test with the line up mark on the plug tightened down past the mark on the cylinder.
Under these conditions, the plug may not unscrew from the cylinder without damaging the
threads or plug handles.
6. Slide the thermocouple through the test cell plug into the slurry cup paddle shaft. Start
the threads of the sealing gland into the test cell plug, but do not tighten the thermocouple
at this time. Verify that the thermocouple is plugged in.
7. Next, fill the test cell with oil. To accomplish this, close the Pressure Release Valve, turn
the CYLINDER control switch to the FILL position. When oil escapes from the top
thermocouple high-pressure fitting, tighten the sealing gland with a 5/8” wrench. Leave
the CYLINDER control switch in the FILL position during the test.
8. To apply the initial pressure to the test cell on a Model 8040D10 without pressure control,
turn the Pump Switch to the MANUAL position. When the pressure reaches the desired
level turn the Pump Switch to the OFF position. Adjust the pressure as required
throughout the test by turning the Pump Switch to MANUAL to increase pressure or by
slowly cracking open the Pressure Release Valve to relieve pressure. Use care to open the
Pressure Release Valve slowly when attempting to bleed pressure.
9. Turn the Heater Switch to the ON position, the Pump Switch to the AUTO position
(Model 8040D10 with pressure control), and start the timer. (The heater and pump will
not start until the program start up is initiated through the controller.)
10. To begin the test, the Temperature Controller and Pressure Controller (if equipped)
programs must be started as follows.
11. Press the Advance Key to display the Control Mode (AUTO, OFF or MAN). Press the
Up or Down keys to select AUTO. Press the Infinity Key to return to the main
screen. Press the EZ1 button to start the program. The “1” light should begin flashing
indicating the control output to the heater and or pump.