manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. KROHNE
  6. •
  7. Measuring Instrument
  8. •
  9. KROHNE SD 200 W User manual

KROHNE SD 200 W User manual

Quick Start
For initial set up, we strongly recommend the use of the relevant manuals in addition!
02/2022 – 4009189401 - QS SD 200 R01 en 1
SD 200 W/R Multiparameter indicator for
analytical and other parameters
2 Electrical connection
Connection with junction box SJB 200 W
SJB 200 W with SMARTPAT sensor, integrated HART® resistor and indicator (display) SD 200 W/R (left side). SJB
200 W with SMARTPAT sensor, indicator (display) SD 200 W/R and without integrated HART® resistor (right side).
Connection without junction box SJB 200 W (only with wall mounting version SD 200 W)
 Field device
 SD 200 W
 Power supply / Control room
 U+
 Sensor+
 Sensor-
 U-
Dimensions [mm] Dimensions [inch]
a 145 5.71
b 120 4.72
c 130 5.12
d 30 1.18
e 75 2.95
f 60 2.36
g 112 4.41
Installation, assembly, start-up and maintenance may only be performed by appropriately trained personnel.
Check the nameplate for correct operating conditions.
For use in hazardous areas, special codes and regulations are applicable. Instruments must not be
connected to power supply before reading instructions described in the supplementary manual.
The responsibility as to the suitability, intended use and corrosion resistance of the used materials against
the measured fluid of this device rests solely with the operator.
1 Installation
Rack mount
Wall mount
Fasten the indicator securely to the wall by using the two marked holes (h)
Dimensions [mm] Dimensions [inch]
a 72 2.38
b 144 5.67
c 30 1.18
d 16 0.63
e 68 2.68
f 138 5.43
Danger:
All work on the electrical connections may only be carried out with the power disconnected.
Take note of the voltage data on the nameplate.
Observe the national regulations for electrical installations.
Specific conditions of use: Yes as shown below:
When the enclosure of the indicator is made of aluminium alloy, when used in a potentially explosive
atmosphere requiring apparatus of equipment protection level Ga, the indicator shall be installed so, that even
in the event of rare incidents, an ignition source due to impact or friction sparks between the enclosure and
iron/steel is excluded.
For EPL Da the ambient temperature Ta shall not exceed 50 °C.
Sensor+
Sensor-
Display+
Display-
Loop- w/Display
Loop+
Loop+
Sensor+
Sensor-
Display+
Display-
Loop- w/Display
02/2022 – 4009189401 - QS SD 200 R01 en 2
Grounding and equipotential bonding
For connection with junction box SJB 200 W:
For connection without junction box SJB 200 W:
Setup
Configuration of the SD 200 W/R indicators is done at SETUP-level.
SETUP-level is reached by pressing the PROG/ENTER key for 7 seconds; at which time, the indication "SETUP" will be
displayed.
In order to return to the operator level, PROG will have to be pressed for three seconds. Alternatively, if no keys are
pressed for 2 minutes, the unit will exit SETUP automatically.
SETUP can be reached at all times while the indicator remains fully operational.
Press the key for 7 seconds to enter the SETUP-level
Selection of function-group and function: SETUP is devided into several function-groups and functions.
Select function-group with
Select function with
Each function has a unique number, which is displayed beside the word "SETUP" at the bottom of the indicator. This
indicator has 2 function-groups: 1 FUNCTION, 2 OTHERS. Press SELECT to select a function.
After selecting a sub-function, the next main function is selected by scrolling through all "active" sub-functions (e.g. 1,
11, 12, 13, 14, 1, 2 etc.).
Change or select a value:
 Press briefly, PROG will start flash
 Select a value by pressing or
 Press to confirm the value / selection
To change a value, use  to select the digits and  to increase that value. To select a setting, both  and  can be
used.
When data is changed but ENTER is not pressed, then the modification can still be cancelled by waiting for 20
seconds or by pressing ENTER for three seconds: the PROG-procedure will be left automatically and the former value
reinstated.
To return to OPERATOR-level press for 3 seconds.
In order to return to the operator level, PROG will have to be pressed for three seconds. Also, when no keys are
pressed for 2 minutes, SETUP will be left automatically.
Setup function and variables
1 Function
11 Unit (measurement unit to be displayed)
12 Decimals (number of digits following the decimal point)
13 Offset (value at minimum signal)
14 Span (measuring range from 4 mA (offset) to 20 mA)
15 Filter (response time on step change of analog value)
16 Calibrate low (re-calibration of input value for 4 mA)
17 Calibrate high (re-calibration of input value for 20 mA)
2 Others
21 Model (reflects the model)
22 Type (reflects the product specific functionality: SD 200 W/R)
23 Software version (for support and maintenance)
24 Serial No. (for support and maintenance)
25 Password (up to 4 digits can be programmed; disabled with 0000)
26 Tag number (for identification; max. 7 digits)
Download documents and software
Scan the code on the nameplate or scan the
following code and enter the serial number.
Contact
Select your country from the region / language selector to
view your local KROHNE contact details on:
www.krohne.com
Danger:
The housing must be connected to the equipotential bonding of the hazardous area. Please use the
ground connection facility (terminal FE) inside the housing.
The sensor cable shield must be connected to the connection facility (terminal S).
Consider the information mentioned in the supplementary instruction of the SMARTPAT sensor.
Information:
In case of performance problems: Connect the external ground with the cable shield internally via luster
terminal.
Information:
A password may be required to enter SETUP. Without this password access to SETUP is denied.
Information:
Modification will only be set after ENTER has been pressed!

This manual suits for next models

1

Other KROHNE Measuring Instrument manuals

KROHNE OPTIWAVE 3500 C User manual

KROHNE

KROHNE OPTIWAVE 3500 C User manual

KROHNE AF-E 400 User manual

KROHNE

KROHNE AF-E 400 User manual

KROHNE OPTISONIC 6300 User manual

KROHNE

KROHNE OPTISONIC 6300 User manual

KROHNE BM 26 BASIC Wiring diagram

KROHNE

KROHNE BM 26 BASIC Wiring diagram

KROHNE GA24 Wiring diagram

KROHNE

KROHNE GA24 Wiring diagram

KROHNE BATCHFLUX 3200 C Wiring diagram

KROHNE

KROHNE BATCHFLUX 3200 C Wiring diagram

KROHNE DK32 Series Wiring diagram

KROHNE

KROHNE DK32 Series Wiring diagram

KROHNE BM 70 A User manual

KROHNE

KROHNE BM 70 A User manual

KROHNE WATERFLUX 3070 User manual

KROHNE

KROHNE WATERFLUX 3070 User manual

KROHNE OPTISONIC 6300 User manual

KROHNE

KROHNE OPTISONIC 6300 User manual

KROHNE OPTISWITCH 3000 C User manual

KROHNE

KROHNE OPTISWITCH 3000 C User manual

KROHNE DK32 Series Wiring diagram

KROHNE

KROHNE DK32 Series Wiring diagram

KROHNE OPTIWAVE 7300 C User manual

KROHNE

KROHNE OPTIWAVE 7300 C User manual

KROHNE ALTOFLUX M900 User manual

KROHNE

KROHNE ALTOFLUX M900 User manual

KROHNE OPTIWAVE 7300 C User manual

KROHNE

KROHNE OPTIWAVE 7300 C User manual

KROHNE OPTISWIRL 4200 Wiring diagram

KROHNE

KROHNE OPTISWIRL 4200 Wiring diagram

KROHNE IFC 020 K User manual

KROHNE

KROHNE IFC 020 K User manual

KROHNE CORIMASS MFM 4085 K/F User manual

KROHNE

KROHNE CORIMASS MFM 4085 K/F User manual

KROHNE DWM 2000 D User manual

KROHNE

KROHNE DWM 2000 D User manual

KROHNE OPTIWAVE 6300 C User manual

KROHNE

KROHNE OPTIWAVE 6300 C User manual

KROHNE OPTIFLUX 5000 User manual

KROHNE

KROHNE OPTIFLUX 5000 User manual

KROHNE OPTISONIC 1400 User manual

KROHNE

KROHNE OPTISONIC 1400 User manual

KROHNE ALTOSONIC V User manual

KROHNE

KROHNE ALTOSONIC V User manual

KROHNE DK32 Series Wiring diagram

KROHNE

KROHNE DK32 Series Wiring diagram

Popular Measuring Instrument manuals by other brands

DODODUCK Q63 instruction manual

DODODUCK

DODODUCK Q63 instruction manual

Unitech UT-30X instruction manual

Unitech

Unitech UT-30X instruction manual

Badger Meter Dynasonics UHC100 user manual

Badger Meter

Badger Meter Dynasonics UHC100 user manual

VDO CONTISYS OBD - Quick reference guide

VDO

VDO CONTISYS OBD - Quick reference guide

Campbell 020C instruction manual

Campbell

Campbell 020C instruction manual

Anritsu Site Master user guide

Anritsu

Anritsu Site Master user guide

Beckhoff EP3356-0022 Documentation

Beckhoff

Beckhoff EP3356-0022 Documentation

newsteo LGR Series installation guide

newsteo

newsteo LGR Series installation guide

IS Technologies SLM600 Plus Series user manual

IS Technologies

IS Technologies SLM600 Plus Series user manual

Polytec TMS-350 TopMap In.Line Hardware manual

Polytec

Polytec TMS-350 TopMap In.Line Hardware manual

National Instruments PXI-5152 CALIBRATION PROCEDURE

National Instruments

National Instruments PXI-5152 CALIBRATION PROCEDURE

Knightsbridge TE8 Installation & maintenance manual

Knightsbridge

Knightsbridge TE8 Installation & maintenance manual

Dräger DrugTest 5000 Technical manual

Dräger

Dräger DrugTest 5000 Technical manual

Sekonic L-98 Instruction

Sekonic

Sekonic L-98 Instruction

Kreg Precision KMS7101 Instructional manual

Kreg

Kreg Precision KMS7101 Instructional manual

Shimpo DT-107A instruction manual

Shimpo

Shimpo DT-107A instruction manual

Polymetron 9210 user manual

Polymetron

Polymetron 9210 user manual

ASHA Beam Profile PMB-201 manual

ASHA

ASHA Beam Profile PMB-201 manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.