Only designated bottled purified water, distilled water or deionized water can be used.
Tap water, mineral water and water filtered by the water purifier cannot be used.
Sincerely thank you for purchasing this hydrogen inhaler.
● Before the first use, please read the Operating Manual carefully to use the product correctly and safely.
● Please properly keep it for future reference.
● Please understand that the company will not be liable for any failure and accident caused by incorrect
installation or use of this product.
Operating Manual
Hydrogen Inhaler
H300
H300 吸氢机(单吸) 中性 英文 说明书 2020-12-24 单色印刷
无
100g双胶纸 0100600300007
Aquavolta® Cavallo 300ml

Before installation and using, please read the Operating Manual carefully and
pay attention to our recommendations for correct installation and proper maintenance,
so as to maximize the performance of the equipment.
If you have any questions in the use, please read the Operating Manual carefully.
You are welcome to provide valuable advice to our products and services.
Dear customer:
Thank you for purchasing the H300 hydrogen inhaler produced by our company.
Contents
1 Product Overview.....................................................................01
1.1 Structure...............................................................................01
1.2 Application............................................................................01
1.3 Performance indexes.............................................................01
2 Components.............................................................................01
3 Panel Buttons and Corresponding Operation.............................02
4 Instructions for Installation and Use.......................................... 03
4.1 Satisfactory purified water is requried....................................03
4.2 Operating steps for the first use............................................. 03
4.3 Wearing method of nasal suction pipe.................................... 04
4.4 When to drain water.............................................................. 04
4.5 Drainage method.................................................................. 04
5 Precautions for Use..................................................................05
6 Safety Precaution.....................................................................06
7 Service and Maintenance.........................................................07
8 Packing List and Compulsory Accessories.................................07

140W
3.5Kg
4Kg
H300
282X221X375mm
~220V
300±5% ml/min
Product name Hydrogen inhaler(single inhaler)
Product model
Rated power
Hydrogen flow
Net weight
Gross weight
Maintenance Machine is
too hot
Water in the tank
is at low level
Water quality is abnormal
Water in the tank
is too full
Corresponding indicator will
turn red in case of failure.
Corresponding indicator will
turn white after the failure is
eliminated.
Hydrogen flow indicator
The machine timing indicator
is measured in hours.
Press "Time" after starting
up, and then select among
0.5h, 1.0h, 1.5h and 2.0h.
No time is set when the
machine is turned on as the
default time is 1 hour.
In the Start/Pause process
of hydrogen production, the
Time key is inoperative.
Flow: Hydrogen
production by machine
Available in
100/200/300
Intercyclic selection.
Power: Power on and enter standby
mode, and then the indicator turns red;
keep pressing it for 2s to turn it on, and
then the power indicator turns white;
and keep pressing it for 2s to turn it off,
and then the power indicator turns red.
Start/Pause: After the machine has set the
time and flow,
Press this button to start the machine to
produce hydrogen.During hydrogen production,
Press this button to pause hydrogen production.
Press this button to resume hydrogen production
while the machine is paused.
Water tank cover
Humidification
bottle (clear water)
1.1 Structure
The product consists of power cord, shell, circuit board, PEM pure water
electrolytic cell, water tank and so on.
1.2 Application
This product is only for home and personal use and not for other purposes
than hydrogen inhalation.
1.3 Performance indexes
Components
Hydrogen outlet
Humidification
bottle cover
Fixing clamp for
humidification bottle
Aluminum alloy
housing
Water inlet Nasal suction
pipe outlet
Water outlet Power line
0201
Product Overview Panel Buttons and Corresponding Operations
Input voltage
Product Measurement

4. Open the pure water tank cover on the equipment, put the
watered cup at the water filler and pour a proper amount of
purified water or distilled water.
5. Add water and wait for 10 minutes, then open the drain
port at the back of the equipment, empty the water tank,
plug the silicone plug,and tighten the drain cap;
6. Put the watered cup at the water filler of equipment again,
add bottled purified water or distilled water in the equipment
at about the middle level of perspective window;
7.Insert the power adapter in the power outlet of the
equipment and the other end in the mains supply socket;
8.Insert the nasal suction pipe
9.Keep pressing "Power" for 2s to turn on the equipment;
press "Time" to set the working time of equipment and click
"Start" to start the equipment to produce hydrogen.
Wearing method of nasal suction pipe
The nasal suction pipe shall be worn as per the method
in the right figure and the other end is inserted in the
hydrogen outlet of humidification bottle.
When to drain water
1. For the first use, soak the new equipment in water for 10 minutes and then
drain water;
2.When the equipment indicates low water level or abnormal water quality in
the water tank, drain residual waste water in the water tank and add new
purified water;
3.When the equipment indicates full water, drain some water till the full water
indicator turns white;
4.When the equipment is used for 14 days or not used for over 14 days, add
pure water to clean the whole waterway.
Drainage method
1. Place the water container below the drain outlet.
2.Unscrew the drain outlet nut.
3. Take out the silica gel plug
4.After the water tank is emptied, plug the silica gel plug
again and tighten the drain nut.
Instructions for Installation and Use
Satisfactory purified water is requried
To ensure the normal operation of equipment, please purchase bottled
distilled water or purified water (bottled water with mineral water and purified
water labels) in advance. To prolong the service life of the equipment,
distilled water and deionized water are preferably recommended.
For Example:
Watsons distilled water First class high purity
ionized water for
laboratory use
C'estbon purified
water for drinking
●Tap water should not be used.
●Water produced by household water purifier should not be used.
●Mineral water or natural water on the market should not be used.
●Equipment failure caused by incorrect addition of non-conforming
water source will result in warranty disqualification.
Operating steps for the first use
1. Install the humidification bottle clamp on the equipment
with a screw.
2.Open the humidification bottle cover and then add drinking
water to the high water level in the humidification bottle.
(To cool the hydrogen add ice water.)
Tips: Add water to the low water level to effectively lower
the noise of bubble blasting.
3.Insert the nasal suction pipe in the air outlet of humidification
bottle, connect the air inlet of humidification bottle and
hydrogen outlet of equipment with a short pipe and then
put the humidification bottle in the fixing clamp.
(Note: The humidification bottle has separate air inlet
and outlet. Please do not install them reversely.)
High water line
Low water line
03 04

Precautions for Use
●Never open the pure water tank cover when the equipment is working.
●For healthy, the nasal suction pipe is dedicated to specific person.
●When the machine is used for a long time, the outer surface of the machine
will heat up, which is a normal phenomenon.
●When the equipment has not been used for more than 14 days, please
empty water in the pure water tank and add new pure water.
●When the "Pure water" indicator is red, it means that the water quality
is wrong or deteriorated in the pure water tank. Please empty the water
tank and add new pure water.
●When the "Low water level" indicator is red, it means low water level in
the pure water tank. At this time, please drain surplus water in the tank
and do not refill water directly.
●When the "Full water" indicator is red, it means that there is too much
water in the water purification tank. At this time, drain part water till the
full water indicator is white.
●When the "Maintenance" indicator is red, it means that the whole
waterway needs to be cleaned. At this time, please drain water in the
water tank and flush the water tank with new purified water for several
times. If the maintenance indicator is still on, it should be returned to
factory for repairing.
●If the equipment is in continuous use for more than 14 days, please
drain water to prolong its service life.
●In the using process, if the equipment is tilted for 45 degrees, the
equipment will be stopped automatically.
0605
●Never unplug the power plug with wet hand, otherwise it may cause
electric shock.
●Stop use immediately in case of exception or failure, e.g. there is
scorched flavor in the equipment.
●Never use combustibles near this equipment.
●Never use this equipment at uneven place or near infants. Otherwise,
the equipment may be overturned by vibration, resulting in personal
injury or equipment failure.
●Never plug the outlet and the hydrogen intake port, otherwise it may
cause equipment failure.
●When cleaning the equipment, never directly sprinkle water on the
equipment parts, otherwise it may cause short circuit or failure.
●Never soak the equipment in water or sprinkle water on the equipment
to add water, otherwise it may cause short circuit or failure.
● Never use this product near fire, otherwise it may cause explosion.
●Never keep the hydrogen outlet of this equipment near the lit cigarette,
otherwise it may cause explosion.
Safety Precautions

Service and Maintenance
Humidification bottle
After cleaning the humidification bottle, the humidification bottle cover
and the catheter in the humidification bottle with neutral detergent, rinse
and wipe off the remaining water.
Equipment
Avoid rinsing the equipment directly with water, which may cause leakage,
electric shock and other dangerous faults.
When the equipment is not used, please unplug the adapter power.
If the equipment has not been used for more than 14 days, please empty
the water in the water purification box and add new purified water for use.
When the equipment is in continuous use for more than 14 days, please
drain the equipment once and then add new purified water for use.
H300 Hydrogen inhaler
Humidification bottle
Fixing clamp for
humidification bottle
Nasal suction pipe (gift)
Operating manual
and warranty card
1Pcs
1Pcs
1Pcs
2Pcs
1Pcs
07
Packing List and Compulsory Accessories
Table of contents
Popular Medical Equipment manuals by other brands

Morita
Morita veraviewepocs 2d Instructions for use

Cardinal Health
Cardinal Health Kangaroo Connect Enteral Feeding Pump Directions for use

Breas
Breas NIPPY Quick reference guide

CROSSTEX
CROSSTEX SYCLONE Technical installation guide

devibliss
devibliss D100F-S manual

Utah Medical Products
Utah Medical Products Filshie Sterishot II Inservice Reference Guide

EDAN INSTRUMENTS
EDAN INSTRUMENTS SE-12 Express Notes

Acteon
Acteon XMIND TRIUM Operator's manual

Halyard
Halyard MIC J FEEDING TUBE WITH ENFit CONNECTOR PATIENT CARE INSTRUCTIONS

Winncare
Winncare XO3R manual

BILT-RITE
BILT-RITE 10-65200 manual

Avery Weigh-Tronix
Avery Weigh-Tronix Eyecon 9430 Quick reference guide