190 821 250 reagent tests, 1 pack of 250 reagent tests acc. to DPD1
DPD1
250 reagent tests, 1 pack of 250 reagent tests acc. to DPD1
190 823 Kit DPD4 + 1 pack of 250 reagent tests acc. to DPD3
(DPD4 obtained by DPD1 + DPD3)
190 839 Test tubes Pack of 2 glass test tubes with cap, for
photometer MW
Code Reference Designation
MES
190-11/1
POCKET PHOTOMETERS
MW line
28-04-2016 190 M1 11 D
Instructions manual
Instructions manual
Pocket photometers
MW line
FREE CHLORINE MEAS REMENT (MW-10)
Turn the meter O by pressing the button. After
all segments are displayed, “C.1”, “Add” appears
with "Press" blinking, the meter is ready.
Fill the test tube with 10 mL of unreacted sample
and tap it with the cap. Place it into the tube
holder and close the meter cap.
Press the button. When the display shows
“Add”, “C.2” with “Press” blinking, zero is
calibrated.
Remove the test tube, open it and add the
content of one DPD1 reagent test. Replace the
cap and shake gently for 20 seconds. Replace
the tube into the meter.
Wait for 1 minute and then press and hold the
button until the timer is displayed on the LCD.
The instrument directly displays the concentration
of free chlorine in ppm. The meter automatically
turns off after 10 seconds following the measu-
rement display.
PACK CONTENT
Each photometer is supplied with:
• 2 test tubes with cap
• 1 battery 1.5 V, type AAA
• 1 instruction manual
TECHNICAL FEAT RES
Range: 0.05 to 2.50 ppm
Resolution: 0.01 ppm
Accuracy: ± 0,05 ppm ± 3% of reading (25°C)
EMC Deviation: ± 0,01 ppm
Light source: LED at 525 nm
Light detector: Silicon photocell
Method: Adaptation of USEPA method 330.5.
The reaction between chlorine and the DPD
reagent causes a pink tint in the sample.
Temperature: +1...+50°C
Humidity: 85% RH, non condensing
Power supply: 1 battery 1.5 V type AAA
Auto switch-off: After 2 min
Dimensions – Mass: 87,5 x 63,7 x 35,2 mm – 65 g
REPLACEMENT PARTS
TOTAL CHLORINE MEAS REMENT (MW-11)
Turn the meter O by pressing the button. After
all segments are displayed, “C.1”, “Add” appears
with "Press" blinking, the meter is ready.
Fill the test tube with 10 mL of unreacted sample
and tap it with the cap. Place it into the tube
holder and close the meter cap.
Press the button. When the display shows
“Add”, “C.2” with “Press” blinking, zero is
calibrated.
Remove the test tube, open it and add the
content of one DPD1 and one DPD3 reagent
tests. Replace the cap and shake gently for 20
seconds. Replace the tube into the meter.
Wait for 2 minutes and 30 seconds then press
and hold the button until the timer is displayed
on the LCD.
The instrument directly displays the concentration
of total chlorine in ppm. The meter automatically
turns off after 10 seconds following the measu-
rement display.
Dust cover
Test tube (with cap)
Tube holder
LCD
Touch button