Initial Meter Setup
You can set up the meter using the TFX-500w keypad or the SoloCUE Flow Device Manager software. This document
addresses procedures using the TFX-500w keypad. To use SoloCUE, see the "SoloCUE Flow Device Manager Installation Guide"
available at www.badgermeter.com.
When you start the meter for the first time, you must select a language, press ENTER, then press MENU/BACK to get to
BASIC SETUP.
In BASIC SETUP, program the parameters in the table below using the transmitter's keypad. Enter the pipe characteristics,
transducer and mounting in the SETUP > METER submenus. For integral mount meters with DTTS/C transducers, these
parameters are already set at the factory and you can skip these steps.
For in-depth parameter programming, see “Parameter Descriptions by Menu” on page18.
Submenu Parameter Action
PIPE
PIPE MATERIAL Select the material of the pipe.
PIPE TYPE
Select the pipe schedule for ANSI pipes or manual entry of the outside diameter (O.D.) in
millimeters or inches. If you select manual entry, you will need to enter the pipe wall thickness.
The available options are based on the pipe material selected. If you do not see a valid option,
check the pipe material setting.
PIPE SIZE NOMINAL When you select an ANSI pipe schedule, you need to select the nominal pipe size in inches. If
you do not see a valid option, check the pipe type.
PIPE SIZE and WALL
THICKNESS
When you select MANUAL for Pipe Type, enter the outer diameter and wall thickness of the
pipe. The units are based on whether MANUAL INCHES or MANUAL MM was selected for the
Pipe Type.
LINER THICKNESS
and LINER MATERIAL
(optional)
If there is a liner in the pipe, enter the liner thickness and select the liner material. The units are
based on the Pipe Type. If you do not see a valid option, check the Pipe Type.
I.D. SIZE The calculated inner diameter based on settings.
TRANSDUCER
DTTN/DTTR 1 MHZ
DTTSU 2 MHZ
DTTS/DTTC 2 MHZ
DTTJ/K EASYRAIL 1 MHZ
If the meter was ordered as a single part number, the transducers are configured at the
factory. Otherwise, select the transducer model from the list. The model is marked on one of
the transducer heads. If you do not see the transducer model in the list, select a transducer
with the same frequency.
MOUNTING
OR
Z-PATH
V-PATH
W-PATH
Select the mounting path to match the required setup. See the transducer user manual to
select the best path.
DTTS/DTTC TYPE DTTS/DTTC TYPE substituted for MOUNTING when TRANSDUCER > DTTS/DTTC is selected.
SPACING SPACING CALCULATED View to see the correct spacing for the transducers.
UNITS See “Setup > Units” on
page18.Select the units and format of flow rate, total and velocity.
FLOW SETUP
See “Setup > Meter
> Flow Setup” on
page21.
Select flow direction, low flow cutoff, signal cutoffs and filtering.
CALIBRATION
See “Setup > Meter
> Calibration” on
page24.
Check that FACTOR MODE is set to FIELD. Enter the CAL FACTOR from the transducer pair into
the SCALE FACTOR.
SET ZERO Confirmation screen
1. Check that the pipe is full of liquid and not owing. Flow must be absolutely zero.
2. Securely close any valves and allow time for settling to occur.
3. Select SET ZERO and press OK to set the new zero.
Menu Map