User Instructions Low Temperature Freezers Model UNI/LT 6
EVK 012
Thermoregulator for general purposes
18 Wiring diagram
With reference to the wiring diagrams:
•the polarity of relay K1 depends on parameter
u9; according to the default setting, the relay
is activated when the regulator is turned off
and deactivated when the regulator is turned
on
•the polarity of relay K2 depends on parameter
u3; according to the default setting, the relay
is activated during the normal operation and
deactivated during an alarm
•the serial port (by request) is the port for the
communication with the supervision system
(through a serial interface, via TTL, with MOD-
BUS communication protocol) or with the
programming key; the port must not be used
at the same time for the same purposes.
Additional information for electrical connec-
tion:
•do not operate on the terminal blocks with
electrical or pneumatic screwers
•if the instrument has been moved from a cold
location to a warm one, the humidity could
condense on the inside; wait about an hour
before supplying it
•test the working power supply voltage,
working electrical frequency and working
electrical power of the instrument; they must
correspond with the local power supply
•disconnect the local power supply before
servicing the instrument
•do not use the instrument as safety device
•for repairs and information on the instrument
please contact Evco sales network.
19 USER INTERFACE
19.1Turning on/off the instrument
To turn on the instrument you have to supply
it; to turn it off it is enough to cut off the power
supply.
19.2 Locking/unlocking the keyboard
To lock the keyboard:
•make sure no procedure is running
•press and 2 s: the display will show
“Loc” 1 s.
If the keyboard is locked, you will not be al-
lowed to:
•modify the working setpoint with the proce-
dure related in paragraph 4.1 (you also can
modify the working setpoint through param-
eter SP).
These operations provoke the visualization of
the label “Loc” 1 s.
To unlock the keyboard:
•press and 2 s: the display will show
“UnL” 1 s.
19.3 Silencing the buzzer
•make sure no procedure is running
•press a button (the first pressure of the button
does not provoke its usual effect).
20 SETTINGS
20.1 Setting the working setpoint
•make sure the keyboard is not locked and no
procedure is running
•press LED out 1 will flash
•press or in 15 s; also look at param-
eters r1, r2 and r3
•press or do not operate 15 s.
You also can modify the working setpoint
through parameter SP.
20.2 Setting configuration parameters
To gain access the procedure:
•make sure no procedure is running
•press and 4 s: the display will show
“PA”
•press
•press to set “-19”
•press
•press and 4 s: the display will show
“SP”.
To select a parameter:
•press or
To modify a parameter:
•press
•press or
•press or do not operate 15 s.
8