API 100A User manual

INSTRUCTION MANUAL
MODEL 100A
SULFUR DIOXIDE ANALYZER
TELEDYNE INSTRUMENTS
ADVANCED POLLUTION INSTRUMENTATION DIVISION
(T-API)
6565 NANCY RIDGE DRIVE
SAN DIEGO, CA 92121-2251
TOLL-FREE: 800-324-5190
FAX: 858-657-9816
TEL: 858-657-9800
WEB SITE: www.teledyne-api.com
02164
REV. G
Copyright 1994 API Inc. 12/12/98


Teledyne API Model 100A SO2Analyzer Instruction Manual, 02164, Rev. G
Table of Contents
1 HOW TO USE THIS MANUAL..................................................................1-1
1.1 SAFETY MESSAGES.................................................................................................... 1-2
2 GETTING STARTED.................................................................................2-1
2.1 INSTALLATION ............................................................................................................ 2-1
2.2 ELECTRICAL AND PNEUMATIC CONNECTIONS................................................................ 2-1
2.3 INITIAL OPERATION..................................................................................................... 2-6
3 SPECIFICATIONS, AGENCY APPROVALS, WARRANTY .....................3-1
3.1 SPECIFICATIONS ........................................................................................................ 3-1
3.2 EPA EQUIVALENCY DESIGNATION ............................................................................... 3-2
3.3 WARRANTY................................................................................................................ 3-3
4 THE M100A SO2ANALYZER ...................................................................4-1
4.1 PRINCIPLE OF OPERATION .......................................................................................... 4-1
4.2 INSTRUMENT DESCRIPTION ......................................................................................... 4-4
4.2.1 Sensor Module, Reaction Cell, Detector............................................................ 4-4
4.2.2 Pneumatic Sensor Board................................................................................... 4-4
4.2.3 Computer Hardware and Software .................................................................... 4-4
4.2.4 V/F Board .......................................................................................................... 4-5
4.2.5 Front Panel ........................................................................................................ 4-5
4.2.6 Power Supply Module........................................................................................ 4-8
4.2.7 Pneumatic System............................................................................................. 4-8
5 SOFTWARE FEATURES..........................................................................5-1
5.1 INDEX TO FRONT PANEL MENUS ................................................................................. 5-1
5.1.1 Sample Menu .................................................................................................... 5-3
5.1.2 Set-Up Menu ..................................................................................................... 5-4
5.2 SAMPLE MODE........................................................................................................... 5-7
5.2.1 Test Functions ................................................................................................... 5-7
5.2.2 CAL, CALS, CALZ, Calibration Functions........................................................ 5-10
5.3 SET-UP MODE ......................................................................................................... 5-12
5.3.1 Configuration Information (CFG) ..................................................................... 5-12
5.3.2 Automatic Calibration (AutoCal)....................................................................... 5-12
5.3.3 Data Acquisition System (DAS) ....................................................................... 5-12
5.3.4 Range Menu .................................................................................................... 5-13
5.3.5 Password Enable............................................................................................. 5-17
5.3.6 Time of Day Clock ........................................................................................... 5-17
5.3.7 Diagnostic Mode.............................................................................................. 5-17
5.3.8 Communications Menu .................................................................................... 5-17
5.3.9 Variables Menu (VARS)................................................................................... 5-18
i

Teledyne API Model 100A SO2Analyzer Instruction Manual, 02164, Rev. G
6 OPTIONAL HARDWARE AND SOFTWARE............................................6-1
6.1 RACK MOUNT OPTIONS .............................................................................................. 6-1
6.2 ZERO/SPAN VALVES................................................................................................... 6-1
6.3 INTERNAL ZERO/SPAN (IZS)........................................................................................ 6-3
6.4 AUTOCAL - SETUP OF IZS AND ZERO/SPAN VALVES ..................................................... 6-3
6.5 PERMEATION TUBE (USED WITH IZS OPTION) ............................................................... 6-6
6.6 4-20 MA, CURRENT LOOP OUTPUT ............................................................................. 6-9
6.7 STATUS OUTPUT ........................................................................................................ 6-9
6.8 RS-232 INTERFACE ................................................................................................. 6-10
6.8.1 Setting Up the RS-232 Interface...................................................................... 6-10
6.8.2 Command Summary........................................................................................ 6-13
6.8.3 TEST Commands and Messages.................................................................... 6-17
6.8.4 WARNING Commands And Messages ........................................................... 6-18
6.8.5 CALIBRATION Commands and Messages ..................................................... 6-19
6.8.6 DIAGNOSTIC Commands and Messages....................................................... 6-20
6.8.7 DAS Commands and Reports ......................................................................... 6-21
6.8.8 VARIABLES Commands and Messages ......................................................... 6-24
7 CALIBRATION AND ZERO/SPAN CHECKS ...........................................7-1
7.1 MANUAL ZERO/SPAN CHECK OR CALIBRATION THROUGH THE SAMPLE PORT .................. 7-4
7.2 MANUAL ZERO/SPAN CHECK OR CALIBRATION WITH ZERO/SPAN VALVES OPTION ........... 7-6
7.3 MANUAL ZERO/SPAN CHECK WITH IZS OPTION ............................................................ 7-7
7.4 AUTOMATIC ZERO/SPAN CHECK .................................................................................. 7-7
7.5 AUTOMATIC ZERO/SPAN CALIBRATION ......................................................................... 7-7
7.6 CALIBRATE ON SO2PERMEATION TUBE ....................................................................... 7-8
7.7 USE OF IZS OR ZERO/SPAN VALVES WITH REMOTE CONTACT CLOSURE ........................ 7-8
7.8 EPA PROTOCOL CALIBRATION .................................................................................... 7-9
7.8.1 Calibration of Equipment ................................................................................... 7-9
7.8.2 Calibration Gas Sources.................................................................................. 7-10
7.8.3 Data Recording Device.................................................................................... 7-12
7.8.4 Dynamic Multipoint Span Calibration ............................................................... 7-12
7.8.5 SO2Calibration Procedure............................................................................... 7-13
7.8.6 Calibration Frequency...................................................................................... 7-19
7.8.7 Other Quality Assurance Procedures .............................................................. 7-19
7.8.8 Summary of Quality Assurance Checks .......................................................... 7-21
7.8.9 ZERO and SPAN Checks ................................................................................ 7-22
7.8.10 Recommended Standards for Establishing Traceability ................................ 7-23
7.9 SPECIAL CALIBRATION REQUIREMENTS FOR DUAL RANGE OR AUTO RANGE.................. 7-24
7.10 CALIBRATION QUALITY............................................................................................ 7-25
7.11 REFERENCES......................................................................................................... 7-26
ii

Teledyne API Model 100A SO2Analyzer Instruction Manual, 02164, Rev. G
8 MAINTENANCE ........................................................................................8-1
8.1 MAINTENANCE SCHEDULE........................................................................................... 8-1
8.2 REPLACING THE SAMPLE PARTICULATE FILTER ............................................................. 8-2
8.3 REPLACING THE PERMEATION TUBE (OPTION)............................................................... 8-4
8.4 REPLACING THE IZS ZERO AIR SCRUBBER (OPTION)..................................................... 8-4
8.5 CLEANING ORIFICE AND ORIFICE FILTER ...................................................................... 8-5
8.6 CLEANING THE REACTION CELL ................................................................................... 8-6
8.7 PNEUMATIC LINE INSPECTION...................................................................................... 8-6
8.8 LEAK CHECK PROCEDURE .......................................................................................... 8-9
8.9 LIGHT LEAK CHECK PROCEDURE ............................................................................... 8-10
8.10 EPROM REPLACEMENT PROCEDURE ..................................................................... 8-10
9 DIAGNOSTIC, TROUBLESHOOTING......................................................9-1
9.1 OPERATION VERIFICATION - M100A DIAGNOSTIC TECHNIQUES ..................................... 9-3
9.1.1 Fault Diagnosis with TEST Variables................................................................. 9-3
9.1.2 Fault Diagnosis with WARNING Messages ....................................................... 9-9
9.1.3 Fault Diagnosis using DIAGNOSTIC Mode ..................................................... 9-11
9.1.4 M100A Internal Variables ................................................................................ 9-18
9.1.5 Test Channel Analog Output ........................................................................... 9-20
9.1.6 Factory Calibration Procedure (Quick Cal) ...................................................... 9-21
9.2 PERFORMANCE PROBLEMS ....................................................................................... 9-23
9.2.1 AC Power Check ............................................................................................. 9-23
9.2.2 Flow Check...................................................................................................... 9-24
9.2.3 No Response to Sample Gas .......................................................................... 9-24
9.2.4 Negative Concentration Display....................................................................... 9-25
9.2.5 Excessive Noise .............................................................................................. 9-25
9.2.6 Unstable Span ................................................................................................. 9-26
9.2.7 Unstable Zero .................................................................................................. 9-27
9.2.8 Inability to Span ............................................................................................... 9-27
9.2.9 Inability to Zero ................................................................................................ 9-28
9.2.10 Non-Linear Response.................................................................................... 9-28
9.2.11 Slow Response.............................................................................................. 9-29
9.2.12 Analog Output Doesn't Agree With Display Concentration ............................ 9-29
9.3 ELECTRONIC SUBSYSTEM TROUBLESHOOTING AND ADJUSTMENTS............................... 9-30
9.3.1 Computer, Display, Keyboard.......................................................................... 9-30
9.3.2 RS-232 Communications................................................................................. 9-33
9.3.3 Voltage/Frequency (V/F) Board ....................................................................... 9-36
9.3.4 Status/Temp Board.......................................................................................... 9-42
9.3.5 Power Supply Module...................................................................................... 9-45
9.3.6 Flow/Pressure Sensor ..................................................................................... 9-49
9.3.7 Reaction Cell Temp ......................................................................................... 9-53
9.3.8 Preamp Board ................................................................................................. 9-53
9.3.9 PMT Cooler ..................................................................................................... 9-53
9.3.10 HVPS (High Voltage Power Supply) .............................................................. 9-56
iii

Teledyne API Model 100A SO2Analyzer Instruction Manual, 02164, Rev. G
9.4 OPTICAL SENSOR MODULE TROUBLESHOOTING.......................................................... 9-58
9.4.1 PMT ................................................................................................................. 9-58
9.4.2 UV Lamp Adjust or Replacement .................................................................... 9-58
9.4.3 UV Filter Replacement..................................................................................... 9-59
9.4.4 Shutter System ................................................................................................ 9-59
9.5 PNEUMATIC SYSTEM TROUBLESHOOTING ................................................................... 9-60
9.5.1 Leak Check...................................................................................................... 9-60
9.5.2 Pump ............................................................................................................... 9-60
9.5.3 Kicker............................................................................................................... 9-60
9.5.4 Z/S Valves & IZS Permeation Tube Oven ....................................................... 9-62
10 TELEDYNE API M100A SPARE PARTS LIST.....................................10-1
APPENDIX A MAINTENANCE SCHEDULE FOR M100A ......................... A-1
APPENDIX B ELECTRICAL SCHEMATICS .............................................. B-1
iv

Teledyne API Model 100A SO2Analyzer Instruction Manual, 02164, Rev. G
List of Figures
FIGURE 2-1: REMOVAL OF SHIPPING SCREWS ...................................................................2-3
FIGURE 2-2: REAR PANEL ................................................................................................2-4
FIGURE 2-3: REAR PANEL PNEUMATIC RECOMMENDATIONS ...............................................2-5
FIGURE 2-4: FRONT PANEL ............................................................................................2-10
FIGURE 2-5: ASSEMBLY LAYOUT.....................................................................................2-11
FIGURE 4-1: M100A SULFUR DIOXIDE ANALYZER..............................................................4-3
FIGURE 4-2: FRONT PANEL DISPLAY.................................................................................4-7
FIGURE 5-1: SAMPLE MENU .............................................................................................5-1
FIGURE 5-2: SETUP MENU TREE ......................................................................................5-2
FIGURE 6-1: PERMEATION TUBE INSTALLATION..................................................................6-8
FIGURE 7-1: MODEL 100A CALIBRATION SETUP ................................................................7-3
FIGURE 7-2: DIAGRAM OF CALIBRATION SYSTEM.............................................................7-15
FIGURE 7-3: DIAGRAM F PERMEATION CALIBRATION SYSTEM ...........................................7-17
FIGURE 8-1: REPLACING THE PARTICULATE FILTER............................................................8-3
FIGURE 8-2: REACTION CELL ...........................................................................................8-7
FIGURE 8-3: PNEUMATIC DIAGRAM ...................................................................................8-8
FIGURE 9-1: CPU BOARD JUMPER SETTINGS..................................................................9-32
FIGURE 9-2: RS-232 PIN ASSIGNMENTS.........................................................................9-34
FIGURE 9-3: V/F BOARD DIP SWITCH SETTINGS..............................................................9-41
FIGURE 9-4: POWER SUPPLY MODULE LAYOUT ...............................................................9-46
FIGURE 9-5: ELECTRICAL BLOCK DIAGRAM .....................................................................9-47
FIGURE 9-6: FLOW/PRESSURE SENSOR ..........................................................................9-50
FIGURE 9-7: SO2SENSOR MODULE................................................................................9-51
FIGURE 9-8: SO2SENSOR MODULE................................................................................9-52
FIGURE 9-9: PMT COOLER SUBSYSTEM .........................................................................9-55
FIGURE 9-10: HIGH VOLTAGE POWER SUPPLY ................................................................9-57
FIGURE 9-11: KICKER LEAK CHECK ................................................................................9-61
v

Teledyne API Model 100A SO2Analyzer Instruction Manual, 02164, Rev. G
List of Tables
TABLE 2-1: FINAL TEST AND CALIBRATION VALUES ..........................................................2-12
TABLE 2-1: FINAL TEST AND CALIBRATION VALUES (CONTINUED)......................................2-13
TABLE 4-1: SYSTEM MODES DISPLAY ...............................................................................4-6
TABLE 4-2: FRONT PANEL STATUS LED'S.........................................................................4-7
TABLE 5-1: M100A SAMPLE MENU STRUCTURE................................................................5-3
TABLE 5-2: M100A SETUP MENU STRUCTURE ..................................................................5-4
TABLE 5-3: M100A MENU STRUCTURE - SETUP MENU #2 .................................................5-5
TABLE 5-4: M100A MENU STRUCTURE - SETUP MENU #3 .................................................5-6
TABLE 5-5: CALIBRATE, SETUP PASSWORDS...................................................................5-17
TABLE 6-1: ZERO/SPAN VALVE OPERATION .......................................................................6-2
TABLE 6-2: SETUP AUTOMATIC ZERO/SPAN CALIBRATION ..................................................6-4
TABLE 6-3: STATUS OUTPUT PIN ASSIGNMENTS ................................................................6-9
TABLE 6-4: RS-232 PORT SETUP - FRONT PANEL...........................................................6-11
TABLE 6-5: RS-232 SWITCHING FROM TERMINAL MODE TO COMPUTER MODE..................6-12
TABLE 6-6: RS-232 TERMINAL MODE EDITING KEYS .......................................................6-13
TABLE 6-7: RS-232 COMMAND SUMMARY ......................................................................6-14
TABLE 6-8: RS-232 COMMAND SUMMARY ......................................................................6-15
TABLE 6-9: RS-232 INTERFACE COMMAND TYPES ..........................................................6-16
TABLE 6-10: RS-232 TEST MESSAGES ..........................................................................6-17
TABLE 6-11: RS-232 WARNING MESSAGES....................................................................6-18
TABLE 6-12: RS-232 CALIBRATION MESSAGES...............................................................6-19
TABLE 6-13: RS-232 CALIBRATION COMMANDS ..............................................................6-20
TABLE 6-14: RS-232 DIAGNOSTIC COMMAND SUMMARY .................................................6-21
TABLE 6-15: RS-232 DAS COMMANDS..........................................................................6-21
TABLE 6-16: RS-232 OPERATING MODES ......................................................................6-24
TABLE 7-1: TYPES OF ZERO/SPAN CHECKS AND CALIBRATIONS ..........................................7-2
TABLE 7-2: MANUAL ZERO CALIBRATION PROCEDURE - ZERO GAS THROUGH SAMPLE
PORT............................................................................................................7-4
TABLE 7-3: ENTER EXPECTED SPAN GAS CONCENTRATIONS PROCEDURE...........................7-5
TABLE 7-4: MANUAL SPAN CALIBRATION PROCEDURE - SPAN GAS THROUGH SAMPLE
PORT............................................................................................................7-5
TABLE 7-5: MANUAL ZERO CALIBRATION PROCEDURE - Z/S VALVES...................................7-6
TABLE 7-6: MANUAL SPAN CALIBRATION PROCEDURE - Z/S VALVES...................................7-6
TABLE 7-7: IZS OR Z/S VALVES MODES WITH REMOTE CONTACT CLOSURE........................7-9
TABLE 7-8: ACTIVITY MATRIX FOR CALIBRATION EQUIPMENT & SUPPLIES..........................7-11
TABLE 7-9: ACTIVITY MATRIX FOR CALIBRATION PROCEDURE ...........................................7-12
TABLE 7-10: EPA ZERO CALIBRATION PROCEDURE.........................................................7-14
TABLE 7-11: EPA EXPECTED SPAN GAS CONCENTRATION PROCEDURE ...........................7-15
TABLE 7-12: EPA SPAN CALIBRATION PROCEDURE.........................................................7-16
TABLE 7-13: DEFINITION OF LEVEL 1 AND LEVEL 2 ZERO AND SPAN CHECKS .....................7-20
TABLE 7-14: ACTIVITY MATRIX .......................................................................................7-21
TABLE 7-15: NIST-SRM'S AVAILABLE FOR TRACEABILITY OF CALIBRATION AND AUDIT
GAS STANDARDS ......................................................................................7-24
TABLE 7-16: CALIBRATION QUALITY CHECK ....................................................................7-25
vi

Teledyne API Model 100A SO2Analyzer Instruction Manual, 02164, Rev. G
TABLE 8-1: PREVENTATIVE MAINTENANCE SCHEDULE ........................................................8-1
TABLE 9-1: TEST FUNCTIONS ...........................................................................................9-3
TABLE 9-1: TEST FUNCTIONS (CONTINUED).......................................................................9-4
TABLE 9-1: TEST FUNCTIONS (CONTINUED).......................................................................9-5
TABLE 9-1: TEST FUNCTIONS (CONTINUED).......................................................................9-6
TABLE 9-1: TEST FUNCTIONS (CONTINUED).......................................................................9-7
TABLE 9-1: TEST FUNCTIONS (CONTINUED).......................................................................9-8
TABLE 9-2: FRONT PANEL WARNING MESSAGES ...............................................................9-9
TABLE 9-2: FRONT PANEL WARNING MESSAGES (CONTINUED).........................................9-10
TABLE 9-3: SUMMARY OF DIAGNOSTIC MODES ................................................................9-12
TABLE 9-4: DIAGNOSTIC MODE - SIGNAL I/O ...................................................................9-13
TABLE 9-5: MODEL 100A VARIABLES..............................................................................9-19
TABLE 9-6: TEST CHANNEL OUTPUT ...............................................................................9-20
TABLE 9-7: POWER SUPPLY MODULE SUBASSEMBLIES.....................................................9-45
TABLE 9-8: POWER SUPPLY MODULE LED OPERATION....................................................9-48
TABLE 10-1: TELEDYNE API M100A SPARE PARTS LIST .................................................10-1
TABLE 10-1: TELEDYNE API M100A SPARE PARTS LIST (CONTINUED) .............................10-2
TABLE 10-1: TELEDYNE API M100A SPARE PARTS LIST (CONTINUED) .............................10-3
TABLE 10-1: TELEDYNE API M100A SPARE PARTS LIST (CONTINUED) .............................10-4
TABLE 10-2: TELEDYNE API M100A LEVEL 1 SPARES KIT...............................................10-4
TABLE 10-3: TELEDYNE API M100A 37 MM EXPENDABLES KIT ........................................10-5
TABLE 10-4: TELEDYNE API M100A 47 MM EXPENDABLES KIT ........................................10-5
TABLE 10-5: TELEDYNE API M100A EXPENDABLES KIT - IZS..........................................10-5
TABLE 10-6: TELEDYNE API M100A SPARES KIT FOR 1 UNIT..........................................10-6
vii


Teledyne API Model 100A SO2Analyzer Instruction Manual, 02164, Rev. G
1 HOW TO USE THIS MANUAL
The Model 100A has been designed with serviceability, reliability and ease of operation in mind.
The M100A's microprocessor continually checks operating parameters such as temperature,
flow, and critical voltages. The instruments modular design uses captive screws to facilitate
repair and ease of access. If you encounter any difficulty refer to Section 9 General
Troubleshooting Hints.
We recognize that the need for information in this manual changes as time passes. When the
instrument first arrives, it is necessary to get it up and running quickly and verify its correct
operation. As time passes, more detailed information is often required on special configurations,
calibration alternatives and other operational details. Finally there is the need for periodic
maintenance and to quickly troubleshoot problems to assure maximum uptime and data integrity.
To address these needs, we have created three indexes to the information inside. They are:
Table of Contents:
Outlines the contents of the manual in the order the information is presented. This is a good
overview of the topics covered in the manual. There is also a list of Tables and a list of Figures.
Index to M100A Front Panel Menus:
The Menu Index (Figure 5-1 and Figure 5-2, and Table 5-2) briefly describes the front panel
menus and refers you to other sections of the manual that have a detailed explanation of each
menu selection.
Troubleshooting Section 9
The Troubleshooting Section, outlined in the Table of contents, allows you to diagnose and repair
the instrument based on variables in the TEST menu, the results of DIAGNOSTIC tests, and
performance faults such as excessive noise or drift. The troubleshooting section also explains the
operation, adjustment, diagnosis and testing of each instrument subsystem.
If you are unpacking the instrument for the first time, please refer to Getting Started in
Section 2.
1-1

Teledyne API Model 100A SO2Analyzer Instruction Manual, 02164, Rev. G
1.1 Safety Messages
Your safety and the safety of others is very important. We have provided many important safety
messages in this manual. Please read these messages carefully.
A safety message alerts you to potential hazards that could hurt you or others. Each safety
message is associated with a safety alert symbol. These symbols are found in the manual and
inside the instrument. The definition of these symbols is described below:
GENERAL WARNING/CAUTION: Refer to the instructions for details on the
specific danger.
CAUTION: Hot Surface Warning
CAUTION: Electrical Shock Hazard
Technician Symbol: All operations marked with this symbol are to be performed
by qualified maintenance personnel only.
CAUTION
The analyzer should only be used for the purpose
and in the manner described in this manual.
If you use the analyzer in a manner other than that for which
it was intended, unpredictable behavior could ensue with
possibly hazardous consequences.
1-2

Teledyne API Model 100A SO2Analyzer Instruction Manual, 02164, Rev. G
2 GETTING STARTED
2.1 Installation
CAUTION
To avoid personal injury, always use two persons to
lift and carry the Model 100A.
8. Verify that there is no apparent shipping damage. If damage has occurred please advise shipper
first, then Teledyne API.
1. Before operation it is necessary to remove the shipping hold-down screws. Remove the
instrument cover, then refer to Figure 2-1 for screw location.
Note:
Save the shipping screws and re-install them whenever the unit is shipped to another location.
2. While the instrument cover is removed, please check the voltage and frequency label on
the cover of the power supply module and compare that to your local power before plugging
in the M100A.
9. Check for internal shipping damage, and generally inspect the interior of the instrument to make
sure all circuit boards and other components are in good shape.
3. Replace the instrument cover.
4. When installing the M100A, allow at least 4” (10 cm) clearance at the back and at least
1” (2.5 cm) clearance at each side for proper venting.
2.2 Electrical and Pneumatic Connections
10. Refer to Figure 2-2 to locate the rear panel electrical and pneumatic connections.
11. The pressure of the sample gas at the inlet port should be at atmospheric pressure (Refer to Figure
2-3 and Figure 7-1 for pneumatic system connection).
1. Connect the analyzer exhaust to a suitable vent at atmospheric pressure. (See Figure 2-3
for exhaust line venting recommendations.)
2-1

Teledyne API Model 100A SO2Analyzer Instruction Manual, 02164, Rev. G
2. If desired, attach the analog output connections to a strip chart recorder and/or
datalogger. Refer to Figure 9-3 for jumper settings for the desired analog output voltage
range. Factory default setting is 0-5VDC.
3. Connect the power cord to the correct voltage line, then turn to Section 2.3 Initial
Operation.
CAUTION
High voltages present inside case.
DO NOT LOOK AT THE UV LAMP, UV LIGHT
COULD CASE EYE DAMAGE.
ALWAYS USE SAFETY GLASSES
(PLASTIC GLASSES WILL NOT DO).
Connect the exhaust fitting on the rear panel to a
suitable vent outside of the room.
CAUTION
DO NOT OPERATE WITH COVER OFF.
Before operation check for correct line voltage and
frequency on Serial Number Sticker.
Do not plug in the power cord if the voltage or
frequency is incorrect.
Do not operate without proper chassis grounding.
Do not defeat the ground wire on power plug.
Turn off analyzer power before disconnecting or
connecting eletrical subassemblies.
Always replace shipping screws when transporting the Analyzer.
2-2

Teledyne API Model 100A SO2Analyzer Instruction Manual, 02164, Rev. G
Figure 2-1: Removal of Shipping Screws
2-3

Teledyne API Model 100A SO2Analyzer Instruction Manual, 02164, Rev. G
Figure 2-2: Rear Panel
2-4

Teledyne API Model 100A SO2Analyzer Instruction Manual, 02164, Rev. G
Figure 2-3: Rear Panel Pneumatic Recommendations
2-5

Teledyne API Model 100A SO2Analyzer Instruction Manual, 02164, Rev. G
2.3 Initial Operation
12. Turn on the instrument power.
1. The display should immediately light, displaying the instrument type (M100A) and the
CPU memory configuration. If you are unfamiliar with the M100A, we recommend that you
read the overview Section 4 before proceeding. A diagram of the software menu trees is in
Figure 5-1 and Figure 5-2.
13. The M100A requires about 60 minutes for all internal components to come up to temperature.
2. While waiting for instrument temperatures to come up, you can check for correct
operation by using some of the M100A's diagnostic and test features.
3. Examine the TEST functions by comparing the values listed in Table 2-1 to those in the
display. Remember that as the instrument warms up the values may change until they reach
their final values. If you would like to know more about the meaning and utility of each
TEST function refer to Table 9-1. Table 2-1 also contains the list of options. Section 6 covers
setting up the options.
4. When the instrument is warmed up, re-check the TEST functions against Table 2-1. All
of the readings should compare closely with those in the Table. If they do not, see Section
9.1.1.
NOTE
Do not calibrate the analyzer within 60 minutes after the power reset.
2-6

Teledyne API Model 100A SO2Analyzer Instruction Manual, 02164, Rev. G
The next task is to calibrate the analyzer. There are several ways to do a calibration; they are
summarized in Table 7-1. For a preliminary checkout we recommend calibration with zero air and
span gas coming in through the sample port. The procedure is:
Step 1 - Enter the expected SO2span gas concentration
Step Number Action Comment
1. Press CAL-CONC This key sequence causes the M100A to prompt for the
expected SO2concentration. Enter the SO2span concentration
value by pressing the key under each digit until the expected
value is set.
2. Press ENTR ENTR stores the expected SO2span value. This value will be
used in the internal formulas to compute subsequent SO2
concentration values.
3. Press EXIT Returns instrument to SAMPLE mode.
4. Press SETUP-
RNGE-SET
If necessary you may want to change ranges. Normally the
instrument is shipped with range set at 500 ppb.
5. Press EXIT Returns the instrument to SAMPLE mode.
2-7

Teledyne API Model 100A SO2Analyzer Instruction Manual, 02164, Rev. G
Step 2 - Calibrate the instrument:
Zero/Span Calibration Procedure
Step Number Action Comment
1. Input Zero gas Allow Zero gas to enter the sample port on the rear of the
instrument.
2. Press CAL The M100A enters the calibrate mode from sample mode.
3. Wait 10 min Wait for reading to stabilize at the zero value. (If you wait less
than 10 minutes the final zero value may drift.)
4. Press ZERO The ZERO button will be displayed.
5. Press ENTR Pressing ENTR actually changes the calculation equations and
zeroes the instrument.
6. Press EXIT M100A returns to the CAL menu. Now switch gas streams to
span gas.
7. Wait 10 min Wait for reading to stabilize at the span value. (If you wait less
than 10 minutes the final span value may drift.)
8. Press SPAN The SPAN button should be displayed. If it is not, check the
Troubleshooting Section 9.2.8 for instructions on how to
proceed. In certain circumstances at low span gas
concentrations (<100ppb), both the ZERO and SPAN buttons
will appear.
9. Press ENTR Pressing ENTR actually changes the calculation equations so
that the concentration displayed is the same as the expected
span concentration you entered above, thus spanning the
instrument.
10. Press EXIT Pressing EXIT returns the instrument to SAMPLE mode.
2-8
Table of contents