If using a soak tube* (pictured, left):
-Insert the tip of the ULT-PC-30 Submersible Conductivity Probe to the bottom of the tube.
-Fill the tube to an appropriate level with cleaning solution, saline, or water.
-Place the transducer tip of the SonoSite Probe into the tube, and submerge to the 60cm mark
(minimum) as indicated on the probe shaft.
If using a soak tray (pictured, bottom-right):
-Hang the ULT-PC-10 Conductivity Probe on the edge of the tray.
-Fill the tray with cleaning solution, saline, or water until the Conductive Probe contacts are
sufficiently submerged (approximately 1 inch, or more).
-Place the transducer tip of the SonoSite Probe into the tray, and submerge between 60cm and 90cm
as indicated on the probe shaft.
*If using an automated TEE Probe Disinfector see Appendices for details.