Operation
PRESSURE SENSORS
The HPC40 Series is equipped with one or two pressure sensors. In addition, a Crystal APM (Advanced Pressure Module) may also be used. For information on
the Crystal APM, please refer to the APM CPF Series Operation Manual. Be sure to choose the proper sensor based on working pressures and accuracy.
! WARNINGS: The following warnings apply to any sensor used with the HPC40 — whether internal or external:
•Pressure sensors may be damaged and/or personal injury may occur due to improper application of pressure. Please refer to the
Ranges and Resolutions table for information on overpressure ratings.
•The calibrator will display
+
OL when an inappropriate pressure is applied. If
+
OL is displayed, the pressure should be reduced or vented immediately
to prevent damage or possible personal injury.
Note:
+
OL is displayed when the pressure exceeds 110% of the nominal range of the sensor.
Pressure Connection
Crystal CPF System: Medium Pressure Female (MPF) (1/4”medium pressure tube system with 7/16-20 threads). See our CPF Brochure for further information.
CPF o-ring size and material: AS568A-012, Viton 80 or 90 durometer (P/N 3981).
For most applications CPF Fittings can be hand tightened (no tools required). Wrench tightening is recommended (to achieve a metal to metal cone seal) for
applications where chemical compatibility of the process uid and the o-ring are a concern. Cone seals require only moderate assembly torque to seal up to
10 000 PSI (700 bar). We recommend a tightening torque of 120 in-lbs ±20 in-lbs for our CPF ttings. Please note this is only a fraction of the typical torque
required to seal a 1/4”NPT tting. If a torque wrench isn’t practical to use, the ttings can be assembled as follows: Hand tighten tting fully
until the cone has bottomed out, then tighten an additional 20° using a hex wrench. Apply a small amount of media-compatible lubricant to the gland
threads and male cone to increase tting life, reduce the likelihood of galling, and promote sealing.
! CAUTION: To achieve CPF maximum allowable working pressures no o-ring substitutions are allowed. See our CPF brochure and
CES-003 CPF Safety Guide available from the website at crystalengineering.net for further detail.
Measuring Vacuum
All versions of the HPC40 can be used to measure moderate vacuum.
When measuring pressure less than ambient barometric conditions, a minus (-) sign will appear.
! CAUTION: The HPC40 is not recommended for continuous use at high vacuum.