Setup ↓Scale →Options →Security →Serial A →Serial B →
BCD Binary Coded Decimal which allows transmission to certain printers. This paramter appears only when the
SC-150 is enabled.
Setup ↓Scale →Options →Security →Serial A →Serial B →BCD →Analog ↓
Output This lets you specify which weight the analog output will follow. Choices are: off, displayed weight, gross
weight, or net weight.
Setup ↓Scale →Options →Security →Serial A →Serial B →BCD →Analog ↓Output →
Units Analog output is independent of the unit of measure selected for the display. This menu lets you specify
which unit of measure is used for the analog output. Choices are: pounds, 1000 g, or gallons.
Setup ↓Scale →Options →Security →Serial A→Serial B →BCD →Analog ↓Output →Units →
Full When selected, the indicator will display the last value used or the default value. You enter the full capacity of
the analog output which may be less than or greater than the capacity of the scale. For example, the capacity
of the indicator may be 5000 lb, but it may be desirable to have 3000 lb as the full capacity of the analog
output. In any case, the analog output has nominal under range and over range limits of 20%.
Setup ↓Scale →Options →Security →Serial A →Serial B →BCD →Analog ↓Output →Units →Full →Adjust ↓
Zero →Full →End
The choices present under
ADJUST
allow the zero and the span of the analog outputs to be adjusted without
actually putting weights on and off the scale.
Selecting
ZERO
lets you adjust the zero of the analog output for a zero weight reading. This is done by
pressing the 0, 1, 2, 3, or 4 key to increase the output and by pressing the 5, 6, 7, 8, or 9 key to decrease the
output. The number on the display gives a visual representation of the zero setting, with 00.000 being the
nominal value. The zero adjustment has a ±10% range, -10.000 to +10.000 on the display.
Selecting
FULL
lets the operator adjust the span of the analog output for the full capacity weight reading. The
keys and the number on the display function like the zero adjustment above, with 100.000 as the nominal full
capacity value. The span has a +/- 10% range, 90.000 to 110.00 on the display. Weight does not have to be
on the scale to perform this task.
When
ZERO
,
FULL
, or
END
are displayed, the analog output follows the value selected under
OUTPUT
and
UNITS
. The only time the value is not output is while actually adjusting zero or full.
Setup ↓Scale →Options →Security →Serial A →Serial B →BCD →Analog →
Sleep This option, which is available only on the battery pack version, lets you set the time to elapse before sleep
mode begins. Set 0 to disable the sleep mode. Up to 99 hours, 99 minutes and 99 seconds are program-
mable. The sleep timer is reset each time a button is pressed or scale motion more than twice the stability
window is seen.
Setup ↓Scale →Options →Security →Serial A →Serial B →BCD →Analog →Sleep →
Seal AllWhen switch S1-1 is in the OFF position, if you choose the
YES
option, all items under configuration are
sealed . If
NO
is selected, units, capacity, division, zero range, stability, AZT, tare, layout, zero, span, linear-
ity, and seal all are sealed.
Setup →Adjust ↓
Zero, Span, Linear., Display
This option lets you set the zero, span, and linearity of the indicator. The following pages contain specific
instructions for setting these parameters.