argon audio CLEANER 15 Manual

1
English
All directions should be read before use
pages
Directions For Use in Peripheral Vasculature ....... 2-5
Directions For Use in Grafts and Fistulae.............. 6-9
Caution: Federal (USA) law restricts this device to sale by, or on the order of physician.
DEVICE DESCRIPTION:
The CLEANER 15™ Rotational Thrombectomy System is a 7Fr percutaneous catheter
based system (single piece construction). A disposable, hand-held battery operated
rotator drive unit is attached to a sinuous wire which rotates at approximately 4000 RPM.
The wire and atraumatic soft tip are radiopaque for fluoroscopic visualization.
Sliding Lever
(Sinuous Wire Cover/Uncover) Battery Operated Hand Held
Rotator Drive Unit
Sinuous Wire
7 Fr Catheter
Figure 1
On/Off Switch
Catheter Lumen
Flush/Infusion Port
& Stopcock

2
English
Peripheral Vasculature
INDICATIONS FOR USE:
The CLEANER 15™ Rotational Thrombectomy System is indicated for mechanical
declotting of acute symptomatic deep vein thrombosis and controlled and selective
infusion of physician-specified fluids, including thrombolytics, in the peripheral
vasculature.
CONTRAINDICATIONS:
The CLEANER 15™ Rotational Thrombectomy System is contraindicated in the following:
-
This system is contraindicated when, in the medical judgment of the physician, such
a procedure may compromise the patient’s condition.
-
Not designed for peripheral vasculature dilation purposes.
-
This system is not intended for the infusion of blood or bloodproducts.
-
Refer to the product insert of the therapeutic solution of choice for indications,
contraindications, side effects, and precautions.
-
In native vessels smaller than 6mm in diameter
WARNINGS AND PRECAUTIONS:
-
Operation of a thrombectomy catheter may cause embolization of some
thrombus and/or thrombotic particulate. The potential for pulmonary
thromboembolism should be carefully considered when Cleaner is
used to break up and remove peripheral venous thrombus. Physician
discretion with regard to the use of the distal protection in patients
with DVT is advised.
-
Prior to use, read all package insert warnings, precautions, and instructions. Failure
to do so may result in severe patient injury anddeath.
-
These procedures should only be performed by physicians and staff familiar with the
equipment and techniques involved. The device has been sterilized by EtO and is
sterile unless the package is opened or damaged. The package should be examined
before use; if damaged, DO NOT USE. The device is intended for single patient use
only; DO NOT REUSE OR RE-STERILIZE.
-
Prior to use, carefully examine the CLEANER 15™ Rotational Thrombectomy System
to verify that it has not been damaged during shipment. If the product components
show any sign of damage, DO NOTUSE.
-
Due to the risk of exposure to HIV (Human Immunodeficiency Virus) or other blood
borne pathogens, health care workers should routinely use universal blood and body
fluid precautions in the care of all patients. Sterile technique must be strictly adhered
to during any handling of the device.
-
Do not advance if resistance is met without first determining the cause of resistance
under fluoroscopy and taking any necessary remedial action. Excessive force against
resistance may result in damage to the device orvasculature.
-
Prior to introduction, and anytime the CLEANER 15™ is removed from the vascular
system, the catheter should be flushed.
-
If the catheter or guidewire becomes kinked or otherwise damaged during use,
discontinue use and replace.
-
Practitioners must be aware of potential complications associated with peripheral
vasculature thrombolysis including:
Vessel perforation or rupture Distal embolization of blood clots
Vessel spasm Hemorrhage
Hematoma Pain and tenderness

3
English
Peripheral Vasculature
Sepsis/Infection Thrombophlebitis
Intimal disruption Arterial dissection
Vascular thrombosis Drug reactions
Allergic reaction to contrast medium Arteriovenous fistula
Thromboembolic episodes Amputation
-
Potential fatigue failure of the CLEANER 15™ sinuous wire may occur with prolonged
activation of the CLEANER 15™ device. A withdrawal rate of 1-2 cm/second is
recommended when sharp radii are encountered (i.e. radius of a loop graft or fistula,
radii < 3 cm).
A SUGGESTED PROCEDURE:
Use sterile technique.
Patient Preparation:
1.
Premedicate with appropriate anxiolytic, analgesic and/or antibiotic per hospital
protocol.
Device Performance Testing:
2.
Remove the CLEANER 15™ Rotational Thrombectomy System from the package.
Press the ON/OFF switch to ensure that the sinuous wire spins freely (refer to
Figure 2). Release the switch to stop the rotator. Precaution: Do not use the
device if it does not activate immediately when the switch is pressed, and
deactivate immediately when the switch is released.
Figure 2
3.
Attach a syringe of therapeutic agent or heparinized saline to the catheter flush port
and flush the CLEANER 15™ catheter. Be sure that fluid exits from the most distal
catheter hole which is located near the end of the catheter. Close the stopcock to
“lock” the therapeutic agent or heparin in the catheter and remove thesyringe.
Treatment:
The exact treatment procedure is to be determined by the physician. The following
option describes how the procedure may be performed.
Sliding Lever
Sinuous Wire
On/Off Switch

4
English
Peripheral Vasculature
4.
Complete the procedure under continuous fluoroscopy. Do not initiate sinuous wire
rotation (device activation) unless proper device positioning is confirmed within the
peripheral vasculature.
5.
Prepare and drape the puncture site as required.
6.
Administer local anesthetic at the puncture site for venous sheathinsertion.
7.
Select an appropriately sized sheath to accommodate the CLEANER 15™ catheter
and other devices/ catheters that may be used during the procedure. Maximum
guidewire size will be dependent upon the introducer sheath/dilator assembly
chosen. If crossing the iliac bifurcation, a long reinforced sheath should beused.
8.
Prepare and place the venous introducer sheath per hospital protocol. The venous
sheath should be placed in the venous limb of the peripheral vasculature, and
directed toward the treatment site. The venous sheath placement can be optional
depending on the clot burden in the vessel.
9.
Place the device in the covered position by pushing the sliding lever to the distal
position and rotating the sliding lever to lock in the covered position (refer to Figure
3). When in the covered position, only the flexible tip of the sinuous wire should
extend from the catheter. Warning: The device should not be activated in the
covered position.
10.
Support the flexible tip between the thumb and index finger during insertion through
the sheath valve. Insert the covered device through the venous sheath and into the
venous limb of the peripheral vasculature.
11.
Attach a 10cc syringe containing the therapeutic agent intended for infusion to the
stopcock. Open the stopcock to enable flow. Warning: Do not draw blood back
into the catheter.
Note: Follow the manufacturer’s instructions regarding reconstitution and
dilution of specified fluids.
Sliding Lever Locked in
Covered Position
Covered Sinuous Wire
(Do Not Activate Device
in Covered Position)
Figure 3

5
English
Peripheral Vasculature
12.
Advance the flexible tip up to the distal most extent of the clot. Uncover the sinuous
wire by unlocking, fully retracting the sliding lever and rotating the sliding lever until
an audible “click” is heard (refer to Figure 4). Confirm device positioning within the
treatment site via fluoroscopy. Press the ON/OFF switch to activaterotation.
Sliding Lever Locked In Uncovered Position (Deployed)
Figure 4
13.
With the device activated, slowly withdraw the rotating sinuous wire along the
treatment site while infusing therapeutic agent through the infusion port to break up
the clot. Warning: A withdrawal rate of 1-2 cm/second is recommended when
sharp radii are encountered. When the sinuous wire reaches the tip of the venous
sheath, release the switch to turn off the rotating dispersionwire.
14.
Cover the device and remove it from the peripheral vasculature. Flush the catheter
lumen with heparinized saline and manually remove any accumulated fibrin from the
sinuous wire.
15.
Aspirate the macerated clot via the sheath and discard the aspirate. Precaution:
Continued unsuccessful aspiration may collapse the sheath andvessel.
16.
Inject a small amount of contrast via the venous sheath to assess the degree of
thrombus removal accomplished. Warning: Avoid over-injection of contrast to
minimize the risk of embolization. Treat the residual thrombus by repeating steps
11-15 until acceptable thrombus removal is achieved.
17.
When the thrombus removal is complete, treat any underlying disease or stenosis
per hospital protocol.
18.
Perform the final angiogram.
19.
Remove the sheaths from the peripheral vasculature.
20.
Achieve hemostasis at the puncture site(s) per hospitalprotocol.
Deployed Position
(Uncovered For Device Activation)
Table of contents
Languages:
Other argon audio Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Getinge
Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Mettler Electronics
Mettler Electronics Sonicator 730 Maintenance manual

Pressalit Care
Pressalit Care R1100 Mounting instruction

Denas MS
Denas MS DENAS-T operating manual

bort medical
bort medical ActiveColor quick guide

AccuVein
AccuVein AV400 user manual