Tighten Screwdriver Stopper (Option)
Threaded Portion
Handpiece
E-type Spray Nozzle
PANA SPRAY Plus/PANA SPRAY
Irrigation Nozzle
Irrigation Tube
4. The Bur
6. Irrigation Nozzle
7. Post-use Maintenance
CAUTION
After washing with thermo-disinfector and prior to lubrication, dry the product until all
internal moisture is thoroughly removed. Thermo-disinfector moisture remaining inside
the product could reduce the effect of lubrication and could cause corrosion inside of
the product.
WARNING
CAUTION
Fig. 4
Fig. 3
Fig. 5
5. Checking the Handpiece Before Each Use
The SGS/SGA series handpieces are designed for use with surgical burs (ø2.35mm, approx.
70mm in shank length).
Follow the check below before use. If any abnormalities are found, stop using the handpiece
immediately and contact your Authorized NSK Dealer.
1) Mount the Bur. (Refer to “3. Mounting and Removing the Bur”)
2) Adjust the handpiece rotation speed in accordance with the allowable maximum speed
for the attached bur.
3) Rotate the handpiece and check for abnormal rotation, vibration, noise and overheating.
If any of these conditions are found, stop using the handpiece and
contact your
Authorized NSK Dealer.
4) If there are no immediate apparent abnormalities, continue to rotate the handpiece for
about 1 minute. Then stop rotation and check for abnormal heating on the outer
surface. If abnormal heating is detected, stop using the handpiece and
contact your
Authorized NSK Dealer.
5) Operate the handpiece if no abnormalities are observed.
To Mount the Stopper (ONLY for SGS-ES, SGA-ES)
When using a bur designed for dental straight handpieces in implant surgery, etc., attach
the Stopper (Option) to the handpiece.
1) Remove the bur from the handpiece. (Refer to “3. Mounting and Removing the Bur”)
2) Turn the Bur Lock Ring to the OPEN position. (Fig. 2)
3) Insert the threaded portion of the Stopper into the chuck then screw it in with the
provided screwdriver.
Stopper: Order Code Y900160
Post-use Maintenance includes cleaning, lubricating, and sterilization.
Insert the Irrigation Nozzle (Option) into the handpiece. Then push the Irrigation Tube into
the nozzle securely.
Perform periodical maintenance checks every three months, referring to the check sheet
below. If any abnormalities are found, contact your Authorized NSK Dealer.
Irrigation Nozzle
for SGA-ES, SGA-E2S, SGS-ES, SGS-E2S: Order Code H263018
for SGA-EM, SGA-E2M, SGS-EM, SGS-E2M: Order Code H263023
for SGA-EL, SGA-E2L, SGS-EL, SGS-E2L: Order Code H263024
Immediately after a surgery, the handpiece should be cleaned, lubricated and
sterilized. Failure to properly maintain the handpiece can lead to contaminants
remaining inside the handpiece. Unremoved blood may result in product failure
due to the internal blood coagulation. Improper maintenance may also lead to
overheating, causing burn injuries, product failure, or difficulty in attaching and
detaching the bur.
8. Periodical Maintenance Checks
CAUTION
The handpieces stated below are not suitable for the stopper. Use only with a surgical bur.
SGS-EM,SGS-EL,SGS-E2S, SGS-E2M, SGS-E2L, SGA-EM, SGA-EL, SGA-E2S,
SGA-E2M, SGA-E2L
(1) Cleaning
1) Clean debris from the surface of the handpiece with running water and wipe to remove
moisture.
2) Insert the Irrigation Tube firmly into the Irrigation Nozzle and clean the inside of the
Irrigation Nozzle with clean water.
This icon denotes that the product can be washed via thermo-disinfector.
Refer to the thermo-disinfector manual.
(2) Lubrication
1) Push E-type spray nozzle attachment over the
lubricant spray nozzle until it firmly seats.
2) Remove the bur from the handpiece.
3) Insert the spray nozzle in the rear of the
handpiece and supply lubricant into the
handpiece 2-3 times for 2-3 seconds each time.
Repeat lubrication until the oil is free of blood
and debris.
ŘThe handpiece should be lubricated with lubricant spray (PANA SPRAY Plus/PANA
SPRAY).
Ř+ROGWKHVSUD\FDQXSULJKW
Ř,IWKHRLOGULSSLQJIURPWKHKDQGSLHFHKHDGLV127FOHDQHYHQDIWHUUHSHDWHG
lubrication, internal parts may be worn out. Contact your
Authorized NSK Dealer.
Ř%HVXUHWRKROGWKHKDQGSLHFHƂUPO\WRSUHYHQWVOLSSLQJRXWRIKDQGE\WKHVSUD\
pressure when lubricating.
Ř6XSSO\OXEULFDQWXQWLOLWFRPHVRXWRIWKHKDQGSLHFHKHDG
CAUTION
ŘBe sure to clean and lubricate the handpiece prior to autoclaving. If blood remains on
the internal surface it can become clotted and cause product failure.
Ř'RQRWDXWRFODYHWKHSURGXFWZLWKRWKHULQVWUXPHQWVHYHQZKHQLWLVLQDSRXFK7KLVLV
to prevent possible discoloration and damage to the product from chemical residue on
other instruments.
Ř
Do not use high acid water or sterilizing solutions to wipe, immerse or clean the product.
Ř,IWKHVWHULOL]HUFKDPEHUWHPSHUDWXUHPD\H[FHHGr&GXULQJWKHGU\F\FOHWKHQ
delete the dry cycle.
Ř.HHSWKHSURGXFWLQVXLWDEOHDWPRVSKHULFSUHVVXUHWHPSHUDWXUHKXPLGLW\YHQWLODWLRQ
and sunlight. The air should be free from dust, salt and sulphur.
Ř'RQRWWRXFKWKHSURGXFWLPPHGLDWHO\DIWHUDXWRFODYLQJDVLWZLOOEHYHU\KRWDQGPXVW
remain in a sterile condition.
(3) Sterilization
Sterilize the handpiece by autoclave sterilization.
Sterilization is required after each patient as noted below.
Autoclave Procedure
1) Insert into an autoclave pouch. Seal the pouch.
$XWRFODYDEOHXSWRPD[r&
([$XWRFODYHIRUPLQDWr&
3) The handpiece should remain in the autoclave pouch until required for use.
6WHULOL]DWLRQDWr&IRUPRUHWKDQPLQXWHVLVUHFRPPHQGHGE\(1RUEN
ISO17665-1.
Specifications are subject to change without notice. 2014.12.20 06 M
The EU directive 93/42/EEC was applied in the design and production of this medical device.
Points to check
Bur mounting/removing
Rotation
Details
Make sure thatthe bur can be properly mounted/removed.
After mounting, make sure that the bur is firmly fixed by pulling
and pushing it.
Rotate the handpiece and check for abnormalities such as
abnormal rotation, vibration, noise, and overheating.