Explanation of Graphical Symbols :
Sterilization of Fiber-Optic Probe :
After each treatment, the fiber-optic probe should be disinfected or sterilized. Wrap
completely in an autoclave sterilization pouch, then autoclave at 132°C/75 psi for 4
minutes by using a pre-vacuum sterilizer to get maximum results. Allow the fiber-optic
probe to air-dry for 20-30 minutes in the sterilization pouch before removing.
IMPORTANT :
Disconnect the fiber-optic probe from the handpiece and autoclave by itself.
Product Cleaning:
If there is any dirt or stains on the surface of control panel, handpiece or other
components, please use a cloth to wipe clean. If it is hard to clean, please use
a neutral cleansing agent. 70% isopropyl alcohol is recommended.
IMPORTANT :
Never use a solvent or volatile oil for cleaning; this can lead to fading
of the print on the casing or the metal ring falling off from the probe.
Class II equipment
ISO 7000-1641 : Follow operating instructions or consult instructions
for use
Cooling System :
>> This device is designed with a built-in cooling fan which automatically activates
whenever the device is in use, and continues to run for a few seconds after the
cycle has ended.
>> If the temperature gets exceeds 45°C, the fan will continue to run until the
internal temperature has cooled down.
>> If the temperature gets exceeds 65°C, the device will automatically shut off and
the fan will continue to run. Until the internal temperature has cooled down, the
device will be able to restart.
>> If the unit needs to be used for continuous curing, please decrease the curing time.
>> Under the premise of not touching the oral tissue, when using the curing light or
switching the treatment area or switching any operation settings, it is necessary
to cool down for seconds before starting the next curing cycle in case the curing
light being used for over 15 seconds during a curing cycle for more cycles.
IEC 60601-1 : Applied part classification - Complies with standards
protecting against electric shock for type B equipment
IEC 60417-5031 : Direct current.
IEC 60417-5009 : STAND-BY
IMPORTANT :
Every component of this device is reusable, but only the fiber-optic probe
needs to be disinfected or sterilized. Other components can not be disinfected
or sterilized.
Disclaimer:
1. First of all, we would like to thank you for purchasing BA Ultimate
BASE280 LED Curing Light. Please carefully read the user manual and
operate the device accordingly. Please keep this user manual well
preserved for future reference.
•
•
•
•
•
•
Content:
•
Intended Use.......................................................................................
Explanation of Graphical Symbols.......................................................
Parts & Functions.................................................................................
Installation............................................................................................
Panel operation....................................................................................
Charger signals....................................................................................
Troubleshooting...................................................................................
3
3
4
4
4
8
9
2. BA International shall not assume any responsibility if any malfunction,
damage, or accident resulting in bodily injury caused by improper removal,
modification, maintenance, or repair of the product conducted by unauthorized
personnel.
3. BA International shall not assume any responsibility if any malfunction,
damage, or accident resulting in bodily injury is caused by the use of this
product in conjunction with other regulated products.
4. BA International shall not assume any responsibility if any malfunction,
damage, or accident resulting in bodily injury is caused by improper
removal, modification, maintenance, or repair of the product with
components not supplied by the manufacturer.
5. BA International shall not assume any responsibility if any malfunction,
damage, or accident of bodily harm is caused by improper operation that is
incompliant with the manual instructions.
6. BA International shall not assume any responsibility if any malfunction,
damage, or accident resulting in bodily harm is caused by inappropriate
modifications to the power supply, installation environment, or any other
safety-regulated operating conditions as specified in this manual.
7. BA International shall not assume any responsibility if any malfunction,
damage, or accident resulting in bodily harm is caused by natural disasters,
environmental causes, or acts of God.
•Safety precautions............................................................................... 4
•Specifications........................................................................................ 9
•EMC Tables.......................................................................................... 9